This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Riedel, S.
Right arrow Articles by Doern, G. V.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Riedel, S.
Right arrow Articles by Doern, G. V.

 Previous Article  |  Next Article 

Journal of Clinical Microbiology, June 2006, p. 2262-2264, Vol. 44, No. 6
0095-1137/06/$08.00+0     doi:10.1128/JCM.00635-06
Copyright © 2006, American Society for Microbiology. All Rights Reserved.

Comparison of the BACTEC 9240 and BacT/Alert Blood Culture Systems for Detection of Bacterial Contamination in Platelet Concentrates

Stefan Riedel, Gregory Siwek,{dagger} Susan E. Beekmann, Sandra S. Richter, Thomas Raife, and Gary V. Doern*

Divisions of Clinical Microbiology and Transfusion Medicine, Department of Pathology, University of Iowa College of Medicine, Iowa City, Iowa 52242

Received 24 March 2006/ Returned for modification 4 April 2006/ Accepted 6 April 2006

The BACTEC 9240 blood culture system with a standard aerobic medium, Plus Aerobic/F, was compared to the BacT/Alert system with a platelet-specific medium, BPA, as a means for detecting bacterial contamination of platelet preparations. One hundred thirteen platelet units seeded with low levels of different bacteria were examined with both systems. In 93 instances, growth was detected first in the BACTEC system; in 12 cases, growth registered first in the BacT/Alert system. Among all comparisons, growth was detected, on average, 1.7 h sooner with the BACTEC system. The differences in length of time to detection were statistically significant.


* Corresponding author. Mailing address: Clinical Microbiology Laboratories, University of Iowa Hospital, C606-GH Boyd Tower, Iowa City, IA 52242. Phone: (319) 356-8616. Fax: (319) 356-4916. E-mail: gary-doern{at}uiowa.edu.

{dagger} Present address: Health Partners Specialty Center, Infectious Diseases, 410 Phalen Boulevard, St. Paul, MN 55101.


Journal of Clinical Microbiology, June 2006, p. 2262-2264, Vol. 44, No. 6
0095-1137/06/$08.00+0     doi:10.1128/JCM.00635-06
Copyright © 2006, American Society for Microbiology. All Rights Reserved.




This article has been cited by other articles:

  • Savini, V., Bianco, A., Catavitello, C., Balbinot, A., Pompilio, A., Piccolomini, R., Di Bonaventura, G., Iacone, A., D'Antonio, D. (2009). Meticillin-heteroresistant Staphylococcus pasteuri from an apheresis platelet product. J Med Microbiol 58: 1527-1528 [Full Text]  
  • Riedel, S., Junkins, A., Stamper, P. D., Cress, G., Widness, J. A., Doern, G. V. (2009). Comparison of the Bactec 9240 and BacT/Alert Blood Culture Systems for Evaluation of Placental Cord Blood for Transfusion in Neonates. J. Clin. Microbiol. 47: 1645-1649 [Abstract] [Full Text]  
  • Nanua, S., Weber, C., Isgriggs, L., Dunne, W. M. Jr. (2009). Performance Evaluation of the VersaTREK Blood Culture System for Quality Control Testing of Platelet Units. J. Clin. Microbiol. 47: 817-818 [Abstract] [Full Text]