JCM Figure table search 04
Home Help [Feedback] [For Subscribers] [Archive] [Search] [Contents]
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Cox, M E
Right arrow Articles by Mangels, J I
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Cox, M E
Right arrow Articles by Mangels, J I

 Previous Article  |  Next Article 

J Clin Microbiol. 1976 July; 4(1): 40-45

Improved chamber for the isolation of anaerobic microorganisms.

M E Cox and J I Mangels

ABSTRACT

A small portable chamber for the recovery of anaerobic bacteria is described. This rigid chamber is constructed of clear acrylic with dimensions of 30 inches (ca. 76.2 cm) wide, 18 inches (ca. 44.7 cm) deep, and 18 inches (ca. 44.7 cm) high. Conventional bacteriological techniques can be used inside the chamber to efficiently isolate strict anaerobic organisms. An adapter allows the attachment of a standard anaerobic jar to the outside of the chamber. The jar can be used to store reduced media. Once the jar is attached to the chamber and the media is removed to the interior of the chamber, the jar is available to receive inoculated media. The anaerobic jar can then be removed from the chamber, without contaminating the jar or chamber with oxygen, and be placed in a conventional 37degreesC incubator. This chamber also allows the microbiologist to process cultures without wearing gloves as was necessary with previous anaerobic chambers. Air-tight latex rubber sleeves seal around the microbiologists arms and to the armport flange of the chamber to prevent the introduction of oxygen into the chamber. Anaerobic conditions are maintained by circulating a 80% N2, 10% H2, 10% CO2 gas mixture through alumina pellets coated with palladium. This study indicates that anaerobic conditions obtained in this chamber are sufficient for recovery of obligate anaerobes.


J Clin Microbiol. 1976 July; 4(1): 40-45