Previous Article | Next Article ![]()
Journal of Clinical Microbiology, May 1998, p. 1277-1284, Vol. 36, No. 5
Abbott Diagnostics Division, Abbott
Laboratories, North Chicago, Illinois
Received 11 November 1997/Returned for modification 29 December
1997/Accepted 18 February 1998
In the present study, we examined the feasibility of using
recombinant antibodies containing murine variable regions and human constant regions as calibrators or controls in immunoassays. As a model
system, we chose the Abbott IMx Toxo immunoglobulin M (IgM) and Toxo
IgG assays designed to detect antibodies to Toxoplasma gondii. Two mouse monoclonal antibodies were selected based on their reactivity to the T. gondii antigens P30 and P66.
Heavy- and light-chain variable-region genes were cloned from both
hybridomas and transferred into immunoglobulin expression vectors
containing human kappa and IgG1 or IgM constant regions. The constructs
were stably transfected into Sp2/0-Ag14 cells. In the IMx Toxo IgG assay, immunoreactivity of the anti-P30 chimeric IgG1 antibody paralleled that of the positive human plasma-derived assay calibrators. Signal generated with the anti-P66 chimeric IgG1 antibody was observed
to plateau below the maximal reactivity observed for the assay
calibrator. Examination of the IgM chimeric antibodies in the IMx Toxo
IgM assay revealed that both the anti-P30 and anti-P66 antibodies
matched the assay index calibrator manufactured with human Toxo
IgM-positive plasma. When evaluated with patient samples, the
correlation between results obtained with the chimeric antibody
calibrators and the positive human plasma calibrators was Serological immunoassays designed to
detect specific antibody present in patient samples provide a rapid and
sensitive method to monitor for infectious agents, allergy, and
autoimmunity (2). Diagnostic immunoassays and kits designed
to measure antibodies typically include one or more components
containing the specific antibody being measured that function as
calibrators (standards) and/or a positive control (3).
Calibrators are used to establish calibration curves in quantitative
assays for interpolation of antibody concentration in the patient
sample, or alternatively, a single index calibrator may be used to
establish the assay cutoff in a qualitative assay. The positive control
is used to establish assay performance characteristics and is a useful
indicator of the integrity of the reagents. Calibrators and positive
controls are usually prepared by spiking known quantities of specific
antibody derived from seropositive plasma or serum into the negative
control reagent.
However, the use of plasma or serum has several significant drawbacks,
including increasing difficulty in sourcing large volumes with high
titer and specificity, lot-to-lot variability in immunoglobulin G (IgG)
or IgM specificity and affinity, limitations with respect to
characterization, and cost.
An alternative method for the manufacture of calibrators and controls
that circumvents the use of human plasma or serum would represent a
significant advance. Hybridoma technology (12) offers an
indefinite supply of monoclonal antibodies; however, murine antibodies
do not react in the assay format required to measure specific human IgG
or IgM and therefore cannot be used to standardize assays to measure
human antibodies. With the development of recombinant DNA technology,
it has become possible to combine the heavy (H)- and light (L)-chain
variable (V) regions of a desired mouse monoclonal antibody with human
constant (C) regions creating hybrid (chimeric) antibody molecules
(1, 15). Chimeric antibodies can be reproducibly generated
in virtually unlimited quantities and are homogeneous in specificity
and affinity.
In the present study, we developed mouse-human chimeric IgM and IgG1
antibodies specific for Toxoplasma gondii and evaluated their performance as calibrators in the Abbott IMx Toxo IgM and IMx
Toxo IgG assays (13, 16).
Cell lines.
The hybridomas 1-706-139 (H1) and 5-465-210 (H5)
were established by fusion of immunized Swiss Webster spleen cells to
Sp2/0-Ag14 cells (Abbott Laboratories, Abbott Park, Ill.; American Type
Culture Collection, Rockville, Md.).
Isolation of immunoglobulin V regions.
PCR cloning of
immunoglobulin V regions was performed as previously described
(6) with primers designed by Jones and Bendig (9)
with altered restriction sites (Table 1).
M-IgG2b and M-IgG2a were used for specific cDNA synthesis of
heavy-chain V regions from H1 and H5, respectively. MK-REV was used for
cDNA priming of kappa (
0095-1137/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.
