J Clin Microbiol. 1985 September; 22(3): 462-464
ABSTRACT
Rotazyme II, which is a shorter version of Rotazyme (less than 3 h), was compared with electron microscopy and Rotazyme for sensitivity and specificity on 229 human stool specimens. Compared with electron microscopy, the newer assay was found to be 99.4% sensitive and 97.3% specific for an overall agreement of 98.7%. After resolution of discordant results by blocking tests, the Rotazyme II enzyme immunoassay was shown to be more sensitive than electron microscopy and, therefore, 100% specific.