We have comparatively evaluated Quantiplex version 3.0 and version
2.0 on 133 plasma samples and a repetitive dilution series. Version 3.0 yielded higher human immunodeficiency virus RNA values, and the ratio
of version 3.0 results to version 2.0 results decreased from 3.47 below
1,000 copies/ml to 1.97 above 50,000 copies/ml [linear regression, log
(version 3.0) = 0.915 + 0.871 × log (version 2.0);
r2 = 0.952].
0095-1137/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
Comparative Evaluation of Two Branched-DNA Human Immunodeficiency
Virus Type 1 RNA Quantification Assays with Lower Detection Limits of
50 and 500 Copies per Milliliter