Journal of Clinical Microbiology, July 1999, p. 2376-2377, Vol. 37, No. 7
Clinical Pathology, Westchester Medical
Center, Valhalla, New York 10595,1 and
Biology Department, College of Mt. St. Vincent, Bronx, New York
10471-10462
Received 30 November 1998/Accepted 12 April 1999
Growth of microaerophilic bacteria in the AnaeroPack Campylo
(Mitsubishi Gas Chemical America, Inc., New York, N.Y.) atmosphere generation system was compared to growth in the CampyPak Plus jar and
CampyPak pouch (Becton-Dickinson Microbiology Systems [BDMS],
Cockeysville, Md.). Growth in the AnaeroPack Campylo system was
considered equivalent to or better than growth obtained in the CampyPak
Plus and CampyPak pouch systems for 48 of the 50 Helicobacter
pylori strains and for all 28 Campylobacter species tested. All of the 78 organisms tested were recovered in each system in
equivalent colony counts. Two strains of H. pylori grown in
the AnaeroPack Campylo system were observed to have colony morphology
growth discrepancies when compared to growth in the two BDMS systems.
Atmosphere failure with the AnaeroPack Campylo was not detected with
Campylobacter jejuni ATCC 33291 used as a growth control.
The AnaeroPack Campylo system is easy to use and supports the growth of
campylobacters and H. pylori.
Helicobacter and
Campylobacter often cause human gastrointestinal infections
(3, 4, 7). These organisms require reduced oxygen
concentrations of approximately 5 to 10% and enhanced CO2 concentrations of approximately 10% for optimal growth (3, 4, 7). These organisms are relatively slow growers and
require atmosphere generation systems that rapidly achieve and
maintain the appropriate microaerobic atmosphere. Culture methods to
obtain a reduced-oxygen atmosphere have often relied on microaerobic gas generator systems and gas evacuation replacement methods (4, 7). The AnaeroPack Campylo system (Mitsubishi Gas
Chemical America, Inc., New York, N.Y.) is a new microaerobic
gas generator system for the culture of microaerophilic microorganisms.
The AnaeroPack Campylo sachet is placed directly into a jar or
resin pouch without the need either for catalyst or for the
addition of water. Upon exposure to air, the sachet rapidly removes
oxygen to produce a final atmosphere of approximately 6% ± 2% oxygen and 14% ± 2% CO2 (6).
We compared the AnaeroPack Campylo sachet in a 2.5-liter jar system to
the CampyPak Plus envelope in a GasPak jar and to the CampyPak pouch
(Becton-Dickinson Microbiology Systems, Cockeysville, Md.) for the
ability of each system to support growth of Helicobacter pylori and Campylobacter species.
(A preliminary report of this work was presented previously
[9].)
A total of 78 microaerophilic strains were tested for growth in a
2.5-liter Mitsubishi AnaeroPack jar with an AnaeroPack Campylo sachet,
in a 2.5-liter GasPak jar with a CampyPak Plus envelope, and in a
CampyPak pouch. The test organisms included 50 recent clinical isolates
of H. pylori, 24 recent clinical isolates of Campylobacter jejuni (17 obtained from Children's
Hospital, Ontario, Canada), C. jejuni ATCC 33291, Campylobacter fetus subsp. venerealis ATCC 33561, Campylobacter lari ATCC 35221, and Campylobacter
coli ATCC 43477. All clinical isolates were identified by
conventional methods (4, 8). C. jejuni ATCC 33291 was used as an atmosphere control indicator in each test system.
A suspension of each freshly grown organism was prepared in sterile
saline to a turbidity measurement equivalent to a 0.5 McFarland
standard. Each suspension was further diluted in saline to
achieve a final organism concentration of approximately 103
to 104 CFU/ml. For each organism tested, the inoculum
was plated to the appropriate blood-based medium with a 10-µl
calibrated loop to achieve an approximate final inoculum of 30 to 500 CFU per plate. All H. pylori organisms were plated to
brucella blood agar with TVP (Remel, Lenexa, Kans.) and all
campylobacters were plated to Campy Blood agar (Remel). All tests were
performed in duplicate. Test plates were placed into each of the three
atmosphere-generating systems, and the systems were charged according
to the manufacturer's instructions. A plate of C. jejuni
ATCC 33291 was added to each system as an atmosphere control indicator.
