Received 21 July 1998/Returned for modification 28 September
1998/Accepted 8 December 1998
The ultrasensitive Amplicor HIV-1 Monitor test (Roche Diagnostic
Systems) was evaluated for precision, linearity, and sensitivity and
was compared to the standard Amplicor assay. The ultrasensitive assay
reliably quantified samples in the range from 50 to 50,000 human
immunodeficiency virus type 1 RNA copies/ml with acceptable correlation
with the standard Amplicor test.
 |
TEXT |
Quantitative viral load measurement
of human immunodeficiency virus type 1 (HIV-1) has become a
valuable tool for the management of HIV-1-infected patients.
Levels of HIV-1 RNA in plasma have been shown to be useful as markers
of prognosis, for the assessment of disease progression, and for
monitoring the efficacy of antiretroviral drug therapy (3, 5,
7). The recent advent of combination drug therapies allows viral
load suppression for many patients to a level below the quantitation
limit of the currently available HIV-1 RNA assays. New methods for
measuring HIV-1 RNA at levels below 400 to 500 HIV-1 RNA copies/ml have
been developed, and recent studies have begun to evaluate the clinical
significance of viral levels in this range. Early reports suggest that
suppression to <50 HIV-1 RNA copies/ml compared to 50 to 400 HIV-1 RNA copies/ml is associated with a more sustained response
(2, 6). These reports support the use of ultrasensitive
HIV-1 RNA quantitation methods for evaluating and optimizing
antiretroviral therapy.
Viral load measurement at extremely low levels is subject to
constraints imposed by the statistics of viral particle sampling as
well as the inherent variability of the assay technology. A clinically
useful ultrasensitive assay will need excellent reproducibility and
reliability, with an appropriate sample volume. In order to assess the
performance characteristics of the ultrasensitive assay with the
Amplicor HIV-1 Monitor test, precision, linearity, and sensitivity were
evaluated and comparison with other methods was made.
The ultrasensitive modification of the Amplicor HIV-1
Monitor test was performed according to the protocol provided
by Roche Molecular Systems, Inc. (Somerville, N.J.). Specimens were
EDTA plasma samples submitted for HIV-1 viral load measurement to ARUP Laboratories, Salt Lake City, Utah. Specimens were drawn in EDTA tubes,
and the plasma was removed from the cells within 30 min. The specimens
were frozen at 
20°C within 2 h of separation and shipped frozen.
Within-run precision was determined with five samples analyzed in
replicates of four or five in one run. The samples had low viral loads
ranging from 30 to 500 HIV-1 RNA copies/ml. The mean percent
coefficient of variation (% CV) for HIV-1 RNA copies per milliliter
was 19.4%, and the mean standard deviation of the HIV-1 RNA
log10(copies per milliliter) was 0.09. A summary of the
within-run precision data is presented in Table
1.
The between-run precision was determined with in-house controls
prepared by using clinical specimens diluted in HIV-1-seronegative plasma. Precision was evaluated over 37 separate runs for the low
control and over 38 separate runs for the high control. The mean value
for the low control was 200 HIV-1 RNA copies/ml, and the mean value for
the high control was 17,000 HIV-1 RNA copies/ml. The % CV for HIV-1
RNA copies per milliliter for the low and high controls was 30.8 and
32.2%, respectively. The standard deviation of the HIV-1 RNA
log10(copies per milliliter) for the low and high controls
was 0.13 and 0.12, respectively. A summary of the between-run precision
is presented in Table 1.
Serial dilutions of three clinical specimens were prepared in
HIV-1-seronegative plasma, and the HIV-1 RNA levels were measured in
the ultrasensitive assay. In order to describe the wide range of values
obtained in the dilution series, the data for percent recoveries are
presented as the percent observed of total expected for the
log10 of the HIV-1 RNA copies per milliliter. The raw data
and percent recoveries are presented in Table
2. The range of recoveries was 81%, for
the dilution with 40 HIV-1 RNA copies/ml expected, to 105%, for the
dilution with 80 HIV-1 RNA copies/ml expected. A plot of the observed
values versus the expected values is presented in Fig.
1. The equation for the regression line
is y = 0.958x + 0.054, r2 = 0.994, r = 0.997, for n = 15 points. These
data show that the ultrasensitive assay is linear over the range from
50 to 50,000 HIV-1 RNA copies/ml.

View larger version (20K):
[in this window]
[in a new window]
|
FIG. 1.
Scatter plot of observed versus expected HIV-1 RNA
log10(copies per milliliter) for the data presented in
Table 2. The equation for the regression line (solid line) is
y = 0.958x + 0.054, r2 = 0.994, r = 0.997. The diagonal dashed line is the line of unity. The
upper and lower horizontal dashed lines are 50,000 and 50 copies/ml,
respectively.
|
|
Viral load levels were determined for 32 clinical samples by both the
ultrasensitive assay and the standard Amplicor HIV-1 Monitor test. The
range of samples tested was from 170 to 60,000 HIV-1 RNA copies/ml as
determined in the ultrasensitive assay and 400 to 50,000 HIV-1 RNA
copies/ml as determined in the standard assay. The regression equation
for the correlation was y = 1.088x
0.343, r2 = 0.904, r = 0.951 for n = 32 samples. The reportable range for the ultrasensitive assay is
50 to 50,000 HIV-1 RNA copies/ml and 400 to 750,000 HIV-1 RNA copies/ml
for the standard assay. Several samples which were >400 copies/ml in
the standard assay measured >50 but <400 copies/ml in the
ultrasensitive assay. This variation may be a consequence of the
imprecision at low concentrations, since all samples were run in
singlet on both assays. As shown in Fig.
