Journal of Clinical Microbiology, July 1998, p. 2052-2056, Vol. 36, No. 7
Department of Pathology, University of
Alabama at Birmingham, Birmingham, Alabama,1
and
Dade MicroScan, Inc., West Sacramento,
California2
Received 4 December 1997/Returned for modification 16 March
1998/Accepted 24 April 1998
Studies were conducted on a method of direct inoculation of
MicroScan dried overnight and of rapid panels with positive aerobic blood cultures obtained from the BacT/Alert to determine antimicrobial susceptibilities. Inocula were limited to specimens that appeared unimicrobic on Gram stain. Results were compared to those obtained from
panels inoculated following subculture. For 133 gram-negative bacilli,
there were 94.7 and 93.5% categorical agreements between direct and
standard methods for all drugs tested with overnight and rapid panels,
respectively. For 104 gram-positive cocci, there were 93.2 and 93.1%
categorical agreements for overnight and rapid panels,
respectively. The major error (false resistance) rate for gram
negatives was 1.4% for overnight versus 0.7% for rapid panels. The
very major error (false susceptibility) rate was 2.7% for
overnight versus 8.1% for rapid panels. The total error rates were
1.6% for overnight panels and 1.5% for rapid panels. The major error
rates for gram-positive direct susceptibility tests were 2.6% for
overnight and 2.5% for rapid panels. The very major error rates were
8.8 and 7.2% for overnight and rapid panels, respectively. Total error
rates were 3.6% for overnight and rapid gram-positive panels. These
findings suggest that susceptibility results obtained from directly
inoculated gram-negative overnight panels have the greatest correlation
to those obtained by standard methods. When discrepant results
occur with direct-susceptibility testing, they are more likely to show
false susceptibility than false resistance.
Results of antimicrobial
susceptibility tests are essential to guide clinicians in the
selection of the most appropriate and cost-effective treatment for
persons with bacteremia or other serious infections. Availability of
these data as soon as possible after infection is confirmed may result
in reduced costs for the pharmacy, laboratory, and other general
charges. These data may also lead to timely changes to more effective
therapy in the event the infecting organism is resistant to coverage
provided by empiric antimicrobial choices (3, 14).
Continuously monitoring automated blood culture systems such as the
BacT/Alert (Organon Teknika Corporation, Durham, N.C.) minimize the
length of time required for detection of positive blood cultures from
bacteremic patients, often indicating a positive culture within 24 h following the initial incubation, and have emerged as the new
standard in blood culture technology (16). Conventional
methodology requires instrument-positive bottles to be
subcultured to solid media and incubated overnight to produce
bacterial colonies, which are then used to prepare standardized
suspensions for species identification and susceptibility testing
(8). Direct inoculation from positive blood culture bottles into MicroScan panels (Dade MicroScan, Inc., West
Sacramento, Calif.) is a potential alternative to decrease
further the length of time from initial inoculation of blood culture
media to reporting of susceptibility results. We performed a
prospective study in which fluid from positive blood culture bottles
from the BacT/Alert were inoculated directly into MicroScan rapid
and overnight panels for gram-negative and gram-positive bacteria and
compared susceptibility results with those obtained in standard fashion
with each panel type.
Specimens tested.
Positive aerobic blood cultures collected
between September 1995 and April 1996 from patients suspected of having
bacteremia at the University of Alabama at Birmingham Hospital were
examined by Gram stain. Specimens that appeared to contain a single
organism were included in the study. Specimens that were unimicrobic on Gram stain but yielded more than one isolate after subculture were
excluded from analysis. Only aerobic and facultative bacteria were
included. Haemophilus spp., Neisseria spp.,
Streptococcus pneumoniae, yeasts, and anaerobes were
excluded.
MicroScan instrumentation and panels.
The WalkAway/40
instrument with Version 20.30 of the Data Management System was used
for all reading and interpretation of panel results. MicroScan Rapid
Neg Combo Type 2 panels and MicroScan Dried Overnight Neg Combo Type 15 panels were used for gram-negative isolates. MicroScan Rapid Pos Combo
Type 1 panels and MicroScan Overnight Pos Combo Type 6 panels were used
for gram-positive isolates.
