Previous Article | Next Article 
Journal of Clinical Microbiology, August 1998, p. 2183-2186, Vol. 36, No. 8
0095-1137/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.
Evaluation of the Biostar Chlamydia OIA Assay with
Specimens from Women Attending a Sexually Transmitted Disease
Clinic
Mitchell S.
Pate,1
Paula B.
Dixon,1
Kim
Hardy,1
Mark
Crosby,2 and
E. W.
Hook III1,3,*
Department of Medicine, Division of
Infectious Diseases, University of Alabama at
Birmingham,1 and
Jefferson County
Department of Health,3 Birmingham, Alabama
35294, and
Biostar, Inc., Boulder, Colorado
803012
Received 4 November 1997/Returned for modification 30 December
1997/Accepted 24 April 1998
 |
ABSTRACT |
Chlamydia trachomatis infections are the most prevalent
sexually transmitted diseases (STDs) in the United States. In
acute-care settings such as clinics and emergency rooms, a desirable
chlamydia screening assay should exhibit good sensitivity and good
specificity and should provide test results while the patient is still
present. The Biostar Chlamydia OIA (Biostar, Inc., Boulder, Colo.) is
an optical immunoassay (OIA) that provides test results in less than 30 min and that uses a test format that allows office-based testing. This
assay is performed entirely at room temperature without the need for
rotators or other specialized equipment. The goal of this study was to
compare the performance of the Biostar Chlamydia OIA for the detection
of C. trachomatis with the performance of cell culture,
direct fluorescent-antibody (DFA) assay (Syva MicroTrak; Syva Co., Palo
Alto, Calif.), and PCR (Roche Amplicor Chlamydia trachomatis; Roche, Branchburg, N.J.) for the detection of
C. trachomatis infections in women attending an urban STD
clinic. For calculations of relative test performance (sensitivity,
specificity, and positive and negative predictive values), patient
specimens that yielded positive results by two or more of the four
assays (cell culture, DFA assay, PCR, and OIA) were classified as
"true infections." By these criteria, 42 of 306 total specimens
were classified as positive for C. trachomatis (positive
prevalence, 13.7%), 11 (3.6%; 10 by PCR and 1 by DFA assay) were
positive by a single assay, and 253 (82.7%) were negative by all four
tests. All culture-positive specimens were also positive by at least one other assay. Among the culture-negative specimens, 14 (5%) specimens were positive by two of the three non-culture-based assays
used. By using the criterion that positivity by at least two of the
tests indicated a true infection, the relative sensitivities were as
follows: culture and PCR, 92.9% each; Biostar Chlamydia OIA, 73.8%;
and DFA assay, 59.5%.
 |
INTRODUCTION |
Chlamydia trachomatis
infections are the most prevalent sexually transmitted diseases (STDs)
in the United States and are a major preventable cause of infertility,
ectopic pregnancy, and chronic pelvic pain in women (3). In
addition, because the signs and symptoms of infection are often mild or
even absent, laboratory testing plays a central role in efforts to
control chlamydia.
Optimally, tests for the detection of chlamydia should be sensitive and
specific and should provide results quickly to guide patient
management. Until recently, the standard for the diagnosis of C. trachomatis infection has been cell culture (8, 9). However, in most clinical settings, antigen or nonamplified nucleic acid detection tests have been preferentially used for testing for
chlamydia because of their logistical advantages and lower costs,
despite the observation that these tests are generally less sensitive
than cell culture (8). More recently, evaluations of newer
nucleic acid amplification assays for C. trachomatis diagnosis such as PCR or ligase chain reaction have shown the true
sensitivity of cell culture for chlamydia detection to be 65 to 85%
(15, 18). Despite their greater sensitivities, amplified nucleic acid detection tests are not used in many clinical settings, in
part because of their substantially higher costs compared to those of
nonamplification assays. In addition, like most currently available
tests for the diagnosis of C. trachomatis infection, the
nucleic acid amplification tests usually do not provide test results at
the time of screening but typically require turnaround times of a day
or more, a characteristic that may introduce delays in translating test
results into treatment (6, 16). In some settings there are
demonstrable advantages to tests that can provide results while
patients are still present in clinical settings, particularly if the
performance of such tests is comparable to the performance of alternate
tests (6, 16).
