Reproducibility and Performance of the
Second-Generation Branched-DNA Assay in Routine Quantification of
Human Immunodeficiency Virus Type 1 RNA in Plasma
Donald G.
Murphy,1,2,*
Patrick
Gonin,1 and
Micheline
Fauvel1
Laboratoire de Santé Publique du
Québec, Sainte-Anne-de-Bellevue,1 and
Département de Microbiologie et Immunologie,
Université de Montréal,
Montréal,2 Québec, Canada
Received 16 July 1998/Returned for modification 28 September
1998/Accepted 18 November 1998
 |
ABSTRACT |
We examined the reproducibility of a second-generation branched-DNA
(bDNA) assay (Quantiplex HIV RNA 2.0) for quantification of human
immunodeficiency virus type 1 (HIV-1) RNA in plasma by retesting 325 specimens on separate runs and on different lots. The performance of
the bDNA test was also assessed by data analysis obtained during
routine testing of 15,365 specimens. Upon retesting, 96 and 86% of
specimens displaying RNA levels above 5,000 and between 500 and 5,000 copies/ml, respectively, showed less than a 0.3 log10
(twofold) difference with their initial values. Assay variability was
found to increase as viral load decreased. Overall, the bDNA version
2.0 assay was found to be a reproducible and efficient test for routine
quantification of HIV-1 RNA in plasma.
 |
TEXT |
Human immunodeficiency virus (HIV)
RNA levels and CD4+ T-lymphocyte counts are important
biological markers for the management of HIV-infected patients
(1). HIV RNA levels in plasma have been shown to be a strong
determinant of disease progression and serve as a guideline to
determine when antiviral therapy should be initiated and changed
(5, 6, 8). The clinical significance of HIV RNA levels will
depend on the biological variability within individuals and on the
variability of the test employed. In clinically stable patients, HIV
RNA levels measured by branched-DNA (bDNA) technology have been found
to vary by 0.4 log10 over a 48-h period (2).
Several limited studies have shown that the reproducibility of the
currently available viral load assays can vary by as much as 0.5 log10 (4, 9, 10). However, the actual variation and the performance of these assays in routine testing have not been
substantially documented.
We performed HIV viral load measurements on a routine basis for the
follow-up of HIV-infected patients in the Canadian province of Quebec.
EDTA-plasma samples were obtained from hospitals and private clinics.
Specimens were frozen at
70°C within 6 h of collection and
transported on dry ice to the Laboratoire de Santé Publique du
Québec, where they were stored at
70°C until testing. In most
cases, two aliquots of at least 2.0 ml were obtained for each specimen.
Viral RNA levels were measured by a second-generation bDNA assay
(Quantiplex HIV RNA 2.0; Chiron Corporation, Emeryville, Calif.)
according to the manufacturer's instructions (3). Briefly, in this assay clinical specimens, positive and negative controls, and
DNA standards are each processed in duplicate. The positive control
consists of human plasma containing inactivated HIV type 1 (HIV-1).
Following a series of nucleic acid hybridization steps, alkaline
phosphatase-labeled probes are incubated with a chemiluminescent substrate. The amount of light emitted is measured in a luminometer and
reported as relative luminescent units (RLU). The quantity of HIV-1 RNA
is determined from a standard curve defined by light emission
from four DNA standards. The DNA standards are calibrated against an
HIV RNA transcript standard as follows (in RNA copies/milliliter): A,
800,000; B, 54,000; C, 6,000; D, 250. The result reported as RNA
copies/milliliter is derived from the mean of the two readings. The
extent of the divergence between duplicate readings is provided by the
coefficient of variation (CV). The quantification limit of the test is
set at 500 RNA copies/ml. Specimens were retested, if sufficient volume
was available, when yielding >500 RNA copies/ml with a CV above 35%
or when yielding <500 RNA copies/ml with a CV above 35% if the RLU of
one of the readings was above the RLU of one of the D standards, as
recommended by Chiron personnel. Retesting of specimens was performed
within 5 working days on a fresh aliquot.
Between April 1, 1997, and March 31, 1998, an average of 301 specimens
were received weekly. This volume of testing required 2.5 operators per
week working a 7-hour day, 5 days a week. The mean turnaround time from
reception of specimens to result printout was 6 days (including weekend
delays). Operator tasks included unpacking of specimens on dry ice,
specimen labeling, revision of patient data, testing, entering of
results, and data collection. Tests were performed by a pool of four
trained operators.
