JCM Figure table search 04
Home Help [Feedback] [For Subscribers] [Archive] [Search] [Contents]
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Spanu, T.
Right arrow Articles by Fadda, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Spanu, T.
Right arrow Articles by Fadda, G.
Journal of Clinical Microbiology, September 2003, p. 4259-4263, Vol. 41, No. 9
0095-1137/03/$08.00+0     DOI: 10.1128/JCM.41.9.4259-4263.2003
Copyright © 2003, American Society for Microbiology. All Rights Reserved.

Use of the VITEK 2 System for Rapid Identification of Clinical Isolates of Staphylococci from Bloodstream Infections

Teresa Spanu,* Maurizio Sanguinetti, Daniela Ciccaglione, Tiziana D'Inzeo, Lucio Romano, Fiammetta Leone, and Giovanni Fadda

Institute of Microbiology, Catholic University of the Sacred Heart, Rome, Italy

Received 10 February 2003/ Returned for modification 14 March 2003/ Accepted 24 June 2003

Staphylococci are an increasing cause of bloodstream infections. Rapid reliable identification of these organisms is essential for accurate diagnosis and prompt effective treatment. We evaluated the ability of the VITEK 2 system (bioMérieux, Inc, Hazelwood, Mo.) to identify these organisms rapidly and accurately. A total of 405 clinically relevant nonduplicate staphylococcal isolates (Staphylococcus aureus, n = 130; coagulase-negative staphylococci, n = 275) collected from blood cultures were tested. VITEK 2 results were considered correct when they were identical to those furnished by the comparison method based on the ID 32 STAPH system (bioMérieux, Marcy l'Etoile, France) plus supplementary manual testing. When discrepancies occurred, isolate identity was verified by molecular typing. The VITEK 2 correctly identified 387 (95.6%) isolates at the species level: 379 (including all but one [99.2%] of 130 S. aureus isolates and 249 of 275 [90.5%] coagulase-negative isolates) were identified by the automated reading; for the other eight, supplemental tests suggested by the manufacturer had to be used. Only one strain (0.2%) was misidentified (Staphylococcus hominis as Staphylococcus epidermidis), and four (1%), all S. epidermidis, were not identified. For the remaining 13 strains (including 10 S. hominis), the VITEK 2 system was unable to discriminate among two species, and no supplemental tests were suggested for conclusive identification. Over 90% of results were obtained within 4 h. These results suggest that the VITEK 2 system can provide rapid, accurate, and reliable species-level identification of staphylococci responsible for bloodstream infections, although there is room for improvement in the identification of certain coagulase-negative species, especially S. hominis.


* Corresponding author. Mailing address: Istituto di Microbiologia, Università Cattolica del Sacro Cuore, Largo F. Vito, 1, 00168 Rome, Italy. Phone: 39 06 3015 4218. Fax: 39 06 305 1152. E-mail: tspanu{at}rm.unicatt.it.


Journal of Clinical Microbiology, September 2003, p. 4259-4263, Vol. 41, No. 9
0095-1137/03/$08.00+0     DOI: 10.1128/JCM.41.9.4259-4263.2003
Copyright © 2003, American Society for Microbiology. All Rights Reserved.