Previous Article | Next Article 
Journal of Clinical Microbiology, October 1999, p. 3102-3107, Vol. 37, No. 10
0095-1137/99/$04.00+0
Copyright © 1999, American Society for Microbiology. All rights reserved.
Multicenter Evaluation of the Abbott LCx
Mycobacterium tuberculosis Ligase Chain Reaction
Assay
Richard
Lumb,1,*
Kirsten
Davies,2
David
Dawson,3
Robert
Gibb,3
Thomas
Gottlieb,2
Clair
Kershaw,4
Katherine
Kociuba,4
Graeme
Nimmo,3
Norma
Sangster,1
Michele
Worthington,4 and
Ivan
Bastian1
Infectious Diseases Laboratories, Institute
of Medical and Veterinary Science, Adelaide, South
Australia,1 Department of Microbiology
and Infectious Diseases, Concord Hospital, Concord, New South
Wales,2 Queensland Health Pathology
Services, Brisbane, Queensland,3 and
Department of Microbiology and Infectious Diseases, South
Western Area Pathology Services, Liverpool, New South
Wales,4 Australia
Received 2 April 1999/Returned for modification 10 May
1999/Accepted 28 June 1999
 |
ABSTRACT |
Four Australian hospital laboratories evaluated the performance of
the Abbott LCx Mycobacterium tuberculosis assay with 2,347 specimens (2,083 respiratory and 264 nonrespiratory specimens) obtained
from 1,411 patients. A total of 152 specimens (6.5%) were culture
positive for Mycobacterium tuberculosis complex (MTBC); of
these, 79 (52%) were smear positive. After resolution of discrepant data, the overall sensitivity, specificity, and positive and negative predictive values for the LCx assay were 69.7, 99.9, 99.1, and 97.7%
respectively. For smear-positive respiratory specimens that were
culture positive for MTBC, the values were 98.5, 100, 100, and 98.4%,
respectively, while the values for smear-negative respiratory specimens
were 41.5, 99.9, 96.4, and 98%, respectively. Relative operating
characteristic curves were constructed to demonstrate the relationship
between sensitivity and specificity for a range of possible cutoff
values in the LCx assay. These graphs suggested that the assay
sensitivity for respiratory samples could be increased from 70.2 to
78.6%, while the specificity would be reduced from 99.9 to 99.4% by
inclusion of a grey zone (i.e., LCx assay values of between 0.2 and
0.99). An algorithm is presented for the handling of specimens with LCx
assay values within this grey zone.
 |
INTRODUCTION |
Culture is usually required for the
laboratory confirmation of tuberculosis. The resulting isolate is also
necessary to identify the organism to the species level, to determine
drug susceptibility, and to obtain a molecular profile for
epidemiological purposes. Unfortunately, substantial time delays occur
(21) because conventional methods may require up to 8 weeks
for cultures to become positive, although the radiometric systems
(2) and the newer, nonradiometric, continuous monitoring
systems (4, 29, 37, 44) have reduced the time to culture
positivity. During this delay, patients with suspected tuberculosis and
smears negative for acid-fast bacilli may be subjected to bronchoscopy
or other invasive procedures to obtain a diagnosis or may be commenced
on antituberculosis therapy. The worldwide reemergence of tuberculosis,
the rise of multidrug-resistant Mycobacterium tuberculosis
(12, 33), and the ongoing transmission of tuberculosis
within and between high-risk groups (16, 40) have
accelerated the search for more sensitive and rapid diagnostic
laboratory methods.
The development of PCR and other nucleic acid amplification techniques
has led to the introduction of in-house and commercial assays for the
detection of M. tuberculosis complex (MTBC) directly from
processed clinical samples. The advantages of commercial systems are
that they are optimized and validated tests, they specifically identify
the amplified product, and they use simplified protocols with greater automation.
The Abbott LCx Mycobacterium tuberculosis (LCx) assay
(Abbott Diagnostics Division, Abbott Park, Ill.) uses the ligase chain reaction for the amplification of a segment of the single-copy gene
that encodes protein antigen b. The gene is specific for members of the
MTBC (38). The LCx assay was designed for use with processed
respiratory specimens, although two studies have evaluated the assay
with nonrespiratory specimens (18, 31). The aims of the
present study were to evaluate the performance of the LCx assay with
respiratory and nonrespiratory specimens and to review the setting of
the sample rate/cutoff value, presently set at a value of 1.0, to
determine whether the cutoff value may be reduced to improve test
sensitivity without unduly compromising specificity.
 |
MATERIALS AND METHODS |
Clinical specimens and patients.
Four experienced
mycobacteriology laboratories in Australia evaluated the LCx assay
(Abbott Laboratories). Respiratory and nonrespiratory specimens were
collected from patients under investigation for tuberculosis. Once the
specimens had reached the laboratory, they were stored at 4°C until
they were processed. After processing, an aliquot of sample was stored
at
20 or
70°C until testing by the LCx assay. For specimens to be
included in the study, microscopy, culture, and LCx assay results plus
details about the patients including antituberculosis therapy were
required. Specimens for which cultures were discontinued due to
contamination were excluded from the study.
Decontamination and culture protocols.
