JCM Figure table search 04
Home Help [Feedback] [For Subscribers] [Archive] [Search] [Contents]
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Pfaller, M A
Right arrow Articles by Rhoden, D
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Pfaller, M A
Right arrow Articles by Rhoden, D
J Clin Microbiol. 1991 July; 29(7): 1422-1428

Comparison of the autoSCAN-W/A rapid bacterial identification system and the Vitek AutoMicrobic system for identification of gram-negative bacilli.

M A Pfaller, D Sahm, C O'Hara, C Ciaglia, M Yu, N Yamane, G Scharnweber and D Rhoden

Iowa City Veteran's Affairs Medical Center, Iowa.

ABSTRACT

The autoSCAN-W/A (W/A; Baxter MicroScan, West Sacramento, Calif.) with the new fluorometric Rapid Neg Combo 1 (RNC) panel is a fully automated fluorometric system for identification of both enteric and nonenteric gram-negative bacilli within 2 h. We compared the W/A with the Vitek AutoMicrobic System (Vitek AMS; Vitek Systems, Inc., Hazelwood, Mo.) for identification of 383 clinical isolates of gram-negative bacilli. The API 20E (Analytab Products, Plainview, N.Y.) and conventional biochemical testing were used as the reference systems. The W/A correctly identified 336 isolates (87.7%) to the species level and classified an additional 29 isolates (7.6%) as correct with low probability (overall identification = 95.3%); the Vitek AMS correctly identified 355 isolates (92.7%) to the species level and classified an additional 8 isolates (2.1%) as correct with low probability (overall identification = 94.8%). A common set of 134 isolates of gram-negative bacilli was tested in both participating laboratories as a means of assessing interlaboratory agreement with both the W/A and the Vitek AMS. The overall agreements between the two laboratories were 86% with the W/A and 92% with the Vitek AMS. The W/A performed comparably to the Vitek AMS for identification of most gram-negative bacilli, actually exceeding the Vitek AMS for identification of nonenteric bacilli. Rapid time to identification and a high level of automation make the W/A an attractive system for clinical microbiology laboratories.


J Clin Microbiol. 1991 July; 29(7): 1422-1428