Journal of Clinical Microbiology, June 1998, p. 1601-1603, Vol. 36, No. 6
Division of Laboratory Medicine, Washington
University, St. Louis, Missouri1;
Microbiology Laboratory, Rochester General Hospital,
Rochester, New York2;
DeKalb Medical
Center, Decatur, Georgia3; and
Denver Health Medical Center, Denver,
Colorado4
Received 2 December 1997/Accepted 11 March 1998
The overall recovery of organisms and time to detection with the
BACTEC 9050 and BACTEC 9240 systems were compared in a multicenter evaluation. In the first phase of the study, a total of 4,383 compliant
aerobic (Plus Aerobic/F) blood culture sets were processed. There was
no significant difference in the recovery of individual groups of
organisms with the two systems, with the exception of Streptococcus pneumoniae which was isolated more frequently
with BACTEC 9050. False-positive signals occurred more often with
BACTEC 9240 (58 cultures) than with BACTEC 9050 (43 cultures), but
false-negative cultures were uncommon with both systems (3 cultures for
each system). Time to detection of positive cultures of clinically significant organisms was essentially the same with both instruments. In the second phase of the study, 2,431 compliant anaerobic (Plus Anaerobic/F) blood culture sets were processed. There was no
significant difference in the recovery of organisms with BACTEC 9050 compared with BACTEC 9240. Significantly (P < 0.03)
more false-positive signals occurred with BACTEC 9240 (15 cultures)
than with BACTEC 9050 (4 cultures). Likewise, more false-negative
cultures occurred with BACTEC 9240 (11 cultures) than with BACTEC 9050 (8 cultures). Time to detection of positive cultures of clinically
significant organisms was essentially the same with both systems with
the exception of anaerobes (N = 10), which were
recovered earlier (P < 0.01) with BACTEC 9240 (35.0 h) than with BACTEC 9050 (61.4 h).
The BACTEC 9000 series of blood
culture systems (Becton Dickinson Microbiology Systems, Sparks, Md.)
are fluorogenic, automated, noninvasive blood culture systems. Two
models, BACTEC 9240 (240-bottle capacity) and BACTEC 9120 (120-bottle
capacity), are essentially identical in design and differ only in their
bottle capacity. Recently, a third model was introduced for use in
small laboratories or clinics that process relatively few blood
cultures. BACTEC 9050, with a capacity of 50 bottles, is a small,
self-contained, automated system designed for processing three to five
blood cultures per day. In addition to the difference in capacity,
BACTEC 9050 differs from the larger systems by agitating the bottles
continuously (versus intermittently for the other systems) and rotating
the bottles to be read by one of three detectors (versus a dedicated detector for each stationary bottle in the large systems). The computer
used to monitor the BACTEC 9050 bottles is contained within the system.
Previous evaluations of the larger BACTEC 9000 instruments demonstrated
excellent recovery and time of detection of positive blood cultures
(1, 2, 4-7). In this evaluation, the performance properties
of BACTEC 9050, using Plus Aerobic/F and Plus Anaerobic/F blood culture
bottles, were compared with those of BACTEC 9240.
Study participants.
This study was performed in four
clinical laboratories: Barnes-Jewish Hospital in St. Louis, Mo.;
Rochester General Hospital in Rochester, N.Y.; DeKalb Medical Center,
Decatur, Ga.; and Denver Health Medical Center, Denver, Colo.
Study design.
The performance of the BACTEC 9050 system was
compared with that of the BACTEC 9240 system. In phase 1 of the study,
the performance of Plus Aerobic/F medium was evaluated in the two
systems. A total volume of 14 to 22 ml of blood was collected
aseptically, divided equally into two Plus Aerobic/F bottles, and
transported to the study laboratories. One bottle was placed into each
instrument after it was determined that an equal volume of 7 to 11 ml
of blood was inoculated into each bottle. Compliance was determined by
measuring blood volumes when the bottles were received in the laboratories. All bottles were incubated for a minimum of 5 days according to the manufacturer's protocol. When a positive signal was
obtained, the bottles were removed and an aliquot of the broth was Gram
stained and processed for organism identification. Bottles with a
false-positive signal were returned to the system for further incubation and testing. All negative bottles were subcultured blindly
to chocolate agar plates and incubated aerobically at the end of the
incubation period.
0095-1137/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.