Blood culture instrumentation and media.
Blood culture
testing was performed on the Organon Teknika BacT/Alert instrument with
standard aerobic media.
Direct susceptibility testing.
Preliminary studies were
performed to determine the optimum method of direct inoculation of
MicroScan panels by using seeded cultures (1) and actual
patient specimens (18). Serum separator tubes
(Becton-Dickinson Vacutainer Systems, Rutherford, N.J.) containing 0.2 ml of Triton X-100 (Sigma, St. Louis, Mo.) were inoculated with 9.5 ml
of each blood specimen from positive BacT/Alert bottles with a 22-gauge
needle. The tubes were centrifuged at 1,400 × g for 10 min at room temperature. Bacteria were harvested from the surface of
the silicon layer by using a cotton swab to make an inoculum suspension
equivalent to a 0.5 McFarland standard, confirmed by a MicroScan
turbidity meter. This technique has been shown previously to be an
acceptable means for preparation of bacterial suspensions for direct
susceptibility testing (16). All blood culture specimens
were tested the same day that the instrument flagged them as positive.
MicroScan panels were inoculated and incubated in the MicroScan
WalkAway/40 according to the manufacturer's instructions, by using the
suspensions prepared as described above.
Standard susceptibility testing.
A small amount of the blood
culture fluid was inoculated onto Trypticase soy agar with 5% sheep
blood (BBL, Cockeysville, Md.) and a MacConkey agar plate (BBL) with
crystal violet and lactose. The blood agar plate was incubated in 5%
CO2, and the MacConkey agar plate was incubated in air for
18 to 24 h at 35°C to produce bacterial colonies for
inoculation. Rapid gram-negative panels were inoculated from MacConkey
agar. All other panels were inoculated from blood agar plates according
to the manufacturer's instructions. Panels were then incubated in the
same manner as in the direct susceptibility test panels.
Quality control.
Appropriate identification and
susceptibility quality control organisms for the MicroScan rapid and
overnight panels as defined by the manufacturer were tested weekly, and
all results were acceptable. All other daily maintenance for the
MicroScan WalkAway/40 and the BacT/Alert was performed in accordance
with each manufacturer's instructions.
Data analysis.
Direct testing on MicroScan rapid panels was
compared to standardized testing on rapid panels. Direct testing on
overnight panels was compared to standardized testing on overnight
panels. Drug-organism combinations which have not been approved by the Food and Drug Administration for testing with MicroScan were not included in data analysis. Only antimicrobial agents considered appropriate by the National Committee for Clinical Laboratory Standards
guidelines were evaluated by using their published susceptibility breakpoints (11). Gram-positive isolates resistant to
oxacillin were considered resistant to other beta-lactam
antimicrobials. All drug-organism contraindications were based on the
direct test method identification. Organism identification used for
summary purposes was that obtained with the standard method.
0095-1137/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.
Direct Susceptibility Testing with Positive BacT/Alert Blood
Cultures by Using MicroScan Overnight and Rapid Panels
![]()
ABSTRACT
Top
Abstract
Introduction
Materials & Methods
Results
Discussion
References
![]()
INTRODUCTION
Top
Abstract
Introduction
Materials & Methods
Results
Discussion
References
![]()
MATERIALS AND METHODS
Top
Abstract
Introduction
Materials & Methods
Results
Discussion
References
| |
RESULTS |
|---|
|
|
|---|
A total of 253 specimens appeared unimicrobic based on initial Gram-stained smears. However, 16 (6.3%) specimens were eventually shown to be polymicrobic when subcultured and excluded from analysis, leaving 237 evaluable specimens. Among these, there were 133 gram-negative bacilli and 104 gram-positive cocci, broken down into species in Table 1.
