Journal of Clinical Microbiology, September 1998, p. 2495-2498, Vol. 36, No. 9
Projet RETRO-CI, CHU Treichville, Abidjan,
Côte d'Ivoire1;
Division of
HIV/AIDS Prevention, National Center for HIV, STD, and TB
Prevention, Centers for Disease Control and Prevention, Atlanta,
Georgia3; and
Institute of Tropical
Medicine, Antwerp, Belgium2
Received 23 March 1998/Returned for modification 27 April
1998/Accepted 11 June 1998
We compared the sensitivity and accuracy of the NucliSens assay and
those of both the standard and modified (addition of a new primer
set, primer mix 1, supplied by Roche) Amplicor HIV Monitor assays to
quantify human immunodeficiency virus type 1 (HIV-1) RNA in persons
infected with HIV-1 subtype A in Abidjan, Côte d'Ivoire.
Seventy-one plasma samples from HIV-1-seropositive persons at different
stages of HIV infection and 15 samples from HIV antibody-negative
persons were analyzed. The HIV-1 genetic subtype was determined either
by DNA sequencing or by a restriction fragment length polymorphism
assay. Of the 71 samples, 70 (98%) were subtype A and 1 was subtype G. Of the 70 subtype A samples, the proportion of RNA-positive plasma
samples and mean HIV-1 RNA levels were significantly higher by the
modified HIV Monitor assay (n = 67 [96%]; mean RNA
levels, 5.2 log10 HIV-1 RNA copies/ml) than the NucliSens
assay (n = 56 [80%]; 4.3 log10 HIV-1
RNA copies/ml) or the standard HIV Monitor assay (n = 44 [63%]; mean RNA levels, 3.8 log10 HIV-1 RNA
copies/ml) (all P values were <0.05). The HIV-1 RNA levels
by the modified HIV Monitor assay correlated significantly with those
by the NucliSens assay (r = 0.76; P < 0.001) and the standard HIV Monitor assay (r = 0.57; P < 0.001), as did the RNA levels by the
NucliSens and the standard HIV Monitor assays (r = 0.60; P < 0.001). Lower CD4 cell counts were
significantly correlated with higher HIV-1 RNA levels by all three
assays (r = The development of commercial
nucleic acid-based assays for the quantification of human
immunodeficiency virus type 1 (HIV-1) RNA represents a major advance in
the clinical management and follow-up of persons with HIV infection.
These assays, which are based on reverse transcriptase PCR, nucleic
acid sequence-based amplification (NASBA), or the branched DNA
technique, were developed and validated with persons infected with
HIV-1 subtype B, the predominant HIV-1 subtype in North America
and Europe. In several comparative evaluations, the three
commercially available nucleic acid-based techniques have proven
to be efficient for the quantification of the HIV-1 RNA load in persons
infected with HIV-1 subtype B viruses and are routinely used in
clinical practice (2, 3, 4, 6, 10-13). However, the high
degree of sequence diversity that characterizes the HIV-1 subtypes may
impair the sensitivity of currently available nucleic acid-based
assays. To date, limited data exist regarding the sensitivities and
accuracies of these assays for the detection of other known subtypes of
HIV group M (subtypes A to J) or HIV-1 group O. One study
indicates that the detection of HIV-1 subtype A, the predominant
subtype in Africa (8), is particularly difficult,
particularly by the Roche Amplicor HIV-1 Monitor and NASBA assays
(1). The poor sensitivities of these assays for the
detection of HIV-1 subtype A limits their clinical application in
studies of natural history, pathogenesis, and antiretroviral drug and
therapeutic vaccine trials in areas where this subtype predominates.
Recently, the Amplicor HIV-1 Monitor test has been modified in an
effort to improve its sensitivity for the quantification of non-B
subtypes of HIV-1, and a newer version of the NASBA assay, the
NucliSens assay, has been produced. In this study, we assessed the
sensitivities and accuracies of the standard and modified versions of
the Amplicor HIV-1 Monitor and the NucliSens assays for the
quantification of the HIV-1 RNA viral load in persons infected with
HIV-1 subtype A in Abidjan, Côte d'Ivoire.
Patient population.