Multicenter Comparison of BACTEC 9050 and BACTEC
9240 Blood Culture Systems
![]()
ABSTRACT
Top
Abstract
Introduction
Materials & Methods
Results & Discussion
References
![]()
INTRODUCTION
Top
Abstract
Introduction
Materials & Methods
Results & Discussion
References
![]()
MATERIALS AND METHODS
Top
Abstract
Introduction
Materials & Methods
Results & Discussion
References
Data analysis. The modified chi-square test described by McNemar (3) was used to assess the statistical significance of differences observed between the two culture systems. When appropriate, the Yates correction for small numbers of observations was used.
| |
RESULTS AND DISCUSSION |
|---|
|
|
|---|
In the phase 1 aerobic comparison, 4,382 compliant blood culture sets were received and processed. A total of 585 (13.4%) positive cultures were detected with one or both BACTEC systems, a false-positive result (positive result with the BACTEC system but negative results for Gram stain and subculture) was recorded for 83 (1.9%) cultures, and 3,714 (84.7%) cultures had negative results. From the positive culture bottles, 651 organisms were isolated, including 366 judged to be clinically significant, 242 clinically insignificant, and 43 of unknown significance. Overall recovery of clinically significant organisms in the BACTEC 9050 and BACTEC 9240 systems is summarized in Table 1. The total numbers of organisms isolated with the BACTEC 9050 and BACTEC 9240 systems were 320 and 303, respectively. There was no significant difference in the recovery of individual groups of organisms in the two systems, with the exception of Streptococcus pneumoniae. Of the 29 cultures positive with S. pneumoniae, 8 were detected only with BACTEC 9050 and 1 was detected only with BACTEC 9240 (P < 0.05).
|
Equal numbers of clinically insignificant organisms were recovered with the two systems (144 isolates with each system), with the coagulase-negative staphylococci and Corynebacterium species the most common contaminants. A total of 43 organisms of unknown clinical significance was isolated in this phase of the study, with 30 and 26 recovered with the BACTEC 9050 and BACTEC 9240 systems, respectively.
False-positive results were observed in 83 cultures, including 17 cultures with both bottles positive, 40 cultures positive only with the BACTEC 9240 system, and 26 cultures positive only with the BACTEC 9050 system.
All negative blood culture bottles were blindly subcultured as described above. An additional six positive bottles, all with isolates of clinically significant organisms, were detected. Three BACTEC 9240 bottles were positive with Candida glabrata (the companion BACTEC 9050 bottles were negative), and three BACTEC 9050 bottles were positive for individual isolates of C. glabrata, Corynebacterium jeikeium, and Streptococcus constellatus (the companion BACTEC 9240 bottles were positive for C. jeikeium and S. constellatus but not for C. glabrata).
In the second phase of the study, 2,431 compliant blood culture sets were processed in the Plus Anaerobic/F medium. A total of 284 (11.7%) positive cultures were detected, a false-positive result was observed for 19 (0.8%) cultures (including 3 cultures with a true-positive companion bottle), and 2,131 (87.7%) cultures had negative results. From the positive culture bottles, 316 organisms were recovered, including 197 clinically significant isolates, 101 clinically insignificant isolates, and 18 organisms of unknown significance. The recovery of clinically significant organisms in the two systems is summarized in Table 2. There was no significant difference in the overall recovery of organisms with BACTEC 9050 (total recovery, 177 isolates) or BACTEC 9240 (170 isolates) or in the recovery of any individual group of organisms.
|
A total of 101 clinically insignificant organisms were recovered in this phase of the study, 60 with BACTEC 9050 and 61 with BACTEC 9240. In addition, the recovery of organisms with unknown clinical significance with the two systems was not significantly different (i.e., 14 with BACTEC 9050 and 10 with BACTEC 9240).
A total of 19 false-positive cultures were observed in the second phase of the study. There were 15 false-positive cultures with the BACTEC 9240 system and 4 with the BACTEC 9050 system (P < 0.03).
A total of 19 organisms were detected in 15 cultures by blind subculture. Eleven organisms including four clinically significant ones (two Clostridium spp., one Pseudomonas aeruginosa, and one unidentified anaerobic gram-negative bacillus) were present in BACTEC 9240 cultures and eight organisms including two clinically significant ones (one P. aeruginosa isolate and one Staphylococcus aureus isolate) were present in BACTEC 9050 cultures. Only two clinically significant isolates (one S. aureus and one Clostridium spp.) were detected by a positive signal in the companion bottle.
