Journal of Clinical Microbiology, November 2005, p. 5808-5810, Vol. 43, No. 11
0095-1137/05/$08.00+0 doi:10.1128/JCM.43.11.5808-5810.2005
Copyright © 2005, American Society for Microbiology. All Rights Reserved.
Evaluation of a Fully Automated System (RAISUS) for Rapid Identification and Antimicrobial Susceptibility Testing of Staphylococci
K. Kanemitsu,*
H. Kunishima,
M. Hatta,
K. Inden,
T. Saga,
H. Ouchi,
K. Ishizawa,
H. Harigae,
H. Takemura, and
M. Kaku
Department of Infection Control and Laboratory Diagnostics, Tohoku University Graduate School of Medicine, Sendai, Miyagi 980-8574, Japan
Received 18 May 2005/
Returned for modification 10 July 2005/
Accepted 20 August 2005
RAISUS is a system for rapid bacterial identification and antimicrobial susceptibility testing. RAISUS and VITEK showed 97.8% and 75.9% agreement in identification of 45 Staphylococcus aureus strains and 58 coagulase-negative staphylococci (CoNS), respectively, and RAISUS and CLSI (formerly NCCLS) methods showed 87.2% and 87.9% agreement in the MICs for S. aureus and CoNS, respectively. RAISUS provided these data within 3.75 h, suggesting its utility for clinical bacteriological laboratories.
* Corresponding author. Mailing address: Department of Infection Control and Laboratory Diagnostics, Tohoku University Graduate School of Medicine, 1-1 Seiryo-machi, Aobaku, Sendai, Miyagi 980-8574, Japan. Phone: 81 (22) 717-7373. Fax: 81 (22) 717-7390. E-mail: kane2{at}mail.tains.tohoku.ac.jp.
Journal of Clinical Microbiology, November 2005, p. 5808-5810, Vol. 43, No. 11
0095-1137/05/$08.00+0 doi:10.1128/JCM.43.11.5808-5810.2005
Copyright © 2005, American Society for Microbiology. All Rights Reserved.
Copyright © 2005 by the American Society for Microbiology. All rights reserved.