This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Stevens, W.
Right arrow Articles by Scott, L. E.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Stevens, W.
Right arrow Articles by Scott, L. E.

 Previous Article  |  Next Article 

Journal of Clinical Microbiology, April 2007, p. 1244-1249, Vol. 45, No. 4
0095-1137/07/$08.00+0     doi:10.1128/JCM.01540-06
Copyright © 2007, American Society for Microbiology. All Rights Reserved.

Evaluation of the Performance of the Automated NucliSENS easyMAG and EasyQ Systems versus the Roche AmpliPrep-AMPLICOR Combination for High-Throughput Monitoring of Human Immunodeficiency Virus Load{triangledown}

W. Stevens, P. Horsfield, and L. E. Scott*

Department of Molecular Medicine and Hematology, University of the Witwatersrand and the National Health Laboratory Service, Johannesburg, South Africa

Received 25 July 2006/ Returned for modification 26 September 2006/ Accepted 19 January 2007

This study presents the data of an evaluation of the automated Nuclisens easyMAG and EasyQ systems versus the Roche AmpliPrep-AMPLICOR combination for testing of high-volume human immunodeficiency virus (HIV) load. This represents a follow-up of a previous study investigating the performance of the real-time Nuclisens assay using the semiautomated NucliSENS miniMAG extraction procedure. Three hundred eighteen patient samples were analyzed using both methods. The easyMAG-EasyQ HIV type 1 system has a higher sensitivity and broader dynamic range than the Cobas AmpliPrep-AMPLICOR system when the standard Roche assay is used alone, 25 to 3,000,000 IU/ml versus 400 to 750,000 HIV RNA copies/ml, respectively. There was significant correlation between the assays (0.93; P < 0.0001), with good accuracy (percent similarity mean µ = 96%), good precision (percent similarity standard deviation = 4.97%), and overall good agreement with a low percent similarity coefficient of variation of 5.17 to 6.11%. Bland-Altman analysis revealed that the AMPLICOR assay generated higher values than the EasyQ combination, with 95% of results within clinically acceptable limits. The throughput of samples was greatly improved using the easyMAG-EasyQ system, allowing 144 samples to be processed within 6 h. The potential for contamination has been dramatically reduced using the automated extraction system. Additional negative controls have been added to the kit to monitor for contamination based on the South African experience. This assay thus presents a real option for monitoring HIV load assays in high-volume testing environments.


* Corresponding author. Mailing address: Department of Molecular Medicine and Hematology, University of the Witwatersrand and the National Health Laboratory Service, Johannesburg, South Africa. Phone: 27 11 489 7567. Fax: 27 11 484 5812. E-mail: lscott{at}witshealth.co.za

{triangledown} Published ahead of print on 31 January 2007.


Journal of Clinical Microbiology, April 2007, p. 1244-1249, Vol. 45, No. 4
0095-1137/07/$08.00+0     doi:10.1128/JCM.01540-06
Copyright © 2007, American Society for Microbiology. All Rights Reserved.




This article has been cited by other articles:

  • Greengrass, V. L., Plate, M. M., Steele, P. M., Denholm, J. T., Cherry, C. L., Morris, L. M., Hearps, A., Crowe, S. M. (2009). Evaluation of the Cavidi ExaVir Load Assay (Version 3) for Plasma Human Immunodeficiency Virus Type 1 Load Monitoring. J. Clin. Microbiol. 47: 3011-3013 [Abstract] [Full Text]  
  • Scott, L. E., Noble, L. D., Moloi, J., Erasmus, L., Venter, W. D. F., Stevens, W. (2009). Evaluation of the Abbott m2000 RealTime Human Immunodeficiency Virus Type 1 (HIV-1) Assay for HIV Load Monitoring in South Africa Compared to the Roche Cobas AmpliPrep-Cobas Amplicor, Roche Cobas AmpliPrep-Cobas TaqMan HIV-1, and BioMerieux NucliSENS EasyQ HIV-1 Assays. J. Clin. Microbiol. 47: 2209-2217 [Abstract] [Full Text]  
  • Chan, K. H., Yam, W. C., Pang, C. M., Chan, K. M., Lam, S. Y., Lo, K. F., Poon, L. L. M., Peiris, J. S. M. (2008). Comparison of the NucliSens easyMAG and Qiagen BioRobot 9604 Nucleic Acid Extraction Systems for Detection of RNA and DNA Respiratory Viruses in Nasopharyngeal Aspirate Samples. J. Clin. Microbiol. 46: 2195-2199 [Abstract] [Full Text]  
  • Ginocchio, C. C., Manji, R., Lotlikar, M., Zhang, F. (2008). Clinical Evaluation of NucliSENS Magnetic Extraction and NucliSENS Analyte-Specific Reagents for Real-Time Detection of Human Metapneumovirus in Pediatric Respiratory Specimens. J. Clin. Microbiol. 46: 1274-1280 [Abstract] [Full Text]