Comparative Evaluation of the BD Phoenix and VITEK 2 Automated Instruments for Identification of Isolates of the Burkholderia cepacia Complex
Sylvain Brisse,1* Stefania Stefani,2 Jan Verhoef,1 Alex Van Belkum,3 Peter Vandamme,4 and Wil Goessens3
Eijkman-Winkler Institute, Utrecht University, 3584 CX Utrecht,1
Department of Medical Microbiology and Infectious Diseases, Erasmus University Medical Center Rotterdam, 3015 GD Rotterdam, The Netherlands,3
Department of Microbiology, University of Catania, 95124 Catania, Italy,2
Laboratory for Microbiology, University of Ghent, B-9000 Ghent, Belgium4
Received 17 December 2001/
Returned for modification 18 January 2002/
Accepted 19 February 2002
We evaluated two new automated identification systems, the BD Phoenix (Becton Dickinson) and the VITEK 2 (bioMérieux), for identification of isolates of the Burkholderia cepacia complex (BCC). The test sample included 42 isolates of the highly virulent and epidemic genomovar III, 45 isolates of B. multivorans, and 47 isolates of other members of the BCC. Rates of correct identification by the BD Phoenix and VITEK 2 were similar when all BCC isolates were considered (50 and 53%, respectively) but differed markedly for genomovar III (71 and 38%; P < 0.01) and for B. multivorans (58 and 89%; P < 0.001). For the BD Phoenix as well as the VITEK 2, taking all 134 isolates of the BCC together, rates of correct identification of clinical isolates (56 and 55%, respectively; n = 85) were higher than those of environmental isolates (21 and 39%, respectively; n = 28). Clinical isolates of genomovar III (n = 27) showed correct identification rates of 81% (BD Phoenix) and 48% (VITEK 2) (P < 0.01). Rates of misidentification for BD Phoenix and VITEK 2 were 9 and 17% for genomovar III, 22 and 7% for B. multivorans, and 36 and 13% for the other BCC members (P < 0.01), respectively. More than half of the isolates misidentified by each instrument were identified as Ralstonia pickettii, Ralstonia paucula (CDC IV C-2 group), Alcaligenes faecalis, Achromobacter spp., or, for the VITEK 2, "various nonfermenters." This study reemphasizes that confirmatory identification of BCC, preferably by molecular methods, is highly recommended.
* Corresponding author. Mailing address: UMC Utrecht, Heidelberglaan 100, AZU G04.614, 3584 CX Utrecht, The Netherlands. Phone: 31 30 250-7625. Fax: 31 30 254-1770. E-mail: sbrisse{at}lab.azu.nl.
Journal of Clinical Microbiology, May 2002, p. 1743-1748, Vol. 40, No. 5
0095-1137/02/$04.00+0 DOI: 10.1128/JCM.40.5.1743-1748.2002
Copyright © 2002, American Society for Microbiology. All Rights Reserved.
This article has been cited by other articles:
-
Snyder, J. W., Munier, G. K., Johnson, C. L.
(2008). Direct Comparison of the BD Phoenix System with the MicroScan WalkAway System for Identification and Antimicrobial Susceptibility Testing of Enterobacteriaceae and Nonfermentative Gram-Negative Organisms. J. Clin. Microbiol.
46: 2327-2333
[Abstract]
[Full Text]
-
Weinberg, J. B., Alexander, B. D., Majure, J. M., Williams, L. W., Kim, J. Y., Vandamme, P., LiPuma, J. J.
(2007). Burkholderia glumae Infection in an Infant with Chronic Granulomatous Disease. J. Clin. Microbiol.
45: 662-665
[Abstract]
[Full Text]
-
Menozzi, M. G., Eigner, U., Covan, S., Rossi, S., Somenzi, P., Dettori, G., Chezzi, C., Fahr, A.-M.
(2006). Two-Center Collaborative Evaluation of Performance of the BD Phoenix Automated Microbiology System for Identification and Antimicrobial Susceptibility Testing of Gram-Negative Bacteria. J. Clin. Microbiol.