A total of 15,365 specimens from 5,339 patients were tested: 58.4% of
specimens showed RNA levels above 500 copies/ml, while 41.6% displayed
RNA levels below this value. A CV of >35% was observed in 3.3 and
9.8% of specimens yielding RNA levels above and below 500 copies/ml,
respectively. The positive control displayed a CV above 35% in 12.4%
of runs. Accordingly, it varied more than the clinical specimens with
>500 RNA copies/ml (P < 0.001). A total of 415 runs
were performed, and 13 (3.1%) were rejected. The kit negative control
yielded a quantifiable result in 6 (1.4%) runs, and the positive
control was outside the prescribed range in 7 (1.7%) runs.
To evaluate the in-house reproducibility of the bDNA assay, a second
aliquot of a previously tested specimen showing a CV below 35% was
included in subsequent assay runs. Each specimen was retested on a
separate run, on a different lot, and in different positions of the
microwell plate. Overall, 325 specimens were retested: 255 had initial
RNA levels above 500 copies/ml, and 70 had levels below this value. The
former were selected as to cover the dynamic range of the standard
curve and, upon repeat testing, 243 of 255 yielded results above 500 RNA copies/ml. Figure 1 illustrates the
difference obtained between duplicate tests. Variance analysis
indicates that this difference increased as the viral load decreased
(P < 0.005). For clinical and comparative reasons, the
data was stratified according to RNA levels (Table 1). The mean difference between duplicate
tests was within 0.3 log10 (twofold). Nonetheless, several
specimens displayed a difference of >0.3 log10, indicating
that results can occasionally vary by more than twofold. The 12 specimens which initially showed >500 RNA copies/ml but whose RNA
levels became undetectable upon repeat testing were excluded from this
analysis as they would have minimized the actual variation. Only one
showed an initial viral load above 1,000 RNA copies/ml. Finally, upon
repeat testing of the 70 specimens displaying <500 RNA copies/ml and a
CV below 35%, 8 (11.4%) showed viral loads above this value. Of these
8 specimens, 4 displayed RNA values between 500 and 999 copies/ml and 4 showed values between 1,000 and 1,999 copies/ml. This observation is
consistent with variation being greater towards the lower limit of the
assay.

View larger version (20K):
[in this window]
[in a new window]
|
FIG. 1.
Log10 difference between duplicate tests
against their means for 243 clinical specimens. The regression line is
included in the figure.
|
|
The results obtained upon repeat testing of specimens that showed more
than 500 RNA copies/ml and a CV above 35% were compared to their
initial values. Table 2 shows the data
for the 141 of 203 specimens that maintained RNA levels above 500 copies/ml after retesting. Specimens displaying a CV between 35 and
50%, of which 40% had RNA levels below 5,000 copies/ml, showed a mean
difference of 0.13 log10, with a standard deviation (SD) of
0.11 (Table 2). This value is similar to that observed for specimens
showing a CV below 35% (Table 1). An increase in the proportion of
specimens displaying more than a 0.3-log10 difference upon
repeat testing was observed for those showing a CV above 50%. Of the
62 (30.5%) specimens whose RNA levels became undetectable upon repeat
testing, 53 (85.5%), 7 (11.3%), and 2 (3.2%) displayed initial
values between 500 and 1,499, 1,500 and 2,499, and 2,500 and 3,499 RNA
copies/ml, respectively.
View this table:
[in this window]
[in a new window]
|
TABLE 2.
Comparison of initial and repeat tests for specimens
displaying initial values of >500 RNA copies/ml and a CV of >35%
|
|
Table 3 shows the results obtained
upon repeat testing of 280 specimens displaying initial RNA levels
below 500 copies/ml and a CV of >35% when the RLU of one of the
readings was above the RLU of one of the D standards. The frequency of
specimens yielding >500 RNA copies/ml was lower than that observed for
specimens with a CV below 35% (6.1 versus 11.4%). A possible
explanation is that a CV above 35% is likely to be observed more
frequently for specimens with extremely low HIV RNA levels than for
specimens with RNA levels just below the 500-copies/ml assay detection
limit. Upon repeat testing, the latter specimens are more likely than the former to yield results above 500 copies/ml due to assay
characteristics.
