Journal of Clinical Microbiology, September 2000, p. 3445-3447, Vol. 38, No. 9
0095-1137/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
Department of Clinical Virology, Christian Medical College and Hospital, Vellore 632 004, India
Received 24 January 2000/Returned for modification 10 May 2000/Accepted 12 June 2000
Two rapid human immunodeficiency virus (HIV) screening assays, HIV TRI-DOT and HIV-SPOT were compared with standard enzyme-linked immunosorbent assays according to a testing algorithm. Sensitivities and specificities in the real-time evaluation were 99.5 and 99.9% for TRI-DOT and 98.2 and 99.7% for HIV-SPOT, respectively. These two tests are suitable for use where facilities and laboratory expertise are limited.