Recombinant Mouse-Human Chimeric Antibodies as
Calibrators in Immunoassays That Measure Antibodies to
Toxoplasma gondii
and
![]()
ABSTRACT
Top
Abstract
Introduction
Materials & Methods
Results
Discussion
References
0.985.
These data demonstrate that chimeric mouse-human antibodies are a
viable alternative to high-titer positive human plasma for the
manufacture of calibrators and controls for diagnostic assays.
![]()
INTRODUCTION
Top
Abstract
Introduction
Materials & Methods
Results
Discussion
References
![]()
MATERIALS AND METHODS
Top
Abstract
Introduction
Materials & Methods
Results
Discussion
References
) light-chain variable regions. H1
VH and V
regions were amplified with primer
combinations MHV-9/M-IgG2b and MKV-1/MK-REV, respectively. For H5,
VH and V
regions were amplified with
MHV-7/M-IgG2a and MKV-2/MK-REV, respectively. Amplifications consisted
of 30 cycles of 94°C for 30 s, 52°C for 30 s, and 72°C
for 1 min. PCR-derived products were isolated by the Promega (Madison,
Wis.) Magic PCR Preps DNA purification system, digested
(SalI and BglII), and cloned into pUC18 (BRL Life
Technologies, Gaithersburg, Md.).
TABLE 1.
Oligonucleotides used to generate chimeric
antibody constructs
Chimeric IgM expression constructs.
The immunoglobulin
expression vector pdHL2 (5) (Fig.
1A) was modified by replacing the human
IgG1 constant region (C
1) with a human IgM constant
region (Cµ). First, the oligonucleotides 5'
GGAACTAGTGGAGC 3' and 5' TCCACTAGTTCCGC 3' were
annealed and cloned into the SacII site in pdHL2 (upstream
of C
1) to introduce a SpeI site. Next, the
secretory portion of the Cµ gene (membrane exons
excluded) was PCR amplified by using 1 ng of a plasmid containing a
genomic clone of human Cµ (pN ·
-µTNP DNA;
provided by Marc Shulman, University of Toronto, Toronto, Ontario,
Canada) as template, 50 pmol each of primers A (Table 1)
(SpeI cloning site) and B (EcoRV cloning site),
and 24 cycles of 94°C for 30 s, 63°C for 30 s, and 72°C
for 90 s. Gel-isolated Cµ product (spanning from 128 nucleotides [nt] upstream of Cµ1 exon to 152 nt
downstream of the polyadenylation signal after Cµ4 [positions 17 to 2278 in reference 20]) was ligated
into pGEM-T (Promega) and sequenced. Finally, the
SpeI-EcoRV fragment containing Cµ
was excised and cloned into SpeI-PvuII-digested
pdHL2, replacing C
1.
|
and
VH are shown in Table 1. The H5 V
and
VH gene fragments were amplified with primer combinations
VK5-5'/VK5-3' and VH5-5'/VH5-3', respectively. The 100-µl-volume
reaction mixtures included 50 pmol of each primer and 1 ng of plasmid
containing a V-region gene (see above). Amplification consisted of 22 cycles of 94°C for 30 s, 62°C for 30 s, and 72°C for 1 min. V
product was digested with
XbaI-BamHI and cloned into the vector.
Subsequently, VH product was introduced into the
V
-containing vector by using
XhoI-HindIII, generating pJH2-24-95B1 (Fig.
1B).
The H1 V
and VH gene transfer was performed
in the same manner by using primer combinations VK1-5'/VK1-3' and
VH1-5'/VH1-3', respectively, yielding pJH3-19-95A.
Transfer of V regions into the IgG1 expression vector pdHL2.
An XbaI-HindIII fragment containing both
V
and VH was excised from each of the IgM
vectors and introduced into
XbaI-HindIII-digested pdHL2, yielding
pJH9-14-94/4.1 (H1 set) and pdHL-2/5-465 (H5 set).
Sequencing. The plasmid template for sequencing was prepared with Magic Miniprep or Maxiprep DNA purification systems (Promega), the EasyPrep Plasmid Prep kit (Pharmacia), or by CsCl banding. Nucleotide sequences were determined manually by using Sequenase version 1.0 with a Sequenase 7-deaza-dGTP DNA sequencing kit (Amersham Life Science, Inc., Arlington Heights, Ill.) or by automated sequencing with the Pharmacia AutoRead sequencing kit and ALF DNA sequencer.
