JCM Figure table search 04
Home Help [Feedback] [For Subscribers] [Archive] [Search] [Contents]
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Tarrand, J J
Right arrow Articles by Rolston, K V
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Tarrand, J J
Right arrow Articles by Rolston, K V

 Previous Article  |  Next Article 

J Clin Microbiol. 1991 October; 29(10): 2245-2249

Clinical comparison of the resin-containing BACTEC 26 Plus and the Isolator 10 blood culturing systems.

J J Tarrand, C Guillot, M Wenglar, J Jackson, J D Lajeunesse and K V Rolston

Laboratory Medicine, University of Texas, M. D. Anderson Cancer Center, Houston 77030.

ABSTRACT

The new resin-containing BACTEC 26 Plus blood culturing system (Becton Dickinson Diagnostic Instrument Systems, Towson, Md.) was compared with the Isolator 10 system (Wampole Laboratories, Cranbury, N.J.). Blood samples were drawn by syringe, and equal 10-ml volumes were evaluated in each blood culture system by the recommended methods. Both systems were incubated aerobically with 5% CO2. Of 11,506 acceptable study specimens, 1,788 aerobic isolates were recovered. Overall, recoveries was similar for the two systems, with 626 bacteria or fungi recovered in the BACTEC 26 Plus system only, 499 recovered in the Isolator system only, and 663 recovered in both systems. Of 345 gram-negative rods, 62 grew in the BACTEC system only and 109 grew in the Isolator system only (P less than 0.001). Thirty-three of these Isolator-only gram-negative organisms were Acinetobacter spp. Of 209 yeasts, 38 grew in BACTEC only and 81 grew in Isolator only (P less than 0.001). Of 200 streptococci and enterococci, 98 were recovered in BACTEC only and 26 grew in Isolator only (P less than 0.001). Two hundred twenty-eight independent episodes of gram-negative rod bacteremia occurred. Isolator was the first system positive in 59 of 197 episodes, compared with 45 of 197 for BACTEC when Acinetobacter episodes were excluded. Times to detection were similar for the two systems. High colony counts correlated with repeat positive blood cultures. Isolator and BACTEC had similar overall recoveries, with individual merits and deficiencies for both systems. The additional quantitative information derived from Isolator had utility in our institution.


J Clin Microbiol. 1991 October; 29(10): 2245-2249