Previous Article | Next Article 
Journal of Clinical Microbiology, April 1998, p. 862-865, Vol. 36, No. 4
0095-1137/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.
Multicenter Evaluation of the COBAS AMPLICOR HCV Assay, an
Integrated PCR System for Rapid Detection of Hepatitis C Virus RNA
in the Diagnostic Laboratory
J.
Albadalejo,1
R.
Alonso,1
R.
Antinozzi,2
M.
Bogard,3
A.-M.
Bourgault,4
G.
Colucci,5,*
T.
Fenner,6
H.
Petersen,6
E.
Sala,2
J.
Vincelette,4 and
C.
Young7
Servicio de Microbiologia Clinica y
Enfermedades Infecciosas, Hospital General Universitario Gregorio
Maranon, Madrid, Spain1;
Laboratorio di
Patologia Clinica, Ospedale S. Anna, Como,
Italy2;
Laboratoire de Biologie
Moleculaire, Hopital de Meaux, Meaux, France3;
PCR Unit, Roche Diagnostic Systems, Basel,
Switzerland5;
PCR Labor,
Gemeinschaftspraxis Dres Fenner, Hamburg,
Germany6;
Hopital St. Luc, Montreal,
Quebec, Canada4; and
Roche Molecular
Systems, Somerville, New Jersey7
Received 8 September 1997/Returned for modification 19 November
1997/Accepted 6 January 1998
 |
ABSTRACT |
The benefits shown by the recent introduction of PCR for the in
vitro diagnosis of hepatitis C virus (HCV) infection has prompted the
development of standardized, ready-to-use assays that can be
implemented in routine clinical laboratories. We have evaluated the clinical performance of COBAS AMPLICOR HCV
(COBAS), the first instrument system that allows the
automation of HCV RNA amplification and detection, to determine its
performance in the routine laboratory setting. More than 2,000 specimens collected at five centers were analyzed in parallel by the
COBAS and the manual AMPLICOR HCV (AMPLICOR) tests, and the
results were compared with the results for biochemical and serological
markers of HCV. In this study the two PCR systems showed the same
accuracy, with a concordance rate of 99.8%. As expected, the
correlation between serology and PCR was not absolute because the
presence of anti-HCV antibodies may be associated with a latent or past
infection. On the other hand, if the presence of confirmed anti-HCV
antibodies and elevated alanine aminotransferase levels are taken as
the "gold standard," indicating an active, ongoing infection,
the COBAS and AMPLICOR tests show high and comparable
sensitivities (100%) and specificities (98%), with positive and
negative predictive values of 100 and 97%, respectively. During the
study no false-positive reactions were detected. The use of an internal
control allowed the identification of inhibitory substances that
prevented amplification for 0.3 and 0.4% of samples tested by the
COBAS and AMPLICOR tests, respectively. Compared to the manual
system, the COBAS system allowed a significant reduction of
hands-on time and could improve the overall laboratory work flow. In
conclusion, these results support the use of the COBAS and
AMPLICOR tests for the molecular diagnosis of active HCV
infections.
 |
INTRODUCTION |
The lack of virus isolation
techniques and antigen detection assays has made nucleic acid-based
amplification the method of choice for the direct identification of
hepatitis C virus (HCV). Indeed, by PCR it is possible to assess the
status of the infection, to detect viral replication in seropositive
patients, and to diagnose the infection in immunocompromised patients
and during the window that precedes seroconversion (2, 3, 15, 18,
21).
The implementation of PCR in routine diagnostic laboratories, however,
requires a level of standardization that was not originally provided by
in-house methods. Additionally, by in-house methods the occurrence of
false-positive results due to amplified product contamination is not
always preventable and false-negative results due to inhibition are not
always easily detectable (5, 17, 19, 20, 24). Also, the
hands-on time required by a manual in-house assay for reagent
preparation as well as the testing of samples from patients limits its
utility in the routine diagnostic laboratory. To address these
technical concerns, a dedicated instrument, the COBAS AMPLICOR
HCV (COBAS) analyzer, that automates amplification and detection
has recently been introduced (6, 10). The system provides amplification and detection with standardized,
ready-to-use reagents based on the reagents formerly developed
for the corresponding AMPLICOR HCV manual format
(AMPLICOR). The amplification reagents include the enzyme
uracil-N-glycosylase (AmpErase) for the prevention of
carryover contamination (12, 22) and an internal
control (IC) for the identification of processed specimens containing substances that may interfere with amplification (16).