Transfection of murine myeloma cells. Sp2/0-Ag14 cells were transfected by electroporation as previously described (6). IgM constructs were transfected by using a modified version of the protoplast fusion technique (19). Briefly, Sp2/0-Ag14 cells were resuspended in the protoplast suspension, transferred to a 60-mm-diameter dish, and centrifuged at 650 × g for 8 min. The supernatant was aspirated, and 1.5 ml of 50% (wt/vol) polyethylene glycol (Sigma Hybri-Max) in phosphate-buffered saline (PBS) prewarmed to 37°C was added. The dish was centrifuged at 110 × g until 90 s had elapsed after polyethylene glycol addition. The cells were resuspended by pipetting in two 5-ml volumes and one 10-ml volume of prewarmed wash solution (Dulbecco Modified Eagle medium [DMEM] supplemented with 1% fetal bovine serum, 50 U of penicillin per ml, and 50 µg of streptomycin per ml), which were added to a 50-ml centrifuge tube containing 15 ml of the wash solution. After centrifugation at 225 × g for 7.5 min, the cells were resuspended in plating medium (DMEM supplemented with 10% fetal bovine serum, 50 U of penicillin per ml, 50 µg of streptomycin per ml, and 100 µg of kanamycin per ml [Sigma]), plated in one 96-well dish, and incubated at 37°C. After 24 h (day 1), 100 µl of selective medium (plating medium supplemented with 0.1 µM methotrexate [MTX; Adria Laboratories, Columbus, Ohio]) was added. On day 5, 50 µl of the selective medium was added per well. After 2 days, 100 µl was removed from each well and 100 µl of fresh selective medium was added per well. MTX-resistant colonies were tested for secretion of chimeric antibody by an enzyme-linked immunosorbent assay (ELISA). Transfectants secreting chimeric antibody were cloned by limiting dilution and passaged in medium containing 0.1 µM MTX.
Assays for chimeric antibody production. Transfectants were assayed by ELISA for production of chimeric IgG1 as previously described (6). To assay for chimeric IgM, plates were coated with 0.36 µg of goat anti-human IgM (Fc5µ-specific; Jackson ImmunoResearch, West Grove, Pa.) per µl in Dulbecco's phosphate-buffered saline (D-PBS). Peroxidase-labeled goat anti-human IgM (Fc5µ; Jackson ImmunoResearch) at 0.8 µg/ml was used as the enzyme antibody conjugate, with ChromPure human IgM (Jackson ImmunoResearch) as a standard.
Antibody concentrations were determined by radial immunodiffusion (RID). Seven- to 14-day cell culture supernatants were vortexed, and then 5- and 10-µl samples for IgG and IgM, respectively, were applied to each well of human IgG or human IgM RID plates (The Binding Site, San Diego, Calif.). Standards (25, 50, 75, 100, and 150 µg/ml) were run on each plate. The IgM and IgG1 standards were ChromPure human IgM and protein A-purified anti-P66 IgG1, respectively. Plates were incubated in an inverted position at 35 to 37°C for 16 to 22 h. Ring diameter was determined by using a calibrated RID electronic plate reader (The Binding Site).Purification of chimeric antibodies. To purify chimeric IgG1 antibodies, culture supernatants were dialyzed overnight against 0.1 M sodium phosphate (pH 8.2) with 0.1% sodium azide, filtered (0.2-µm-pore-size filter), and passed through a column containing PerSeptive Biosystems (Cambridge, Mass.) Poros 50A resin by using a Bio-Rad low-pressure chromatography system. Antibody was eluted with 0.1 M citrate-0.15 M NaCl (pH 3.0), and pooled fractions were dialyzed against PBS (10 mM sodium phosphate, 150 mM NaCl [pH 7.2]).