The time to detect positive cultures of clinically significant organisms is summarized in Tables 3 and 4. The time for detection using the Plus Aerobic/F bottles was essentially the same for the BACTEC 9050 system (mean detection time, 19.3 h) and the BACTEC 9240 system (19.1 h). All groups of organisms were detected rapidly, with the exception of the coagulase-negative staphylococci, Corynebacterium spp., yeasts, and Mycobacterium fortuitum (one isolate only). It is interesting to note that the time for detecting growth of yeasts was influenced by the species, with relatively rapid detection observed for Candida albicans (five isolates; mean detection time, 49.2 h with BACTEC 9050 and 44.0 h with BACTEC 9240) and Candida tropicalis (two isolates; 25.2 and 33.4 h) and slower detection of C. glabrata (two isolates; 134.1 and 149.6 h) and Cryptococcus neoformans (two isolates; 79.2 and 78.8 h).
|
|
The time for detection using the Plus Anaerobic/F bottles was similar for both systems (i.e., 17.8 h with the BACTEC 9050 system and 15.7 h with the BACTEC 9240 system). The observed difference in overall detection time was attributed to the slower recovery of five isolates of non-Bacteroides anaerobic bacteria (P < 0.01) with BACTEC 9050 (mean detection time, 81.7 h) than with BACTEC 9240 (mean, 30.5 h).
In summary, the overall recovery of organisms and time for detection of positive cultures with the BACTEC 9050 and 9240 blood culture systems were essentially the same. Minor differences between the two systems include better recovery of S. pneumoniae, fewer false-positive aerobic and anaerobic bottles, and fewer false-negative anaerobic bottles with the BACTEC 9050 system, and faster detection of anaerobic bacteria in anaerobic bottles incubated in the BACTEC 9240 system. During this evaluation, no technical failures were observed with the multiple BACTEC 9050 instruments used in the four study sites, and all study participants found the system to be easy to use. The BACTEC 9050 system is well suited for laboratories that process relatively few blood culture bottles and should offer significant advantages over the manual or older semiautomated blood culture systems that are used in these laboratories.
| |
ACKNOWLEDGMENTS |
|---|
We thank the technologists in our laboratories for their assistance.
Support for this study was provided by Becton Dickinson Microbiology Systems.
| |
FOOTNOTES |
|---|
* Corresponding author. Mailing address: Department of Pathology, Washington University School of Medicine, 660 S. Euclid Ave., St. Louis, MO 63110. Phone: (314) 362-1547. Fax: (314) 362-1308. E-mail: murray{at}labmed.wustl.edu.
| |
REFERENCES |
|---|
|
|
|---|
| 1. | Cockerill, F. R., G. S. Reed, J. G. Hughes, C. A. Torgerson, E. A. Vetter, W. S. Harmesen, J. C. Dale, G. D. Roberts, D. M. Ilstrup, and N. K. Henry. 1997. Clinical comparison of BACTEC 9240 Plus Aerobic/F resin bottles and the Isolator aerobic culture system for detection of bloodstream infections. J. Clin. Microbiol. 35:1469-1472[Abstract]. |
| 2. | Jorgensen, J. H., S. Mirrett, L. C. McDonald, P. R. Murray, M. P. Weinstein, J. Fune, C. W. Trippy, M. Masterson, and L. B. Reller. 1997. Controlled clinical laboratory comparison of BACTEC Plus Aerobic/F resin medium with BacT/Alert aerobic FAN medium for detection of bacteremia and fungemia. J. Clin. Microbiol. 35:53-58[Abstract]. |
| 3. | McNemar, Q. 1962. Psychological statistics, 3rd ed., p. 209-239. John Wiley & Sons, Inc., New York, N.Y. |
| 4. |
Nolte, F. S.,
J. M. Williams,
R. C. Jerris,
J. A. Morello,
C. D. Leitch,
S. Matushek,
L. D. Schwabe,
F. Dorigan, and F. E. Kocka.
1993.
Multicenter clinical evaluation of a continuous monitoring blood culture system using fluorescent-sensor technology (BACTEC 9240).
J. Clin. Microbiol.
31:552-557 |
| 5. | Pohlman, J. K., B. A. Kirkley, K. A. Easley, B. A. Basille, and J. A. Washington. 1995. Controlled clinical evaluation of BACTEC Plus Aerobic/F and BacT/Alert aerobic FAN bottles for detection of bloodstream infections. J. Clin. Microbiol. 33:2856-2858[Abstract]. |
| 6. | Pohlman, J. K., B. A. Kirkley, K. A. Easley, and J. A. Washington. 1995. Controlled clinical comparison of Isolator and BACTEC 9240 Aerobic/F resin bottle for detection of bloodstream infections. J. Clin. Microbiol. 33:2525-2529[Abstract]. |
| 7. | Smith, J. A., E. A. Bryce, J. H. Ngui-Yen, and F. J. Roberts. 1995. Comparison of BACTEC 9240 and BacT/Alert blood culture systems in an adult hospital. J. Clin. Microbiol. 33:1905-1908[Abstract]. |
This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»