The Biostar Chlamydia OIA (Biostar, Inc., Boulder, Colo.) is an optical
immunoassay (OIA) designed to rapidly detect chlamydia infection in
women, providing test results in less than 30 min in a test format that
allows office-based testing. To evaluate the performance of the Biostar
Chlamydia OIA for the detection of C. trachomatis, we
compared the performances of four assays, i.e., cell culture, an
immunofluorescence antigen detection assay (Syva MicroTrak; Syva Co.,
Palo Alto, Calif.), PCR (Roche Amplicor Chlamydia
trachomatis; Roche, Branchburg, N.J.), and the Biostar Chlamydia
OIA, for the detection of C. trachomatis infections in women
attending an urban STD clinic.
 |
MATERIALS AND METHODS |
Patient population and specimen collection.
Between 2 November 1994 and 30 December 1994, women attending the Jefferson
County Department of Public Health STD Clinic in Birmingham, Ala., were
asked to participate in the present investigation. After swab specimens
for Neisseria gonorrhoeae cultures were taken, three
endocervical swab specimens were collected from consenting participants
for C. trachomatis testing. The first specimen collected
from each patient for chlamydia testing was always obtained with a
Dacron-tipped plastic shaft swab placed in 1.5 ml of 0.2 M sucrose
phosphate chlamydia cell culture transport medium containing 2% fetal
bovine serum and antibiotics (gentamicin, mycostatin, and vancomycin).
Next, a specimen was collected with a Dacron polyester swab that was
placed in the OIA transport container. Finally, a swab specimen was
collected for direct-fluorescent antibody (DFA) evaluation and was
rolled over a slide, which, upon drying, was immediately fixed with
methanol and placed in the transport unit. PCR testing was performed
with an aliquot of the Biostar Chlamydia OIA antigen extract solution.
All specimens were maintained at 3 to 5°C and were transported to the
laboratory within 18 h of collection, where they were stored
before evaluation. Cell culture vials were stored for up to 72 h
at
70°C before cultivation (2). The swab for OIA and the
slide for the DFA evaluation were stored at 3 to 5°C for no more than
24 h, batched, and processed for determination of the test
results. The technologists who were processing the specimens were
unaware of the results of the other chlamydial assays which had been
performed with specimens from the same patient.
Cell culture method.
Cell culture for C. trachomatis was performed as described previously (11).
Briefly, chlamydia culture medium was inoculated, in triplicate, into a
96-well microtiter format culture system. The cultures were incubated
at 37°C, and if no chlamydia inclusions were noted at 48 to 72 h with
two different fluorescein-conjugated antibody reagents (Syva MicroTrak
[Syva Co.] and Kallestad Pathfinder [Sanofi Diagnostics Pasteur,
Chaska, Minn.]), a blind passage was performed.
PCR assay.
The Roche Amplicor Chlamydia
trachomatis PCR assay was performed twice weekly with previously
collected and assayed samples in the OIA transport medium. A 100-µl
aliquot of the sample processed for OIA (see below) was added to 900 µl of the Roche Amplicor swab transport medium. From this point
forward, the diluted sample was processed by the PCR assay performed
according to the manufacturer's package insert.
DFA assay.
For DFA testing, methanol-fixed slides were
allowed to come to room temperature, placed in a humidified chamber,
and stained according to the manufacturer's directions with the Syva
MicroTrak Chlamydia trachomatis Direct Specimen Test stain
(Syva Co.). The slides were then examined with a ×100 oil immersion
lens with a fluorescence microscope (Zeiss, Flushing, N.Y.). Each slide was graded and interpreted as adequate according to the manufacturer's directions. Adequate specimens had at least 10 columnar and/or cuboidal
epithelial cells. By using the manufacturer's criteria for positivity,
slides were considered positive for C. trachomatis if 10 elementary bodies (EBs) were present and negative if fewer than 10 EBs
were observed.
OIA assay.
