Journal of Clinical Microbiology, January 1999, p. 233-234, Vol. 37, No. 1
Institut für Medizinische Mikrobiologie
und Immunologie, Universitäts-Krankenhaus Eppendorf,
20246 Hamburg, Germany
Received 21 January 1998/Returned for modification 6 May
1998/Accepted 13 October 1998
The rate of false-positive hepatitis C virus enzyme immunoassay
results was determined to be at least 10% among 1,814 reactive serum
samples based on (i) negative results in an independent confirmation
assay, (ii) negative PCR results, and (iii) no patients developing
clinical or biochemical signs of hepatitis during a 1-year follow-up.
In daily laboratory routine,
reliable diagnosis of hepatitis C virus (HCV) infection is not always
possible by sole use of an HCV enzyme immunoassay (EIA), since it is
well known that for a number of patients this assay produces
false-positive results (4, 8, 17). Therefore, results
obtained by EIA need to be confirmed by additional testing. However,
the commercially available assay RIBA 2.0 (Chiron Corporation) does not
fulfill the criteria defining a confirmation assay since it consists of recombinant proteins identical to those in the EIA (1, 5, 9). HCV PCR cannot be used for confirmation of positive EIA results, since a negative PCR result does not exclude the possibility of HCV infection with low-level viremia (below the limit of detection). Furthermore, PCR is too laborious and expensive to be used regularly as
a confirmatory assay. Therefore, we have established an HCV strip
immunoblot assay (SIA) (Universitäts-Krankenhaus Eppendorf [UKE] SIA) consisting of four recombinant proteins, derived from the
core and three nonstructural regions (NS3, NS4, and NS5) of HCV, which
are different from those used in the HCV EIA (5).
In the present study we compared the results of a second-generation HCV
EIA with those of the UKE SIA for 2,283 serum samples. The aim was to
assess the significance of positive results in the HCV EIA to define
criteria for the performance of further tests to reliably diagnose HCV
infection in the daily laboratory routine. Sera were drawn from 2,283 persons living in northern Germany around the city of Hamburg. They
were sent to our laboratory under suspicion of HCV infection due to
either elevated liver enzyme values (alanine aminotransferase, >45
U/liter) or clinical signs of hepatitis (jaundice and upper abdominal
pain) or risk factors for parenterally transmitted diseases, such as
chronic hemodialysis, blood transfusion, or intravenous drug use. At
the time of investigation they tested negative for acute infection with
HAV (anti-HAV immunoglobulin M antibodies) and HBV (hepatitis B surface
antigen). Repeated examinations were performed as follow-up every 3 months for 1 year. For serological screening a second-generation HCV
EIA (Abbott Laboratories, North Chicago, Ill.) was performed. For
confirmation of HCV EIA results, sera were tested in parallel by the
UKE SIA as previously described (5). The immunoblot assay
was considered positive when antibodies to at least two different
recombinant proteins were detectable. Reactivity against only a single
protein was rated as an indeterminate result. For detection of HCV RNA
reverse transcription-PCR was performed as previously described
(6, 7).
The HCV EIA was negative for 469 samples, of which 456 (97%) were also
negative by UKE SIA. For 13 samples the UKE SIA was considered
indeterminate. All 469 of these sera were negative by HCV PCR, and none
of the patients developed clinical or biochemical signs of hepatitis
during the follow-up.
The HCV EIA was reactive for 1,814 samples, of which 1,394 (77%) were
also positive by the UKE SIA (Table 1).
However, in 240 cases (13%) the reactivity in the HCV EIA could
not be confirmed by UKE SIA. Suitable specimens for HCV PCR were
available for 193 of these 240 samples, and a positive PCR result was
obtained with 13 samples. Of these, nine became positive by UKE SIA
when retested after 3 months, which suggests that these patients had acquired HCV infection shortly prior to the first examination. In the
remaining four patients, who repeatedly tested PCR positive despite a
negative result by UKE SIA, immunosuppressing conditions could be
found. One had a B-cell lymphoma, one was chronically hemodialyzed, and
two practiced intravenous drug use. It has been shown earlier that in
patients with immunosuppressive conditions, serological response is low
or even absent (10, 14, 15). This could lead to negative or
indeterminate results in serological assays although the individual
suffers from infection with HCV (13). Therefore, for
patients with known immunosuppressive disorders PCR should always be
performed. The 180 initially PCR-negative subjects remained negative by
UKE SIA and HCV PCR in repeated examinations during the follow-up.
