Journal of Clinical Microbiology, March 2006, p. 928-933, Vol. 44, No. 3
0095-1137/06/$08.00+0     doi:10.1128/JCM.44.3.928-933.2006

Evaluation of the Phoenix 100 ID/AST System and NID Panel for Identification of Enterobacteriaceae, Vibrionaceae, and Commonly Isolated Nonenteric Gram-Negative Bacilli

Caroline M. O'Hara*

Division of Healthcare Quality Promotion, Centers for Disease Control and Prevention, Atlanta, Georgia 30333

Received 20 April 2005/ Returned for modification 25 May 2005/ Accepted 3 January 2006

The Phoenix 100 ID/AST system (Becton Dickinson Co., Sparks, Md.) is an automated system for the identification and antimicrobial susceptibility testing of bacterial isolates. This system with its negative identification (NID) panel was evaluated for its accuracy in the identification of 507 isolates of the family Enterobacteriaceae, 57 other nonenteric gram-negative isolates that are commonly isolated in clinical microbiology laboratories, and 138 isolates of the family Vibrionaceae. All of the isolates had been characterized by using approximately 48 conventional tube biochemicals. Of the 507 isolates of the Enterobacteriaceae, 456 (89.9%) were correctly identified to the genus and species levels. The five isolates of Proteus penneri required an off-line indole test, as suggested by the system to differentiate them from Proteus vulgaris. The identifications of 20 (3.9%) isolates were correct to the genus level but incorrect at the species level. Two (0.4%) isolates were reported as "no identification." Misidentifications to the genus and species levels occurred for 29 (5.7%) isolates of the Enterobacteriaceae. These incorrect identifications were spread over 14 different genera. The most common error was the misidentification of Salmonella species. The shortest time for a correct identification was 2 h 8 min. The longest time was 12 h 27 min, for the identification of a Serratia marcescens isolate. Of the 57 isolates of nonenteric gram-negative bacilli (Acinetobacter, Aeromonas, Burkholderia, Plesiomonas, Pseudomonas, and Stenotrophomonas spp.), 48 (84.2%) were correctly identified to the genus and species levels and 7 (12.3%) were correctly identified to the genus level but not to the species level. The average time for a correct identification was 5 h 11 min. Of the Vibrionaceae spp., 123 (89.1%) were correctly identified at the end of the initial incubation period, which averaged 4 h. Based on the findings of this study, the Phoenix 100 ID/AST system NID panel falls short of being an acceptable new method for the identification of the Enterobacteriaceae, Vibrionaceae, and gram-negative nonenteric isolates that are commonly encountered in many hospital microbiology laboratories.


* Corresponding author. Mailing address: Division of Healthcare Quality Promotion, Centers for Disease Control and Prevention, Mailstop C16, Atlanta, GA 30333. Phone: (404) 639-2316. Fax: (404) 639-3822. E-mail: cmo1{at}cdc.gov.


Journal of Clinical Microbiology, March 2006, p. 928-933, Vol. 44, No. 3
0095-1137/06/$08.00+0     doi:10.1128/JCM.44.3.928-933.2006




This article has been cited by other articles:

  • Farber, J., Moder, K.-A., Layer, F., Tammer, I., Konig, W., Konig, B. (2008). Extended-Spectrum Beta-Lactamase Detection with Different Panels for Automated Susceptibility Testing and with a Chromogenic Medium. J. Clin. Microbiol. 46: 3721-3727 [Abstract] [Full Text]  
  • Snyder, J. W., Munier, G. K., Johnson, C. L. (2008). Direct Comparison of the BD Phoenix System with the MicroScan WalkAway System for Identification and Antimicrobial Susceptibility Testing of Enterobacteriaceae and Nonfermentative Gram-Negative Organisms. J. Clin. Microbiol. 46: 2327-2333 [Abstract] [Full Text]  
  • Carroll, K. C., Glanz, B. D., Borek, A. P., Burger, C., Bhally, H. S., Henciak, S., Flayhart, D. (2006). Evaluation of the BD Phoenix Automated Microbiology System for Identification and Antimicrobial Susceptibility Testing of Enterobacteriaceae.. J. Clin. Microbiol. 44: 3506-3509 [Abstract] [Full Text]