To purify chimeric IgM antibodies, culture supernatants were diluted 1:1 with H2O and loaded on a column containing 175 ml of PBS-equilibrated DEAE FastFlow resin (Pharmacia). Antibody was eluted with 10 mM sodium phosphate-300 mM NaCl at pH 7.2.IMx evaluation of chimeric calibrators. Reactivity of chimeric IgG1 and IgM antibodies was measured with the Abbott IMx Toxo IgG 2.0 antibody assay (Abbott Laboratories) and the Abbott IMx Toxo IgM 2.0 antibody assay, respectively. Testing was done as described in the assay package insert, except where chimeric antibodies were substituted for calibrators in the assay. Reactivity rates are expressed as counts per second per second.
Chimeric IgG and IgM antibodies were diluted into anti-Toxoplasma antibody-negative plasma to evaluate their reactivity in the assays and to prepare calibrator sets. Point-to-point data reduction was used to generate calibration curves and determine patient specimen IgG values. Index values in the IgM assay are generated by dividing the patient specimen assay rate by the index calibrator assay rate.Nucleotide sequence accession numbers.
GenBank accession
numbers for V region sequences are as follows: for 1-706-139 V
, AF031633; for 1-706-139 VH, AF031634; for
5-465-210 V
, AF031635; and for 5-465-210 VH,
AF031636.
| |
RESULTS |
|---|
|
|
|---|
Selection of monoclonal antibodies. A panel of murine monoclonal antibodies specific for a variety of T. gondii antigens was established. Two with potential utility as calibrators and controls were selected based on a combination of antigen specificity and performance in competitive binding studies with pooled human plasma reactive to T. gondii. The monoclonal antibody H5 is specific for the major surface antigen of T. gondii, P30, which is thought to be a key antigen for identification of individuals infected with T. gondii. A second monoclonal antibody, H1, reactive with P66, was also evaluated since this appears to be a potentially important protein for measurement of the IgM response to T. gondii (unpublished observations). In competitive binding experiments with the IMx Toxo IgM index calibrator, each of these monoclonal antibodies showed some inhibition of human IgM antibody binding (data not shown).
Cloning and sequencing of antibody variable regions.
To
generate chimeric mouse-human antibodies, the H- and L-chain V regions
were cloned from H5 and H1. The V genes were amplified by PCR using a
combination of degenerate primers annealing to conserved VH
and V
gene leader sequences and C region-specific primers (9). Multiple clones of each V gene product were
sequenced to monitor for AmpliTaq DNA polymerase-induced errors.
gene from the murine subgroup
III (10), and a J
2 minigene. The
J
2 segment contains four mismatches relative to the germ
line sequence of J
2 isolated from the BALB/c mouse
strain (10). The observed mismatches may reflect somatic mutations or represent differences in the germ line sequence since the
hybridoma was generated from a Swiss Webster strain mouse.
|
gene from
the murine subgroup II (10), and a J
1
minigene.
|
Construction of mouse-human chimeric antibodies. The VH and VL gene fragments were modified to form cassettes (Fig. 2 and 3) suitable for expression with any C region containing a functional splice acceptor site. As a result of RNA processing, the V and C regions are fused in the cDNA clones. Therefore, adaptation to the cassette format involved recreation of a splice donor site at the end of the V region and addition of flanking cloning sites. A leader sequence was already present in the cDNA clones. However, in cases where degenerate positions were present in the amplification primers, the sequence was modified, if necessary, based on a comparison to leader sequences in the database.
To produce chimeric mouse-human IgG1, the V gene cassettes were cloned into the immunoglobulin expression vector pdHL2 (5) (Fig. 1A). This vector contains genomic clones of the human kappa (C
) and IgG1 (C
1) C-region genes. Both
loci are controlled by a metallothionein I promoter and a mouse
immunoglobulin H-chain enhancer. An altered dihydrofolate reductase
gene serves as a selectable marker. Introduction of the murine
VH and V
regions results in expression of
chimeric mouse-human IgG1 antibody.
To produce chimeric mouse-human IgM antibodies, pdHL2 was modified by
replacement of the C
1 gene with a genomic clone of the
human IgM (Cµ) gene. Sequence analysis of the
Cµ clone revealed only a single base change relative to
the published germ line sequence (20), the insertion of a T
between residues 1583 and 1584 in the
Cµ3-Cµ4 intron. This alteration would not
be expected to affect expression of the Cµ gene since it
is in the intron. The H1 and H5 VH and V
cassettes were cloned into the IgM expression vector to generate
chimeric IgM antibody specific for P66 and P30, respectively.