The closed system provides additional protection against carryover
contamination. In the present study we evaluated the performance of the
COBAS test with a collection of more than 2,000 clinical specimens
and compared it with that of the corresponding manual AMPLICOR test and the results for other serological and biochemical parameters.
 |
MATERIALS AND METHODS |
We analyzed 2,292 consecutive serum specimens collected and
separately tested at five centers. The five centers provided results for 484, 655, 495, 498, and 160 serum specimens, respectively. Serum
was separated from whole blood and was immediately stored at
80°C
in several aliquots to preserve the integrity of the HCV RNA.
Serology.
All specimens were tested for the presence of
anti-HCV antibodies by enzyme-linked immunosorbent assay (ELISA; Abbott
Diagnostics, North Chicago, Ill.; Ortho Diagnostic Systems, Raritan,
N.J.) and at most centers by RIBA (Chiron Corp., Emeryville, Calif.). This confirmatory assay was performed with 2,056 (89%) samples. In
addition, biochemical markers of liver inflammation and necrosis including aspartate aminotransferase and alanine aminotransferase (ALT)
levels were evaluated.
PCR analysis.
Serum specimens were analyzed by the COBAS
test (Roche Diagnostic Systems, Somerville, N.J.) by following the
manufacturer's instructions. Briefly, 100 µl of serum was incubated
with a lysis buffer, and the RNA was then precipitated with isopropanol
by centrifugation, washed once with ethanol, and resuspended in 1 ml of
specimen diluent. An IC RNA was also introduced into the specimen with
the diluent and served as an amplification control for each
individually processed specimen. The HCV IC consists of a synthetic RNA
transcript with primer binding regions identical to those of the
HCV target sequence, a randomized internal sequence with a length and a
base composition similar to those of the HCV target sequence, and a
unique probe binding region that differentiates the IC from the target
amplicon.
Each specimen or control was amplified in the thermal cycler section of
the COBAS analyzer with primers KY78 (biotinylated) and KY80, which
identify a 244-bp sequence of the highly conserved 5' untranslated
region of the HCV genome.
After amplification, the analyzer automatically denatured the
double-stranded, biotinylated amplified products and captured
them
with a suspension of magnetic particles coated with an oligonucleotide
probe specific for HCV (or IC). The unbound material was washed
away,
and the biotinylated amplicon was detected with an avidin-horseradish
peroxidase conjugate-tetramethylbenzidine-H
2O
2
colorometric reaction.
The absorbance (660 nm) of each sample was
recorded and was compared
with a predefined cutoff value to determine
positive or negative
results for both HCV and IC.
An aliquot of each specimen was tested in parallel by the AMPLICOR
assay (Roche Diagnostic Systems) as described in the manufacturer's
instructions.
The results of the PCR assay as well as all other clinical,
serological, and biochemical data were collected and analyzed.
Discrepancies between PCR, both automated and manual, and between
PCR
and serology were resolved by repeat testing of frozen aliquots.
Repeat
testing was also done with specimens with no IC signal
or a low IC
signal to ascertain the presence of substances inhibitory
to the PCR.
At two of the participating centers a workload analysis was carried out
by the method of the College of American Pathologists
(
4).
To determine core timing values, data were collected for
the following
steps of the procedure: (i) initial handling of
the specimen, (ii)
specimen testing, (iii) daily and periodic
activities, and (iv)
recording of test data from the instrument.
Individual times for each
step were summed, and "hands-off" times
were also calculated to
give the total time from the initiation
of the test to retrieval of the
final test result.
 |
RESULTS |
The 2,292 consecutive specimens obtained from patients at risk of
HCV infection were simultaneously analyzed by the COBAS and the
AMPLICOR tests. The concordance between the two systems was 98%,
but it rose to 99.8% after repeat testing (Fig.
1). Discrepant results were observed for
4 (0.2%) of all samples.

View larger version (11K):
[in this window]
[in a new window]
|
FIG. 1.
Overall performance of COBAS and AMPLICOR tests.