|
Standardized susceptibility results were available for all 133 gram-negative bacteria tested on overnight panels. For 114 of 133 (85.7%) gram-negative isolates there were rapid panel MIC results. For 102 of 104 (98.1%) gram-positive cocci there were standardized MIC results on overnight panels, whereas for 81 of 104 (77.9%) there were rapid panel results. Two gram-positive cocci tested by overnight panels were excluded from analysis because of contamination and erroneous readings by the standard method, which were not detected until after the samples had been evaluated. All gram-negative and gram-positive isolates that did not give rapid panel results were due to insufficient growth. Six of 133 (4.5%) gram-negative isolates were retested due to five or more very major or major errors on initial testing. Four were due to errors in the overnight panel, and one was due to errors in the rapid panel. Five of 104 (4.8%) of the gram-positive cocci were retested due to five or more major or very major errors on initial testing. Four were retested due to errors in the overnight panel and one was retested because of errors in the rapid panel. After repeat testing, one coagulase-negative Staphylococcus sp. (CNS), two Escherichia coli spp., and one Pseudomonas aeruginosa sp. still had five major errors.
One hundred fifteen of 133 (86.4%) gram-negative isolates showed no major or very major errors for any drugs on the overnight panel, compared to 98 of 114 (85.9%) on the rapid panel. Complete categorical agreement, i.e., no very major, major, or minor errors for any drug tested, was observed with 94.7 and 93.5% of drugs tested on overnight and rapid gram-negative panels, respectively (Table 2). Seventy-seven of 102 (75.5%) gram-positive cocci showed no major or very major errors on overnight panels, versus 64 of 81 (79.0%) on rapid panels. Complete concordance was observed with 93.2 and 93.1% of drugs tested on overnight and rapid gram-positive panels, respectively.
|
The major error rate for all gram-negative direct susceptibility tests was 1.4% for overnight panels versus 0.7% for rapid panels. The very major error rate was 2.7% for overnight versus 8.1% for rapid panels. The total error rates for gram-negative bacteria, obtained by combining very major and major errors and dividing by the total number of those bacteria for which MICs indicated susceptibility and resistance, were 1.6% for overnight panels and 1.5% for rapid panels.
The major error rates for gram-positive direct susceptibility tests were 2.6% for overnight panels and 2.5% for rapid panels. The very major error rates were 8.8 and 7.2% for overnight and rapid panels, respectively. Total error rates were 3.6% for both overnight and rapid gram-positive panels.
Analysis of errors detected by direct susceptibility testing with overnight and rapid panels according to individual antimicrobials and bacterial species is shown in Table 3. Among drugs on the gram-negative panels, only amikacin and gentamicin had no major or very major errors. Among those on gram-positive panels, narrow-spectrum cephalosporins and gentamicin synergy had no major or very major errors. Aztreonam had the most errors (seven on overnight and one on rapid panels), followed by mezlocillin (two on overnight and five on rapid panels), and cefuroxime (five on overnight and one on rapid panels) among agents tested on gram-negative panels. Drugs with the most errors on gram-positive panels were clindamycin (four on overnight and eight on rapid panels), oxacillin (four on overnight and seven on rapid panels), and trimethoprim-sulfamethoxazole (nine on overnight panels).
|
One P. aeruginosa and two E. coli isolates produced 15 of 25 major errors on overnight gram-negative panels. One CNS isolate gave 5 of 12 major errors on rapid panels. Overall, CNS gave 42 of 66 errors on both panel types combined.
| |
DISCUSSION |
|---|
|
|
|---|
Direct susceptibility testing on positive blood cultures has been recommended to facilitate early confirmation of antimicrobial susceptibility (2, 4, 6-10, 12, 13). Automated blood culture systems such as BacT/Alert and rapid MIC panels provided by automated bacterial identification and susceptibility testing systems such as MicroScan can potentially make direct susceptibility testing even more clinically valuable, allowing preliminary results in a matter of hours after growth is first detected in the blood culture bottle. Although the concept of direct inoculation of fluid from positive blood culture bottles into MIC panels was investigated in the early 1980s (6), no prospective evaluations by using the current MicroScan instrumentation, panels, software, and database, newer antimicrobials, or with more recently developed continuously monitoring blood culture instruments such as the BacT/Alert have been published.