All four laboratories
used digestion and decontamination protocols for specimens likely to
contain contaminating organisms. The processing and culture protocols
used in each of the four participating laboratories are summarized in
Table 1. BACTEC 12B vials containing 0.1 ml of PANTA-plus supplement were inoculated according to the
manufacturer's recommendations (0.5 ml for decontaminated specimens;
up to 1.0 ml for specimens from usually sterile sites) onto
Löwenstein-Jensen slants, which received 0.2 to 0.4 ml of specimen. MB/BacT vials containing 0.5 ml of antibiotic supplement including vancomycin) were inoculated according to the manufacturer's recommendations (up to 0.5 ml for all specimen types). Inoculated BACTEC 12B and MB/BacT vials were incubated for a minimum of 6 weeks,
although smear-positive specimens were incubated for an additional 4 weeks. Löwenstein-Jensen media were incubated for a minimum of 10 to 12 weeks, regardless of the smear status. Samples from sterile sites
were cultured without decontamination. Tissues were macerated in
sterile saline, and material from swabs was resuspended in sterile
saline. Cultures of specimens obtained from superficial sites were
incubated at 30 to 32°C and at 35 to 36°C. Laboratories that used
the fluorochrome staining technique confirmed the results for all new
smear-positive specimens by overstaining with the Ziehl-Neelsen stain.
Smears were quantified by a recognized reporting scheme
(22).
Identification of mycobacteria.
All isolates were checked
for acid fastness. Conventional identification procedures based on
physical and biochemical properties (22), multiplex PCR
(13), commercial probes for culture confirmation (Accuprobe;
Gen-Probe, Inc., San Diego, Calif.) (17, 25), or 16S rRNA
gene sequence analysis (35) were used for the identification of mycobacterial or nocardial isolates.
LCx assay protocol.
The LCx assay was performed according to
the manufacturer's recommendations and comprised three stages:
specimen preparation to remove potential amplification inhibitors,
amplification by LCR technology, and detection via a microparticle
enzyme immunoassay (MEIA). Duplicate negative and calibrator controls
are included in each run. The MEIA result for each sample is compared
with the averaged LCx assay calibrator control value and is expressed as the ratio of the sample rate/cutoff (S/CO) value multiplied by 0.3. Samples with an S/CO value of 1.0 or greater were considered positive.
The four participating laboratories also included a positive control
(e.g., M. bovis BCG) and a negative control (e.g., an
atypical mycobacterium) in each run to confirm the assay performance.
Resolution of discrepant results.
A discrepancy occurred
when the results of culture and the LCx assay differed from each other.
In these situations, a review of the patient's medical records was
undertaken to determine a clinical diagnosis of tuberculosis or disease
caused by MTBC on the basis of symptoms, signs, X rays, results of
other laboratory investigations, and response to antituberculosis
treatment. In this analysis for resolution of discrepant results, the
LCx assay results were compared with the combined results of the
"gold standard" of culture and the final clinical diagnosis.
Statistical analysis.
The data were tabulated and analyzed
with a commercial computer program (Microsoft Excel 97; Microsoft
Corporation, Redmond, Wash.). Epi Info (version 6.04b; Centers for
Disease Control and Prevention, Atlanta, Ga.) was used to perform
chi-square analyses to compare the performance characteristics of the
LCx assay in each of the participating laboratories and to calculate
the confidence intervals surrounding various estimates of sensitivity
and specificity. Relative operating characteristic (ROC) curves were
also constructed by plotting the sensitivity and specificity of the LCx
assay for a range of cutoff values (5).
 |
RESULTS |
Specimen and patient analysis.
A total of 2,347 specimens were
examined by microscopy, culture, and the LCx assay. Of these, 2,083 specimens were of respiratory origin (sputum or tracheal aspirates,
1,659; bronchoscopy specimens, 420; and gastric aspirates, 4). There
were 264 nonrespiratory specimens (biopsy specimens, 98; pleural fluid,
49; lymph node, 36; other sterile site aspirates, 30; pus, 14;
cerebrospinal fluid, 11; urine, 8; feces, 3), plus another 15 specimens
from miscellaneous sites. The specimens were obtained from 1,411 patients; for 1,332 (94.4%) of the patients three or fewer specimens
were tested in the study. The mean numbers of samples tested from each
patient were 1.6 (standard deviation [SD], 1.1), 1.7 (SD, 1.1), and
1.2 (SD, 0.7) for all specimens combined, for respiratory samples, and
for samples from nonrespiratory sites, respectively.
Culture and smear results.
Of the 2,347 specimens, 152 (6.5%)
specimens (119 respiratory and 33 nonrespiratory specimens) from 98 patients were culture positive for MTBC (M. tuberculosis,
150; M. bovis BCG, 2). Three or fewer specimens were tested
for 94% of the patients with MTBC culture-positive specimens. Of the
MTBC culture-positive specimens, 79 (52%) had a positive smear result
(respiratory specimens, 66; nonrespiratory specimens, 13).
Atypical mycobacteria were isolated from 112 specimens. The species
isolated included M. avium complex (n = 67),
M. abscessus (n = 17), M. fortuitum (n = 5), M. chelonae
(n = 2), M. kansasii (n = 1), M. terrae (n = 1), and M. ulcerans (n = 1). A further 18 single isolates
were not identified. Nocardia asteroides was recovered from
a single respiratory specimen. None of these specimens were positive by
the LCx assay.
Initial analysis of Abbott LCx assay.
The results of the LCx
assay and microscopy were compared with those of culture and then with
the results of the revised gold standard at four Australian hospital
laboratories. No significant differences were noted in the performance
of the LCx assay between the four laboratories (data not shown). For
example, there was no statistical difference between the four estimates
of the sensitivity of the LCx assay when culture was used as the gold
standard (
2 = 1.71, degrees of freedom = 3, P = 0.63). The pooled results from the four
laboratories are therefore presented in the following analyses. For all
specimens, when culture was used as the gold standard, microscopy had a
sensitivity, specificity, positive predictive value (PPV), and negative
predictive value (NPV) of 52, 96.2, 48.8, and 96.7%, respectively.