44: 4085-4094
[Abstract]
[Full Text]
-
Bosshard, P. P., Zbinden, R., Abels, S., Boddinghaus, B., Altwegg, M., Bottger, E. C.
(2006). 16S rRNA Gene Sequencing versus the API 20 NE System and the VITEK 2 ID-GNB Card for Identification of Nonfermenting Gram-Negative Bacteria in the Clinical Laboratory. J. Clin. Microbiol.
44: 1359-1366
[Abstract]
[Full Text]
-
O'Hara, C. M.
(2006). Evaluation of the Phoenix 100 ID/AST System and NID Panel for Identification of Enterobacteriaceae, Vibrionaceae, and Commonly Isolated Nonenteric Gram-Negative Bacilli.. J. Clin. Microbiol.
44: 928-933
[Abstract]
[Full Text]
-
Fontana, C., Favaro, M., Pelliccioni, M., Pistoia, E. S., Favalli, C.
(2005). Use of the MicroSeq 500 16S rRNA Gene-Based Sequencing for Identification of Bacterial Isolates That Commercial Automated Systems Failed To Identify Correctly. J. Clin. Microbiol.
43: 615-619
[Abstract]
[Full Text]
-
O'Hara, C. M.
(2005). Manual and Automated Instrumentation for Identification of Enterobacteriaceae and Other Aerobic Gram-Negative Bacilli. Clin. Microbiol. Rev.
18: 147-162
[Abstract]
[Full Text]
-
Horstkotte, M. A., Knobloch, J. K.-M., Rohde, H., Dobinsky, S., Mack, D.
(2004). Evaluation of the BD PHOENIX Automated Microbiology System for Detection of Methicillin Resistance in Coagulase-Negative Staphylococci. J. Clin. Microbiol.
42: 5041-5046
[Abstract]
[Full Text]
-
Brisse, S., Cordevant, C., Vandamme, P., Bidet, P., Loukil, C., Chabanon, G., Lange, M., Bingen, E.
(2004). Species Distribution and Ribotype Diversity of Burkholderia cepacia Complex Isolates from French Patients with Cystic Fibrosis. J. Clin. Microbiol.
42: 4824-4827
[Abstract]
[Full Text]
-
Donay, J.-L., Mathieu, D., Fernandes, P., Pregermain, C., Bruel, P., Wargnier, A., Casin, I., Weill, F. X., Lagrange, P. H., Herrmann, J. L.
(2004). Evaluation of the Automated Phoenix System for Potential Routine Use in the Clinical Microbiology Laboratory. J. Clin. Microbiol.
42: 1542-1546
[Abstract]
[Full Text]
-
Segonds, C., Paute, S., Chabanon, G.
(2003). Use of Amplified Ribosomal DNA Restriction Analysis for Identification of Ralstonia and Pandoraea Species: Interest in Determination of the Respiratory Bacterial Flora in Patients with Cystic Fibrosis. J. Clin. Microbiol.
41: 3415-3418
[Abstract]
[Full Text]
-
Fahr, A.-M., Eigner, U., Armbrust, M., Caganic, A., Dettori, G., Chezzi, C., Bertoncini, L., Benecchi, M., Menozzi, M. G.
(2003). Two-Center Collaborative Evaluation of the Performance of the BD Phoenix Automated Microbiology System for Identification and Antimicrobial Susceptibility Testing of Enterococcus spp. and Staphylococcus spp.. J. Clin. Microbiol.
41: 1135-1142
[Abstract]
[Full Text]
-
Leverstein-van Hall, M. A., Fluit, A. C., Paauw, A., Box, A. T. A., Brisse, S., Verhoef, J.
(2002). Evaluation of the Etest ESBL and the BD Phoenix, VITEK 1, and VITEK 2 Automated Instruments for Detection of Extended-Spectrum Beta-Lactamases in Multiresistant Escherichia coli and Klebsiella spp.. J. Clin. Microbiol.
40: 3703-3711
[Abstract]
[Full Text]
Copyright © 2002 by the American Society for Microbiology. All rights reserved.