Numbers indicate the numbers of samples with the indicated test
results.
|
|
Samples presenting negative IC values due to the presence of
inhibitory substances that prevented amplification of the IC were
resolved by retesting another aliquot of the specimen. After resolution of the results, invalid results were obtained for 0.3 and 0.4% of all specimens analyzed by the COBAS and AMPLICOR
tests, respectively. The values obtained prior to repeat testing were 3 and 2% for the COBAS and AMPLICOR tests, respectively.
When compared to the patients' serological status, as assessed by
anti-HCV ELISA and RIBA, the results of PCR showed an overall concordance rate of 84% (Fig. 2).

View larger version (9K):
[in this window]
[in a new window]
|
FIG. 2.
Comparison between COBAS and AMPLICOR test
results and the HCV serological status as defined by ELISA and RIBA.
Numbers indicate the numbers of samples with the indicated test
results.
|
|
To better define the relationship between serological and molecular
data for our patients, we compared the results obtained with the
COBAS and AMPLICOR tests with those of ALT level and RIBA
determinations. The combination of the latter two parameters served to
identify active HCV infection. By this approach the concordance rate
increased to 98% (Fig. 3), indicating
that most anti-HCV-positive, PCR-negative discrepant results were due
to past or latent infections. Moreover, if PCR results are compared to
the active infection status, with use of active infection status as the "gold standard," the COBAS and AMPLICOR tests
show sensitivities and specificities of 96 and 100% and of 95 and
100%, respectively. The corresponding positive predictive and negative
predictive values were 100 and 98%, respectively, for the COBAS
test and 100 and 97%, respectively, for the AMPLICOR test.

View larger version (9K):
[in this window]
[in a new window]
|
FIG. 3.
Comparison between COBAS and AMPLICOR test
results and the presence or absence of an active HCV infection as
defined by the detection of specific antivirus antibodies and elevated
ALT levels. Numbers indicate the numbers of samples with the indicated
test results.
|
|
The turnaround time for the complete COBAS test including specimen
preparation and the amplification and detection of HCV and IC was
5 h for a batch of 20 specimens. The operator was required to be
present for less than 2 h to perform manual steps that included sample and reagent preparation and loading of the amplification tubes
in the thermal cyclers and in the detection racks. The overall workload
time was about 5 min/specimen, compared to 7 min/specimen for the
manual test. By using the parallel-run feature of the instrument, the
throughput can be increased to three to four runs per day, including an
unattended overnight run.
 |
DISCUSSION |
HCV infection is a typical example of a disease in which direct
detection of the virus is essential for a correct diagnosis. In
contrast to the other available in vitro assays, PCR has the potential
for high diagnostic value because it offers a definitive identification
of HCV. However, the poor reproducibility of in-house-developed assays,
shown in previous studies, has indicated the need for standardized and
more accurate procedures (5, 24). The recent introduction of
an automated instrument for the amplification and detection of HCV RNA
has prompted us to evaluate its performance with a large collection of
clinical specimens obtained from different institutions. The results of
this study have shown that the COBAS test has the same performance
as the established manual AMPLICOR test because only 0.2% of all
results remained discordant after repeat testing. Initial discrepancies
were due mostly to invalid results caused by low values for the IC.
The COBAS and AMPLICOR tests also showed the same sensitivities
and specificities compared to those of serology and clinical diagnosis,
which were taken as the gold standards. We used clinical diagnosis
because serology alone does not necessarily reflect the presence of an
active infection but reflects only a previous exposure to the virus
(2, 21). On the other hand, the presence of anti-HCV
antibodies in conjunction with elevated ALT levels and histological
evidence of chronic hepatitis better represent the features of an
ongoing, clinically relevant infection (1, 3, 13, 21). In
this regard, the samples with discrepant results that were anti-HCV
positive and HCV RNA negative were from patients who had normal ALT
levels and hence a latent, inactive infection. The few patients (2%)
who had a clinical diagnosis of type C hepatitis but who were PCR
negative had borderline ALT level elevations and probably had HCV RNA
concentrations below the analytical sensitivity of the AMPLICOR
system (7).
We did not compare the COBAS test with nested PCR techniques
because the manual AMPLICOR test, which served as the
reference for the automated system in this study, had been previously
extensively analyzed (8, 9, 14, 23).
An interesting feature of the COBAS test is the optional detection
of an IC that is coamplified with the target and that serves to
identify the possible presence of interfering substances that inhibit
DNA polymerization, causing false-negative reactions (16).