When stratified on the basis of specimen site, the sensitivities of
microscopy for respiratory and nonrespiratory specimens were 55.5 and
39.4%, respectively.
When the LCx assay was compared to culture, the sensitivity,
specificity, PPV, and NPV for all specimens were 75, 98.9, 80.9,
and
98.3%, respectively. When stratified by specimen site, the
values for
respiratory specimens were 77.3, 98.9, 80.7, and 98.6%,
respectively,
and the values for nonrespiratory specimens were
66.7, 97.8, 81.5, and
95.4%, respectively (Table
2). The
performance
of the LCx assay for smear-positive and smear-negative
respiratory
specimens is presented in Table
2. Notably, the sensitivity
for
smear-positive respiratory specimens was 98.5%, but it was only
50.9% for those that were smear negative. A similar analysis by
smear
status was not performed for the nonrespiratory specimens
because of
the smaller sample size.
Discrepant analysis of Abbott LCx assay.
Laboratory and
clinical data for specimens with discordant results were reviewed. This
analysis for resolution of discrepant results involved the exclusion
and reclassification of some respiratory and nonrespiratory specimens:
42 specimens from patients on treatment longer than 7 days were
excluded (24 LCx assay-positive and culture-negative specimens, 17 LCx
assay-negative and culture-negative specimens, and 1 LCx assay-positive
and culture-positive specimen), 2 culture-negative specimens with high
LCx assay values were reclassified as MTBC positive on the basis of
clinical criteria, as were 12 LCx assay-negative and culture-negative
samples. Of the 14 specimens reclassified as MTBC positive, 9 were from
patients from whom samples were collected from the same or a similar
site and whose samples were culture positive for MTBC. The remaining
five specimens came from patients without laboratory confirmation of
tuberculosis but whose clinical presentation, radiographic findings,
and response to treatment resulted in a final diagnosis of
tuberculosis. In the analysis of the data for resolution of discrepant
results, the sensitivity, specificity, PPV, and NPV of the LCx assay
were 69.7, 99.9, 99.1, and 97.7%, respectively. For respiratory
specimens, the values were 70.2, 99.9, 98.9, and 98.0%, respectively,
and for nonrespiratory specimens, the values were 67.6, 100, 100, and
95.3%, respectively (Table 3).
View this table:
[in this window]
[in a new window]
|
TABLE 3.
Assessment of LCx assay results after analysis for
resolution of discrepant results (culture result plus clinical
diagnosis)
|
|
The analysis for resolution of discrepant results also found that the
LCx assay performed better for smear-positive respiratory
specimens
than for smear-negative respiratory specimens (Table
3). For
smear-positive respiratory specimens, the values were
98.5, 100, 100, and 98.4%, respectively, and for smear-negative
respiratory specimens,
the values were 41.5, 99.9, 96.4, and 98.0%,
respectively.
The estimated sensitivities of the LCx assay for smear-positive and
smear-negative nonrespiratory samples were 84.6% (95%
confidence
interval, 53.7 to 97.2%) and 57.1% (95% confidence
interval, 34.4 to
77.4%), respectively. These confidence intervals
are wide because only
34 nonrespiratory specimens had a final
diagnosis of tuberculosis
(Table
3).
ROC curve analysis.
In the LCx assay, samples with results
that fall above or below an established cutoff value are defined as
positive or negative, respectively. The cutoff value is the mean rate
for the LCx assay calibrator duplicates multiplied by 0.30. When the
S/CO value is 1.0 or greater, the LCx assay result is interpreted as
positive, while S/CO values of less than 1.0 indicate a negative result.
In addition to evaluating the performance of the LCx assay at this one
predetermined cutoff value, ROC curves were constructed
for respiratory
and nonrespiratory specimens by using the resolved
data set to show the
correlation of sensitivity and specificity
over a range of cutoff
values. Inspection of Fig.
1 suggested
that lowering of the cutoff to 0.2 could improve sensitivity with
only
a marginal reduction in specificity for both specimen types.
This
impression was confirmed by reanalyzing the LCx results for
respiratory
specimens (of which this study had a large cohort
of 2,083 specimens)
and comparing the sensitivity and specificity
of the assay at different
cutoff values (Table
4). Inclusion
of a
grey zone of LCx assay values of from 0.2 to 0.99 improved
the
sensitivity by 8.4%, while it reduced the specificity by only
0.5%.
Overall, 21 respiratory samples, including 11 positive for
MTBC by
analysis for resolution of discrepant results, and 3 nonrespiratory
samples, including 2 positive for MTBC by analysis for resolution
of
discrepant results, produced LCx assay values within this grey
zone. Of
the 13 specimens whose values fell within this grey zone
but that were
ultimately classified as MTBC positive, only one
respiratory specimen
was smear positive.

View larger version (23K):
[in this window]
[in a new window]
|
FIG. 1.
ROC curve for LCx assay for respiratory and
nonrespiratory specimens. These ROC curves were constructed on the
basis of the performance of the Abbott LCx assay in the analysis for
resolution of discrepant results for isolates from respiratory ( )
and nonrespiratory ( ) specimens. The boxed area represents the
performance of the assay for cutoff values of 0.1 to 1.0. Table 4
describes the sensitivity and specificity of the assay for various
cutoff values within this range.
|
|
The use of this grey zone was then investigated by reviewing the S/CO
values for both respiratory and nonrespiratory specimens.