Each system was closed securely and incubated at 35°C for all
organisms except for the 25 C. jejuni isolates, which were
incubated at 42°C. The H. pylori isolates were incubated
for at least 5 days up to a maximum of 7 days. The
Campylobacter species were incubated for at least 48 h
up to a maximum of 96 h.
Upon removal from each test system, and prior to examination and
scoring, all media were marked with a code designating the appropriate
test system to minimize any reading bias. All plates were examined and
interpreted independently by two microbiologists. Organism colony
counts were obtained for each plate incubated in each system. Colony
counts were considered discrepant between systems only if there was
greater than a 90% reduction of CFU in one system when compared to the
system with the highest CFU for the strain tested. The quality of
growth for each organism was scored as described previously
(10) and based on typical colony morphology characteristics.
A growth characteristic value (GCV) from 0 (no growth) to 5 (most
typical morphology) was obtained based on the observed quality of
growth scored by the average of two independent readings. The quality
of growth in one system was considered equivalent to that of the other
systems if the average of the GCV readings for growth in one system was
within 2 GCV of the highest average GCV scored for growth in another system.
The colony counts for all 50 H. pylori strains and all 28 Campylobacter species were considered equivalent among the
three systems tested (Table 1). None of
the H. pylori or campylobacter strains failed to grow in any
of the test systems. A total of five strains (three H. pylori and two C. jejuni) exhibited growth quality
discrepancies (Table 1). Only two of those five were observed with the
AnaeroPack Campylo system (Table 2). The
quality of growth of H. pylori 14265 and 10715 was scored
lower (GCV, 2.5) in the AnaeroPack Campylo system than in the two
CampyPak systems (both GCVs, 5). The quality of growth of H. pylori 51687 was observed to be optimal in both the AnaeroPack
Campylo and CampyPak Plus jar systems (GCV, 5) while the quality of
growth in the CampyPak pouch was observed to be lower (GCV, 3).
The growth qualities of H. pylori 25094 (GCV,
2.5) and C. jejuni 8396 (GCV, 3) were observed to be
lower in the CampyPak Plus jar than in the AnaeroPack Campylo system
and the CampyPak pouch system.
0095-1137/99/$04.00+0
Copyright © 1999, American Society for Microbiology. All rights reserved.
Evaluation of the AnaeroPack Campylo System for
Growth of Microaerophilic Bacteria
![]()
ABSTRACT
Top
Abstract
Text
References
![]()
TEXT
Top
Abstract
Text
References
TABLE 1.
Number of H. pylori strains and
Campylobacter spp. yielding equivalent growth in the three
microaerobic atmosphere systemsa
TABLE 2.
Growth discrepancies between H. pylori and
C. jejuni strains in three microaerobic systems
To our knowledge, this is the first extensive evaluation of the AnaeroPack Campylo microaerobic gas generation system for growth of microaerophilic organisms. The AnaeroPack Campylo sachet for a pouch system was not evaluated in this study. All three atmosphere generation systems appear to satisfactorily support the growth of microaerophilic organisms, since none of the strains failed to grow in any of the systems. Atmosphere failures, as indicated by growth of the C. jejuni ATCC 33291 indicator control strain, were not observed in any test system. The previously reported failures of the GasPak jar systems with anaerobic atmosphere generators (1, 2, 5, 10) were not observed in this study with the CampyPak Plus system and could have been due to the anaerobic generator itself. It is also possible that leakage of small amounts of oxygen into a jar system may not affect the microaerophilic organisms as much as it might affect the growth of anaerobes. This is more likely, since GasPak jar failures with an AnaeroPack sachet added for the growth of anaerobes have been reported (1, 10).