The Biostar Chlamydia OIA assay was performed
daily with batched samples in the laboratory. This assay is performed
entirely at room temperature without the need for rotators or other
specialized equipment. Reagents, detection devices, and swab specimens
were allowed to warm to room temperature before testing began. The test
was performed as follows. The swabs on which the specimens had been
collected were initially placed in the manufacturer's polypropylene
extraction tube, and 2 drops of an extraction reagent (reagent 1A) was
added, followed by brief agitation, and the tubes were allowed to stand
for 2 min. At this point, 6 drops of a second extraction reagent
(reagent 1B) were added and another 2-min incubation was carried out.
Following this step, six drops of a neutralizing reagent (reagent 2)
were added and the swab used for specimen collection was used to
vigorously mix the reagent solution. Next, the swab was expressed
against the walls of the tube to remove excess liquid and was
discarded. Then, with a disposable pipet, 1 drop of the final extracted
solution was placed directly onto the center of the reflective optical
surface of the test device and the remaining extraction solution was
refrigerated for subsequent PCR testing. The liquid specimen droplet
was allowed to remain on the surface of the test device for 5 min.
Then, 1 drop of Biostar murine monoclonal antibody reagent (reagent 3)
was added directly to the sample specimen on the OIA device and the
mixture was allowed to incubate for an additional 5 min. The test
surface was then washed vigorously by squirting a wash solution
(reagent 4) across the reflective optical surface, and the remaining
wash solution was blotted from the optical surface with the blotter
located in the test device lid. One drop of the substrate (reagent 5) was placed onto the center of the device and was allowed to remain in
place for 10 mins. The surface was again washed with reagent 4 and the
remaining wash solution was absorbed. The lid was then opened and the
result was observed in reflected light. An internal procedural control
(a 1- to 2-mm-diameter blue-purple dot) contained in each test unit
became visible following proper processing, allowing the technician to
confirm that all steps of the detection process had been followed for
all samples. Negative samples were identified by observation of the
internal control dot only. Positive assays were apparent as a large
blue to purple filled circle on the yellow reflective surface. If the
result was positive, the more intensely colored internal control dot
was visible within the larger test specimen area. Positive and negative
control samples provided by the manufacturer were included with each
run. The results were recorded, and all test devices were stored for
confirmation of the results by a third party if necessary. The result
obtained with the device is stable for an unlimited period at room
temperature.
DFA discrepant analysis.
For further evaluation of the
results with specimens from patients which yielded only a single
positive test result, the questionable specimens were analyzed by the
DFA assay. Briefly, the 0.2 M sucrose phosphate culture transport
medium was centrifuged, the pellet was resuspended, and the suspension
was placed on a slide, stained with a DFA stain (Syva MicroTrak), and
evaluated for the presence of chlamydial EBs as described previously
(11).
Analyses.
Recent studies have suggested that cell culture
fails to detect a substantial proportion of C. trachomatis
infections, and that, other than cell culture, all non-culture-based
C. trachomatis detection assays have the potential to give
false-positive results (8). At the same time, efforts to
resolve differences in test results through further testing of
specimens yielding discrepant results (i.e., specimens from the same
patient yielding different results when tested by different assays)
have been criticized because of concerns that this approach introduces
biases which favor the new test under consideration (4, 5).
For this study, for assessment of test performance, patient specimens
that yielded positive results by any two assays (cell culture, DFA assay, PCR, and OIA) were considered "true infections" for
calculation of sensitivity, specificity, and positive and negative
predictive values. Thus, for our analyses any specimen that yielded
only a single positive result when all four assays were used to test the specimen was considered "false positive." While these
definitions may underestimate the true proportion of infected study
participants (see Discussion), this approach permits internally
consistent analyses and comparison of the relative performance of the
four assays.
 |
RESULTS |
The age, race or ethnicity, and reasons for clinic attendance of
the study population (n = 306) were typical of those of
female patients attending the Birmingham STD clinic. The age range of the participants was 15 to 52 years (median, 27 years); 87% were African-American, 13% were white, and one subject was of Hispanic ethnicity; and 56% acknowledged symptoms including genital discharge, dysuria, or abdominal pain.
