Journal of Clinical Microbiology, 09 1997, 2331-2336, Vol 35, No. 9
G Icardi, P Bonanni, AM Raffo, A Masini, L Orione, M De Martini and P Crovari
The performance of a new automated analyzer for the processing and
interpretation of the RIBA Strip Immunoblot Assay (SIA), used in the
diagnosis of hepatitis C virus (HCV) infection, was evaluated. Laboratory
performance of the RIBA SIA was compared with that of two manually
processed supplementary anti-HCV tests (RIBA HCV 3.0 SIA and INNO-LIA HCV
Antibody III). Specificity of the automated processing of SIA was 100% for
90 selected anti-HCV-negative samples. On the other hand, 119 of 120
(99.2%) previously confirmed anti-HCV-positive samples were also positive
when assayed on the automated processor. Results for all specimens except
one (51 of 52) were concordant for manual and automated RIBA, while 15 of
68 sera tested with automated RIBA and the INNO-LIA assay showed different
patterns of reactivity. Three HCV sensitivity panels and one seroconversion
panel were also compared. The results show a high sensitivity for SIA NS3-
and NS5-encoded antigens. Moreover, data obtained for the anti-HCV
seroconversion panel and for samples with borderline or discordant anti-HCV
enzyme-linked immunosorbent assay results suggest that bands with a
relative intensity of >0.5 on the automated analyzer (theoretically
negative) should be evaluated with care. Coefficients of variability ranged
from 9 to 14.8% in an interassay reproducibility study. Overall, the
performance of the automated analysis of SIA is comparable to that of the
manual RIBA assay. The new automated processor for SIA bands proved to be
sensitive and specific. Its use makes the optical scoring of bands
unnecessary by indicating relative intensity values, which could be
particularly useful in the follow-up care of anti-HCV-positive patients
receiving antiviral therapy.
Copyright © 1997 by the American Society for Microbiology. All rights reserved.
Evaluation of performance of the RIBA processor system for automated analysis of the strip immunoblot assay for detection of antibodies to hepatitis C virus
Institute of Hygiene and Preventive Medicine, University of Genoa, Italy.
| Antimicrob. Agents Chemother. | Clin. Microbiol. Rev. |
|---|---|
| Clin. Vaccine Immunol. | ALL ASM JOURNALS |
|---|