The mean S/CO
value for specimens positive for MTBC by analysis
for resolution of
discrepant results was 3.05 (range, 0.03 to
7.56). Of the 50 specimens
that produced false-negative LCx assay
results, 13 had values within
the grey zone. Adequate volumes
remained for retesting of five of these
specimens; the grey zone
LCx assay values were reproducible for all
five specimens, with
the results of repeat LCx tests ranging from 0.45 to 1.16.
The mean S/CO value for 2,140 specimens negative for MTBC by analysis
for resolution of discrepant results was 0.06 (range,
0.03 to 1.08).
Only 11 of these MTBC-negative specimens had S/CO
values between 0.2 and 0.99. Sufficient volumes remained for nine
of the specimens to be
retested; eight returned an S/CO value
of 0.04 or less. The other
specimen had an initial S/CO value
of 0.73 and a value of 0.66 on
repeat testing. This sputum specimen
was negative by microscopy and
culture, and an additional sputum
sample was collected and was negative
by all assays, including
the LCx
assay.
Finally, one sputum specimen reproducibly gave an LCx assay value of
1.08 that was defined as a false-positive result. The
specimen was
collected from an elderly nursing home resident who
had been exposed to
a smear-positive tuberculosis patient. This
individual who had contact
with a patient with tuberculosis has
been monitored for more than 12 months, and the results of all
other microbiological and radiological
investigations have remained
negative.
 |
DISCUSSION |
This collaborative study is the most extensive evaluation of the
LCx assay published to date and was based on the testing of 152 culture-positive samples from among 2,083 respiratory and 264 nonrespiratory specimens. After resolution of discrepant results, the
sensitivity, specificity, PPV, and NPV for respiratory specimens were
70.2, 99.9, 98.9, and 98.0%, respectively. These findings agree with
previous evaluations of the LCx assay, which reported values of 77.1 to
90.2, 98.4 to 100, 72.8 to 100%, and 90.5 to 99.5%, respectively
(3, 20, 24, 28, 31, 36, 39, 45). The assay sensitivities for
smear-positive and smear-negative respiratory specimens were 98.5 and
41.5%, respectively (Table 3). Again, these values are similar to
those reported previously (i.e., 92.1 to 100 and 36.8 to 73.3%, respectively).
On the basis of sensitivity and specificity, the performance of the LCx
assay is comparable to those of other commercially available nucleic
acid amplification tests (NAATs). For resolved data, the sensitivities
for smear-positive and smear-negative pulmonary specimens were 93 to
100 and 43 to 81%, respectively, for the Gen-Probe amplified
Mycobacterium tuberculosis direct test (8, 9, 14, 26,
30, 41, 42), 89 to 97.6 and 42.9 to 74%, respectively, for the
Roche Mycobacterium tuberculosis Amplicor test (6, 7,
10, 14, 15, 27), and 90 to 96 and 48 to 68%, respectively, for
the Cobas Amplicor test (32, 34). Although the LCx assay is
not licensed for use with nonrespiratory specimens, the sensitivities
were 84.6 and 57.1% for smear-positive and smear-negative specimens,
respectively, in the present evaluation; these findings were comparable
to those of other studies investigating the performance of the LCx
assay with nonrespiratory specimens (81.8 to 100 and 35.3 to 71.1%,
respectively (18, 31).
However, comparison of NAATs for the detection of MTBC nucleic acid
directly from clinical specimens is problematic for several reasons.
The proportion of smear-positive and culture-positive specimens in a
study cohort affects the estimate of assay sensitivity because the
ability of an NAAT to detect MTBC DNA is directly related to the
concentration of bacilli in the specimen (3, 27, 28). For
example, Moore and Curry (28) found that all specimens with
>500 CFU of MTBC per ml were LCx assay positive, but only 44% were
positive when specimens contained <500 CFU of MTBC per ml. Ausina et
al. (3) found that all 18 specimens with <100 CFU of MTBC
per ml were negative by the LCx assay. Similar findings have been noted
for the Gen-Probe AMTDT and Roche Amplicor assays (27).
Hence, studies with a high rate of smear-positive and culture-positive
samples have found the LCx assay to be about 90% sensitive (3,
24, 39), while studies with lower rates of strongly positive
results for samples have reported sensitivities of about 78% (20,
28).
NAAT evaluations may also be biased as a result of the inclusion of
multiple specimens from individual patients (11). More than
one specimen from each patient should be tested to provide sufficient
opportunity for the detection of MTBC (24, 26, 32),
especially in specimens smear negative for acid-fast bacilli but
culture positive for MTBC. However, when more than three specimens are
tested from a given patient, bias may be introduced (11). The present study has attempted to avoid these biases. Consecutive samples submitted to the participating laboratories were tested; the
sample cohort was not enriched with positive samples; hence, the
prevalence of culture-positive samples was relatively low (i.e., 6.5%)
and the sensitivity of the present study is among the lower of
published estimates (3, 20, 24, 28, 31, 36, 39, 45).
Furthermore, an average of only 1.6 specimens were collected per
patient, and the 152 culture-positive specimens were collected from 98 patients (for 94% of these patients three or fewer samples were
collected, with an average of 1.5 specimens collected from each patient).
Studies have also varied in their handling of discrepant results.
Several evaluations have defined culture-negative, LCx assay-positive specimens from patients on antituberculosis treatment as having false-negative culture results (3, 18, 20, 24, 36, 39). At
present, there is no clinical utility in detecting MTBC DNA or RNA in
specimens from currently (or previously) treated patients.
Amplification tests may remain positive for extended periods of time
and are not useful for the monitoring of patients undergoing treatment
(3, 10, 14, 19, 43). Specimens from patients who had
received antituberculosis therapy for 7 days or longer were therefore
excluded from the analysis for resolution of discrepant results.