Individuals with positive test results by two or more of the four
assays were classified as "infected" for calculation of the
relative performance of the assays. By using these criteria, specimens
from 42 of 306 total participants were classified as positive for
C. trachomatis (positive prevalence, 13.7%), 11 (3.6%; 10 by PCR and 1 by DFA assay) were positive by a single assay, and 253 (82.7%) were negative by all four tests (Table
1). All culture-positive specimens were
also positive by at least one other assay. Among the culture-negative
specimens, specimens from 14 (5%) participants were positive by two of
the three non-culture-based assays used in this study (most often both
the PCR and the OIA were positive). By using the criterion of
positivity by at least two of the tests, the sensitivities of culture
and PCR were each 92.9% (Table 2). The
sensitivity of the Biostar Chlamydia OIA was 73.8%, and the
sensitivity of the DFA assay was 59.5%.
View this table:
[in this window]
[in a new window]
|
TABLE 1.
Results of all diagnostic tests evaluated for detection
of C. trachomatis by using positive results by at least two
tests as the criterion for inclusiona
|
|
View this table:
[in this window]
[in a new window]
|
TABLE 2.
Relative performance of each assay compared to that when
the criterion was positivity by at least two of
the testsa
|
|
The culture transport medium from the 11 specimens obtained from
participants with a single positive non-culture-based assay result was
tested for the presence of chlamydial EBs by the DFA assay. Chlamydial
EBs were seen in specimens from 7 of 10 PCR-positive patients, while
EBs were not seen on testing of the single specimen which was positive
only by the DFA assay. If the seven PCR-positive specimens for which
EBs were seen in culture transport medium were recorded as true
positives, the revised sensitivities of culture, OIA, and the DFA assay
decreased to 79.6, 63.2, and 50%, respectively, while the sensitivity
of PCR would increase to 93.8%.
 |
DISCUSSION |
In the present investigation the Biostar Chlamydia OIA, an antigen
detection assay suitable for on-site use in acute-care settings such as
clinics and emergency rooms, accurately detected 74 to 63% of C. trachomatis infections, depending on how sensitivity was
calculated. Although many published studies have evaluated non-culture-based tests for the diagnosis of chlamydial infections, few
of these studies have compared the performance of non-culture-based tests to each other (17). When compared to cell culture and the Roche Amplicor PCR, the Biostar Chlamydia OIA was less sensitive for the detection of chlamydia infection, while when compared to direct
immunofluorescence microscopy, the OIA appeared to be somewhat more
sensitive. The sensitivities reported in Table 2, however, reflect the
relative performance of the four assay methods used in this study. For
10 patients, only the PCR-based tests were positive, and 7 of these
patients were subsequently judged to most likely have a true infection
on the basis of the demonstration of chlamydial EBs by
immunofluorescence microscopic examination of the culture transport
medium. Thus, the performance reported in Table 2 likely underestimates
the true sensitivity of the PCR-based assay and somewhat inflates the
relative sensitivities of cell culture, the MicroTrak DFA assay, and
the OIA for the detection of chlamydial infection.
The sensitivity of the Biostar Chlamydia OIA in this study is less than
the 83.8% sensitivity reported in the manufacturer's package insert,
a finding that most likely reflects differences in analytic approach
rather than variations in test sensitivity. The studies used to
calculate the test performance reported in the manufacturer's package
insert were performed, in part, by our laboratory; however, the U.S.
Food and Drug Administration's practice for considering the
performance of new tests is to consider only the results of initial
cell culture, without blind passage. This practice helps to explain the
differences in sensitivities between our findings and those in the
package insert, as well as to illustrate the complexities of evaluating
the performance of non-culture-based tests for the detection of
chlamydia.