A total of 86 plasma samples were tested
in this study, including 71 samples obtained from HIV-1-seropositive
individuals (47 tuberculosis patients, 17 commercial sex workers, and 7 pregnant women) and 15 samples obtained from HIV antibody-negative
commercial sex workers. Whole blood was drawn from these individuals
into VACUTAINER CPT tubes (Becton Dickinson, Rutherford, N.J.)
containing sodium citrate gel and density gradient media. Plasma was
separated from cells within 6 h of the time of collection by
centrifugation at 200 × g prior to aliquoting and
storage at HIV-1 RNA assays.
HIV-1 RNA levels in plasma were quantified
by the NucliSens assay (Organon Teknika, Boxtel, The Netherlands) and
by both the standard (hereinafter referred to as the standard HIV
Monitor test) and modified (hereinafter referred to as the modified HIV Monitor test) Amplicor HIV-1 Monitor tests (Roche Diagnostic Systems, Branchburg, N.J.). The modification of the HIV Monitor test consists of
20 µl of a new primer set (SK145 and SK151 [primer mix 1], provided
by Roche Diagnostics) that is added to the master mixture prior to
amplification. For all assays, HIV-1 RNA extraction, amplification, and
detection were performed according to the manufacturer's instructions.
The internal controls for the NucliSens assay consisted of three
synthetic RNAs (Qa, Qb, and Qc) of known high, medium, and low
concentrations, respectively. These RNAs serve as internal calibrators,
each differing from the HIV-1 wild-type RNA by only a small sequence.
The HIV Monitor assay uses three internal controls provided by the
manufacturer, including a negative and low- and high-positive controls,
respectively.
Analysis of data.
All RNA-negative samples were assigned a
viral load value equal to the detection limits of the assay: 400 HIV-1
RNA copies/ml (log10 2.6 RNA copies/ml) for the NucliSens
assay and 200 copies/ml (log10 2.3 RNA copies/ml) for the
HIV Monitor test. All HIV-1 RNA copy numbers were log transformed
before statistical analysis. However, the absolute numbers of HIV-1 RNA
copies per milliliter were used to calculate coefficients of
variations. HIV serologic status was used as a "gold standard" for
defining the sensitivities and specificities of the assays. Comparison
of the means were carried out by the paired Student t test,
and correlation was assessed with the Spearman rank correlation
coefficient. Statistical analyses were carried out with EPIINFO,
version 6 (Centers for Disease Control and Prevention, Atlanta, Ga.).
Sensitivities and specificities of the quantitative assays for
RNA.
Of the 71 HIV-1-positive samples, 70 (98%) were subtype A
and 1 was subtype G. Among the 70 subtype A samples, a higher
proportion of samples were RNA positive by the modified HIV Monitor
test (n = 67 [96%]) than by the NucliSens assay
(n = 56 [80%]) and the standard HIV Monitor assay
(44 [63%]) (all P values were <0.05). The three samples
that were not detected by the modified HIV Monitor assay were also
missed by both the standard HIV Monitor and the NucliSens assays. Of
the samples that were positive by the corresponding assays, the mean
viral load was significantly higher by the modified HIV Monitor assay
than by the NucliSens assay (n = 56; 5.1 versus 4.3 log10 HIV-1 RNA copies/ml; P < 0.001) and
the standard HIV Monitor assay (n = 44; 5.2 versus 3.8 log10 HIV-1 RNA copies/ml; P < 0.001). Of
the 42 samples that were positive by the NucliSens and standard
Amplicor Monitor assays, the mean viral load was significantly higher
by the NucliSens assay than by the standard Monitor HIV assay (4.5 versus 3.8 log10 HIV-1 RNA copies/ml; P < 0.001). In the one subtype G-positive plasma sample, viral RNA levels
were 3.9, 4.0, and 5.5 log10 copies/ml by the standard and
modified HIV Monitor assays and the NucliSens assay, respectively.
0095-1137/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.
Comparison of NucliSens and Amplicor Monitor Assays for
Quantification of Human Immunodeficiency Virus Type 1 (HIV-1) RNA in
Plasma of Persons with HIV-1 Subtype A Infection in Abidjan,
Côte d'Ivoire
![]()
ABSTRACT
Top
Abstract
Introduction
Materials & Methods
Results
Discussion
References
0.47 for the NucliSens assay,
0.45 for
the standard HIV Monitor assay, and
0.62 for the modified HIV Monitor
assay). These results indicate that the modified HIV Monitor assay has
the highest sensitivity and efficiency at quantifying the levels of RNA
in persons infected with HIV-1 subtype A and thus constitutes
a valuable tool for the monitoring of RNA levels in areas
of Africa were HIV-1 subtype A is predominant.