This study demonstrated that MicroScan overnight and rapid panels can be used to provide direct susceptibility test results for unimicrobic gram-negative blood culture isolates detected by the BacT/Alert with 94.7 and 93.1% overall categorical agreement for MICs in comparison to standardized methodology by using the same instrumentation. For gram-positive bacteria, direct and standard testing with overnight and rapid panels provided 93.2 and 93.1% categorical agreement.
The definitive evaluation of direct susceptibility testing would involve a third reference test method for arbitration of discrepancies between the standardized result and the direct result. However, this was not attempted in the present evaluation, mainly because the objective of the study was to determine whether direct susceptibility testing was an acceptable alternative for use with an automated instrument already present in our laboratory. MicroScan susceptibility test capability has already been compared with reference methods numerous times (7, 17).
Only 6.3% of the specimens tested that appeared unimicrobic were later found to be polymicrobic and unsuitable for direct susceptibility testing by this method, numbers which are in agreement with previously published rates for polymicrobic bacteremia (10). Verification of direct inoculum purity could be made following incubation of overnight panels and prior to reporting results if a subculture to agar is performed at the time of direct inoculation of susceptibility panels. This would not be possible for rapid panels for which MIC data would be available prior to verification of culture purity if rapid direct susceptibility data are to be reported as soon as results are available.
False susceptibility occurred with direct susceptibility tests with a greater frequency than false resistance in overnight and rapid panels for gram-negative as well as gram-positive bacteria. When considering the total error rates, there was minimal difference between overnight versus rapid gram-negative panels and none between overnight versus rapid gram-positive panels. Total error rates for overnight gram-negative panels were less than the corresponding values for overnight gram-positive panels.
Among gram-positive cocci, CNS were responsible for the most errors. Of major concern, however, was false susceptibility in directly inoculated panels for oxacillin in two isolates of Staphylococcus aureus, one with an overnight panel and the other with a rapid panel. This finding was not surprising since oxacillin-resistant S. aureus is often present in heterogeneous populations and tends to grow more slowly.
In comparative evaluations of susceptibility testing procedures, very
major errors should occur in <1.5% of all tests, and the overall
agreement between tests and the reference method should be
95%
(12). Rigorous application of these criteria to data from
the present evaluation would mean that direct susceptibility testing
with either overnight or rapid panels would be inappropriate, mainly
because of the high false-susceptibility rates. However, it has been
suggested that these criteria may be too restrictive (12).
Even though automated systems have reduced the time required to detect microorganisms in bloodstream infections, the relatively low number of bacteria in culture bottles may compromise the inoculum concentration, if not standardized, for susceptibility tests. This may partially account for the relatively high error rates for direct susceptibility testing with MIC methods reported by some studies evaluating blood cultures performed in automated instruments (5, 12), but others (2, 6, 13) yielded more favorable comparisons.
Kiehn et al. (6) found only 0.7% categorical discrepancies between MicroScan MIC panels inoculated directly and by standard methodology from nonautomated blood culture systems. More recently, Pettigrew et al. (13) compared results of direct inoculation from positive BacT/Alert bottles to MicroScan gram-positive and gram-negative rapid panels. Complete agreements for 1,403 antimicrobial agent-organism combinations were 96.9% for gram-negative bacilli and 96.8% for 1,083 gram-positive coccus combinations, slightly more favorable results than obtained in our investigation, which was performed in a similar manner.
Zimmer et al. (18) also evaluated direct susceptibility testing by using MicroScan rapid and dried overnight panels, testing three different inoculum preparation methods and employing multiple media from three different blood culture systems. Due to a small number of positive cultures, results from all media and blood culture systems were combined in data analysis. Overall categorical agreement rates similar to those reported in the present study were described. They also detected false susceptibility by using direct testing more often than they detected false resistance, consistent with our findings.
Direct susceptibility testing of bacteria from positive blood culture bottles by agar disk diffusion is a relatively inexpensive procedure, not tied to bacterial identification, in contrast to direct testing with an automated system such as MicroScan. Agar disk diffusion has been the most widely studied technique of direct susceptibility testing and has the greatest correlation with standardized methods, according to most studies (9).
