JCM Figure table search 04
Home Help [Feedback] [For Subscribers] [Archive] [Search] [Contents]
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Dennehy, P H
Right arrow Articles by Tente, W E
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Dennehy, P H
Right arrow Articles by Tente, W E
J Clin Microbiol. 1988 September; 26(9): 1630-1634

Comparison of nine commercial immunoassays for the detection of rotavirus in fecal specimens.

P H Dennehy, D R Gauntlett and W E Tente

Department of Pediatrics, Rhode Island Hospital, Providence.

ABSTRACT

One hundred fecal specimens obtained from patients with acute gastroenteritis were tested for rotavirus with nine commercial immunoassays to evaluate the sensitivity, specificity, predictive value, and diagnostic accuracy of these assays. Kits evaluated included two monoclonal antibody-based enzyme immunoassays (EIAs) (Rotaclone and Pathfinder Rotavirus), three polyclonal antibody-based EIAs (Rotavirus Immunoassay, Rotazyme II, and Wellcozyme Rotavirus), and four latex agglutination assays (Rotastat, Virogen Rotatest, Meritec-Rotavirus, and The Wellcome Latex Test). Thirty-eight of the 100 specimens were found to contain rotavirus by a reference microplate EIA. The accuracy of the reference assay was determined by RNA electrophoresis and a blocking assay on discordant specimens. The two monoclonal antibody EIAs had superior sensitivities (100%) and identified two positive specimens which were negative by the reference method but positive by the blocking assay. Among the polyclonal EIAs, all had sensitivities of greater than 90%, but specificities were variable; Rotazyme II, with a specificity of 50%, showed considerable discrepancy from other polyclonal EIAs. The latex tests had sensitivities ranging from 70 to 90% and specificities of 80 to 100%. Latex agglutination tests were more rapid than EIAs and did not require expensive equipment. The final choice of assay system will depend on the cost, speed, and accuracy requirements of the clinical laboratory.


J Clin Microbiol. 1988 September; 26(9): 1630-1634