In the bDNA version 2.0 assay, specimens are tested in duplicate. As a
consequence, the divergence (%CV) between the two readings could be
used to determine when a test should be repeated. However, no
information is provided in the manufacturer's instructions as to what
degree of discordance between duplicate readings is acceptable. The
data in Table 2 support retesting of specimens displaying >500 RNA
copies/ml only when the CV is above 50%. However, the data excludes
specimens whose RNA levels became undetectable after repeat testing,
most of which (60 of 62) displayed initial RNA levels between 500 and
2,499 copies/ml. Of specimens displaying initial RNA levels between 500 and 2,499 copies/ml and a CV between 35 and 50%, 34.7% showed levels
that became undetectable after repeat testing. Therefore, we recommend
retesting of specimens with RNA levels between 500 and 2,500 copies/ml
when the CV is above 35% and retesting of specimens with >2,500
copies/ml when the CV is >50%. Table 3 shows the results obtained
upon retesting of specimens whose levels were initially below the
detection limit according to initial RLU values. For these specimens
the %CV is not useful in determining if a specimen should be retested.
The data indicates that specimens had a tendency to yield >500 RNA copies/ml upon repeat testing only when both of their RLU values were
above at least one of the D standard RLU (P < 0.001),
irrespective of whether the CV was above or below 35%.
The HIV RNA level is an indicator of disease progression and is
monitored for the management of infected patients. The variability and
performance of the viral load assays can have an impact on therapeutic
decisions. Thus, it is important that clinicians be informed on the
reproducibility and performance of the viral load assays. In this
study, the reproducibility of the bDNA version 2.0 assay was evaluated
at a single site by retesting specimens on different assay runs and
lots and in the majority of cases by different operators. This
represents the variability likely to be encountered in the day-to-day
practice, as most patients are likely to be monitored by the same
clinician and to have their viral loads determined in the same
laboratory. It remains to be determined if testing at different sites
would further affect the quality of results (10). The data
presented in this report should guide clinicians on the use of viral
load measures obtained by the bDNA version 2.0 assay in patient management.
In conclusion, the bDNA version 2.0 assay was found to be a
reproducible and efficient test for routine quantification of HIV-1 RNA
in plasma. It is noteworthy that 41.6% of specimens tested in this
study showed RNA levels below 500 copies/ml. An assay with a lower
detection limit would thus provide more informative data for these specimens.
 |
ACKNOWLEDGMENTS |
We thank Claire Sauvé, Danielle Sasseville, Jasmine
Chamberland, and Micheline Lortie for performing the viral load tests and data collection.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: Laboratoire de
Santé Publique du Québec, 20045 chemin Sainte-Marie,
Sainte-Anne-de-Bellevue, Québec, Canada H9X 3R5. Phone: (514)
457-2070. Fax: (514) 457-6346. E-mail: dmurphy{at}lspq.org.
 |
REFERENCES |
| 1.
|
Carpenter, C. C. J.,
M. A. Fischl,
S. M. Hammer,
M. S. Hirsch,
D. M. Jacobsen,
D. A. Katzenstein,
J. S. G. Montaner,
D. D. Richman,
M. S. Saag,
R. T. Schooley,
M. A. Thompson,
S. Vella,
P. G. Yeni, and P. A. Volberding for the International AIDS Society-USA.
1996.
Antiretroviral therapy for HIV infection in 1996.
JAMA
276:146-154[Abstract].
|
| 2.
|
Deeks, S. G.,
R. L. Coleman,
R. White,
C. Pachl,
M. Schambelan,
D. N. Chernoff, and M. B. Feinberg.
1997.
Variance of plasma human immunodeficiency virus type 1 RNA levels measured by branched DNA within and between days.
J. Infect. Dis.
176:514-517[Medline].
|
| 3.
|
Kern, D.,
M. Collins,
T. Fultz,
J. Detmer,
S. Hamren,
J. J. Peterkin,
P. Sheridan,
M. Urdea,
R. White,
T. Yeghiazarian, and J. Todd.
1996.
An enhanced-sensitivity branched-DNA assay for quantification of human immunodeficiency virus type 1 RNA in plasma.
J. Clin. Microbiol.
34:3196-3202[Abstract].
|
| 4.
|
Lin, H. J.,
L. Pedneault, and F. B. Hollinger.
1998.
Intra-assay performance characteristics of five assays for quantification of human immunodeficiency virus type 1 RNA in plasma.
