Previous Article | Next Article 
Journal of Clinical Microbiology, February 1998, p. 387-390, Vol. 36, No. 2
0095-1137/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.
Automated RIBA Hepatitis C Virus (HCV) Strip
Immunoblot Assay for Reproducible HCV Diagnosis
P.
Martin,1,*
F.
Fabrizi,1
V.
Dixit,1
S.
Quan,2
M.
Brezina,1
E.
Kaufman,2
, K.
Sra,2
R.
DiNello,2
A.
Polito,2 and
G.
Gitnick1
Department of Medicine, School of Medicine,
University of California at Los Angeles,1 Los
Angeles, California, and
Chiron Corporation,2
Emeryville, California
Received 12 May 1997/Returned for modification 11 July
1997/Accepted 31 October 1997
 |
ABSTRACT |
A comparison between the CHIRON RIBA hepatitis C virus (HCV)
processor and manual systems was performed by using 88 specimens repeatedly reactive by the second-generation HCV enzyme-linked immunosorbent assay (ELISA) (HCV 2.0 ELISA) and 111 random specimens from volunteer donors. For the second-generation RIBA HCV strip immunoblot assay (SIA) (RIBA HCV 2.0 SIA), test results correlated strongly between the manual and the automated runs (kappa value, 0.937). For the RIBA HCV 3.0 SIA, the correlation of the test results
was also high (kappa value, 0.899). Among the specimens with positive
results by RIBA HCV 2.0 and 3.0 SIAs, there was a very strong
concordance of the test results between the manual and the automated
runs with regard to the reactive bands. Nine samples had discordant
results between the manual and the automated runs; this was probably
attributable to increased variability in antigen scores close to the
cutoff values for both tests. Run-to-run and within-run testing by the
CHIRON RIBA HCV Processor System showed a very low rate of conflicting
values. In conclusion, the CHIRON RIBA HCV Processor System is capable
of performing RIBA HCV 2.0 and 3.0 SIAs accurately with minimal
operator involvement. In addition, the CHIRON RIBA HCV Processor System
shows excellent reproducibility, with the potential for
operator-to-operator and site-to-site variability being greatly
reduced. Our data indicate that this novel methodology may be very
useful for supplemental anti-HCV testing of specimens repeatedly
reactive by ELISA in routine clinical assessments and epidemiologic
evaluations.
 |
INTRODUCTION |
The discovery of the genome
(4) sequence of hepatitis C virus (HCV), the causative agent
of non-A, non-B hepatitis (NANBH) (1, 2), has resulted in
the development of a variety of diagnostic assays for HCV antibodies
(9). The earliest anti-HCV assays had important limitations,
notably, a high rate of false-positive and false-negative results
(12). To increase both sensitivity and specificity, in these
diagnostic assays a greater number of HCV-encoded antigens are now
included, allowing more specific antibody detection. Second-generation
and now third-generation enzyme-linked immunosorbent assays
(ELISAs) (ELISA-2 and ELISA-3, respectively) are also available.
The anti-HCV ELISA-3 includes antigens coded by the putative core and
NS3, NS4, and NS5 regions of the HCV genome. ELISA-3 is widely used to
screen donor blood (11, 16), and despite its high
specificity (99.7%), false-positive results may occur (14).
The positive predictive value of the test depends on the prevalence of
HCV antibodies in the donor population, which is very low. In view of
this, anti-HCV ELISA reactivity should be tested with a supplemental
assay. Recombinant immunoblot assays (15) (RIBA; Chiron
Corporation) and synthetic peptide assays (Inno-Lia; Innogenetics)
(7) have been developed as supplemental tests for
discriminating between true- and false-positive results for samples
repeatedly reactive by ELISAs (12). The second- and
third-generation RIBA HCV strip immunoblot assays (SIAs) (HCV 2.0 and
3.0 SIAs, respectively) are established methods for supplemental
testing of samples repeatedly reactive by HCV ELISAs. Since 1992, the
third-generation RIBA system (RIBA HCV 3.0 SIA) has been widely used in
Europe for supplementary testing of samples repeatably reactive by HCV
ELISA. The RIBA HCV 3.0 SIA has four significant differences from its
predecessor (RIBA HCV 2.0 SIA): it includes a recombinant antigen from
the NS5 region, the recombinant c22-3 antigen is replaced by a
four-epitope core synthetic peptide c22(p), the 5-1-1 and c100-3
recombinant antigens are replaced by a mixture of synthetic peptides
c100(p), and more c33-c recombinant antigen is used. Various
investigators have shown that this new combination of antigens has
greater specificity and sensitivity than its predecessor (3, 5, 8,
13, 17).