The atmosphere for isolation of microaerophilic organisms is considered ideal at 5 to 10% oxygen and 10% CO2 (3, 4, 7). The production of approximately 14% CO2 by the AnaeroPack Campylo system does not appear to adversely affect the growth of either H. pylori or the Campylobacter species tested, since all strains grew well, with only two discrepant H. pylori strains observed. High humidity is also reported to be required for good growth of H. pylori (3), and recent AnaeroPack system instructions recommend adding moisture for better growth. In this study, methods to add moisture were not used. Based on these results, added moisture may not be necessary for growth of H. pylori and Campylobacter species in the AnaeroPack Campylo system. Further studies with variable levels of humidity during incubation might be warranted to obtain optimal incubation conditions for these organisms.
The AnaeroPack Campylo sachet, when used in a 2.5-liter jar, is an acceptable method for the growth of H. pylori and Campylobacter species. The system is easy to use and does not require the addition of water, reagents, or a catalyst. Since this study challenged the AnaeroPack Campylo system only with known organisms, future studies might include direct testing of clinical specimens. Further testing to include the AnaeroPack Campylo pouch system is also needed.
| |
ACKNOWLEDGMENTS |
|---|
This study was funded in part by a grant from Mitsubishi Gas Chemical America, Inc., New York, N.Y.
We thank Frank Chan of Children's Hospital, Eastern Ontario, Ottawa, Canada, for supplying the additional clinical strains of C. jejuni.
| |
FOOTNOTES |
|---|
* Corresponding author. Mailing address: Clinical Pathology, 1J04, Westchester Medical Center, Grasslands Rd., Valhalla, NY 10595. Phone and Fax: (914) 493-7502. E-mail: kgvanhorn{at}juno.com.
| |
REFERENCES |
|---|
|
|
|---|
| 1. | Delaney, M. L., and A. B. Onderdonk. 1997. Evaluation of the AnaeroPack system for growth of clinically significant anaerobes. J. Clin. Microbiol. 35:558-562[Abstract]. |
| 2. | Imhoff, A., and I. Heinzer. 1996. Continuous monitoring of oxygen concentrations in several systems for cultivation of anaerobic bacteria. J. Clin. Microbiol. 34:1646-1648[Abstract]. |
| 3. | Jerris, R. C. 1995. Helicobacter, p. 492-498. In P. R. Murray, E. J. Baron, M. A. Pfaller, F. C. Tenover, and R. H. Yolken (ed.), Manual of clinical microbiology, 6th ed. American Society for Microbiology, Washington, D.C. |
| 4. | Kaplan, R. L., and A. S. Weissfeld. 1994. Campylobacter, Helicobacter, and related organisms, p. 453-460. In B. J. Howard, J. F. Keiser, A. S. Weissfeld, T. F. Smith, and R. C. Tilton (ed.), Clinical and pathogenic microbiology, 2nd ed. The C. V. Mosby Co., St. Louis, Mo. |
| 5. | Miller, P. H., L. S. Wiggs, and J. M. Miller. 1995. Evaluation of AnaeroGen system for growth of anaerobic bacteria. J. Clin. Microbiol. 33:2388-2391[Abstract]. |
| 6. | Mitsubishi Gas Chemical Co. 1996. AnaeroPack product literature. Mitubishi Gas Chemical Co., New York, N.Y. |
| 7. | Nachamkin, I. 1995. Campylobacter and Arcobacter, p. 483-491. In P. R. Murray, E. J. Baron, M. A. Pfaller, F. C. Tenover, and R. H. Yolken (ed.), Manual of clinical microbiology, 6th ed. American Society for Microbiology, Washington, D.C. |
| 8. | On, S. L. W. 1996. Identification methods for campylobacters, helicobacters, and related organisms. Clin. Microbiol. Rev. 9:405-422[Abstract]. |
| 9. | Van Horn, K. 1997. Evaluation of a new microaerophilic atmosphere generation system for growth of Helicobacter pylori, abstr. C-98, p. 137. In Abstracts of the 97th General Meeting of the American Society for Microbiology, Washington, D.C. |
| 10. | Van Horn, K. G., K. Warren, and E. J. Baccaglini. 1997. Evaluation of the AnaeroPack system for growth of anaerobic bacteria. J. Clin. Microbiol. 35:2170-2173[Abstract]. |