Expression of IgM and IgG chimeric antibodies. The chimeric antibody constructs were transfected by electroporation into Sp2/0-Ag14 cells. Multiple MTX-resistant colonies from each transfection were screened for production of chimeric antibody by using an anti-human antibody ELISA. Transfectants that tested positive were expanded and cloned. Clones secreting high levels of chimeric antibody were further subcloned. The observed production levels for the anti-P30 and anti-P66 chimeric IgG1 antibodies were 90 µg/ml or greater. In contrast, IgM transfectants established by electroporation secreted very low levels of antibody. Protoplast fusion is an alternative method for transfection (1, 19) that often results in introduction of higher copy numbers of vector. The IgM constructs were introduced into Sp2/0-Ag14 cells by protoplast fusion in an effort to enhance expression. A clone secreting anti-P30 chimeric IgM antibody that produces 115 µg/ml was isolated. Transfection of the anti-P66 chimeric IgM construct met with more limited success. Transfectants secreting chimeric IgM were obtained, but secretion levels were low (<1 µg/ml). In all but one case, production of recombinant chimeric antibody by the transfectants greatly exceeded that of the hybridomas from which they were derived. Analysis of the IgM chimeric antibody revealed that it is pentameric.
Assay performance of the chimeric antibodies. Purified chimeric IgG1 and IgM antibodies were examined for immunoreactivity to T. gondii in the IMx Toxo IgG and IMx Toxo IgM assays. These assays utilize microparticles coated with formalin-treated T. gondii as the solid phase to capture T. gondii-specific antibodies. The captured antibodies are detected with anti-human IgG or IgM antibody-labeled conjugates.
Reactivity of purified anti-P30 chimeric IgG1 was compared to the reactivity of human anti-T. gondii antibody in the IMx Toxo IgG assay. The calibrators used in this assay are manufactured from pooled anti-T. gondii IgG-positive plasma and are matched to the World Health Organization International Standard for anti-T. gondii antibody at the following levels: 0, 10, 50, 100, 200, and 300 IU/ml. The F calibrator (300 IU/ml) marks the upper end of the dynamic range of the assay. The anti-P30 chimeric IgG1 antibody at 218 µg/ml was found to give the same signal as the F calibrator in the IMx assay. Twofold serial dilutions of the anti-P30 chimeric IgG1 antibody into negative human serum were tested in duplicate. The observed calibration curve for the chimeric IgG1 antibody was similar to the dilution profile of the human plasma-derived assay calibrator (Fig. 4A). At a concentration of 870 µg/ml, the anti-P30 chimeric IgG1 antibody generated a signal of 4,261 cps/s in the assay.
|
|
Evaluation of patient samples by using kits calibrated with chimeric antibody-derived calibrators. The anti-P30 chimeric IgG1 and IgM antibodies were diluted to match human plasma-derived calibrators to demonstrate the utility of the chimeric antibody as a source of calibrators for the IMx Toxo IgG and IMx Toxo IgM assays. Human plasma-derived calibrators for the IMx Toxo IgG assay ranged from 0 to 300 IU/ml. The anti-P30 chimeric IgG1 antibody was diluted in negative human plasma to match each level of the human plasma calibrators within 2% of IMx assay calibrator rates. Calibration curves obtained for each of the kit calibrators and anti-P30 chimeric IgG1-derived calibrators were similar (Fig. 6A). Thirty-six serum specimens with anti-Toxoplasma IgG concentrations less than 60 IU/ml were evaluated in the IMx Toxo IgG assay using the chimeric antibody calibrators as the source of the standard curve, and the results were compared with those of the same specimens run in the assay using human plasma-derived calibrators. The correlation exhibited an r2 value of 0.985 and a slope of 0.991 (Fig. 6B).
|
|
| |
DISCUSSION |
|---|
|
|
|---|
In the present report, we describe the development and utility of mouse-human chimeric antibodies as calibrating reagents in automated immunoassays measuring human antibodies to T. gondii. Murine monoclonal antibodies specific for P30 (11) and P66 (18) were selected for conversion to mouse-human chimeric antibodies. Functional immunoglobulin V region genes were cloned from the hybridoma cells and transferred into expression vectors containing human constant-region genes to generate recombinant mouse-human chimeric IgM and IgG1 antibodies. In all cases, the recombinant chimeric antibodies retained their native specificity.