However, the current manual procedures in the RIBA HCV 2.0 and 3.0 SIAs
are labor-intensive. Both band scoring and result interpretation are
subjective. A CHIRON RIBA HCV Processor System has now been developed
to automate supplemental testing for HCV. The bench-top instrument
provides objective strip scoring and result interpretation; it can
process up to 28 specimens plus controls per run and offers walkaway
operational capability.
The purpose of this study was to evaluate the accuracy of the CHIRON
RIBA HCV Processor System. We compared the RIBA HCV 2.0 and .0 SIAs
performed with the CHIRON RIBA HCV Processor System with those
processed manually for a large cohort of specimens. Additionally, the
run-to-run and within-run precisions of the CHIRON RIBA HCV Processor
System were tested by using a specimen yielding weak to medium
reactivities for all RIBA antigens.
(This study was presented in part at the 46th Annual Meeting of the
American Association for the Study of Liver Disease, Chicago, Ill.
November 1995.)
 |
MATERIALS AND METHODS |
Enzyme immunoassays.
The ELISA-2 Ortho HCV ELISA system
(Ortho Diagnostic Systems, Raritan, N.J.) detects antibodies to a
structural antigen (core antigen) and a fusion of the c100-3 and c33-c
antigens (c200) of HCV. It uses recombinant antigens derived from three
regions of the viral genome (core, NS3, and NS4).
RIBA HCV SIAs.
The RIBA HCV 2.0 strip consists of a
nitrocellulose solid support on which four bands of recombinant HCV
proteins (bands 2 to 5), two bands of high-level and low-level
immunoglobulin G (IgG) (bands 1 and 7, respectively), and a superoxide
dismutase band (band 6) are immobilized. Band 2 contains the 5-1-1 antigen derived from the NS4 region of the HCV genome. Band 3 contains the c100-3 recombinant antigen derived from the NS4 region. Band 4 contains c33-c recombinant antigen derived from the NS3 region of the
HCV genome. Band 5 contains a recombinant antigen (c22-3) derived from
the core region of the HCV genome. The RIBA HCV 3.0 strip is different
from the RIBA HCV 2.0 strip in that in addition to recombinant HCV
antigens it contains peptide segments of the HCV genome. In the RIBA
HCV 3.0 strip the distribution of the bands is as follows. Band 2 contains two synthetic peptides, 5-1-1(p) and c100(p), from the NS4
region of the HCV genome. Band 3 contains c33-c recombinant antigen
derived from the NS3 region. Band 4 contains c22(p) from the core
portion of HCV. Band 5 contains NS5 region of the HCV genome.
The procedures for manual and automated RIBA HCV SIAs were as follows:
20 microliters of each specimen in which the presence or absence of HCV
is to be confirmed was added to a tube containing a RIBA HCV 2.0 or 3.0 strip and 1 ml of RIBA HCV 2.0 or 3.0 SIA specimen diluent. The next
step was a 4-h incubation with the specimen at room temperature,
decantation, a 30-min incubation in specimen diluent, two washes in
wash buffer, a 10-min incubation in conjugate (peroxidase-labeled goat
anti-human IgG [heavy and light chains]) followed by three washes in
wash buffer, and a 15-min incubation in substrate (4-chloro-1-naphthol)
followed by two washes with deionized water.
Algorithm for supplemental testing of HCV by RIBA HCV SIA.
Anti-HCV reactivity in a specimen is determined by comparing the
intensity of each band to the intensity of the human IgG (level I and
level II) internal control band on each strip. The identities of the
antibodies are defined by the specified location of the antigen band on
the strip in the kit. The intensity of the antigen and peptide bands is
scored in relation to the intensities of the internal IgG controls
according to the manufacturer's recommendations. A negative,
indeterminate, or positive result is based on the reaction pattern
present on the strip. Interpretation of the results was performed
according to the manufacturer's instructions.
CHIRON RIBA HCV Processor System.
The CHIRON RIBA HCV
Processor System measures the intensities of the control and antigen
bands of an immunoblot strip on which there are four bands of
immobilized recombinant antigens and/or synthetic peptides. It
illuminates the strip and creates a density of reflectance, measuring
the light differentially reflected from the developed bands and white
background. An algorithm interpolates relative intensity values for
antigen band reactivity, with constants assigned to the control bands.