2, the correlation is acceptable over the
overlapping range of the two assays, 400 to 50,000 HIV-1 RNA copies/ml.

View larger version (22K):
[in this window]
[in a new window]
|
FIG. 2.
Scatter plot of HIV-1 RNA log10(copies per
milliliter) as determined by the ultrasensitive and standard Amplicor
HIV-1 Monitor tests for 32 clinical specimens. The equation for the
regression line (solid line) is y = 1.088x 0.343, r2 = 0.904, r = 0.951. The diagonal
dashed line is the line of unity. The upper and lower horizontal dashed
lines are 50,000 and 400 copies/ml, respectively.
|
|
Clinical specimens below the detection limit (400 HIV-1 RNA copies/ml)
of the standard Amplicor HIV-1 Monitor test were evaluated in the
ultrasensitive assay. The samples chosen for this study were biased
toward a likelihood of being >50 HIV-1 RNA copies/ml. This was done by
selecting samples which were not quantifiable but had optical densities
(OD) at 450 nm (at a dilution factor of 1 in the assay detection step)
which were greater than the expected background of 0.070 OD U, as
specified in the Amplicor package insert. The assumption was that these
samples with OD readings greater than background may actually contain
HIV-1 RNA at levels that are not quantifiable by the standard assay.
Ten samples chosen by these criteria were evaluated in the
ultrasensitive assay. A summary of the data is presented in Table
3. HIV-1 RNA was quantifiable in 60% (6 of 10) of the samples tested. In only one sample, the calculated value
of the ultrasensitive assay was greater than 400 HIV-1 RNA copies/ml,
and the variability of the assay would account for the difference
observed (452 copies/ml for sample 4). No association of the viral load
level obtained in the ultrasensitive assay can be made with either the
OD of the samples or the OD variation above background.
Clinical specimens which were quantified as less than the detection
limit of the Quantiplex HIV RNA 2.0 assay (bDNA) (Chiron Corporation,
Emeryville, Calif.) (500 HIV-1 RNA copies/ml) were also evaluated in
the ultrasensitive assay. The samples chosen for this study were also
biased toward a likelihood of being >50 HIV-1 RNA copies/ml. For the
Quantiplex assay, this was done by selecting samples which were not
quantifiable but had relative luminescence (RL) units in the assay
greater than the RL units for the standard D (STD D) of the test. The
assumption was again that samples with RL readings greater than the STD
D may actually contain HIV-1 RNA at levels that are not measured by the
Quantiplex assay. Ten samples chosen by these criteria were evaluated
in the ultrasensitive assay. A summary of the data is presented in Table 4. The ultrasensitive assay was
able to measure HIV-1 RNA in 100% (10 of 10) of the samples tested. In
7 of 10 samples, the calculated viral load was greater than 500 HIV-1
RNA copies/ml. Other studies have noted a similar disparity between
Quantiplex and Amplicor results (4, 8). No conclusions can
be drawn from these results regarding the accuracy or validity of
either assay.
The widespread utilization of viral load testing for the evaluation and
monitoring of HIV-1-infected patients has led to a rapid generational
improvement in assay technology. A key modification has been the
enhancement of assay sensitivity to levels which approach the limits
imposed by the statistics of sampling, i.e., 20 to 50 HIV-1 RNA
copies/ml. The early clinical experience with ultrasensitive HIV-1
testing supports the ability of these assays to discriminate treatment
outcomes and provides an indication for routine testing (1).
This study supports the reliability of the Amplicor HIV-1 Monitor test
for the measurement of very low viral load levels. The ultrasensitive
assay showed good within-run and between-run precision, and the
dilution studies confirmed as reliable the stated reportable range of
50 to 50,000 HIV-1 RNA copies/ml.
An important issue for users of the Roche HIV-1 quantification system
is the ability of the standard and ultrasensitive tests to give
concordant results within their overlapping but nonidentical dynamic
ranges. Although the two assays utilize essentially the same reagents
for extraction, amplification, and detection, the ultrasensitive test
depends on a highly efficient ultracentrifugation step in order to
match results with the standard assay. This study finds an acceptable
correlation between the ultrasensitive and the standard assays in the
overlapping range between 400 and 50,000 HIV-1 RNA copies. Although
small differences may exist in the absolute detection efficiencies of
the two tests, the intrinsic sample-to-sample variation of either assay
probably overshadows these differences. Careful attention to the
details of sample preparation and proper centrifuge maintenance will
contribute to good test concordance.