![]()
INTRODUCTION
Top
Abstract
Introduction
Materials & Methods
Results
Discussion
References
![]()
MATERIALS AND METHODS
Top
Abstract
Introduction
Materials & Methods
Results
Discussion
References
70°C. HIV antibody status was determined by
enzyme-linked immunosorbent assay testing and was confirmed by either
Western blotting or line immunoassays. CD4 lymphocyte cell counts were
enumerated by standard flow cytometry with a FACScan flow cytometer
(Becton Dickinson, San Jose, Calif.) and commercially available
monoclonal antibodies (Becton Dickinson, San Jose, Calif.). HIV-1
subtyping was determined either by DNA sequencing of the C2V3 region of
the env gene as described previously or by a restriction
fragment length polymorphism assay that is based on the protease gene
(5).
![]()
RESULTS
Top
Abstract
Introduction
Materials & Methods
Results
Discussion
References

View larger version (23K):
[in a new window]
FIG. 1.
Correlation between plasma HIV-1 subtype A RNA levels
obtained by the modified HIV Monitor, NucliSens, and standard HIV
Monitor assays. The diagonal lines indicate the calculated linear
regression curves (solid lines) and the 95% confidence limits (dashed
lines) for the assays. The vertical and horizontal solid lines
represent the detection limits of the assays.
Viral RNA load and CD4 lymphocyte counts.
To assess the
relationship between the level of immunosuppression and HIV-1 RNA
levels as measured by the NucliSens and the HIV Monitor assays, a
subset of 47 persons for whom CD4 counts were available were analyzed.
The median CD4 cell counts in these persons was 482 cells/mm3 (range, 36 to 1,390 cell/mm3). For
all three assays, we observed a similar and significant inverse correlation between decreasing CD4 cell counts and
increasing HIV-1 RNA levels (r =
0.47 for NucliSens
assay,
0.45 for standard HIV Monitor assay, and
0.62 for the
modified HIV Monitor assay (all P values were <0.005) (Fig.
2).
|
Reproducibility and specificity. Because the standard HIV Monitor assay had the lowest sensitivity, the reproducibilities and specificities of only the NucliSens and modified HIV Monitor assays were assessed. The reproducibility included the variation introduced by all steps of assay performance, including sample preparation, reverse transcription, amplification, and detection of PCR products. Reproducibility was determined in 10 different runs of the HIV Monitor assay by using one specimen whose RNA level (2,159 copies/ml) was known from the first test. In the 10 repeat runs, RNA levels in the specimen ranged from 1,548 to 3,602 copies/ml (mean ± standard deviation, 2,406 ± 674 copies/ml), resulting in a coefficient of variation of 28%. Similarly, for the NucliSens assay, one plasma specimen with initial RNA levels of 8,700 copies/ml was tested in 10 runs; RNA levels for this positive sample ranged from 8,400 to 14,000 copies/ml (mean ± standard deviation, 10,983 ± 2,603 copies/ml), with a coefficient of variation of 24%. The specificities of the modified HIV Monitor and NucliSens assays were assessed by testing 15 HIV-antibody negative samples, all of which gave negative results.
| |
DISCUSSION |
|---|
|
|
|---|
The field evaluation described here included a large panel of HIV-1 subtype A samples, the predominant HIV-1 subtype in Africa, and indicates that the modified HIV Monitor assay has a significantly higher sensitivity (96%) than the standard HIV monitor and NucliSens assays for the quantification of virus RNA levels in these samples. The sensitivities that we observed for the standard HIV Monitor (63%) and NucliSens (80%) assays were relatively higher than those reported previously for these assays (44 and 56%, respectively) by Alaeus and colleagues (1). This higher sensitivity may be explained by the fact that more than half of our study population were tuberculosis-positive HIV-1-infected persons, and thus, they had high viral RNA loads, whereas in the previously reported study of 27 persons infected with subtype A, most subjects either were asymptomatic or were undergoing antiretroviral therapy (1).
