JCM Figure table search 04
Home Help [Feedback] [For Subscribers] [Archive] [Search] [Contents]
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Krisher, K K
Right arrow Articles by Koepnick, F E
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Krisher, K K
Right arrow Articles by Koepnick, F E
J Clin Microbiol. 1993 April; 31(4): 793-797

Comparison of the BacT/Alert pediatric blood culture system, Pedi-BacT, with conventional culture using the 20-milliliter Becton-Dickinson supplemented peptone broth tube.

K K Krisher, D R Whyburn and F E Koepnick

Clinical Microbiology Laboratory, Children's Medical Center of Dallas, Texas 75235.

ABSTRACT

The performance of the Pedi-BacT system, the BacT/Alert (Organon Teknika Corp., Durham, N.C.) pediatric blood culture bottle, was compared with that of a conventional 20-ml supplemented peptone broth tube (Becton-Dickinson Corp., Cockeysville, Md.) (BD system) in matched aerobic cultures. The tubes of the BD system were visually examined daily for 7 days and were subcultured during the first 24 h of incubation. Pedi-BacT cultures were mechanically agitated and continuously monitored for growth by the instrument. Of the 6,628 compliant pairs, 331 (5.0%) were positive in both systems, 220 (3.3%) were positive in the Pedi-BacT system only, and 170 (2.6%) were positive in the BD system only. One (0.02%) false-negative culture and 15 (0.2%) false-positive cultures occurred with the Pedi-BacT system while 20 (0.3%) false-negative cultures and 35 (0.5%) false-positive cultures occurred with the BD system. Of 288 clinically significant organisms detected in matched pairs from which a single isolate was recovered, 176 (61%) were recovered from both systems, 83 (29%) were recovered from the Pedi-BacT system only (P < 0.0001), and 29 (10%) were recovered from the BD system only. Members of the family Enterobacteriaceae (P < 0.01), miscellaneous nonfermenters (P < 0.05), and Candida spp. (P < 0.01) were isolated more frequently in the Pedi-BacT system than in the BD system. No significant difference in recovery of other organisms was found between the systems. The average time to detection for the Pedi-BacT system ranged from 11.5 h for streptococci to 29.7 h for enterococci, while that for the BD system ranged from 20.3 h for streptococci to 66.4 h for some nonfermenters. The BacT/Alert system is a reliable, labor-saving alternative to conventional blood culture methods.


J Clin Microbiol. 1993 April; 31(4): 793-797




This article has been cited by other articles:




Home Help [Feedback] [For Subscribers] [Archive] [Search] [Contents]
Antimicrob. Agents Chemother. Clin. Microbiol. Rev.
Clin. Vaccine Immunol. ALL ASM JOURNALS

Copyright © 1993 by the American Society for Microbiology. All rights reserved.