Similarly, six specimens from patients who had completed a course of
antituberculosis treatment and who were considered clinically cured but
who remained LCx assay positive were not included in the analysis for
resolution of discrepant results. The reproducible borderline-positive
LCx assay result obtained for the elderly individual who had contact
with a patient with tuberculosis was defined as false positive on
clinical, radiological, and laboratory grounds. Alonso et al.
(1) have also reported reproducible borderline-positive LCx
assay results when they tested a sputum sample from a person who had
been in contact with a patient with tuberculosis.
Construction of ROC curves suggested that reducing the cutoff value to
0.2 could improve the sensitivity of the LCx assay. Only a marginal
loss of specificity would result because the LCx assay values for
specimens from patients negative for MTBC cluster around a low mean
value (0.06 in the present study). Only 11 (0.5%) of these specimens
had S/CO values within the grey zone of 0.2 to 0.99. Yuen et al.
(45) reported a mean LCx assay value for negative samples of
0.035 and also noted the wide separation of readings for positive and
negative samples in the LCx assay. They suggested the use of a lower
cutoff but did not propose a particular grey zone.
While a larger evaluation of specimens whose results fall in the grey
zone is necessary, a simple algorithm for classification of these
specimens as positive or negative is proposed. Repeat testing of nine
samples from the present study with initial LCx assay values between
0.2 and 0.99 demonstrated that, for specimens from patients without
tuberculosis, repeat testing produced a true-negative result for eight
samples. In contrast, specimens culture positive for MTBC with an
initial LCx assay result between 0.2 and 0.99 repeatedly produced S/CO
values within the grey zone. The preponderance of smear-negative,
culture-positive specimens in this group suggests that inclusion of a
grey zone may facilitate detection of samples containing smaller
numbers of bacilli. Our laboratories now perform repeat tests for any
specimen with LCx assay values between 0.2 and 0.99; those that give a
negative result on repeat testing are reported as such. Specimens that repeatedly produce S/CO values within the grey zone are reported as
"equivocal" (until further experience is gained with these borderline samples), but the clinician is informed personally of the
high likelihood of the presence of MTBC. The handling of specimens with
results in the grey zone highlights the fact that effective
communication between the clinician and the laboratory is even more
important when new laboratory tests are being introduced.
The LCx assay integrates well into the routine diagnostic laboratory.
Because the amplification and detection steps are automated, once
specimen preparation is completed, the assay requires minimal involvement from laboratory staff, with the time to test completion being approximately 5 h. Specimen preparation involves two washing steps and appears to reduce the potential for specimen inhibitors to
interfere with the assay (23). An internal amplification control is not included in the assay, so a second tube should be spiked
with whole MTBC cells to check for the presence of inhibitors. Additionally, the assay may be held at almost any step of the procedure, further enhancing its integration into the laboratory work
flow. The kit instruction sheet and equipment manuals are comprehensive
and easy to follow, making the assay protocol straightforward.
In conclusion, the Abbott LCx assay is a well-presented,
straightforward, rapid, and reliable semiautomated NAAT for the
detection of MTBC directly from respiratory and nonrespiratory
specimens. The LCx assay detects MTBC in almost all smear-positive,
culture-positive and nearly half of smear-negative, culture-positive
specimens. With modification of the cutoff value, a further improvement
in test performance is possible. It is a relevant adjunct test in circumstances in which a rapid diagnosis of tuberculosis may have a
substantial impact upon patient management and public health considerations. Microscopy and culture remain mandatory components of
mycobacterial investigations.
 |
ACKNOWLEDGMENTS |
We thank Abbott Diagnostics for supplying the LCx assay kits. We
acknowledge the technical assistance of Allan Goodwin and Shirley Ellis.
I.B. is supported by a Neil Hamilton Fairley Fellowship (fellowship
987069) awarded by the National Health and Medical Research Council of Australia.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: Mycobacterium
Reference Laboratory, Infectious Diseases Laboratories, Institute of Medical and Veterinary Science, Box 14, Rundle Mall, Adelaide, South
Australia, 5000. Phone: 08 8222 3579. Fax: 08 8222 3543. E-mail:
richard.lumb{at}imvs.sa.gov.au.
 |
REFERENCES |
| 1.
|
Alonso, P.,
A. Orduña,
M. A. Bratos,
A. San Miguel, and A. Rodriguez Torres.
1998.
Clinical evaluation of a commercial ligase-based gene amplification method for detection of Mycobacterium tuberculosis.
Eur. J. Clin. Microbiol. Infect. Dis.
17:371-376[Medline].
|
| 2.
|
Anargyros, P.,
D. S. J. Astill, and I. S. L. Lim.
1990.
Comparison of improved BACTEC and Lowenstein-Jensen media for culture of mycobacteria from clinical specimens.
J. Clin. Microbiol.
28:1288-1291[Abstract/Free Full Text].
|
| 3.
|
Ausina, V.,
F. Gamboa,
E. Gazapo,
J. M. Manterola,
J. Lonca,
L. Matas,
J. R. Manzano,
C. Rodrigo,
P. J. Cardona, and E. Padilla.
1997.
Evaluation of the semiautomated Abbott LCx Mycobacterium tuberculosis assay for direct detection of Mycobacterium tuberculosis in respiratory specimens.
J. Clin. Microbiol.
35:1996-2002[Abstract].
|
| 4.
|
Badak, Z. F.,
D. L. Kiska,
S. Setterquist,
C. Hartley,
M. A. O'Connell, and R. L. Hopfer.
1996.
