Journal of Clinical Microbiology, November 2005, p. 5541-5546, Vol. 43, No. 11
0095-1137/05/$08.00+0 doi:10.1128/JCM.43.11.5541-5546.2005
Copyright © 2005, American Society for Microbiology. All Rights Reserved.
Automated Extraction of Viral-Pathogen RNA and DNA for High-Throughput Quantitative Real-Time PCR
Kurt Beuselinck,*
Marc van Ranst, and
J. van Eldere
Laboratory of Molecular Diagnostics, University Hospital Gasthuisberg, Katholieke Universiteit Leuven, Leuven, Belgium
Received 28 April 2005/
Returned for modification 5 July 2005/
Accepted 7 September 2005
The performance of the m1000 system (Abbott Laboratories, Illinois) as a front-end extraction system for high-throughput "in-house" quantitative real-time PCR assays was analyzed and compared to that of manual extraction of plasma and serum samples (hepatitis C virus [HCV] and hepatitis B virus [HBV]) and EDTA-blood samples (cytomegalovirus [CMV] and Epstein-Barr virus [EBV]). Linearity of extraction was tested on dilution series of HCV and HBV reference materials. The correlation coefficient for standard curves based on repeated extraction runs was 0.97 ± 0.06 for HCV and 0.97 ± 0.03 for HBV, indicating a linear extraction from 100 to 1.0 x 105 HCV IU/ml and from 100 to 1.0 x 106 HBV IU/ml. Intra- and interrun variability was below 0.23 log10 IU/ml for 2.98 to 5.28 log10 HCV IU/ml and 2.70 to 5.20 log10 HBV IU/ml. Correlation between automated and manual extraction was very good. For HCV, the correlation coefficient was 0.91 and the mean difference in viral load was 0.13 log10 HCV IU/ml. For HBV, the correlation coefficient was 0.98 and the mean difference in viral load 0.61 log10 HBV IU/ml. For CMV and EBV, the correlation coefficient was 0.98 and the mean difference in viral load 0.33 log10 copies/ml. Accuracy was confirmed with a reference panel (QCMD, Glasgow, Scotland) for all four assays. No cross-contamination was observed when extracting strongly positive polyomavirus samples (8.10 log10 copies/ml) interspersed with polyomavirus-negative samples. Automated extraction via the m1000 system offers a high reliability of extraction and resulted in a strong reduction of the required extraction hands-on time for high-throughput PCR compared to manual extraction protocols.
* Corresponding author. Mailing address: Laboratory of Molecular Diagnostics, University Hospital Gasthuisberg, Herestraat 49, BE-3000 Leuven, Belgium. Phone: 32-16-347950. Fax: 32-16-347949. E-mail: kurt.beuselinck{at}uz.kuleuven.ac.be.
Journal of Clinical Microbiology, November 2005, p. 5541-5546, Vol. 43, No. 11
0095-1137/05/$08.00+0 doi:10.1128/JCM.43.11.5541-5546.2005
Copyright © 2005, American Society for Microbiology. All Rights Reserved.
This article has been cited by other articles:
-
Dundas, N., Leos, N. K., Mitui, M., Revell, P., Rogers, B. B.
(2008). Comparison of Automated Nucleic Acid Extraction Methods with Manual Extraction. J. Mol. Diagn.
10: 311-316
[Abstract]
[Full Text]
-
Mapes, S., Leutenegger, C. M., Pusterla, N.
(2008). Nucleic acid extraction methods for detection of EHV-1 from blood and nasopharyngeal secretions. Vet Rec.
162: 857-859
[Full Text]
-
Forman, M. S., Valsamakis, A.
(2008). Performance Characteristics of a Quantitative Hepatitis C Virus RNA Assay Using COBAS AmpliPrep Total Nucleic Acid Isolation and COBAS TaqMan Hepatitis C Virus Analyte-Specific Reagent. J. Mol. Diagn.
10: 147-153
[Abstract]
[Full Text]
-
Fafi-Kremer, S., Morand, P., Barranger, C., Bargues, G., Magro, S., Bes, J., Bourgeois, P., Joannes, M., Seigneurin, J.-M.
(2008). Evaluation of the Epstein-Barr Virus R-Gene Quantification Kit in Whole Blood with Different Extraction Methods and PCR Platforms. J. Mol. Diagn.
10: 78-84
[Abstract]
[Full Text]
-
Caliendo, A. M., Ingersoll, J., Fox-Canale, A. M., Pargman, S., Bythwood, T., Hayden, M. K., Bremer, J. W., Lurain, N. S.
(2007). Evaluation of Real-Time PCR Laboratory-Developed Tests Using Analyte-Specific Reagents for Cytomegalovirus Quantification. J. Clin. Microbiol.
45: 1723-1727
[Abstract]
[Full Text]
-
Teughels, W., Sliepen, I., Quirynen, M., Haake, S. K., Van Eldere, J., Fives-Taylor, P., Van Ranst, M.
(2007). Human Cytomegalovirus Enhances A. actinomycetemcomitans Adherence to Cells. J. Dent. Res.
86: 175-180
[Abstract]
[Full Text]
-
Loens, K., Bergs, K., Ursi, D., Goossens, H., Ieven, M.
(2007). Evaluation of NucliSens easyMAG for Automated Nucleic Acid Extraction from Various Clinical Specimens. J. Clin. Microbiol.
45: 421-425
[Abstract]
[Full Text]
-
Stormer, M., Kleesiek, K., Dreier, J.
(2007). High-Volume Extraction of Nucleic Acids by Magnetic Bead Technology for Ultrasensitive Detection of Bacteria in Blood Components. Clin. Chem.
53: 104-110
[Abstract]
[Full Text]
-
Laperche, S., Thibault, V., Bouchardeau, F., Alain, S., Castelain, S., Gassin, M., Gueudin, M., Halfon, P., Larrat, S., Lunel, F., Martinot-Peignoux, M., Mercier, B., Pawlotsky, J.-M., Pozzetto, B., Roque-Afonso, A.-M., Roudot-Thoraval, F., Saune, K., Lefrere, J.-J.
(2006). Expertise of laboratories in viral load quantification, genotyping, and precore mutant determination for hepatitis B virus in a multicenter study.. J. Clin. Microbiol.
44: 3600-3607
[Abstract]
[Full Text]
Copyright © 2005 by the American Society for Microbiology. All rights reserved.