Previous Article | Next Article ![]()
Journal of Clinical Microbiology, May 2009, p. 1559-1561, Vol. 47, No. 5
0095-1137/09/$08.00+0 doi:10.1128/JCM.02473-08
Copyright © 2009, American Society for Microbiology. All Rights Reserved.

Division of Medical Microbiology, Department of Laboratory Medicine and Pathology, University of Alberta, Edmonton,1 National Centre for Streptococcus, Provincial Laboratory for Public Health, Edmonton, Alberta, Canada2
Received 23 December 2008/ Returned for modification 27 January 2009/ Accepted 18 March 2009
The performance of the Pneumoslide test for rapid identification of Streptococcus pneumoniae was evaluated when used directly on positive blood culture specimens. The sensitivity was 75.3%, and the specificity was 98.6%. Pneumoslide test performance accuracy varied depending on the pneumococcal serotype.
Published ahead of print on 25 March 2009.