Inhibition occurred with 3% of all samples and was readily solved by
repeat testing of frozen aliquots. This may indicate that modification
or degradation of inhibitory substances occurred during storage or that
inhibitors are not equally distributed throughout a sample
(16).
On the other hand, the presence of uracil-N-glycosylase
(AmpErase), which destroys a dUTP-containing amplicon from
previous reactions, and the use of proper precautions taken at
the time of specimen preparation prevented false-positive reactions due to carryover contamination or cross contamination from positive to
negative samples (11, 12).
In this respect, an additional benefit of the COBAS system is the
closed containment in which all amplification and detection steps take
place. Amplification products are kept in sealed tubes that are never
opened but that are pierced by the transfer tip during the detection
part of the procedure. Throughout the duration of the study we assessed
the robustness of the instruments and never had failures requiring
external assistance. Taken together, these data suggest that the
COBAS test may be a suitable system for the routine analysis of
samples obtained from anti-HCV-positive individuals.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: Roche Diagnostic
Systems, Bldg. 222/121, 4070 Basel, Switzerland. Phone: 41 61 6873363. Fax: 41 61 6872109. E-mail:
Giuseppe.Colucci.GC1{at}Roche.com.
 |
REFERENCES |
| 1.
|
Alberti, A.,
G. Morsica,
L. Chemello,
D. Cavalletto,
F. Noventa,
P. Pontisso, and A. Ruol.
1992.
Hepatitis C viraemia and liver disease in symptom-free individuals with anti-HCV.
Lancet
340:697-698[Medline].
|
| 2.
|
Alter, H. J.
1995.
To C or not to C: these are the questions.
Blood
85:1681-1695[Free Full Text].
|
| 3.
| Bonino, F., M. R. Brunetto, F. Negro, M. Baldi, G. Saracco, M. L. Abate, A. Fabiano, G. Verme, et al. 1993. Hepatitis C virus infection and disease. Diagnostic problems. J. Hepatol. 17(Suppl. 3):78-82.
|
| 4.
|
College of American Pathologists.
1992.
Workload recording method and personnel management manual.
College of American Pathologists, Northfield, Ill.
|
| 5.
|
Damen, M.,
H. T. M. Cuypers,
H. W. Zaaijer,
H. W. Reesink,
W. P. Schaasberg,
W. H. Gerlich,
H. G. M. Niesters, and P. N. Lelie.
1996.
International collaborative study on the second EUROHEP HCV-RNA reference panel.
J. Virol. Methods
58:175-185[Medline].
|
| 6.
|
DiDomenico, N.,
H. Link,
R. Knobel,
T. Caratsch,
W. Weschler,
Z. G. Loewy, and M. Rosenstraus.
1996.
COBAS AMPLICOR: fully automated RNA and DNA amplification and detection system for routine PCR.
Clin. Chem.
42:1915-1923[Abstract/Free Full Text].
|
| 7.
|
Dusheiko, G. M.,
S. Khakoo,
P. Soni, and L. Grellier.
1996.
A rational approach to the management of hepatitis C infection.
Br. Med. J.
312:357-364[Abstract/Free Full Text].
|
| 8.
|
Gerken, G.,
P. Pontisso,
M. Roggendorf,
M. G. Rumi,
P. Simmonds,
C. Trepo,
S. Zeuzem, and G. Colucci.
1996.
Clinical evaluation of a single reaction, diagnostic PCR assay for the detection of hepatitis C virus (HCV) RNA.
J. Hepatol.
24:33-37[Medline].
|
| 9.
|
Goergen, B.,
K. H. Meyer Zum Buschenfelde, and G. Gerken.
1994.
Assessment of viral replication for routine diagnosis of HCV infection and for monitoring antiviral treatment using Amplicor HCV detection system.
Klin. Lab.
40:1-7.
|
| 10.
|
Jungkind, D.,
S. DiRienzo,
K. G. Beavis, and N. S. Silverman.
1996.
Evaluation of automated COBAS AMPLICOR PCR system for detection of several infectious agents and its impact on laboratory management.
J. Clin. Microbiol.
34:2778-2783[Abstract].
|
| 11.
|
Kwok, S., and R. Higuchi.
1989.