Study samples.
A comparison of RIBA HCV 2.0 and 3.0 SIA
results between the processor and the manual runs was performed with a
cohort of 199 specimens. There were 111 random serum samples from
presumably healthy blood donors and 88 specimens repeatedly reactive by
anti-HCV ELISA-2. Among the anti-HCV ELISA-2-reactive specimens, 4 serum samples were from patients with acute NANBH, 33 serum samples were from patients with chronic NANBH, 26 samples were from 12 patients
with posttransfusion hepatitis C, 19 serum samples were from chronic
hemodialysis patients, and 6 serum samples were from 4 chronic
hemodialysis patients showing seroconversion for HCV infection.
CHIRON RIBA HCV Processor System and RIBA HCV SIA reproducibility
testing.
RIBA HCV 2.0 SIA reproducibility testing was performed
with 109 samples; each specimen was tested by the CHIRON RIBA HCV
Processor System in four runs. RIBA HCV 3.0 SIA reproducibility testing was performed with 168 specimens; each sample was tested by the CHIRON
RIBA HCV Processor System in six runs. Material of the same lot was
used for the RIBA HCV 2.0 and 3.0 SIA reproducibility testing.
Statistical analysis.
The agreement between the reactivity
to HCV antigens shown by the CHIRON RIBA HCV Processor System versus
that shown by the RIBA manual assay was measured by using kappa
statistics (10). The statistical package JMP IN, version
3.1.7 for MacIntosh (SAS Institute Inc.), was used.
 |
RESULTS |
RIBA HCV 2.0 SIA: automated versus manual testing.
For the
RIBA HCV 2.0 SIA the correlation of all test results between the manual
and the automated assays was 98.9% (197 of 199). The kappa value was
very high (0.937). Forty-nine were specimens positive by the manual and
the automated assays; 22 samples showed indeterminate results by the
manual technology and by testing with the RIBA Processor System.
One-hundred twenty-six samples gave negative results. Two samples had
discordant results between the two procedures. Both of them tested
negative by PCR. The first sample (Table
1, specimen A) was negative by the manual
RIBA HCV 2.0 SIA but indeterminate by the automated RIBA HCV 2.0 SIA. The conflicting result concerned band c22-3. The second sample (Table
1, specimen B) was indeterminate by the RIBA manual assay and negative
by testing with the CHIRON RIBA HCV Processor System, with band c33-c
having conflicting results. The discordant results were due to
increased variability of antigen scores close to the cutoff values for
both tests.
Among the specimens positive by RIBA HCV 2.0 SIA, there was a strong
correlation between manual and automated runs with regard
to the number
of reactive bands: 96% (46 of 49) of samples positive
by RIBA HCV 2.0 SIA showed bands with concordant reactivities.
Two specimens (specimens
C and D) showed two reactive bands by
the manual procedure and three
reactive bands by the automated
methodology. One sample (specimen E)
had three reactive bands
by the manual protocol and two reactive bands
by testing with
the CHIRON RIBA HCV Processor System.
RIBA HCV 3.0 SIA: automated versus manual testing.
For the
RIBA HCV 3.0 SIA the correlation of test results was 96.5% (192 of
199) between runs performed with the CHIRON RIBA HCV Processor System
and manually. The kappa value was high (0.899). Sixty-five samples were
positive by the manual assay and by testing with the CHIRON RIBA
Processor System; nine specimens gave indeterminate reactivities by the
manual and the automated technologies. One hundred eighteen specimens
had negative results. Seven samples generated conflicting results
(Table 2). Four samples were
indeterminate by the manual RIBA HCV 3.0 SIA but tested positive by
testing with the CHIRON RIBA HCV Processor System. The discordant
samples concerned 5-1-1(p) c100(p) bands (specimen F), the c22(p) band (specimens G and H), and the c33-c band (specimen I). Specimens G and I
were positive by PCR, and specimens F and H were negative by PCR. Two
samples were negative by manual RIBA HCV 3.0 SIA and gave indeterminate
reactivities by testing with the CHIRON RIBA HCV Processor System. The
conflicting results involved the c33-c (specimen L) and the c22(p)
(specimen M) bands. One specimen (specimen N) was positive by the RIBA
manual assay and gave indeterminate results by testing with the CHIRON
RIBA HCV Processor System. The discordant result involved the c33-c
band. Specimens M and N were found to be PCR positive; sample L gave
negative results by PCR. The discordant results between the two tests
were related to increased variability of antigen scores close to the
cutoff values for both tests.