To date, few published studies have compared the performance of
multiple diagnostic tests to one another (9). Instead, there
have been comparisons of the performance of a non-culture-based test to
that of culture, sometimes supplemented by immunofluorescence microscopy of specimens positive by non-culture-based tests in order to
compensate for possible false-negative culture results. This approach,
however, fails to address the theoretical possibility that for
substantial numbers of specimens results for both culture and the
non-culture-based assay are false negatives (4, 5). For this
study, which included four different tests, we chose a comparative
method that evaluated the performance of the four tests relative to
each other. However, given what appears to be the greater sensitivity
of the PCR test, the positive results for 42 specimens classified as
C. trachomatis positive by using for these analyses the
criterion of positivity by two or more of the four tests may slightly
overestimate the true performance of OIA, DFA assay, or culture.
Several other limitations of this study should be acknowledged. In
particular, the performance of the DFA test for the diagnosis of
chlamydial infection may have been diminished by two potentially confounding factors. The performance of the DFA assay in this study may
have been suboptimal since the specimen collection order was not
randomized and the specimens for DFA testing were collected fourth
(last). Good patient specimen collection with appropriate numbers of
cells and proper slide preparation techniques are essential for the
maximum performance of the DFA assay. Seventy-two percent of all
specimens tested by the DFA assay in the course of this study were
deemed inadequate on the basis of having fewer than 10 columnar and/or
cuboidal epithelial cells, but in the performance calculations the
results for those specimens were considered negative for C. trachomatis. Typical proportions of inadequate specimens associated with routine cervical swab specimen collection for the DFA
assay range between a low of 1% and a high of 79% (7, 14).
Thus, while the number of inadequate specimens obtained during the
course of this study was high, they are within the range reported in
the literature in the testing of endocervical swab specimens. In
addition, the DFA assays in this study were evaluated with a
requirement that the specimen contain 10 EBs to be classified as
positive, according to the manufacturer's instructions. In prior
studies (1, 7, 10, 13), other investigators have
demonstrated that the sensitivity of this assay may be increased by
using a cutoff of one EB as the criterion for a positive assay result.
In our study, 12 specimens were found to contain fewer than 10 EBs
(range, 2 to 7 EBs), and if these were considered positive, the result
would have been a relative sensitivity for the DFA test of 88%.
It should be acknowledged that for PCR testing, although a commercially
available assay was used, the material used for testing was the
specimen processed for OIA rather than a swab specimen. Studies in our
laboratory suggest that the use of this material for testing did not
detract from the performance of the PCR (data not shown).
Finally, another issue which should be acknowledged is that this study
was performed in a research laboratory by experienced laboratory staff.
Thus, the results may not reflect the performance characteristics of
these tests in the office- or clinic-based settings for which they are
designed and where nonlaboratory staff are performing the assay.
Real-world use of the OIA in busy clinical settings may not yield the
same sensitivity reported here. Nonetheless, the performance of the
Biostar Chlamydia OIA in this study is comparable to that in
laboratory-based enzyme immunoassays previously performed in our
laboratory (12). While the sensitivity of the OIA for the
detection of chlamydial infection was not as great as that of cell
culture or PCR, the performance of the OIA in selected settings, with
patients present, could result in the expeditious treatment of
chlamydia-infected individuals. The Biostar Chlamydia OIA is classified
as a moderately complex assay. It is also considered a quick method for
the detection of C. trachomatis in endocervical samples.
In summary, in this prospective study, the relative comparisons of the
four assays (cell culture, PCR, DFA assay, and OIA) showed that there
was a good correlation between the results of PCR and cell culture,
that the relative sensitivity of the OIA was 73.8% and that the
relative sensitivity of the DFA assay was 59.5%. In settings in which
the prevalence of C. trachomatis is high, the Biostar
Chlamydia OIA should be considered a patient point-of-care screening
device and should be a useful addition to the bench-top assays
currently available for the detection of chlamydia.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: Division of
Infectious Diseases, University of Alabama at Birmingham, 1900 University Blvd., THT-229, Birmingham, AL 35294-0006. Phone: (205)
934-5191. Fax: (205) 934-5155. E-mail:
ehook{at}uabid.dom.uab.edu.
 |
REFERENCES |
| 1.
|
Baselski, V. S.,
G. McNeely,
G. Ryan, and M. Robinson.
1987.
A comparison of nonculture-dependent methods for detection of Chlamydia trachomatis infections in pregnant women.