Moreover, these patients did not develop clinical or biochemical signs
of hepatitis. This indicates that in at least these 180 samples (10%),
false-positive results occurred. We must assume that the EIA was also
false positive in the specimens for which no suitable material for PCR
was available, since the UKE SIA remained negative and none of the
patients developed clinical or biochemical signs of hepatitis during
the follow-up. This indicates that as long as no better screening
assays are commercially available every positive HCV EIA result must be
confirmed.
0095-1137/99/$04.00+0
Copyright © 1999, American Society for Microbiology. All rights reserved.
Definition of False-Positive Reactions in Screening
for Hepatitis C Virus Antibodies
![]()
ABSTRACT
Top
Abstract
Text
References
![]()
TEXT
Top
Abstract
Text
References
TABLE 1.
Comparison of results of Abbott second-generation HCV EIA
and UKE SIA for 2,283 serum samples
An indeterminate result in the UKE SIA was observed with 180 of the 1,814 EIA-positive samples (10%). Suitable specimens for HCV PCR were obtained for 134 of these 180 samples, and HCV RNA could be detected in 58 of them. During the follow-up full seroconversion was observed in four patients. All of them initially revealed antibodies directed solely against the NS3 protein of the UKE SIA. In follow-up samples, reactivity against additional recombinant proteins emerged. These results support the previous assumption that antibody reactivity against NS3 plays an important role in the early serological detection of HCV infection (5). Furthermore, a particularly high correlation has been found between HCV viremia and antibody reactivity against the c33c antigen of the commercially available RIBA (2). Samples with a positive result by HCV EIA and an indeterminate result by immunoblot assay must be subjected to PCR, since we detected HCV RNA in 43% of samples (58 of 134). The percentage of indeterminate results by the UKE SIA is remarkably low compared to that by RIBA 2.0 or 3.0 (2, 3, 11, 16). One reason for this might be that local isolates were used to establish the UKE SIA, since serological tests containing recombinant proteins of local isolates have been shown to have better sensitivity and specificity than commercially available assays (5, 12). However, this is unlikely to be the only reason, since the UKE SIA was evaluated with serum samples containing a variety of HCV genotypes as previously described (5).
The diagnosis HCV positive has a deep impact on the life of the afflicted person. Therefore, it must be reached as reliably as possible. Our data indicate that the widely used HCV EIA produces a high percentage (10%) of false-positive results. Compared to other screening assays, e.g., human immunodeficiency virus EIAs, this is unacceptably high. Therefore, confirmation of every positive HCV EIA result by supplemental tests is mandatory. As we have shown with our in-house UKE SIA, one possibility for improving the reliability of HCV diagnosis is to introduce proteins into the confirmation assay which are different from those used in the screening assay.
| |
FOOTNOTES |
|---|
* Corresponding author. Mailing address: Institut für Medizinische Mikrobiologie und Immunologie, Universitäts-Krankenhaus Eppendorf, Martinistr. 52, 20246 Hamburg, Germany. Phone: 49-40.47173159. Fax: 49-40.47174062. E-mail: mschroeter{at}uke.uni-hamburg.de.
| |
REFERENCES |
|---|
|
|
|---|
| 1. | Bresters, D., H. L. Zaaijer, H. T. M. Cuypers, H. W. Reesing, I. N. Winkel, P. J. van Exel-Oehlers, A. A. J. van Drimmelen, P. L. M. Jansen, C. L. van der Poel, and P. N. Lelie. 1993. Recombinant immunoblot assay reaction patterns and hepatitis C virus RNA in blood donors and non-A, non-B hepatitis patients. Transfusion 33:634-638[Medline]. |
| 2. | Chicheportiche, C., J. F. Cantaloube, P. Biagini, P. Aumont, F. Donnadieu, J. Escher, F. Larabi, and J. P. Zepitelli. 1993. Analysis of ELISA hepatitis C virus-positive blood donors population by polymerase chain reaction and recombinant immunoblot assay (RIBA). Comparison of second and third generation RIBA. Acta Virol. 37:123-131[Medline]. |
| 3. | Damen, M., H. L. Zaaijer, H. T. Cuypers, H. Vrielink, C. L. van der Poel, H. W. Reesink, and P. N. Lelie. 1995. Reliability of the third-generation recombinant immunoblot assay for hepatitis C virus. Transfusion 9:745-749. |
| 4. | Dow, B. C., I. Coote, H. Munro, F. McOmish, P. L. Yap, P. Simmonds, and E. A. C. Follet. 1993. Confirmation of hepatitis C virus antibody in blood donors. J. Med. Virol. 41:215-220[Medline]. |
| 5. | Feucht, H. H., B. Zöllner, S. Polywka, and R. Laufs. 1995. Study on reliability of commercially available hepatitis C virus antibody tests. J. Clin. Microbiol. 33:620-624[Abstract]. |
| 6. | Feucht, H. H., B. Zöllner, M. Schröter, H. Altrogge, and R. Laufs. 1995. Tear fluid of hepatitis C virus carriers could be infectious. J. Clin. Microbiol. 33:2202-2203[Abstract]. |
| 7. | Feucht, H. H., B. Zöllner, M. Schröter, A. Hoyer, M. Sterneck, and S. Polywka. 1996. Distribution of genotypes and response to alpha interferon in patients with hepatitis C virus infection in Germany. Eur. J. Clin. Microbiol. Infect. Dis. 15:128-132[Medline]. |