Analysis of the anti-P30 IgG1 chimeric antibody revealed that it was fully functional in the Toxo IgG immunoassay designed to measure human IgG antibody to T. gondii. The signal generated by the anti-P30 IgG1 chimeric antibody at a concentration of 218 µg/ml matched that of the F calibrator which is set at the high end of the dynamic range of the Toxo IgG assay. Moreover, the calibration curve obtained with serial dilutions of the anti-P30 chimeric IgG1 antibody was parallel to that of the human plasma-derived assay calibrator (Fig. 4A). Such parallelism between dilution curves of the test and reference (calibrator) antibodies is required to provide accurate quantitation of specific antibody (3, 8). To establish the utility of the chimeric calibrator, 36 T. gondii-reactive clinical specimens were run in the IMx Toxo IgG assay using the kit calibrators and the resulting values were compared to the values obtained with the anti-P30 IgG1 chimeric antibody calibrators. The observed correlation between the two types of calibrators was excellent. Based on these data, the anti-P30 IgG1 chimeric antibody provides an excellent alternative to the use of positive human plasma in the manufacture of calibrators.
In contrast, although the anti-P66 IgG1 chimeric antibody was reactive in the IMx Toxo IgG assay, it did not reach the maximum rate read of the positive human plasma calibrators. Attempts to increase the signal generated by the chimeric antibody by increasing antibody input concentration were unsuccessful. This suggests that the P66 antigen on the surface of the microparticles is limiting. Unless the available P66 epitope were to be increased, the anti-P66 IgG1 chimeric antibody alone does not appear to be a viable alternative for manufacture of calibrators.
Since the Toxo IgG calibrator is made with polyclonal T. gondii-reactive human plasma, it would be expected to contain antibodies against many different T. gondii proteins (17, 18) and to multiple epitopes on each antigen. Thus, the observation that the signal of a monoclonal reagent specific for one epitope of a single T. gondii protein is unable to match the cumulative signal generated by the polyclonal calibrator reagent is not necessarily unexpected. It is conceivable that blending of two or more monoclonal reagents may better mimic the polyclonal human antibody calibrator and more closely approximate the routine human test samples. In preliminary tests where the anti-P66 IgG1 chimeric antibody was combined with the anti-P30 chimeric IgG1, a signal exceeding the high end of the dynamic range (F calibrator) was achievable and the dilution profile was similar to that of the human plasma-derived Toxo IgG calibrator (data not shown). The concept of utilizing more than one chimeric antibody to generate a calibrator might become particularly important if one is measuring a response to more than one epitope of an antigen or to more than one antigen when considering issues such as the differential stability of antigen reagents in immunoassay kits. If the epitope recognized by one of the chimeric antibodies was impaired, a second epitope may be required to accurately calibrate the assay. Determination of whether it is advantageous to pool more than one chimeric antibody reagent will be dependent upon the desired performance characteristics for the calibrator and can be determined empirically.
As opposed to the quantitative Toxo IgG assay, the Toxo IgM assay is a qualitative assay that utilizes a single index calibrator to determine the cutoff value. When examined in the Toxo IgM assay, acceptable calibration curves were obtained independently for both the anti-P30 and anti-P66 IgM chimeras. The feasibility of using a chimeric antibody calibrator was examined by comparing the results obtained for 31 T. gondii-reactive specimens in the Toxo IgM assay to those obtained by using the anti-P30 chimeric IgM index calibrator. There was a high correlation between the results generated with the human plasma-derived and chimeric IgM antibody calibrators. These data demonstrate that the anti-P30 IgM chimeric antibody can be substituted for T. gondii-reactive human plasma in the manufacture of an index calibrator.