Obstet. Gynecol.
70:47-52[Medline].
|
| 2.
|
Drew, W. L.,
W. A. Clyde, Jr.,
G. E. Kenny, and J. Schachter.
1984.
Cumitech 19
Laboratory diagnosis of chlamydial and mycoplasmal infections. Coordinating ed., L. W. Drew. American Society for Microbiology, Washington, D.C.
|
| 3.
|
Eng, T. R., and W. T. Butler (ed.).
1997.
The hidden epidemic; confronting sexually transmitted diseases, p. 31.
and 44-45. National Academy Press, Washington, D.C.
|
| 4.
|
Green, T. A.,
C. M. Black, and R. E. Johnson.
1998.
Evaluation of bias in diagnostic test sensitivity and specificity estimates computed by discrepant analyses.
J. Clin. Microbiol.
36:375-378[Abstract/Free Full Text].
|
| 5.
|
Hadgu, A.
1996.
The discrepancy in discrepant analysis.
Lancet
348:592-593[Medline].
|
| 6.
|
Hook, E. W., III,
C. Spitters,
C. A. Reichart,
T. M. Neumann, and T. C. Quinn.
1994.
Use of cell culture and a rapid diagnostic assay for Chlamydia trachomatis screening.
JAMA
272:867-870[Abstract/Free Full Text].
|
| 7.
|
Howard, C.,
D. L. Friedman,
J. K. Leete, and M. L. Christensen.
1991.
Correlation of the percent of positive Chlamydia trachomatis direct fluorescent antibody detection tests with the adequacy of specimen collection.
Diagn. Microbiol. Infect. Dis.
14:233-237[Medline].
|
| 8.
|
Kellogg, J.
1995.
Impact of variation in endocervical specimen collection and testing techniques on frequency of false-positive and false negative chlamydia detection results.
Am. J. Clin. Pathol.
104:554-559[Medline].
|
| 9.
|
Kellogg, J. A.,
J. W. Seiple,
J. L. Klinedinst, and J. S. Levisky.
1991.
Impact of endocervical specimen quality on apparent prevalence of Chlamydia trachomatis infections diagnosed using an enzyme-linked immunosorbent assay method.
Arch. Pathol. Lab. Med.
115:1223-1227[Medline].
|
| 10.
|
Larsen, J. H.,
H. C. Wulf, and A. Friis-Moller.
1986.
Comparison of a fluorescent monoclonal antibody assay and a tissue culture assay for routine detection of infections caused by Chlamydia trachomatis.
Eur. J. Clin. Microbiol.
5:554-558[Medline].
|
| 11.
|
Parks, K. S.,
P. B. Dixon,
C. M. Richey, and E. W. Hook, III.
1997.
Spontaneous clearance of Chlamydia trachomatis infection in untreated patients.
Sex. Transm. Dis.
24:229-235[Medline].
|
| 12.
|
Pate, M. S.,
P. B. Dixon,
F. W. Schubert III, and E. W. Hook, III.
1994.
Comparison of the Sanofi Diagnostics Pasteur chlamydia microplate and Syva MicroTrak II enzyme immunoassay (EIA) to cell culture for the detection of Chlamydia trachomatis in a STD population. Poster presentation, abstr. C-499, p. 579.
In
Abstracts of the 94th General Meeting of the American Society for Microbiology 1994. American Society for Microbiology, Washington, D.C.
|
| 13.
|
Pothier, P., and A. Kazmierczak.
1986.
Comparison of cell culture with two direct chlamydia tests using immunofluorescence or enzyme-linked immunosorbent assay.
Eur. J. Clin. Microbiol.
5:569-572[Medline].
|
| 14.
|
Schachter, J.
1997.
DFA, EIA, PCR, LCR and other technologies: what tests should be used for diagnosis of chlamydia infections?
Immunol. Invest.
26:157-161[Medline].
|
| 15.
|
Schachter, J.,
W. E. Stamm,
T. C. Quinn,
W. W. Andrews,
J. Burczak, and H. H. Lee.
1994.
Ligase chain reaction to detect Chlamydia trachomatis infection of the cervix.