Comparison of the mycobacteria growth indicator tube with BACTEC 460 for detection and recovery of mycobacteria from clinical specimens.
J. Clin. Microbiol.
34:2236-2239[Abstract].
|
| 5.
|
Beck, J. R., and E. K. Shultz.
1986.
The use of relative operating characteristic (ROC) curves in test performance evaluation.
Arch. Pathol. Lab. Med.
110:13-20[Medline].
|
| 6.
|
Bennedsen, J.,
V. E. Thomsen,
G. E. Pfyffer,
G. Funke,
K. Feldmann,
A. Beneke,
P. A. Jenkins,
M. Hegginbothom,
A. Fahr,
M. Hengstler,
G. Cleator,
P. Klapper, and E. G. L. Wilkins.
1996.
Utility of PCR in diagnosing pulmonary tuberculosis.
J. Clin. Microbiol.
34:1407-1411[Abstract].
|
| 7.
|
Bergmann, J. S., and G. L. Woods.
1996.
Clinical evaluation of the Roche AMPLICOR PCR Mycobacterium tuberculosis test for detection of M. tuberculosis in respiratory specimens.
J. Clin. Microbiol.
34:1083-1085[Abstract].
|
| 8.
|
Bodmer, T.,
A. Gurtner,
K. Schopfer, and L. Matter.
1994.
Screening of respiratory tract specimens for the presence of Mycobacterium tuberculosis by using the Gen-Probe amplified Mycobacterium Tuberculosis Direct Test.
J. Clin. Microbiol.
32:1483-1487[Abstract/Free Full Text].
|
| 9.
|
Bodmer, T.,
E. Möckl,
K. Mühlemann, and L. Matter.
1996.
Improved performance of Gen-Probe Amplified Mycobacterium Tuberculosis Direct Test when 500 instead of 50 microliters of decontaminated sediment is used.
J. Clin. Microbiol.
34:222-223[Medline].
|
| 10.
|
Carpentier, E.,
B. Drouillard,
M. Dailloux,
D. Moinard,
E. Vallee,
B. Dutilh,
J. Maugein,
E. Bergogne-Berezin, and B. Carbonnelle.
1995.
Diagnosis of tuberculosis by Amplicor Mycobacterium tuberculosis test: a multicenter study.
J. Clin. Microbiol.
33:3106-3110[Abstract].
|
| 11.
|
Charache, P.
1996.
Comparison of the amplified Mycobacterium tuberculosis (MTB) direct test, Amplicor MTB PCR, and IS6110-PCR for detection of MTB in respiratory specimens.
Clin. Infect. Dis.
23:1107-1108. (Editorial response.)
|
| 12.
|
Cohn, D. L.,
F. Bustreo, and M. C. Raviglione.
1997.
Drug-resistant tuberculosis: review of the worldwide situation and the WHO/IUATLD global surveillance project.
Clin. Infect. Dis.
24:S121-S130.
|
| 13.
|
Cousins, D. V.,
S. D. Wilton,
B. R. Francis, and B. L. Gow.
1992.
Use of polymerase chain reaction for rapid diagnosis of tuberculosis.
J. Clin. Microbiol.
30:255-258[Abstract/Free Full Text].
|
| 14.
|
Dalovisio, J. R.,
S. Montenegro-Jones,
S. A. Kemmerly,
C. F. Genre,
R. Chambers,
D. Freer,
G. A. Pankey,
D. M. Failla,
K. G. Haydel,
L. Hutchinson,
M. F. Lindley,
B. M. Nunez,
A. Praba,
K. D. Eisenach, and E. S. Cooper.
1996.
Comparison of the amplified Mycobacterium tuberculosis (MTB) direct test, Amplicor MTB PCR, and IS6110-PCR for detection of MTB in respiratory specimens.
Clin. Infect. Dis.
23:1099-1106[Medline].
|
| 15.
|
D'Amato, R. F.,
A. A. Wallman,
L. H. Hochstein,
P. M. Colaninno,
M. Scardamaglia,
E. Ardila,
M. Ghouri,
K. Kim,
R. C. Patel, and A. Miller.
1995.
Rapid diagnosis of pulmonary tuberculosis by using Roche AMPLICOR Mycobacterium tuberculosis PCR test.
J. Clin. Microbiol.
33:1832-1834[Abstract].
|
| 16.
|
Dwyer, B.,
K. Jackson,
K. Raios,
A. Sievers,
E. Wilshire, and B. Ross.
1993.
DNA restriction fragment analysis to define an extended cluster of tuberculosis in homeless men and their associates.
J. Infect. Dis.
167:490-494[Medline].
|
| 17.
|
Evans, K. D.,
A. S. Nakasone,
P. A. Sutherland,
L. M. de la Maza, and E. M. Peterson.
1992.
Identification of Mycobacterium tuberculosis and Mycobacterium avium-M. intracellulare directly from primary BACTEC cultures by using acridinium-ester-labeled DNA probes.
J. Clin. Microbiol.
30:2427-2431[Abstract/Free Full Text].
|
| 18.
|
Gamboa, F.,
J. Dominguez,
E. Padilla,
J. M. Manterola,
E. Gazapo,
J. Lonca,
L. Matas,
A. Hernandez,
P. J. Cardona, and V. Ausina.
1998.
Rapid diagnosis of extrapulmonary tuberculosis by ligase chain reaction amplification.
J. Clin. Microbiol.
36:1324-1329[Abstract/Free Full Text].
|
| 19.
|
Gamboa, F.,
J. M. Manterola,
B. Viñado,
L. Matas,
M. Gimenez,
J. Lonca,
J. R. Manzano,
C. Rodrigo,
P. J. Cardona,
E. Padilla,
J. Dominguez, and V. Ausina.
1997.