J. Clin. Microbiol.
36:835-839[Abstract/Free Full Text].
|
| 5.
|
Mellors, J. W.,
C. R. Rinaldo, Jr.,
P. Gupta,
R. M. White,
J. A. Todd, and L. A. Kingsley.
1996.
Prognosis in HIV-1 infection predicted by the quantity of virus in plasma.
Science
272:1167-1170[Abstract].
|
| 6.
|
O'Brien, W. A.,
P. M. Hartigan,
D. Martin,
J. Esinhart,
A. Hill,
S. Benoit,
M. Rubin,
M. S. Simberkoff,
J. D. Hamilton, and the Veterans Affairs Cooperative Study Group on AIDS.
1996.
Changes in plasma HIV-1 RNA and CD4+ lymphocyte counts and the risk of progression to AIDS.
N. Engl. J. Med.
334:426-431[Abstract/Free Full Text].
|
| 7.
|
Revets, H.,
D. Marissens,
S. De Wit,
P. Lacor,
N. Clumeck,
S. Lauwers, and G. Zissis.
1996.
Comparative evaluation of NASBA HIV-1 RNA QT, AMPLICOR-HIV Monitor, and QUANTIPLEX HIV RNA assay, three methods for quantification of human immunodeficiency virus type 1 RNA in plasma.
J. Clin. Microbiol.
34:1058-1064[Abstract].
|
| 8.
|
Saag, M. S.,
M. Holodniy,
D. R. Kuritzkes,
W. A. O'Brien,
R. Coombs,
M. E. Poscher,
D. M. Jacobsen,
G. M. Shaw,
D. D. Richman, and P. A. Volberding.
1996.
HIV viral load markers in clinical practice.
Nat. Med.
2:625-629[Medline].
|
| 9.
|
Schuurman, R.,
D. Descamps,
G. J. Weverling,
S. Kaye,
J. Tijnagel,
I. Williams,
R. van Leeuwen,
R. Tedder,
C. A. B. Boucher,
F. Brun-Vezinet, and C. Loveday.
1996.
Multicenter comparison of three commercial methods for quantification of human immunodeficiency virus type 1 RNA in plasma.
J. Clin. Microbiol.
34:3016-3022[Abstract].
|
| 10.
|
Yen-Lieberman, B.,
D. Brambilla,
B. Jackson,
J. Bremer,
R. Coombs,
M. Cronin,
S. Herman,
D. Katzenstein,
S. Leung,
H. J. Lin,
P. Palumbo,
S. Rasheed,
J. Todd,
M. Vahey, and P. Reichelderfer.
1996.
Evaluation of a quality assurance program for quantitation of human immunodeficiency virus type 1 RNA in plasma by the AIDS Clinical Trials Group virology laboratories.
J. Clin. Microbiol.
34:2695-2701[Abstract].
|
Journal of Clinical Microbiology, March 1999, p. 812-814, Vol. 37, No. 3
0095-1137/99/$04.00+0
Copyright © 1999, American Society for Microbiology. All rights reserved.
This article has been cited by other articles:
-
Damond, F., Descamps, D., Farfara, I., Telles, J. N., Puyeo, S., Campa, P., Lepretre, A., Matheron, S., Brun-Vezinet, F., Simon, F.
(2001). Quantification of Proviral Load of Human Immunodeficiency Virus Type 2 Subtypes A and B Using Real-Time PCR. J. Clin. Microbiol.
39: 4264-4268
[Abstract]
[Full Text]
-
Désiré, N., Dehée, A., Schneider, V., Jacomet, C., Goujon, C., Girard, P.-M., Rozenbaum, W., Nicolas, J.-C.
(2001). Quantification of Human Immunodeficiency Virus Type 1 Proviral Load by a TaqMan Real-Time PCR Assay. J. Clin. Microbiol.
39: 1303-1310
[Abstract]
[Full Text]
-
Murphy, D. G., Côté, L., Fauvel, M., René, P., Vincelette, J.
(2000). Multicenter Comparison of Roche COBAS AMPLICOR MONITOR Version 1.5, Organon Teknika NucliSens QT with Extractor, and Bayer Quantiplex Version 3.0 for Quantification of Human Immunodeficiency Virus Type 1 RNA in Plasma. J. Clin. Microbiol.
38: 4034-4041
[Abstract]
[Full Text]