Journal of Clinical Microbiology, February 1999, p. 436-437, Vol. 37, No. 2
Molecular Diagnostics Laboratory, Institute
of Hygiene, Karl Franzens University Graz, A-8010 Graz,
Austria,1 and
Clinical Microbiology
Laboratory, Thomas Jefferson University, Philadelphia,
Pennsylvania2
Received 17 August 1998/Returned for modification 5 October
1998/Accepted 2 November 1998
The use of AMPLILINK version 1.0 software was evaluated for the
operation and control of one COBAS AMPLICOR instrument and for two
COBAS AMPLICOR instruments run simultaneously to perform and detect
nucleic acid amplification reactions. A total of 3,384 results were
analyzed. The initial accuracy of the results was 99.91%. Three
errors of omission of transfer of data from the COBAS AMPLICOR to the
AMPLILINK system were observed. Two of these errors were from a single
specimen, where both the analyte and internal control results were not
transmitted. These errors did not interfere with the correctness of any
other data. There were no interruptions of runs, and no data were
mixed. AMPLILINK increased convenience, saved labor, and was found to
be a very useful addition for clinical laboratories performing
molecular-diagnostic procedures with the COBAS AMPLICOR system.
0095-1137/99/$04.00+0
Copyright © 1999, American Society for Microbiology. All rights reserved.
Evaluation of AMPLILINK Software for the COBAS
AMPLICOR System
*
Corresponding author. Mailing address: Institute of
Hygiene, KF-University Graz, Universitaetsplatz 4, A-8010 Graz,
Austria. Phone: 43/316/380-4363. Fax: 43/316/380-9651. E-mail:
harald.kessler{at}kfunigraz.ac.at.
| Antimicrob. Agents Chemother. | Clin. Microbiol. Rev. |
|---|---|
| Clin. Vaccine Immunol. | ALL ASM JOURNALS |
|---|