Comparison of LCx with Other Current Viral Load Assays for Detecting and Quantifying Human Immunodeficiency Virus Type 1 RNA in Patients Infected with the Circulating Recombinant Form A/G (CRF02)
Alessandra Amendola, Licia Bordi, Claudio Angeletti, Enrico Girardi, Giuseppe Ippolito, and Maria R. Capobianchi*
National Institute for Infectious Diseases "Lazzaro Spallanzani," Rome, Italy
Received 11 June 2003/
Returned for modification 25 September 2003/
Accepted 11 November 2003
LCx was compared to other assays in measuring human immunodeficiency virus type 1 (HIV-1) CRF02 viremia. LCx showed significant but low correlation with the other methods. Values of <2.60 log10 cp/ml were observed in 29.6% of specimens with LCx and in only 14.8% with bDNA and PCR, suggesting suboptimal performance of LCx with CRF02.
* Corresponding author. Mailing address: Laboratorio di Virologia, Istituto Nazionale per le Malattie Infettive I.R.C.C.S. "Lazzaro Spallanzani," Via Portuense 292, 00149 Rome, Italy. Phone: 39-655170434. Fax: 39-65582346. E-mail:
capobianchi{at}inmi.it.
Journal of Clinical Microbiology, February 2004, p. 811-815, Vol. 42, No. 2
0095-1137/04/$08.00+0 DOI: 10.1128/JCM.42.2.811-815.2004
Copyright © 2004, American Society for Microbiology. All Rights Reserved.
This article has been cited by other articles:
-
Pyne, M. T., Konnick, E. Q., Phansalkar, A., Hillyard, D. R.
(2009). Evaluation of the Abbott Investigational Use Only RealTime HIV-1 Assay and Comparison to the Roche Amplicor HIV-1 Monitor Test, Version 1.5. J. Mol. Diagn.
11: 347-354
[Abstract]
[Full Text]
-
Foulongne, V., Montes, B., Didelot-Rousseau, M.-N., Segondy, M.
(2006). Comparison of the LCx Human Immunodeficiency Virus (HIV) RNA Quantitative, RealTime HIV, and COBAS AmpliPrep-COBAS TaqMan Assays for Quantitation of HIV Type 1 RNA in Plasma.. J. Clin. Microbiol.
44: 2963-2966
[Abstract]
[Full Text]
-
Garcia-Diaz, A., Clewley, G. S., Booth, C. L., Labett, W., McAllister, N., Geretti, A. M.
(2006). Comparative Evaluation of the Performance of the Abbott Real-Time Human Immunodeficiency Virus Type 1 (HIV-1) Assay for Measurement of HIV-1 Plasma Viral Load following Automated Specimen Preparation.. J. Clin. Microbiol.
44: 1788-1791
[Abstract]
[Full Text]
-
Swanson, P., de Mendoza, C., Joshi, Y., Golden, A., Hodinka, R. L., Soriano, V., Devare, S. G., Hackett, J. Jr.
(2005). Impact of Human Immunodeficiency Virus Type 1 (HIV-1) Genetic Diversity on Performance of Four Commercial Viral Load Assays: LCx HIV RNA Quantitative, AMPLICOR HIV-1 MONITOR v1.5, VERSANT HIV-1 RNA 3.0, and NucliSens HIV-1 QT. J. Clin. Microbiol.
43: 3860-3868
[Abstract]
[Full Text]