An advantage of chimeric antibody technology in the manufacture of calibrating reagents is its flexibility. One can readily screen candidate monoclonal antibodies for desired properties such as specificity and affinity prior to the more labor-intensive steps of cloning the V regions and production of chimeric antibody. Binding properties of the chimeric antibody should reflect those of the original monoclonal antibody. In the present study, the choice of antibody specificity was based on competitive inhibition studies using T. gondii-reactive human plasma in an effort to identify monoclonal antibodies reacting with immunodominant epitopes. Presumably, the more closely the epitope reactivity of the chimeric antibodies and human test samples are matched, the better the calibrator. In some cases, this may require the use of more than one chimeric antibody recognizing different antigens or of multiple epitopes on the same antigen. Ideally, the affinity of the chimeric antibodies would be similar to that of the antibodies being monitored in the test samples. Other factors such as epitope stability (i.e., after heat stress or prolonged storage) can also be examined prior to conversion of the monoclonal antibody to the chimeric format.
In addition to epitope specificity, proper selection of heavy-chain isotype is also important. In most cases, immunoassays that monitor specific antibody are designed to detect antibody classes or subclasses with in vivo or diagnostic relevance. In patients acutely infected with T. gondii, IgM is the predominant class of antibody, whereas in patients with acquired or reactivated toxoplasmosis, IgG1, IgG2, IgG3, and IgA antibodies can be detected (4). The IMx Toxo IgG assay uses a polyclonal anti-human IgG as the conjugate reagent. We found that the chimeric IgG1 antibody calibrates the IMx assay accurately over multiple lots of the polyclonal anti-human IgG reagent (data not shown).
The use of chimeric antibodies in preparation of calibrators and positive controls for immunoassays offers several substantial advantages over traditional methods utilizing high-titer plasma or serum, as well as other alternatives such as chemical coupling of human IgG Fc' fragments to a murine monoclonal antibody (14). First, the chimeric antibody can be produced continuously with the same affinity and specificity. Reduction in lot-to-lot variability over time with respect to antibody class composition, titer, specificity, and affinity would be expected to yield a consistent calibrator that should give the most reproducible patient results. Second, virtually unlimited quantities of the chimeric antibodies can be generated. Recombinant cell lines can stably produce chimeric antibodies at manufacturable levels and at a reasonable cost. Finally, the homogeneous nature of the chimeric antibodies allows for better characterization.
Mouse-human chimeric antibodies have previously been shown to be useful as quality control reagents and for quantitation of specific antibody in reference standards by heterologous interpolation (7, 8). The present study establishes the feasibility of using mouse-human chimeric antibodies as calibrator reagents in automated immunoassays for the measurement of human antibodies specific for T. gondii. This is the first demonstration of the utility of chimeric antibody calibrators for homologous interpolation of specific human antibody levels. Chimeric antibody calibrators should be applicable to any diagnostic assay or kit designed to detect the presence of human antibodies specific for a given antigen (e.g., infectious agent, autoantigen, allergen, pharmaceutical).
| |
ACKNOWLEDGMENTS |
|---|
We thank Robert Ziemann, Jeffrey Alder, Joan Tyner, Lawrence Howard, and Michael Sheu for generation, screening, and characterization of the monoclonal antibodies; Odin Cabal for assistance with the protoplast fusions; and Richard Thomas for purification of the chimeric antibodies.
| |
FOOTNOTES |
|---|
* Corresponding author. Mailing address: Abbott Laboratories, AIDS Research and Retrovirus Discovery, D-9NG, Bldg. R1, 1401 Sheridan Rd., North Chicago, IL 60064. Phone: (847) 938-0457. Fax: (847) 937-1401. E-mail: john.hackett{at}add.ssw.abbott.com.
Present address: Hospital Products Division, Abbott Laboratories,
Abbott Park, Ill.
Present address: DGI Biotechnologies, Edison, N.J.