Among specimens positive by RIBA HCV 3.0 SIA, there was a strong
correlation between the manual and the automated runs with
regard to
the number of reactive bands: 92% (60 of 65) of positive
samples gave
bands with concordant reactivities. Three positive
specimens (specimens
O, P, and Q) showed three reactive bands
with the automated methodology
and two reactive bands by the manual
procedure; one specimen (specimen
R) showed four reactive bands
by testing with the CHIRON RIBA HCV
Processor System and two reactive
bands by the manual assay; a fifth
specimen (specimen S) had four
reactive bands by testing with the
CHIRON RIBA HCV Processor System
and three bands by the manual
methodology.
CHIRON RIBA HCV Processor System and RIBA HCV SIA reproducibility
testing.
For the RIBA HCV 2.0 SIA, the precisions of assays run on
the CHIRON RIBA HCV Processor System for the reactive HCV antigen bands
were <16% (range, 7 to 15.9%) for within-run and run-to-run testing
of a single specimen. The mean relative intensity values, standard
deviation, and coefficient of variation for each of four RIBA HCV 2.0 SIA-antigenic bands are reported in Table
3.
For the RIBA HCV 3.0 SIA, the precision of assays run on the CHIRON
RIBA HCV Processor System for the reactive HCV antigen
bands was <10%
(range, 5.2 to 9.6%) for within-run and run-to-run
testing of a single
specimen. The mean relative intensity values,
standard deviation, and
coefficient of variation for each of four
RIBA HCV 3.0 SIA antigenic
bands are presented in Table
4.
 |
DISCUSSION |
RIBA HCV 2.0 and 3.0 SIAs have been established as supplemental
assays for discriminating between true- and false-positive specimens
which are repeatedly reactive in anti-HCV screening ELISAs. RIBA HCV
SIAs are labor-intensive and require subjective scoring of strips and
interpretation of results. However, the RIBA methodology lends itself
to automation, as reported previously (6), and with
automation, it can be used to test large numbers of patient specimens.
Recently, a CHIRON RIBA HCV Processor System has been developed for
automated supplemental testing for HCV.
We describe here the application of this automated technique for
supplemental testing of samples for HCV infection. The RIBA HCV SIA
methodology detects serum antibodies to individual recombinant antigens
or synthetic peptides. Individual HCV proteins or peptides are
immobilized on a nitrocellulose strip, where they react with anti-HCV
antibodies if the antibodies are present in the patient's serum. If
bound conjugate is present, a colorimetric enzyme detection system will
produce an insoluble blue-black reaction product at each specific HCV
antigen, peptide, or control band. For supplemental testing for HCV,
the RIBA strip is then interpreted by the CHIRON RIBA HCV Processor
System, an instrument which illuminates the strip, measures the light
differentially reflected from the developed bands and the white
background, and determines the intensities of the reactive bands in
relation to the intensities of the internal control bands on each
strip. The relative intensity values for antigen band reactivity are
scored by the CHIRON RIBA HCV Processor System, which uses the value
assigned to the control bands. The CHIRON RIBA HCV Processor System is
capable of performing the RIBA HCV 2.0 and HCV 3.0 SIAs with minimal
operator involvement. It provides an objective evaluation of band
scoring and interpretation of results and offers a walkaway operational
capability.
To evaluate the accuracy of the CHIRON RIBA HCV Processor System in
performing supplemental testing by the RIBA HCV 2.0 and 3.0 SIAs, we
compared the results obtained with the CHIRON RIBA HCV Processor System
with those obtained by the manual procedure with specimens from
presumably healthy blood donors and from patients serologically or
clinically confirmed as having HCV infection. For the RIBA HCV 2.0 SIA,
we observed a very high correlation between the manual and the
automated procedures (kappa value, 0.937). For the RIBA HCV 3.0 SIA,
the correlation of test results between the manual and automated tests
was also high (kappa value, 0.899). The discordant results between the
two procedures were likely related to the increased variability of
antigen scores close to the cutoff values for both tests. Moreover, we
found a very strong correlation between manual and automated runs
regarding the number of reactive bands in samples positive by the RIBA
HCV 2.0 and 3.0 SIAs. Ninety-three percent (111 of 119) of specimens positive by the RIBA HCV 2.0 and 3.0 SIAs showed bands with concordant reactivities between the manual and the automated procedures.
