JCM Figure table search 04
Home Help [Feedback] [For Subscribers] [Archive] [Search] [Contents]
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Nijs, A.
Right arrow Articles by Magerman, K.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Nijs, A.
Right arrow Articles by Magerman, K.

 Previous Article  |  Next Article 

Journal of Clinical Microbiology, August 2003, p. 3627-3630, Vol. 41, No. 8
0095-1137/03/$08.00+0     DOI: 10.1128/JCM.41.8.3627-3630.2003
Copyright © 2003, American Society for Microbiology. All Rights Reserved.

Comparison and Evaluation of Osiris and Sirscan 2000 Antimicrobial Susceptibility Systems in the Clinical Microbiology Laboratory

A. Nijs, R. Cartuyvels, A. Mewis, V. Peeters, J. L. Rummens, and K. Magerman*

Department of Clinical Laboratory, Virga Jesseziekenhuis, Stadsomvaart 11, B-3500 Hasselt, Belgium

Received 1 November 2002/ Returned for modification 30 December 2002/ Accepted 2 June 2002

The Osiris and Sirscan 2000 systems are two semiautomated systems that can be used to read and interpret the results on disk diffusion agar plates. They are both used for determination of susceptibility to antimicrobial agents. The present study compared both systems versus the NCCLS standard method of visual reading with a ruler. Both inpatient and outpatient samples with a total of 315 nonfastidious gram-negative strains were obtained. In total, 3,724 organism-antimicrobial agent combinations that fulfilled the NCCLS guidelines for disk diffusion susceptibility testing were evaluated prospectively. The results obtained with both systems in comparison with those obtained by the classical nonautomated means of interpretation were excellent, with correlation coefficients of 0.96 for both systems. The overall agreements for susceptibility interpretation were 96.56 and 96.24% with the Osiris and Sirscan systems, respectively. Very major errors were obtained for 8 (1.07%) and 10 (1.34%) organism-antimicrobial agent combinations with the Osiris and Sirscan systems, respectively. In addition, major errors were obtained for 2 (0.07%) and 6 (0.21%) combinations with the Osiris and Sirscan systems, respectively. Minor errors were obtained for 118 and 124 organism-antimicrobial agent combinations with the Osiris and Sirscan systems, respectively. Overall, both the Osiris system and the Sirscan system are comparable and reliable systems for determination of interpretative categories from the zone diameters of standard disk diffusion test plates.


* Corresponding author. Mailing address: Klinisch Laboratorium, Virga Jesseziekenhuis, Stadsomvaart 11, B-3500 Hasselt, Belgium. Phone: 32 11 30 97 40. Fax: 32 11 30 97 50. E-mail: koen.magerman{at}virgajesse.be.


Journal of Clinical Microbiology, August 2003, p. 3627-3630, Vol. 41, No. 8
0095-1137/03/$08.00+0     DOI: 10.1128/JCM.41.8.3627-3630.2003