Most laboratories which invest in MicroScan technology use the instrument for both bacterial identification and susceptibility testing, typically employing the same biochemical-antimicrobial panel for each organism tested. Due to the costs of panels, it is not practical to perform direct susceptibility testing with this technology if results must be repeated with a standardized inoculum. Likewise, it would be impractical to perform direct susceptibility testing unless direct bacterial identification also proves to be accurate so that confirmation following subculture is unnecessary.
Organism identification evaluated as part of this investigation has been preliminarily described (15). Ninety-six percent of overnight and 72% of rapid panels showed complete agreement between direct and standard methods for identification of gram-negative bacteria, with the highest concordance (99%) occurring with Enterobacteriaceae. As might be expected, nonfermentative gram-negative bacilli proved more difficult to identify by direct inoculation. Significant problems occurred with gram-positive cocci, for both overnight and rapid panels inoculated directly. Only 82% concordance with standard identification occurred for overnight and 52% for rapid panels directly inoculated with gram-positive cocci. These discrepancies were predominantly due to the inability to distinguish among CNS, but problems also occurred with identification of S. aureus in directly inoculated panels. Even though Pettigrew and colleagues (13) had a better correlation in their study with respect to direct versus standardized susceptibility testing by using MicroScan rapid panels, their identification agreement was only 89% for gram-negative bacilli and 70% for gram-positive cocci.
Our practice has been to perform direct susceptibility testing on all blood culture isolates by agar disk diffusion and confirm results for reporting purposes by using MicroScan overnight panels which provide both susceptibility and organism identification. We did not change this practice after performance of this evaluation. If MicroScan panels are to be used for direct susceptibility testing, results should be considered preliminary, mainly because of potential false susceptibility. However, direct susceptibility results obtained from gram-negative overnight panels indicating resistance are likely to be in agreement with standard methods and could be considered for immediate reporting if laboratories choose to use such a direct method to improve turnaround time.
Given the many potential advantages of rapid turnaround times for antimicrobial susceptibility results, periodic evaluations of automated microbiological instrumentation are extremely important. It is particularly relevant to the study of emerging resistant organisms and newer antimicrobials. Reliability of direct susceptibility testing by using any automated method should be tested by individual laboratories before the method is considered for routine use.
| |
ACKNOWLEDGMENTS |
|---|
Financial support for this project was provided by Dade MicroScan, Inc., West Sacramento, Calif., and Organon Teknika Corporation, Durham, N.C.
The technical assistance of Linda Van Pelt and Mary McKinnon of Dade Microscan, Inc., is appreciated.
| |
FOOTNOTES |
|---|
* Corresponding author. Mailing address: Department of Pathology, WP 230, University of Alabama at Birmingham, Birmingham, AL 35233. Phone: (205) 934-6421. Fax: (205) 975-4468. E-mail: Waites{at}path.uab.edu.
| |
REFERENCES |
|---|
|
|
|---|
| 1. | Devitt, K. A., J. M. O'Sullivan, M. L. McKinnon, and B. L. Zimmer. 1995. Feasibility of direct susceptibility testing from various seeded blood culture media into MicroScan rapid and dried overnight susceptibility panels, abstr. C-345, p. 60. In Abstracts of the 95th General Meeting of the American Society for Microbiology 1995. American Society for Microbiology, Washington, D.C. |
| 2. |
Dipersio, J. R.,
S. M. Ficorilli, and F. J. Varga.
1984.
Direct identification and susceptibility testing of gram-negative bacilli from Bactec bottles by use of MS-2 System with updated bacterial identification software.
J. Clin. Microbiol.
20:1202-1204 |
| 3. |
Doern, G. V.,
R. Vautour,
M. Gaudet, and B. Levy.
1994.
Clinical impact of rapid in vitro susceptibility and bacterial identification.
J. Clin. Microbiol.
32:1757-1762 |
| 4. |
Fay, D., and J. E. Oldfather.
1979.
Standardization of direct susceptibility tests for blood cultures.