The run-to-run and within-run precisions offered by the CHIRON RIBA HCV
Processor System were excellent. For reproducibility testing, a
specimen was tested in several replicate runs, and the coefficients of
variation were less than 16 and 10% for the RIBA HCV 2.0 and 3.0 SIAs,
respectively. Thus, the CHIRON RIBA HCV Processor System's results
were highly reproducible.
At present, the serological diagnosis of antibody against HCV remains
the most common method of assessing HCV infection. Although a positive
result by ELISA and RIBA HCV SIA indicates active HCV infection, it
does not distinguish between current infection and previous exposure.
Direct detection of HCV would be more useful. However,
methodologies aimed at the direct detection of the HCV genome,
such as PCR, are expensive and laborious and need appropriate standardization (18). Commercially available kits allowing
HCV RNA detection have only recently been introduced.
In conclusion, (i) the CHIRON RIBA HCV Processor System proved to be
capable of performing RIBA HCV 2.0 and HCV 3.0 SIAs with minimal
operator involvement, was able to objectively interpret the results,
and had a walkaway capability. (ii) The CHIRON RIBA HCV Processor
System was substantially equivalent to the manually performed assay;
for the RIBA HCV 2.0 and 3.0 SIAs, we found a very strong correlation
between the results of the manual and the automated runs. The few
discordant results between the two procedures were due to increased
variability of antigen scores close to the cutoff values for both
tests. (iii) The CHIRON RIBA HCV Processor System offers excellent
reproducibility, and the potential for operator-to-operator and
site-to-site variabilities is greatly reduced.
The CHIRON RIBA HCV Processor System is a very useful tool for
confirming HCV-positive results obtained by ELISA in routine clinical
assessments and in epidemiological surveys.
 |
ACKNOWLEDGMENT |
This work was supported by the 1996 Research Fellowship Award of
the Society of Italian-American Nephrologists (to F.F. at the
Department of Medicine, School of Medicine, University of California at
Los Angeles).
 |
FOOTNOTES |
*
Corresponding author. Mailing address: Dumont-UCLA
Liver Transplant Program, UCLA School of Medicine, 77-123D CHS, 10833 Le Conte Ave., Los Angeles, CA 90024-1749. Phone: (310) 825-5302. Fax:
(310) 852-5302. E-mail: Pmartin{at}surgery.medsch.ucla.edu.
 |
REFERENCES |
| 1.
|
Aach, R.,
C. Stevens,
B. Hollinger,
J. W. Mosley,
D. A. Peterson,
P. E. Taylor,
R. G. Johnson,
L. H. Barbosa, and G. J. Nemo.
1991.
Hepatitis C virus infection in post-transfusion hepatitis an analysis with first- and second-generation assays.
N. Engl. J. Med.
325:1325-1329[Abstract].
|
| 2.
|
Alter, M. J.,
S. C. Hadler,
F. N. Judson,
A. Mares,
J. Alexander,
P. Y. Hu,
J. K. Miller,
L. A. Moyer,
H. A. Fields,
D. W. Bradley, and R. H. Margolis.
1990.
Risk factors for acute non-A, non-B hepatitis in the United States and association with hepatitis C virus infection.
JAMA
264:2231-2235[Abstract/Free Full Text].
|
| 3.
|
Buffet, C.,
N. Charmaux,
P. Laurent-Puig,
S. Chopineau,
J. P. Quichon,
M. J. Briantais, and E. Dussaix.
1994.
Enhanced detection of antibodies to hepatitis C virus by use of a third-generation recombinant imunoblot assay.
J. Med. Virol.
43:259-261[Medline].
|
| 4.
|
Choo, Q. L.,
G. Kuo,
A. Weiner,
L. R. Overby,
D. W. Bradley, and M. Houghton.
1989.
Isolation of a cDNA clone derived from a blood-borne non-A, non-B viral hepatitis genome.
Science
244:359-362[Abstract/Free Full Text].
|
| 5.
|
Damen, M.,
H. L. Zaaijer,
H. T. M. Cuypers,
H. Vrielink,
C. L. van der Poel,
H. W. Reesink, and P. N. Lelie.
1995.
Reliability of the third-generation recombinant immunoblot assay for hepatitis C virus.
Transfusion
36:745-749.
|
| 6.
|
Dixit, V.,
S. Quan,
P. Martin,
D. Larson,
M. Brezina,
R. DiNello,
K. Sra,
J. Y. N. Lau,
D. Chien,
J. Kolberg,
A. Tagger,
G. Davis,
A. Polito, and G. Gitnick.
1995.
