JCM Figure table search 04
Home Help [Feedback] [For Subscribers] [Archive] [Search] [Contents]
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Weinstein, M P
Right arrow Articles by Reller, L B
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Weinstein, M P
Right arrow Articles by Reller, L B

 Previous Article  |  Next Article 

J Clin Microbiol. 1994 September; 32(9): 2103-2106

Controlled evaluation of 5 versus 10 milliliters of blood cultured in aerobic BacT/Alert blood culture bottles.

M P Weinstein, S Mirrett, M L Wilson, L G Reimer and L B Reller

Department of Medicine, University of Medicine and Dentistry of New Jersey-Robert Wood Johnson Medical School, New Brunswick 08901.

ABSTRACT

Bottles developed for use in the BacT/Alert automated blood culture system (Organon Teknika Corp., Durham, N.C.) can accept up to 10 ml of blood without falling below a 1:5 ratio of blood to broth. We compared the yield and speed of detection of microorganisms in 13,128 adequately filled, paired, aerobic bottles inoculated with 5 versus 10 ml of blood at three university hospitals. A total of 798 microorganisms causing sepsis grew in one or both bottles. The overall recovery of microorganisms from 10-ml samples exceeded that from 5-ml samples (P < 0.001); the increased yield attributed to the additional 5 ml in the samples was 7.2%. The increased yield from 10-ml inocula was most marked for Escherichia coli (P < 0.01) and other members of the family Enterobacteriaceae (P < 0.001). Ten-milliliter samples did not yield more gram-positive bacteria, nonfermentative gram-negative rods, or yeasts. When both bottles were positive, the bottles inoculated with 10 ml of blood showed growth sooner (P < 0.001). Earlier detection with 10-ml inocula was especially notable for coagulase-negative staphylococci (P < 0.001), streptococci (P < 0.001), E. coli (P < 0.025), and other members of the family Enterobacteriaceae (P < 0.025). We conclude that an increase in the volume of blood inoculated into BacT/Alert aerobic blood culture bottles from 5 to 10 ml will increase the overall yield and the speed of detection of clinically important blood pathogens.


J Clin Microbiol. 1994 September; 32(9): 2103-2106