| 8. |
Gretch, D.,
W. Lee, and L. Corey.
1992.
Use of aminotransferase, hepatitis C antibody, and hepatitis C polymerase chain reaction RNA assays to establish the diagnosis of hepatitis C virus infection in a diagnostic virology laboratory.
J. Clin. Microbiol.
30:2145-2149 |
| 9. | Lelie, P. N., H. T. Cuypers, H. W. Reesink, C. L. van der Poel, I. N. Winkel, E. Bakker, P. J. van Exel-Oehlers, D. Vallari, J. P. Allain, and L. Mimms. 1992. Patterns of serological markers in transfusion-transmitted hepatitis C virus infection using a second-generation HCV assay. J. Med. Virol. 37:203-209[Medline]. |
| 10. | Lok, A. S., D. Chien, Q. L. Choo, T. M. Chan, E. K. Chien, I. K. Cheng, M. Houghton, and G. Kuo. 1993. Antibody response to core, envelope and nonstructural hepatitis C virus antigen: comparison of immunocompetent and immunosuppressed patients. Hepatology 18:497-502[Medline]. |
| 11. | McHutchison, J. G., J. L. Person, S. Govindarajan, B. Valinluck, T. Gore, S. R. Lee, M. Nelles, A. Polito, D. Chien, R. DiNello, S. Quan, G. Kuo, and A. G. Redeker. 1992. Improved detection of hepatitis C virus antibodies in high-risk populations. Hepatology 15:19-25[Medline]. |
| 12. | Neville, J. A., L. E. Prescott, V. Bhattacherjee, N. Adams, I. Pike, B. Rodgers, A. El-Zayadi, S. Hamid, G. M. Dusheiko, A. A. Saeed, G. H. Haydon, and P. Simmonds. 1997. Antigenic variation of core, NS3, and NS5 proteins among genotypes of hepatitis C virus. J. Clin. Microbiol. 35:3062-3070[Abstract]. |
| 13. |
Pawlotzky, J.-M.,
A. Fleury,
V. Choukroun,
L. Deforges,
F. Roudot-Thoraval,
P. Aumont,
J. Duval, and D. Dhumeaux.
1994.
Significance of highly positive c22-3 "indeterminate" second-generation hepatitis C virus (HCV) recombinant immunoblot assay (RIBA) and resolution by third-generation HCV-RIBA.
J. Clin. Microbiol.
32:1357-1359 |
| 14. | Pereira, B. J. G., E. L. Milford, R. L. Kirkman, and A. S. Levey. 1991. Transmission of hepatitis C virus by organ transplantation. N. Engl. J. Med. 325:454-460[Abstract]. |
| 15. | Schröter, M., H. H. Feucht, P. Schäfer, B. Zöllner, and R. Laufs. 1997. High percentage of seronegative HCV-infections in hemodialysis patients: the need for PCR. Intervirology 40:277-278[Medline]. |
| 16. | Tobler, L. H., M. P. Busch, J. Wilber, R. Dinello, S. Quan, A. Polito, R. Kochesky, C. Bahl, M. Nelles, and S. R. Lee. 1994. Evaluation of indeterminate c22-3 reactivity in volunteer blood donors. Transfusion 2:130-134. |
| 17. | van der Poel, C. L., H. T. Cuypers, H. W. Reesing, A. J. Weiner, S. Quan, R. DiNello, J. J. P. van Boven, I. N. Winkel, D. Mulder-Folkerts, P. J. Exel-Oehlers, W. Schaasberg, A. Leentvar-Kuypers, A. Polito, M. Houghton, and P. N. Lelie. 1991. Confirmation of hepatitis C infection by new four-antigen recombinant immunoblot assay. Lancet 337:317-319[Medline]. |
This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»