Evaluation of a novel serotyping system for hepatitis C virus: strong correlation with standard genotyping methodologies.
J. Clin. Microbiol.
33:2978-2983[Abstract].
|
| 7.
|
Dow, B. C.,
I. Coote,
H. Munro,
F. McOmish,
P. L. Yap,
P. Simmonds, and E. A. C. Follett.
1993.
Confirmation of hepatitis C virus antibody in blood donors.
J. Med. Virol.
41:215-220[Medline].
|
| 8.
|
Fabrizi, F.,
G. Lunghi,
L. Raffaele,
I. Guarnori,
G. Bacchini,
M. Corti,
A. Pagano,
G. Erba, and F. Locatelli.
1997.
Serologic survey for control of hepatitis C in haemodialysis patients: third-generation assays and analysis of costs.
Nephrol. Dial. Transplant.
12:298-303[Abstract/Free Full Text].
|
| 9.
|
Kuo, G.,
Q. L. Choo,
H. J. Alter,
G. L. Gitnick,
A. G. Redeker,
R. H. Purcell,
T. Miyamura,
J. L. Dienstag,
M. J. Alter,
C. E. Stevens,
G. E. Tegmeier,
F. Bonino,
M. Colombo,
W. S. Lee,
C. Kuo,
K. Berger,
J. R. Shuster,
L. R. Overby,
D. W. Bradley, and M. Houghton.
1989.
An assay for circulating antibodies to a major etiologic virus of human non-A, non-B hepatitis.
Science
244:362-364[Abstract/Free Full Text].
|
| 10.
|
Landis, J. R., and C. G. Koch.
1997.
The measurement of observer agreement for categorical data.
Biometrics
33:159-174.
|
| 11.
|
Lee, S. R.,
C. L. Wood,
M. J. Lane,
B. Francis,
C. Gust,
C. M. Higgs,
M. J. Nelles,
A. Polito,
R. DiNello, and D. Achord.
1995.
Increased detection of hepatitis C virus infection in commercial plasma donors by a third-generation screening assay.
Transfusion
35:845-849[Medline].
|
| 12.
|
Skidmore, S.
1990.
Recombinant immunoblot assay for hepatitis C antibody.
Lancet
335:1346[Medline].
|
| 13.
|
Uyttendaele, S.,
H. Claeys,
W. Mertens,
H. Verhaert, and C. Vermylen.
1994.
Evaluation of third-generation screening and confirmatory assays for HCV antibodies.
Vox Sang.
66:122-129[Medline].
|
| 14.
|
van der Poel, C. L.,
H. T. M. Cuypers, and H. W. Reesink.
1994.
Hepatitis C virus six years on.
Lancet
344:1475-1479[Medline].
|
| 15.
|
van der Poel, C. L.,
H. T. M. Cuypers,
H. W. Reesink,
A. J. Weiner,
S. Quan,
R. DiNello,
J. J. P. van Boven,
I. Winkel,
D. Mulder-Folkerts,
P. J. Exel-Oehlers,
W. Schaasberg,
A. Leentvaar-Kuypers,
A. Polito,
M. Houghton, and P. N. Lelie.
1991.
Confirmation of hepatitis C virus infection by new four-antigen recombinant immunoblot assay.
Lancet
337:317-319[Medline].
|
| 16.
|
Vrielink, H.,
H. L. Zaaijer,
H. W. Reesink,
P. N. Lelie, and C. L. van der Poel.
1995.
Comparison of two anti-hepatitis C virus enzyme-linked immunosorbent assays.
Transfusion
35:601-604[Medline].
|
| 17.
|
Zaaijer, H. L.,
H. Vrielink,
P. J. van Exel-Oehlers,
H. T. M. Cuypers, and P. N. Lelie.
1994.
Confirmation of hepatitis C infection: a comparison of five immunoblot assays.
Transfusion
34:603-607[Medline].
|
| 18.
|
Zaajer, H. L.,
H. T. Cuypers,
H. W. Reesink,
I. N. Winkel,
G. Gerken, and P. N. Lelie.
1993.
Reliability of polymerase chain reaction for detection of hepatitis C virus.
Lancet
341:722-724[Medline].
|
Journal of Clinical Microbiology, February 1998, p. 387-390, Vol. 36, No. 2
0095-1137/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.