Avoiding false positive with PCR.
Nature
339:237-238[Medline].
|
| 12.
|
Longo, M. C.,
M. S. Berninger, and H. L. Hartley.
1990.
Use of uracil DNA glycosylase to control carry-over contamination in polymerase chain reactions.
Gene
93:125-128[Medline].
|
| 13.
|
Marin, M. G.,
S. Bresciani,
M. Puoti,
A. Rodella,
A. Gussago,
A. Ravaggi,
G. Pizzocolo,
A. Albertini, and E. Cariani.
1996.
Clinical significance of serum hepatitis C virus (HCV) RNA as marker of HCV infection.
J. Clin. Microbiol.
32:3008-3012[Abstract/Free Full Text].
|
| 14.
|
Prati, D.,
C. Capelli,
P. Bosoni,
F. Mozzi,
A. Zanella, and G. Sirchia.
1994.
Determination of hepatitis C virus RNA in the serum by the Amplicor HCV PCR kit.
Vox Sang.
67:112-114[Medline].
|
| 15.
|
Roggendorf, M.,
M. Lu,
H. Meisel,
M. Riffelmann,
E. Schreier, and S. Viazov.
1996.
Rational use of diagnostic tools in hepatitis C.
J. Hepatol.
24:26-34[Medline].
|
| 16.
| Rosenstraus, M., Z. Wang, S.-Y. Chang, D. DeBoneville, and J. P. Spadoro. An internal control for
routine diagnostic PCR: design, properties, and effect on clinical
performance. J. Clin. Microbiol. 36:191-197.
|
| 17.
|
Rys, P. N., and D. H. Persing.
1993.
Preventing false positives: quantitative evaluation of three protocols for inactivation of polymerase chain reaction amplification products.
J. Clin. Microbiol.
31:2356-2360[Abstract/Free Full Text].
|
| 18.
|
Tedeschi, V., and L. B. Seeff.
1995.
Diagnostic tests for hepatitis C: where are we now?
Ann. Intern. Med.
123:383-385[Free Full Text].
|
| 19.
|
Victor, T.,
A. Jordaan,
R. du Toit, and P. D. Van Helden.
1993.
Laboratory experience and guidelines for avoiding false positive polymerase chain reaction results.
J. Clin. Chem. Clin. Biochem.
31:531-535.
|
| 20.
|
White, T. J.,
R. Madej, and D. H. Persing.
1992.
The polymerase chain reaction: clinical applications.
Adv. Clin. Chem.
29:161-196[Medline].
|
| 21.
|
Wilber, J. C., and A. Polito.
1995.
Serological and virological diagnostic tests for hepatitis C virus infection.
Semin. Gastrointestinal Dis.
6:13-19[Medline].
|
| 22.
|
Young, K. K. Y.,
R. M. Resnick, and T. W. Myers.
1993.
Detection of hepatitis C virus RNA by a combined reverse transcription-polymerase chain reaction assay.
J. Clin. Microbiol.
31:882-886[Abstract/Free Full Text].
|
| 23.
|
Young, K. Y.,
J. J. Archer,
O. Yokosuka,
M. Omata, and R. Resnick.
1995.
Detection of hepatitis C virus RNA by a combined reverse transcription PCR assay: comparison with nested amplification and antibody testing.
J. Clin. Microbiol.
33:654-657[Abstract].
|
| 24.
|
Zaaijer, H. J.,
H. T. Cuypers, and H. W. Reesink.
1993.
Reliability of HCV PCR results.
Lancet
341:722-724[Medline].
|
Journal of Clinical Microbiology, April 1998, p. 862-865, Vol. 36, No. 4
0095-1137/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.
This article has been cited by other articles:
-
Scott, J. D., Gretch, D. R.
(2007). Molecular Diagnostics of Hepatitis C Virus Infection: A Systematic Review. JAMA
297: 724-732
[Abstract]
[Full Text]
-
Germer, J. J., Harmsen, W. S., Mandrekar, J. N., Mitchell, P. S., Yao, J. D. C.
(2005). Evaluation of the COBAS TaqMan HCV Test with Automated Sample Processing Using the MagNA Pure LC Instrument. J. Clin. Microbiol.
43: 293-298
[Abstract]
[Full Text]
-
Forman, M. S., Valsamakis, A.