Direct detection of Mycobacterium tuberculosis complex in nonrespiratory specimens by Gen-Probe Amplified Mycobacterium Tuberculosis Direct Test.
J. Clin. Microbiol.
35:307-310[Abstract].
|
| 20.
|
Garrino, M. G.,
Y. Glupczynski,
J. Degraux,
H. Ninet, and M. Delmée.
1999.
Evaluation of the Abbott LCx Mycobacterium tuberculosis assay for direct detection of Mycobacterium tuberculosis complex in human samples.
J. Clin. Microbiol.
37:229-232[Abstract/Free Full Text].
|
| 21.
|
Huebner, R. E.,
R. C. Good, and J. I. Tokars.
1993.
Current practices in mycobacteriology: results of a survey of state public health laboratories.
J. Clin. Microbiol.
31:771-775[Abstract/Free Full Text].
|
| 22.
|
Kent, P. T., and G. P. Kubica.
1985.
Public health mycobacteriology: a guide to the level III laboratory.
Centers for Disease Control, Atlanta, Ga.
|
| 23.
|
Leckie, G. W.,
D. D. Erickson,
Q. He,
I. E. Facey,
B.-C. Lin,
J. Cao, and F. G. Halaka.
1998.
Method for reduction of inhibition in a Mycobacterium tuberculosis-specific ligase chain reaction DNA amplification assay.
J. Clin. Microbiol.
36:764-767[Abstract/Free Full Text].
|
| 24.
|
Lindbråthen, A.,
P. Gaustad,
B. Hovig, and T. Tønjum.
1997.
Direct detection of Mycobacterium tuberculosis complex in clinical samples from patients in Norway by ligase chain reaction.
J. Clin. Microbiol.
35:3248-3253[Abstract].
|
| 25.
|
Lumb, R.,
J. A. Lanser, and I. S. Lim.
1993.
Rapid identification of mycobacteria by the Gen-Probe Accuprobe system.
Pathology
25:313-315[Medline].
|
| 26.
|
Miller, N.,
S. G. Hernandez, and T. J. Cleary.
1994.
Evaluation of Gen-Probe Amplified Mycobacterium Tuberculosis Direct Test and PCR for direct detection of Mycobacterium tuberculosis in clinical specimens.
J. Clin. Microbiol.
32:393-397[Abstract/Free Full Text].
|
| 27.
|
Moore, D. F., and J. I. Curry.
1995.
Detection and identification of Mycobacterium tuberculosis directly from sputum sediments by Amplicor PCR.
J. Clin. Microbiol.
33:2686-2691[Abstract].
|
| 28.
|
Moore, D. F., and J. I. Curry.
1998.
Detection and identification of Mycobacterium tuberculosis directly from sputum sediments by ligase chain reaction.
J. Clin. Microbiol.
36:1028-1031[Abstract/Free Full Text].
|
| 29.
|
Pfyffer, G. E.,
C. Cieslak,
H. M. Welscher,
P. Kissling, and S. Rüsch-Gerdes.
1997.
Rapid detection of mycobacteria in clinical specimens by using the automated BACTEC 9000 MB system and comparison with radiometric and solid-culture systems.
J. Clin. Microbiol.
35:834-841.
|
| 30.
|
Pfyffer, G. E.,
P. Kissling,
R. Wirth, and R. Weber.
1994.
Direct detection of Mycobacterium tuberculosis complex in respiratory specimens by a target-amplified test system.
J. Clin. Microbiol.
32:918-923[Abstract/Free Full Text].
|
| 31.
|
Piersimoni, C.,
A. Callegaro,
C. Scarparo,
V. Penati,
D. Nista,
S. Bornigia,
C. Lacchini,
M. Scagnelli,
G. Santini, and G. De Sio.
1998.
Comparative evaluation of the new Gen-Probe Mycobacterium tuberculosis amplified direct test and the semiautomated Abbott LCx Mycobacterium tuberculosis assay for direct detection of Mycobacterium tuberculosis complex in respiratory and extrapulmonary specimens.
J. Clin. Microbiol.
36:3601-3604[Abstract/Free Full Text].
|
| 32.
|
Rajalahti, I.,
P. Vuorinen,
M. M. Nieminen, and A. Miettinen.
1998.
Detection of Mycobacterium tuberculosis complex in sputum specimens by the automated Roche Amplicor Mycobacterium Tuberculosis Test.
J. Clin. Microbiol.
36:975-978[Abstract/Free Full Text].
|
| 33.
|
Raviglione, M. C.,
D. E. Snider, and A. Kochi.
1995.
Global epidemiology of Tuberculosis. Morbidity and mortality of a worldwide epidemic.
JAMA
273:220-226[Abstract/Free Full Text].
|
| 34.
|
Reischl, U.,
N. Lehn,
H. Wolf, and L. Naumann.
1998.
Clinical evaluation of the automated COBAS AMPLICOR MTB assay for testing respiratory and nonrespiratory specimens.
J. Clin. Microbiol.
36:2853-2860[Abstract/Free Full Text].
|
| 35.
|
Rogall, T.,
T. Flohr, and E. C. Böttger.
1990.
Differentiation of Mycobacterium species by direct sequencing of amplified DNA.
J. Gen. Microbiol.
40:1915-1920.
|
| 36.
|
Rohner, P.,
E. I. M. Jahn,
B. Ninet,
C. Ionati,
R. Weber,
R. Auckenthaler, and G. E. Pfyffer.
1998.