In addition to the sensitivity, the accuracy of RNA assays is critical. Because no gold standard for HIV-1 RNA quantification is currently available, the assessment of accuracy is difficult. However, several observations suggest that both the NucliSens and the modified HIV Monitor assays can accurately quantify RNA levels in HIV-1 subtype A-infected persons. First, both assays showed similar and significant inverse relationships between viral load and CD4 counts (Fig. 2), which is consistent with what has been reported for persons with HIV-1 subtype B infection (2, 3, 6, 7, 9, 11-13). Second, the RNA levels obtained by both the HIV Monitor and the NucliSens assays were significantly correlated. Lastly, although only a limited number of HIV-negative plasma samples were tested, the specificities of the assays were high (100%), and the high reproducibility observed was within the range reported for other viral RNA load assays. For instance, our observed coefficients of variation of 24 and 28% for the NucliSens and the modified HIV Monitor assays, respectively, are consistent with previously reported values, which ranged from 10 to 50% for the standard HIV Monitor assay, 20 to 30% for the branched DNA assay, and 15 to 60% for the NASBA test (2, 3, 6, 11-13).
An important question for investigators carrying out HIV-related laboratory research in Africa is the choice of the most sensitive and accurate assays for quantifying the viral RNA loads in the plasma of persons infected with the non-subtype B HIV-1 subtypes. A number of factors must be considered in addressing this question. Although sensitivity, specificity, and reproducibility are the most critical parameters, minimum sample volume, commercial availability, cost, ease of performance, shelf life, potential for contamination, turnaround time, and the need for refrigeration of reagents during shipping must also be taken into account. None of the available commercial assays satisfy all the considerations outlined above; however, each assay has its strengths and limitations. Both the NucliSens and HIV Monitor assays have the same ease of performance, use small samples volumes (100 to 200 µl), and have similar turnaround times (5 to 6 h) and high costs (US$ 50 to 60 per sample). Some strengths of the modified HIV Monitor assay include its higher sensitivity and low detection limit of 200 HIV-1 RNA copies/ml. Furthermore, the assay incorporates dUTP and uracil-N-glycosylase to prevent carryover contamination (contamination can easily be achieved by the NucliSens assay, given the numerous openings and closings of the tubes involved in the procedure), can be run on standard PCR and enzyme-linked immunosorbent assay equipment, and does not require refrigeration of reagents during shipment. The NucliSens assay has the advantage that it can be used with serum specimens as well as plasma samples collected on heparin (which is inhibitory to the HIV Monitor test), but some of its reagents require refrigeration during shipment, and its performance necessitates specialized instrumentation.
A possible limitation of our study is that the HIV-1 subtype designations that we used do not exclude the possibility for the occurrence of recombinant viruses, since the subtypes were based on either the env C2V3 region or the protease gene (for the restriction fragment length polymorphism assay), which are different from the gag region used in both the HIV Monitor and NucliSens assays. However, the proportion of recombinant HIV-1 is likely to be very low in Côte d'Ivoire; we have shown previously that more than 95% of HIV-1 infections are caused by subtype A viruses (5, 8).
In summary, we have shown that compared with the NucliSens and standard HIV Monitor tests, the modified HIV Monitor test has a significantly higher sensitivity for the quantification of HIV-1 RNA levels in persons infected with HIV-1 subtype A, thereby constituting a valuable tool for the monitoring of HIV-1 RNA levels in areas of Africa were HIV-1 subtype A is predominant. Similar comprehensive evaluations are needed from other parts of Africa where HIV-1 subtypes C and D predominates.
| |
ACKNOWLEDGMENTS |
|---|
We thank Kevin De Cock for critical review of the manuscript and for helpful suggestions and Joan-Luis Njampo, James Baggs, and Bharat Parekh for technical assistance.
