Two commercially available hypersensitive assays for human
immunodeficiency virus type 1 (HIV-1) RNA quantitation, AMPLICOR HIV-1
Monitor Test 1.5 and Quantiplex HIV RNA 3.0, were compared to detect
and quantify HIV-1 RNA in the cell-free fraction of cervicovaginal
secretions collected by vaginal washing. Three panel specimens were
used: pooled cervicovaginal secretions spiked with HIV-1 subtype A or
HIV-1 subtype B and cervicovaginal lavages from HIV-positive and
HIV-negative women. Compared to the AMPLICOR HIV-1 Monitor Test 1.5 assay, the Quantiplex HIV-1 3.0 assay yielded higher estimates of HIV-1
RNA concentrations in several tested samples spiked with HIV-1 RNA
subtype A, as well as subtype B, particularly samples containing low
amounts of HIV-1 RNA. The sensitivity and specificity of the AMPLICOR
HIV-1 Monitor Test 1.5 assay were 93 and 100%, respectively; the
sensitivity and specificity of the Quantiplex HIV RNA 3.0 assay were 97 and 50%, respectively. In conclusion, in quantifying HIV-1 RNA in
cervicovaginal secretions, the Quantiplex HIV RNA 3.0 may lack
specificity, and the AMPLICOR HIV-1 Monitor Test 1.5 assay, although
highly specific, may lack sensitivity.
0095-1137/01/$04.00+0 DOI: 10.1128/JCM.39.6.2055-2059.2001
Copyright © 2001, American Society for Microbiology. All rights reserved.
Quantitation of Human Immunodeficiency Virus Type 1 (HIV-1)
RNA in Cell-Free Cervicovaginal Secretions: Comparison of Reverse
Transcription-PCR Amplification (AMPLICOR HIV-1 MONITOR 1.5) with
Enhanced-Sensitivity Branched-DNA Assay (Quantiplex 3.0)
*
Corresponding author. Mailing address: Laboratoire de
Virologie, Service de Microbiologie, Hôpital Européen
Georges Pompidou, 20, rue Leblanc, 75 908 Paris Cedex 15, France.
Phone: (331) 56-09-39-58. Fax: (331) 56-09-24-47. E-mail:
ali.simohamed{at}egp.ap-hop-paris.fr.
This article has been cited by other articles:
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»