J. Clin. Microbiol.
9:347-350 |
| 5. | Howard, W. J., B. J. Buschelman, M. J. Bale, M. A. Pfaller, F. P. Koontz, and R. N. Jones. 1996. Vitek GPS card susceptibility testing accuracy using direct inoculation from Bactec 9240 blood culture bottles, abstr. C-311, p. 56. In Abstracts of the 96th General Meeting of the American Society for Microbiology 1996. American Society for Microbiology, Washington, D.C. |
| 6. |
Kiehn, T. E.,
C. Capitelo, and D. Armstrong.
1982.
Comparison of direct and standard microtiter broth dilution of susceptibility testing of blood culture isolates.
J. Clin. Microbiol.
16:96-98 |
| 7. |
McGregor, A.,
F. Schio,
S. Beaton,
V. Boulton,
M. Perman, and G. Gilbert.
1995.
The MicroScan WalkAway diagnostic microbiology system an evaluation.
Pathology
27:172-176[Medline].
|
| 8. | Mirrett, S. 1994. Antimicrobial susceptibility testing and blood cultures. Clin. Lab. Med. 14:171-179[Medline]. |
| 9. |
Mirrett, S., and L. B. Reller.
1979.
Comparison of direct and standard antimicrobial disk susceptibility testing for bacteria isolated from blood.
J. Clin. Microbiol.
10:482-487 |
| 10. |
Moore, D. F.,
S. S. Hamada,
E. Marso, and W. J. Martin.
1981.
Rapid identification and antimicrobial susceptibility testing of gram-negative bacilli from blood cultures by the automicrobic system.
J. Clin. Microbiol.
13:934-939 |
| 11. | National Committee for Clinical Laboratory Standards. 1994. Methods for dilution antimicrobial susceptibility tests for bacteria that grow aerobically. Publication M7-A3, 3rd ed. National Committee for Clinical Laboratory Standards, Villanova, Pa. |
| 12. | Nolte, F. S., P. B. Contestable, D. Lincaslis, and A. Punsalang, Jr. 1986. Rapid, direct antibiotic susceptibility testing of blood culture isolates using the Abbott Advantage System. Am. J. Clin. Pathol. 86:665-669[Medline]. |
| 13. | Pettigrew, J., D. Ballou, and J. Salyer. 1993. Direct susceptibility testing of positive BacT/Alert blood culture systems with AutoSCAN-W/A rapid combo panels, abstr. C-66, p. 457. In Abstracts of the 93rd General Meeting of the American Society for Microbiology 1993. American Society for Microbiology, Washington, D.C. |
| 14. |
Trenholme, G. M.,
R. L. Kaplan,
P. H. Karakusis,
T. Stine,
J. Fuhrer,
W. Landau, and S. Levin.
1989.
Clinical impact of rapid identification in susceptibility testing of bacterial blood culture isolates.
J. Clin. Microbiol.
27:1342-1345 |
| 15. | Waites, K. B., E. S. Brookings, S. A. Moser, M. L. McKinnon, L. Van Pelt, and B. L. Zimmer. Direct bacterial identification from positive BacT/Alert blood culture specimens using MicroScan overnight and rapid panels. Diagn. Microbiol. Infect. Dis., in press. |
| 16. | Wilson, M. L., M. P. Weinstein, and L. B. Reller. 1994. Automated blood culture systems. Clin. Lab. Med. 14:149-169[Medline]. |
| 17. | York, M. K., C. F. Brooks, and E. H. Fiss. 1992. Evaluation of the autoScan W/A rapid system for identification and susceptibility testing of gram-negative fermentative bacilli. J. Clin. Microbiol. 30:2901-2910. |
| 18. | Zimmer, B. L., K. Devitt, M. McKinnon, S. Mirrett, and L. B. Reller. 1995. Direct susceptibility testing from positive blood culture specimens into MicroScan rapid and dried overnight susceptibility panels, abstr. C-346, p. 60. In Abstracts of the 95th General Meeting of the American Society for Microbiology 1995. American Society for Microbiology, Washington, D.C. |
This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Antimicrob. Agents Chemother. | Clin. Microbiol. Rev. |
|---|---|
| Clin. Vaccine Immunol. | ALL ASM JOURNALS |
|---|