Entirely Automated Quantification of Human Immunodeficiency Virus Type 1 (HIV-1) RNA in Plasma by Using the Ultrasensitive COBAS AMPLICOR HIV-1 Monitor Test and RNA Purification on the MagNA Pure LC Instrument
Gabriele Hölzl, Markus Stöcher, Victoria Leb, Herbert Stekel, and Jörg Berg*
Institute of Laboratory Medicine, General Hospital Linz, A-4020 Linz, Austria
Received 15 July 2002/
Returned for modification 7 October 2002/
Accepted 20 December 2002
The ultrasensitive COBAS AMPLICOR HIV-1 Monitor test was complemented with automated RNA purification on the MagNA Pure LC instrument. This enabled entirely automated ultrasensitive assessment of viral loads in human immunodeficiency virus type 1 (HIV-1)-infected individuals. The detection limit of the fully automated assay and the viral load measurements in 80 clinical samples were found to be in good agreement with those of the conventional ultrasensitive COBAS AMPLICOR HIV-1 Monitor test. The fully automated assay showed markedly reduced hands-on time and was found to be suitable for the routine assessment of HIV-1 viral loads in a clinical diagnostic laboratory.
* Corresponding author. Mailing address: Institute of Laboratory Medicine, General Hospital Linz, Krankenhausstr. 9, A-4020 Linz, Austria. Phone: 43-70-7806-1174. Fax: 43-70-7806-1815. E-mail:
joerg.berg{at}akh.linz.at.
Journal of Clinical Microbiology, March 2003, p. 1248-1251, Vol. 41, No. 3
0095-1137/03/$08.00+0 DOI: 10.1128/JCM.41.3.1248-1251.2003
Copyright © 2003, American Society for Microbiology. All Rights Reserved.
This article has been cited by other articles:
-
Paar, C., Palmetshofer, C., Flieger, K., Geit, M., Kaiser, R., Stekel, H., Berg, J.
(2008). Genotypic Antiretroviral Resistance Testing for Human Immunodeficiency Virus Type 1 Integrase Inhibitors by Use of the TruGene Sequencing System. J. Clin. Microbiol.
46: 4087-4090
[Abstract]
[Full Text]
-
Yao, J. D. C., Germer, J. J., Damond, F., Roquebert, B., Descamps, D.
(2008). Plasma Load Discrepancies between the Roche Cobas Amplicor Human Immunodeficiency Virus Type 1 (HIV-1) Monitor Version 1.5 and Roche Cobas AmpliPrep/Cobas TaqMan HIV-1 Assays. J. Clin. Microbiol.
46: 834-834
[Full Text]
-
Espy, M. J., Uhl, J. R., Sloan, L. M., Buckwalter, S. P., Jones, M. F., Vetter, E. A., Yao, J. D. C., Wengenack, N. L., Rosenblatt, J. E., Cockerill, F. R. III, Smith, T. F.
(2006). Real-Time PCR in Clinical Microbiology: Applications for Routine Laboratory Testing. Clin. Microbiol. Rev.
19: 165-256
[Abstract]
[Full Text]
-
Hong, K. M., Najjar, H., Hawley, M., Press, R. D.
(2004). Quantitative Real-Time PCR with Automated Sample Preparation for Diagnosis and Monitoring of Cytomegalovirus Infection in Bone Marrow Transplant Patients. Clin. Chem.
50: 846-856
[Abstract]
[Full Text]
-
Muller, Z., Stelzl, E., Bozic, M., Haas, J., Marth, E., Kessler, H. H.
(2004). Evaluation of Automated Sample Preparation and Quantitative PCR LCx Assay for Determination of Human Immunodeficiency Virus Type 1 RNA. J. Clin. Microbiol.
42: 1439-1443
[Abstract]
[Full Text]
-
Yao, J. D. C., Germer, J., Patel, R., Berg, J., Holzl, G., Stocher, M., Leb, V., Stekel, H.
(2003). Need for Procedural Details in the Protocol for Specimen Processing by the MagNA Pure LC Instrument. J. Clin. Microbiol.
41: 4911-4912
[Full Text]