Although the current generation of commercial HIV-1 viral load tests
achieves good precision over a broad dynamic range, they have not been
cross-standardized for accuracy of RNA quantitation. Therefore, current
recommendations for clinical testing discourage the comparison of
results obtained in different assay systems. The development of
ultrasensitive HIV-1 RNA assays further emphasizes the need for
quantitative standardization. Examination of a subset of
Quantiplex-negative samples with detectable but below-threshold signals
(Table 4) allows the identification and comparison of clinical samples
at the low end of HIV viremia. The data comparing Quantiplex and
ultrasensitive Amplicor results, shown in Table 4, are in poor
agreement and indicate that, at low levels of HIV viremia, different
methodologies may produce very discrepant values. This supports the
need for standardized calibration of viral load assays across the
reportable range of the tests.
Efficient physician utilization of the Roche HIV-1 quantification
assays depends on proper test selection. Since the great majority of
patients who present for an initial HIV-1 viral load determination have
levels above 400 HIV-1 RNA copies/ml, testing may begin with the
standard assay. When the baseline viral load is <50,000 HIV-1 RNA
copies/ml or when that level is achieved in response to therapy,
testing can continue with the ultrasensitive assay without the need to
obtain a new baseline. During the first 8 months of ultrasensitive
testing in our laboratory, approximately 15% of the results reported
for the ultrasensitive assay exceeded 50,000 copies/ml. Careful
attention to the levels and direction of viral load change is key to
good patient management as well as cost-effective utilization of this
test. The future development of ultrasensitive assays with extended
dynamic ranges will simplify test utilization.
| 1.
|
Conway, B.,
A. Shillington,
S. Franses, et al.
1998.
Use of ultra-sensitive PCR technology to provide insight into viral dynamics, abstr. 96.
In
Program and abstracts of the International Workshop on HIV Drug Resistance Treatment, Strategies, and Eradication, St. Petersburg, Fla., 25 to 28 June 1998.
|
| 2.
|
Izopet, J.,
G. Salama,
C. Pasquier,
K. Sandres,
E. Kuhlein,
A. Blancher,
B. Marchou,
P. Massip, and J. Puel.
1998.
Ultra sensitive detection of plasma HIV-1 RNA for predicting the durability of 3-drug antiretroviral therapy, abstr. 326.
In
Program and abstracts of the 5th Conference on Retroviruses and Opportunistic Infections, Chicago, Ill., 2 to 4 February 1998.
|
| 3.
|
Mellors, J. W.,
A. Munoz,
J. V. Giorgi,
J. B. Margolick,
C. J. Tassoni,
P. Gupta,
L. A. Kingsley,
J. A. Todd,
A. J. Saah,
R. Detels,
J. P. Phair, and C. R. Rinaldo.
1997.
Plasma viral load and CD4+ lymphocytes as prognostic markers of HIV-1 infection.
Ann. Intern. Med.
126:946-954[Abstract/Free Full Text].
|
| 4.
|
Nolte, F. S.,
J. Boysza,
C. Thurmond,
W. S. Clark, and J. L. Lennox.
1998.
Clinical comparison of an enhanced-sensitivity branched-DNA assay and reverse transcription-PCR for quantitation of human immunodeficiency virus type 1 RNA in plasma.
J. Clin. Microbiol.
36:716-720[Abstract/Free Full Text].
|
| 5.
|
O'Brien, W. A.,
P. M. Hartigan,
E. S. Daar,
M. S. Simberkoff, and J. D. Hamilton.
1997.
Changes in plasma HIV RNA levels and CD4+ lymphocyte counts predict both response to antiretroviral therapy and therapeutic failure.
Ann. Intern. Med.
126:939-945[Abstract/Free Full Text].
|
| 6.
|
Ruiz, N. M.,
D. J. Manion,
D. F. Labriola,
P. A. Friedman,
K. J. Gorelick,
T. D. Saxton,
A. P. Goldberg,
G. Chao, and P. Jarrett and the DMF 266 Development Team.
1997.
Demographic and laboratory predictors of virologic treatment failures in patients achieving viral load reductions to below quantifiable levels (BQL) by Amplicor assay receiving indinavir +/ DMP 266 (efavirenz), abstr. 921.
In
Program and abstracts of the 6th European Conference on Clinical Aspects and Treatment of HIV Infections, Hamburg, Germany, 11 to 15 October 1997.
|
| 7.
|
Saag, M. S.,
M. Holodniy,
D. R. Kuritzkes,
W. A. O'Brien,
R. Coombs,
M. E. Poscher,
D. M. Jacobsen,
G. M. Shaw,
D. D. Richman, and P. A. Volberding.
1996.
HIV viral load markers in clinical practice.
Nat. Med.
2:625-629[Medline].
|
| 8.
|
Yerkovich, M.,
K. W. Ng,
R. Sampoleo,
J. Dragavon, and R. W. Coombs.
1998.
Low HIV-1 RNA copy number assessed by bDNA and RT-PCR assays: implications for managing patients above or below the viral RNA quantitation limit, abstr. 303.
In
Program and abstracts of the 5th Conference on Retroviruses and Opportunistic Infections, Chicago, Ill., 1 to 5 February 1998.
|