J. Clin. Microbiol.
32:2540-2543[Abstract/Free Full Text].
|
| 16.
|
Schwebke, J. R.,
R. Sadler,
J. M. Sutton, and E. W. Hook, III.
1997.
Positive screening tests for gonorrhea and chlamydial infections fail to lead consistently to treatment of patients attending a sexually transmitted disease clinic.
Sex. Transm. Dis.
24:181-184[Medline].
|
| 17.
|
Thejls, H.,
J. Gnarpe,
H. Gnarpe,
P. G. Larsson,
J. J. Platz-Christensen,
L. Ostergaard, and A. Victor.
1994.
Expanded gold standard in the diagnosis of Chlamydia trachomatis in a low prevalence population: diagnostic efficacy of tissue culture, direct immunofluorescence, enzyme immunoassay, PCR and serology.
Genitourin. Med.
70:300-303[Medline].
|
| 18.
|
Viscidi, R. P.,
L. Bobo,
E. W. Hook III, and T. C. Quinn.
1993.
Transmission of Chlamydia trachomatis among sex partners assessed by polymerase chain reaction.
J. Infect. Dis.
168:488-492[Medline].
|
Journal of Clinical Microbiology, August 1998, p. 2183-2186, Vol. 36, No. 8
0095-1137/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.
This article has been cited by other articles:
-
Gaydos, C. A
(2009). Can we climb out of the "pit" of poorly performing rapid diagnostic tests for chlamydia?. Sex. Transm. Infect.
85: 158-158
[Full Text]
-
Michel, C-E C, Saison, F G, Joshi, H, Mahilum-Tapay, L M, Lee, H H
(2009). Pitfalls of internet-accessible diagnostic tests: inadequate performance of a CE-marked Chlamydia test for home use. Sex. Transm. Infect.
85: 187-189
[Abstract]
[Full Text]
-
Bandea, C. I., Koumans, E. H., Sawyer, M. K., Dover, J., O'Connor, A., Papp, J. R., Unger, E. R., Braxton, J., Black, C. M.
(2009). Evaluation of the Rapid BioStar Optical Immunoassay for Detection of Chlamydia trachomatis in Adolescent Women. J. Clin. Microbiol.
47: 215-216
[Abstract]
[Full Text]
-
Mahilum-Tapay, L., Laitila, V., Wawrzyniak, J. J, Lee, H. H, Alexander, S., Ison, C., Swain, A., Barber, P., Ushiro-Lumb, I., Goh, B. T
(2007). New point of care Chlamydia Rapid Test bridging the gap between diagnosis and treatment: performance evaluation study. BMJ
335: 1190-1194
[Abstract]
[Full Text]
-
Peeling, R W
(2006). Testing for sexually transmitted infections: a brave new world?. Sex. Transm. Infect.
82: 425-430
[Full Text]
-
Dean, G L
(2006). Near-patient testing will not improve the control of sexually transmitted infections. Sex. Transm. Infect.
82: 509-512
[Full Text]
-
Swain, G. R., McDonald, R. A., Pfister, J. R., Gradus, M. S., Sedmak, G. V., Singh, A.
(2004). Decision Analysis: Point-of-care Chlamydia Testing vs. Laboratory-based Methods. Clin Med Res
2: 29-35
[Abstract]
[Full Text]
-
WATSON, E. J., TEMPLETON, A., RUSSELL, I., PAAVONEN, J., MARDH, P.-A., STARY, A., PEDERSON, B. S.
(2002). The accuracy and efficacy of screening tests for Chlamydia trachomatis: a systematic review. J Med Microbiol
51: 1021-1031
[Abstract]
[Full Text]
-
Chernesky, M., Jang, D., Krepel, J., Sellors, J., Mahony, J.
(1999). Impact of Reference Standard Sensitivity on Accuracy of Rapid Antigen Detection Assays and a Leukocyte Esterase Dipstick for Diagnosis of Chlamydia trachomatis Infection in First-Void Urine Specimens from Men. J. Clin. Microbiol.
37: 2777-2780
[Abstract]
[Full Text]