This work was financially supported by the Division of HIV/AIDS Prevention, National Center for HIV, STD, and TB Prevention, Centers for Disease Control and Prevention, Atlanta, Ga.
| |
FOOTNOTES |
|---|
* Corresponding author. Mailing address: Projet RETRO-CI, Virology Laboratory, BP 1712, 01 Abidjan, Côte d'Ivoire. Phone: (225) 25 41 89. Fax: (225) 24 29 69. E-mail: jcn5{at}cdc.gov.
| |
REFERENCES |
|---|
|
|
|---|
| 1. | Alaeus, A., K. Lidman, A. Sönnerborg, and J. Albert. 1997. Subtype-specific problems with quantification of plasma HIV-1 RNA. AIDS 11:859-865[Medline]. |
| 2. | Cao, Y., D. Ho, J. Todd, R. Kokka, M. Urdea, J. Lifson, M. Piatak, S. Chen, B. Hahn, M. Saag, and G. Shaw. 1995. Clinical evaluation of branched DNA signal amplification for quantifying HIV type 1 in human plasma. AIDS Res. Hum. Retroviruses 11:353-361[Medline]. |
| 3. | Dewar, R., H. C. Highbarger, M. Sarmiento, J. Todd, M. B. Vasudevachari, R. Davey, J. Kovacs, N. Salzman, H. C. Lane, and M. S. Urdea. 1994. Application of branched DNA signal amplification to monitor human immunodeficiency virus type 1 burden in human plasma. J. Infect. Dis. 170:1172-1179[Medline]. |
| 4. | Dunne, A. L., and S. M. Crowe. 1997. Comparison of branched DNA and reverse transcriptase polymerase chain reaction for quantifying six different HIV-1 subtypes in plasma. AIDS 11:126-127[Medline]. |
| 5. | Ellenberger, D., D. Pieniazek, J. Nkengasong, C.-C. Luo, C. Maurice, M. Janini, A. Ramos, D. Hu, I.-M. Coulibaly, E. Ekpini, S. Z. Wiktor, A. E. Greenberg, G. Schochetman, and M. Rayfield. Unpublished data. |
| 6. |
Havlir, D. V., and D. D. Richman.
1996.
Viral dynamics of HIV: implication for drug development and therapeutic strategies.
Ann. Intern. Med.
124:984-994 |
| 7. |
Henrard, D. R.,
J. F. Philips,
L. R. Muenz,
W. A. Blattner,
D. Wiesner,
M. E. Eyster, and J. J. Goedert.
1995.
Natural history of HIV-1 cell free viremia.
JAMA
274:554-558 |
| 8. | Janssens, W., A. Buvé, and J. Nkengasong. 1997. The puzzle of HIV-1 subtypes in Africa. AIDS 11:705-712[Medline]. |
| 9. | Mellor, J. W., C. R. Rinaldo, P. Gupta, R. M. White, J. A. Todd, and L. A. Kingsley. 1996. Progression in HIV-1 infection predicted by the quantity of virus in plasma. Science 272:1167-1170[Abstract]. |
| 10. | Pachl, C., J. Todd, D. Kern, P. Sheridan, S.-J. Fong, M. Stempien, B. Hoo, D. Besemer, T. Yeghiazarian, B. Irvine, J. Kolberg, R. Kokka, P. Neuwald, and M. Urdea. 1995. Rapid and precise quantification of HIV-1 RNA in plasma using a branched DNA signal amplification assay. J. Acquired Immune Defic. Syndr. Hum. Retrovirol. 8:446-454[Medline]. |
| 11. | Revets, H., D. Marissens, S. De Wit, P. Lacor, N. Clumeck, S. Lauwers, and G. Zissis. 1996. Comparative evaluation of NASBA HIV-1 RNA QT, Amplicor-HIV monitor, and Quantiplex HIV RNA assay, three methods for quantification of human immunodeficiency virus type 1 RNA in plasma. J. Clin. Microbiol. 34:1058-1064[Abstract]. |
| 12. | Todd, J., T. Yeghiazarian, B. Hoo, J. Detmer, J. Kolberg, R. White, J. Wilber, and M. Urdea. 1994. Quantitation of human immunodeficiency virus plasma RNA by branched DNA and reverse transcriptase coupled polymerase chain reaction assay methods: a critical evaluation of accuracy and reproducibility. Serodiagn. Immunother. Infect. Dis. 6:233-239. |
| 13. | Vandamme, A. M., J. C. Schmit, S. Van Dooren, K. Van Laethem, E. Gobbers, W. Kok, P. Coubau, M. Witvrouw, W. Peetermans, E. De Clercq, and J. Desyter. 1996. Quantification of HIV-1 RNA in plasma: comparable results with the NASBA HIV-1 RNA QT and the Amplicor HIV monitor test. J. Acquired Immune Defic. Syndr. Hum. Retrovirol. 13:127-139[Medline]. |
This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»