Laboratorio di Microbiologia e Virologia,
Ospedale di Careggi, 50139 Florence, Italy
Received 17 February 1998/Returned for modification 24 March
1998/Accepted 11 June 1998
 |
TEXT |
Rapid diagnosis of tuberculosis is
an important component of measures to control the disease
(13), which is an objective made more urgent by the
appearance of multidrug-resistant strains (7).
Nucleic acid amplification techniques, which allow detection of
Mycobacterium tuberculosis directly from samples in a matter of hours, apparently represent the ultimate answer to this issue. They
lack sensitivity for smear-negative specimens, and their use is
generally restricted to selected samples. Culture remains the only
mycobacterial investigation performed, along with microscopy, for the
majority of clinical specimens.
The radiometric method (BACTEC; Becton Dickinson, Towson, Md.) is
at present the fastest and most sensitive cultural tool for the
diagnosis of mycobacterial infections (9). The
sensitivity of the radiometric assay allows a culture to reach a
growth index (GI) of
10, which is considered the positivity
threshold, in a very short time, often within a week. However, a BACTEC
vial flagged as positive does not necessarily harbor mycobacteria, since other organisms may overgrow despite the decontamination procedure. Confirmation of the presence of acid-fast bacilli
is required. Although microscopic tests are very easy to perform, they are characterized by a very low sensitivity, and examination of
smears prepared from BACTEC vials with early positive signals has
proven to be so frustrating that a delay in examination until the GI
becomes greater than 100 is recommended. Several days of further
incubation are often required, thus delaying the detection of
mycobacteria. Furthermore, a microscopically confirmed positive culture
may be due either to M. tuberculosis or to a nontuberculous mycobacterium (MOTT), which poses important dilemmas both for therapy
and for the measures to be adopted to prevent spread of the infection.
There is indeed interest in a procedure that would enable early
recognition of the M. tuberculosis complex as the organism responsible for the positive BACTEC signal; such a method would in
fact be consistent with the recommendations of the Centers for Disease
Control for a timely identification of isolates to the species level
(13).
For this purpose, we investigated the reliability of a commercial
ligase chain reaction (LCx M. tuberculosis; Abbott
Diagnostic Division, Abbott Park, Ill.) (14) to detect
M. tuberculosis in BACTEC vials on the same day on which
they yield the first positive signal.
A total of 150 consecutive BACTEC 12B (Becton Dickinson) vials with GI
scores of
10 at the usual reading were subjected to LCx amplification
following the procedure recommended for clinical specimens. BACTEC
readings were performed according to the recommended schedule, i.e.,
twice a week in the first 14 days of incubation and weekly in the
subsequent 4 weeks.
In short, 500 µl of each BACTEC broth was transferred to a
ready-to-use screw-cap microcentrifuge tube containing respiratory specimen buffer; the tube was subsequently vortexed and centrifuged (1,500 × g for 10 min) before the pellet was washed
and resuspended with resuspension buffer. After inactivation for 20 min
at 95°C in the LCx covered dry bath, the suspension was cooled to
room temperature and lysed for 10 min with the Lysor sonicator
(Abbott); after further centrifugation (9,000 × g for
2 min), 100 µl of supernatant was transferred to the ready-to-use
tube containing the amplification mixture. Amplification was carried
out in a separate area for 37 cycles in a thermal cycler (LCx Thermal
Cycler) as follows: 94°C for 1 s, 64°C for 1 s, and
69°C for 40 s, with maintenance at 25°C at the end of the last
cycle. For each series of tests, the provided negative control and
calibrator were prepared in duplicate and subjected to the same
amplification procedure as the samples. Amplified tubes were
pulse-centrifuged (10 to 15 s) and transferred unopened to the
carousel of the LCx analyzer, which directly detects amplification
products by a microparticle enzyme immunoassay, reporting the results
as fluorescence rates that are compared to the calibrator rate; results
greater than 30% of the average of the calibrator rate were considered
positive.
BACTEC 12B vials presenting a GI of <100 were further incubated and
read daily until a value of
100 was achieved, when an acid-fast smear
was prepared; the smear was immediately made from vials with initial GI
values of >100.
Each broth that was microscopically confirmed as positive for acid-fast
bacilli was subcultured on solid medium, and the mycobacteria were
identified with commercial DNA probes (AccuProbe, San Diego, Calif.) or
by resorting to conventional tests and/or to high-performance liquid
chromatographic analysis of cell wall mycolic acids. Overgrowing contaminants were grossly identified as such on the basis of morphology and Gram staining.
The GIs of the BACTEC vials included in the study ranged from 10 to 513 (average, 70; median, 30) for M. tuberculosis-positive specimens, from 10 to 999 (average, 337; median, 200) for MOTT-positive specimens, and from 14 to 690 (average, 173; median, 86) for
contaminants.
In 106 cases, the LCx gave a positive result and all of the
corresponding cultures grew M. tuberculosis. From
LCx-negative broths, MOTTs were isolated in 28 cases (eight M. avium isolates, seven M. xenopi isolates, six M. intracellulare isolates, two M. gordonae isolates, two
M. simiae isolates, one M. scrofulaceum isolate,
one M. genavense isolate, and one unidentified MOTT), contaminants were isolated in 14 cases (six gram-positive cocci, six
gram-negative bacilli, and two gram-positive bacilli), and M. tuberculosis was isolated in two cases. Sensitivity and
specificity were, therefore, 98.15 and 100%, respectively.
Several previous studies evaluated the possibility of a rapid detection
of M. tuberculosis in early-positive BACTEC vials; in the
majority of these, the commercial AccuProbe was used (1-4, 8, 10,
12). Despite the subjection to hybridization testing of only
strongly positive cultures (in no case were vials with GIs of <100
considered), that were concentrated by means of centrifugation, the
reported sensitivities ranged from 33 to 83%; only in one study, which
had a very low prevalence of M. tuberculosis, were nine
of nine positive cultures detected (2). A commercial
amplification assay (Amplicor MTB; Roche Molecular Systems, Somerville,
N.J.) has also been recently investigated for the same purpose;
although in that study only BACTEC cultures with GIs of
20 were
tested, a sensitivity that was not higher than 93% was obtained
(11). A sensitivity of 100% was observed in two studies
that used in-house PCRs (5, 6), but such procedures are
suitable only for specialized laboratories, since they require
well-trained personnel.
A thorough review of our data revealed that in both of our
false-negative results, the ratio of the LCx signal to cutoff
exceeded 0.50, compared to an average of 0.09 ± 0.15 in
true-negative results (Fig. 1);
furthermore, the GIs were only 10 and gave negative scores on the
following day. The apparent decrease in the GI value is frequent in
BACTEC vials having minimal growth when they are switched from the
regular to the daily reading schedule, because the first positive score
refers to the radiolabeled CO2 accumulated since the
previous reading from 3 or 7 days previously.

View larger version (15K):
[in this window]
[in a new window]
|
FIG. 1.
Correlation between LCx signal-to-cutoff ratio and GIs.
Empty circles, M. tuberculosis-negative cultures; black
triangles, M. tuberculosis-positive cultures.
|
|