(2004). Increased Sensitivity of the Roche COBAS AMPLICOR HCV Test, Version 2.0, Using Modified Extraction Techniques. J. Mol. Diagn.
6: 225-230
[Abstract]
[Full Text]
-
Richter, S. S.
(2002). Laboratory Assays for Diagnosis and Management of Hepatitis C Virus Infection. J. Clin. Microbiol.
40: 4407-4412
[Full Text]
-
Kessler, H. H., Clarici, A. M. K., Stelzl, E., Muhlbauer, G., Daghofer, E., Santner, B. I., Marth, E., Stauber, R. E.
(2002). Fully Automated Detection of Hepatitis C Virus RNA in Serum and Whole-Blood Samples. CVI
9: 1385-1388
[Abstract]
[Full Text]
-
Stelzl, E., Kormann-Klement, A., Haas, J., Daghofer, E., Santner, B. I., Marth, E., Kessler, H. H.
(2002). Evaluation of an Automated Sample Preparation Protocol for Quantitative Detection of Hepatitis C Virus RNA. J. Clin. Microbiol.
40: 1447-1450
[Abstract]
[Full Text]
-
Kessler, H. H., Stelzl, E., Raggam, R. B., Haas, J., Kirchmeir, F., Hegenbarth, K., Daghofer, E., Santner, B. I., Marth, E., Stauber, R. E.
(2001). Effects of Storage and Type of Blood Collection Tubes on Hepatitis C Virus Level in Whole Blood Samples. J. Clin. Microbiol.
39: 1788-1790
[Abstract]
[Full Text]
-
Yu, M.-L., Chuang, W.-L., Dai, C.-Y., Chen, S.-C., Lin, Z.-Y., Hsieh, M.-Y., Wang, L.-Y., Chang, W.-Y.
(2000). Clinical Evaluation of the Automated COBAS AMPLICOR HCV MONITOR Test Version 2.0 for Quantifying Serum Hepatitis C Virus RNA and Comparison to the Quantiplex HCV Version 2.0 Test. J. Clin. Microbiol.
38: 2933-2939
[Abstract]
[Full Text]
-
Louie, M., Louie, L., Simor, A. E.
(2000). The role of DNA amplification technology in the diagnosis of infectious diseases. CMAJ
163: 301-309
[Abstract]
[Full Text]
-
Martinot-Peignoux, M., Le Breton, V., Fritsch, S., Le guludec, G., Labouret, N., Keller, F., Marcellin, P.
(2000). Assessment of Viral Loads in Patients with Chronic Hepatitis C with AMPLICOR HCV MONITOR Version 1.0, COBAS HCV MONITOR Version 2.0, and QUANTIPLEX HCV RNA Version 2.0 Assays. J. Clin. Microbiol.
38: 2722-2725
[Abstract]
[Full Text]
-
Gerken, G., Rothaar, T., Rumi, M. G., Soffredini, R., Trippler, M., Blunk, M. J., Butcher, A., Soviero, S., Colucci, G.
(2000). Performance of the COBAS AMPLICOR HCV MONITOR Test, Version 2.0, an Automated Reverse Transcription-PCR Quantitative System for Hepatitis C Virus Load Determination. J. Clin. Microbiol.
38: 2210-2214
[Abstract]
[Full Text]
-
Hofgärtner, W. T., Kant, J. A., Weck, K. E.
(2000). Hepatitis C Virus Quantitation: Optimization of Strategies for Detecting Low-Level Viremia. J. Clin. Microbiol.
38: 888-891
[Abstract]
[Full Text]
-
Doglio, A., Laffont, C., Caroli-Bosc, F. X., Rochet, P., Lefebvre, J.-C.
(1999). Second Generation of the Automated Cobas Amplicor HCV Assay Improves Sensitivity of Hepatitis C Virus RNA Detection and Yields Results That Are More Clinically Relevant. J. Clin. Microbiol.
37: 1567-1569
[Abstract]
[Full Text]
-
Kessler, H. H., Jungkind, D., Stelzl, E., Direnzo, S., Vellimedu, S. K., Pierer, K., Santner, B., Marth, E.
(1999). Evaluation of AMPLILINK Software for the COBAS AMPLICOR System. J. Clin. Microbiol.
37: 436-437
[Abstract]
[Full Text]