| |
REFERENCES |
|---|
|
|
|---|
| 1. | Boulianne, G. L., N. Hozumi, and M. J. Shulman. 1984. Production of functional chimaeric mouse/human antibody. Nature 312:643-646[Medline]. |
| 2. | Butler, J. E. (ed.). 1991. Immunochemistry of solid-phase immunoassay. CRC Press, Inc., Boca Raton, Fla. |
| 3. | Butler, J. E., and R. G. Hamilton. 1991. Quantitation of specific antibodies: methods of expression, standards, solid-phase considerations, and specific applications, p. 173-198. In J. E. Butler (ed.), Immunochemistry of solid-phase immunoassay. CRC Press, Inc., Boca Raton, Fla. |
| 4. | Derouin, F., G. Sulcebe, and J. J. Ballet. 1987. Sequential determination of IgG subclasses and IgG specific antibodies in primary and reactivating toxoplasmosis. Biomed. Pharmacother. 41:429-433[Medline]. |
| 5. | Gillies, S. D., K.-M. Lo, and J. Wesolowski. 1989. High-level expression of chimeric antibodies using adapted cDNA variable region cassettes. J. Immunol. Methods 125:191-202[Medline]. |
| 6. | Hackett, J., Jr., J. Hoff-Velk, A. Golden, C. Dealwis, D. Ostrow, and W. Mandecki. 1997. The effect of site-specific mutagenesis of a cysteine residue on the stability of a monoclonal antibody, p. 133-157. In W. Hori (ed.), Antibody engineering II, vol. 2. New technology, application and commercialization. International Business Communications, Inc., USA, Southborough, Mass. |
| 7. | Hamilton, R. G. 1990. Engineered human antibodies as immunologic quality control reagents. Ann. Biol. Clin. 48:473-477. |
| 8. | Hamilton, R. G. 1991. Application of engineered chimeric antibodies to the calibration of human antibody standards. Ann. Biol. Clin. 49:242-248. |
| 9. | Jones, S. T., and M. M. Bendig. 1991. Rapid PCR-cloning of full-length mouse immunoglobulin genes. Bio/Technology 9:88-89[Medline]. |
| 10. | Kabat, E. A., T. T. Wu, H. M. Perry, K. S. Gottesman, and C. Foeller. 1991. Sequences of proteins of immunological interest, 5th ed. U.S. Department of Health and Human Services, Bethesda, Md. |
| 11. | Kasper, L. H., J. H. Crabb, and E. R. Pfefferkorn. 1983. Purification of a major membrane protein of Toxoplasma gondii by immunoabsorption with a monoclonal antibody. J. Immunol. 130:2407-2412[Abstract]. |
| 12. | Köhler, G., and C. Milstein. 1975. Continuous cultures of fused cells secreting antibody of predefined specificity. Nature 256:495-497[Medline]. |
| 13. | Liesenfeld, O., C. Press, R. Flanders, R. Ramirez, and J. S. Remington. 1996. Study of Abbott Toxo IMx system for detection of immunoglobulin G and immunoglobulin M toxoplasma antibodies: value of confirmatory testing for diagnosis of acute toxoplasmosis. J. Clin. Microbiol. 34:2526-2530[Abstract]. |
| 14. | Miyachi, J., K. Doi, K. Kitamura, T. Jitsukawa, and H. Watanabe. 1992. Chemically humanized murine monoclonal antibody against a cell nuclear antigen: usefulness in autoimmune diagnostics. J. Clin. Lab. Anal. 6:343-350[Medline]. |
| 15. |
Morrison, S. L.,
M. J. Johnson,
L. A. Herzenberg, and V. T. Oi.
1984.
Chimeric human antibodies: mouse antigen binding domains with human constant regions.
Proc. Natl. Acad. Sci. USA
81:6851-6855 |
| 16. |
Safford, J. W.,
G. G. Abbott,
M. C. Craine, and R. G. MacDonald.
1991.
Automated microparticle enzyme immunoassays for IgG and IgM antibodies to Toxoplasma gondii.
J. Clin. Pathol.
44:238-242 |
| 17. | Santoro, F., H. Charif, and A. Capron. 1986. The immunodominant epitope of the major membrane tachyzoite protein (P30) of Toxoplasma gondii. Parasite Immunol. 8:631-639[Medline]. |
| 18. | Sharma, S. D., J. Mullenax, F. G. Araujo, H. A. Erlich, and J. S. Remington. 1983. Western blot analysis of the antigens of Toxoplasma gondii recognized by human IgM and IgG antibodies. J. Immunol. 131:977-983[Abstract]. |
| 19. | Shin, S., and S. L. Morrison. 1989. Production and properties of chimeric antibody molecules. Methods Enzymol. 178:459-476[Medline]. |
| 20. |
Word, C. J.,
M. B. White,
W. A. Kuziel,
A. L. Shen,
F. R. Blattner, and P. W. Tucker.
1989.
The human immunoglobulin Cµ-C locus: complete nucleotide sequence and structural analysis.
Int. Immunol.
1:296-309 |
This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»