Rapid diagnosis of pulmonary tuberculosis with the LCx Mycobacterium tuberculosis assay and comparison with conventional techniques.
J. Clin. Microbiol.
36:3046-3047[Abstract/Free Full Text].
|
| 37.
|
Rohner, P.,
B. Ninet,
C. Metral,
S. Emler, and R. Auckenthaler.
1997.
Evaluation of the MB/BacT system and comparison to the BACTEC 460 system and solid media for isolation of mycobacteria from clinical specimens.
J. Clin. Microbiol.
35:3127-3131[Abstract].
|
| 38.
|
Sjöbring, U.,
M. Mecklenburg,
Å. B. Andersen, and H. Miörner.
1990.
Polymerase chain reaction for detection of Mycobacterium tuberculosis.
J. Clin. Microbiol.
28:2200-2204[Abstract/Free Full Text].
|
| 39.
|
Tortoli, E.,
F. Lavinia, and M. T. Simonetti.
1997.
Evaluation of a commercial ligase chain reaction kit (Abbott LCx) for direct detection of Mycobacterium tuberculosis in pulmonary and extrapulmonary specimens.
J. Clin. Microbiol.
35:2424-2426[Abstract].
|
| 40.
|
Van Deutekom, H.,
J. J. J. Gerritsen,
D. van Soolingen,
E. J. C. van Ameijden,
J. D. A. van Embden, and R. A. Coutinho.
1997.
A molecular epidemiological approach to studying the transmission of tuberculosis in Amsterdam.
Clin. Infect. Dis.
25:1071-1077[Medline].
|
| 41.
|
Vlaspolder, F.,
P. Singer, and C. Roggeveen.
1995.
Diagnostic value of an amplification method (Gen-Probe) compared with that of culture for diagnosis of tuberculosis.
J. Clin. Microbiol.
33:2699-2703[Abstract].
|
| 42.
|
Vuorinen, P.,
A. Miettinen,
R. Vuento, and O. Hällstrom.
1995.
Direct detection of Mycobacterium tuberculosis complex in respiratory specimens by Gen-Probe Amplified Mycobacterium Tuberculosis Direct Test and Roche Amplicor Mycobacterium Tuberculosis Test.
J. Clin. Microbiol.
33:1856-1859[Abstract].
|
| 43.
|
Wobeser, W. L.,
M. Kradjen,
J. Conly,
H. Simpson,
B. Yim,
M. D'Costa,
M. Fuksa,
C. Hian-Cheong,
M. Patterson,
A. Phillips,
R. Bannatyne,
A. Haddad,
J. L. Brunton, and S. Kradjen.
1996.
Evaluation of Roche Amplicor PCR assay for Mycobacterium tuberculosis.
J. Clin. Microbiol.
34:134-139[Abstract].
|
| 44.
|
Woods, G. L.,
G. Fish,
M. Plaunt, and T. Murphy.
1997.
Clinical evaluation of the Difco ESP culture system II for growth and detection of mycobacteria.
J. Clin. Microbiol.
35:121-124[Abstract].
|
| 45.
|
Yuen, K.-Y.,
W.-C. Yam,
L.-P. Wong, and W.-H. Seto.
1997.
Comparison of two automated DNA amplification systems with a manual one-tube nested PCR assay for diagnosis of pulmonary tuberculosis.
J. Clin. Microbiol.
35:1385-1389[Abstract].
|
Journal of Clinical Microbiology, October 1999, p. 3102-3107, Vol. 37, No. 10
0095-1137/99/$04.00+0
Copyright © 1999, American Society for Microbiology. All rights reserved.
This article has been cited by other articles:
-
Greco, S, Girardi, E, Navarra, A, Saltini, C
(2006). Current evidence on diagnostic accuracy of commercially based nucleic acid amplification tests for the diagnosis of pulmonary tuberculosis. Thorax
61: 783-790
[Abstract]
[Full Text]
-
Piersimoni, C., Scarparo, C.
(2003). Relevance of Commercial Amplification Methods for Direct Detection of Mycobacterium tuberculosis Complex in Clinical Samples. J. Clin. Microbiol.
41: 5355-5365
[Full Text]
-
Mazzarelli, G., Rindi, L., Piccoli, P., Scarparo, C., Garzelli, C., Tortoli, E.
(2003). Evaluation of the BDProbeTec ET System for Direct Detection of Mycobacterium tuberculosis in Pulmonary and Extrapulmonary Samples: a Multicenter Study. J. Clin. Microbiol.
41: 1779-1782
[Abstract]
[Full Text]
-
Piersimoni, C., Scarparo, C., Piccoli, P., Rigon, A., Ruggiero, G., Nista, D., Bornigia, S.
(2002). Performance Assessment of Two Commercial Amplification Assays for Direct Detection of Mycobacterium tuberculosis Complex from Respiratory and Extrapulmonary Specimens. J. Clin. Microbiol.
40: 4138-4142
[Abstract]
[Full Text]
-
Gilpin, C. M., Dawson, D. J., O'Kane, G., Armstrong, J. G., Coulter, C.
(2002). Failure of Commercial Ligase Chain Reaction To Detect Mycobacterium tuberculosis DNA in Sputum Samples from a Patient with Smear-Positive Pulmonary Tuberculosis Due to a Deletion of the Target Region. J. Clin. Microbiol.
40: 2305-2307
[Abstract]
[Full Text]
-
Louie, M., Louie, L., Simor, A. E.
(2000). The role of DNA amplification technology in the diagnosis of infectious diseases. CMAJ
163: 301-309
[Abstract]
[Full Text]