Comparative Evaluation of Semiautomated COBAS AMPLICOR Hepatitis
B Virus (HBV) MONITOR Test and Manual Microwell Plate-Based AMPLICOR
HBV MONITOR Test
Irena J.
Marin,1
Mario
Poljak,1,*
Katja
Seme,1
Jelka
Megli
-Volkar,2
Mojca
Mati
i
,2
Gorazd
Le
ni
ar,3 and
Vladimir
Brinovec2
Institute of Microbiology and Immunology,
Medical Faculty,1 and Department of
Infection Diseases, Clinical Center,2 Ljubljana,
and Department of Infection Diseases, General Hospital,
Celje,3 Slovenia
Received 8 August 2000/Accepted 4 December 2000
Comparative evaluation of the semiautomated COBAS AMPLICOR
hepatitis B virus (HBV) MONITOR Test (COBAS-HBV) and manual AMPLICOR HBV MONITOR Test (AMPLICOR-HBV) on 208 serum samples revealed no
significant difference in the sensitivities of the two assays. Twenty
samples tested HBV DNA negative and 183 samples tested HBV DNA positive
by both assays. Three samples tested positive by COBAS-HBV only and two
samples tested positive by AMPLICOR-HBV only. HBV DNA concentrations
determined by the two assays were significantly related
(n = 183, r = 0.97, P < 0.0001), which indicates that COBAS-HBV could
replace AMPLICOR-HBV. The major inconvenience of COBAS-HBV is the
required performance of appropriate predilutions of high-titer samples
in order to extend the narrow dynamic range of the assay.
*
Corresponding author. Mailing address: Institute of
Microbiology and Immunology, Medical Faculty, Zalo
ka 4, 1105 Ljubljana, Slovenia. Phone: 386 1 543 7453. Fax: 386 1 543 7401. E-mail: mario.poljak{at}mf.uni-lj.si.
Journal of Clinical Microbiology, February 2001, p. 758-761, Vol. 39, No. 2
0095-1137/01/$04.00+0 DOI: 10.1128/JCM.39.2.758-761.2001
Copyright © 2001, American Society for Microbiology. All rights reserved.
This article has been cited by other articles:
-
Laperche, S., Thibault, V., Bouchardeau, F., Alain, S., Castelain, S., Gassin, M., Gueudin, M., Halfon, P., Larrat, S., Lunel, F., Martinot-Peignoux, M., Mercier, B., Pawlotsky, J.-M., Pozzetto, B., Roque-Afonso, A.-M., Roudot-Thoraval, F., Saune, K., Lefrere, J.-J.
(2006). Expertise of laboratories in viral load quantification, genotyping, and precore mutant determination for hepatitis B virus in a multicenter study.. J. Clin. Microbiol.
44: 3600-3607
[Abstract]
[Full Text]
-
Welzel, T. M., Miley, W. J., Parks, T. L., Goedert, J. J., Whitby, D., Ortiz-Conde, B. A.
(2006). Real-time PCR assay for detection and quantification of hepatitis B virus genotypes a to g.. J. Clin. Microbiol.
44: 3325-3333
[Abstract]
[Full Text]
-
Hui, C.-K., Bowden, S., Zhang, H.-Y., Wong, A., Lewin, S., Rousseau, F., Mommeja-Marin, H., Lee, N. P., Luk, J. M., Locarnini, S., Leung, N., Naoumov, N. V., Lau, G. K. K.
(2006). Comparison of Real-Time PCR Assays for Monitoring Serum Hepatitis B Virus DNA Levels during Antiviral Therapy.. J. Clin. Microbiol.
44: 2983-2987
[Abstract]
[Full Text]
-
Gordillo, R. M., Gutierrez, J., Casal, M.
(2005). Evaluation of the COBAS TaqMan 48 Real-Time PCR System for Quantitation of Hepatitis B Virus DNA. J. Clin. Microbiol.
43: 3504-3507
[Abstract]
[Full Text]
-
Konnick, E. Q., Erali, M., Ashwood, E. R., Hillyard, D. R.
(2005). Evaluation of the COBAS Amplicor HBV Monitor Assay and Comparison with the Ultrasensitive HBV Hybrid Capture 2 Assay for Quantification of Hepatitis B Virus DNA. J. Clin. Microbiol.
43: 596-603
[Abstract]
[Full Text]
-
Stelzl, E., Muller, Z., Marth, E., Kessler, H. H.
(2004). Rapid Quantification of Hepatitis B Virus DNA by Automated Sample Preparation and Real-Time PCR. J. Clin. Microbiol.
42: 2445-2449
[Abstract]
[Full Text]
-
Valentine-Thon, E., van Loon, A. M., Schirm, J., Reid, J., Klapper, P. E., Cleator, G. M.
(2001). European Proficiency Testing Program for Molecular Detection and Quantitation of Hepatitis B Virus DNA. J. Clin. Microbiol.
39: 4407-4412
[Abstract]
[Full Text]