Previous Article | Next Article 
Journal of Clinical Microbiology, August 1999, p. 2525-2532, Vol. 37, No. 8
0095-1137/99/$04.00+0
Copyright © 1999, American Society for Microbiology. All rights reserved.
Genotype Dependence of Hepatitis C Virus Load
Measurement in Commercially Available Quantitative Assays
Janet
Mellor,
Anna
Hawkins, and
Peter
Simmonds*
Department of Medical Microbiology,
University of Edinburgh, Edinburgh EH8 9AG, United Kingdom
Received 28 December 1998/Returned for modification 16 March
1999/Accepted 29 April 1999
 |
ABSTRACT |
Standardization and genotype independence of methods used to
quantify hepatitis C virus (HCV) RNA in clinical specimens are necessary for accurate assessment of the role of HCV quantitation as a
prognostic marker for HCV infection and monitoring of the response to
antiviral treatment. Commercially available methods used to measure HCV
loads include PCR-based (Roche Monitor) and hybridization-based
(Quantiplex bDNA-2) methods. Recently, a new version of the Roche
Monitor assay (version 2.0) has become available; it has been modified
to achieve more equal quantitation of different HCV genotypes.
Consistent with previous reports, Roche Monitor version 1.0 substantially underestimated concentrations of RNA transcripts of types
2b, 3a, 4a, 5a, and 6a and virus loads in individuals infected with
genotypes 2 to 6 relative to reference tests. However, version 2.0 achieved equivalent quantitation of each genotype over a narrow
quantitative range (103 to 5 × 105 copies
of RNA/ml) but significantly underestimated RNA concentrations above
this range. The assay showed an equivalent inability to quantify high
levels of HCV RNA in plasma samples, and this was responsible for the
falsely narrow range of virus loads detected in HCV-infected
individuals. In contrast, the Chiron bDNA-2 assay could only measure
RNA concentrations in the upper quantitative range (2 × 105 to 5 × 107 copies of RNA/ml) but
showed equivalent sensitivity for genotypes 1 to 5; however,
concentrations of type 6a RNA transcripts and virus loads in clinical
specimens from individuals infected with type 6a were underestimated by
a factor of 2 to 4. Differences were observed between PCR- and
hybridization-based assays in their relative quantitation of HCV RNA
transcripts and HCV genomic RNA, which may cause problems with the use
of transcripts for interassay calibration.
 |
INTRODUCTION |
Quantitation of hepatitis C virus
(HCV) RNA sequences in plasma has been used extensively as a prognostic
marker for individuals undergoing treatment with interferon and in the
subsequent monitoring of their responses. Several investigators have
found significant differences in virus load associated with infection
with different genotypes of HCV (1, 2, 10, 11, 13, 14, 18),
and it had been argued that this could be one the factors involved in
genotype-specific differences in the outcome of interferon therapy, in
particular, the reduced responses of individuals infected with type 1b
compared to those infected with types 2 and 3 (5, 14, 18).
Proper assessment of the effect of HCV genotype on outcome requires a
method for the quantitation of viral loads in plasma that is equally
sensitive for each genotype.
We have previously investigated the genotype independence of the Chiron
bDNA assays (Quantiplex RNA assay bDNA-1) and Quantiplex bDNA 2, Roche
Monitor assay version 1.0, and nested PCR at limiting dilution by using
samples of genotypes 1, 2, and 3 (8). Significant differences in the efficiency of detection of genotypes 1, 2, and 3 were observed for the bDNA-1 and Roche Monitor assays, whereas the
bDNA-2 assay and the nested PCR at limiting dilution were able to
quantify the RNA sequences of different genotypes with equal
efficiencies. In the current study, we have extended this investigation
to a comparison of virus load measurement for genotypes 4 and 6 in the
bDNA-2, limiting-dilution PCR, and Roche Monitor version 1.0 assays and
retested all of the genotypes using the new version (version 2.0) of
the Roche Monitor assay. This assay had been modified by the
manufacturer with the express intention of achieving more equal
quantitation of different genotypes.
 |
MATERIALS AND METHODS |
Samples.
Ninety-six plasma samples from HCV-infected,
PCR-positive blood donors (20 samples each of genotypes 1, 2, and 3; 19 of genotype 4; 1 of genotype 5; and 16 of genotype 6) were analyzed for
HCV RNA levels. Type 4 samples were from Middle Eastern countries, and
type 6 samples were from Hong Kong. Fresh plasma samples were aliquoted
to ensure that all of the quantitative assays were carried out on
samples that had been frozen and thawed only twice. All samples were
tested with the bDNA-2, limiting-dilution PCR, and Roche Monitor
version 1.0 and 2.0 assays. Previously reported results for type 1, 2, and 3 samples in the bDNA-2, limiting-dilution PCR, and Roche Monitor
version 1.0 assays were included in the current study for comparison,
while RNA transcripts were retested. Samples from individuals infected
with HCV genotypes 4, 5, and 6 were genotyped by restriction fragment
length polymorphism analysis of the 5' noncoding region (5'NCR)
(19). Samples were aliquoted and stored at
40°C and were
only frozen and thawed twice prior to testing.
Quantitation of HCV-infected samples.
Three commercial
assays for the detection of HCV RNA were used, i.e., Quantiplex HCV RNA
assay 2.0 or bDNA-2 and the Roche Monitor assay versions 1.0 and 2.0. These assays were performed in accordance with the manufacturers'
instructions. Unamplified HCV RNA was detected in the bDNA-2 assay by
capture onto the solid phase by hybridization to oligonucleotides
complementary to the 5'NCR and core regions, followed by detection and
signal amplification to labelled probes. All samples were tested with
kit lot no. MMM169.
The Roche Monitor assays were based upon reverse transcription and
amplification of the HCV RNA with primers from the 5'NCR in the
presence of an internal control which shared primers with HCV. Both the
competitor and sample DNAs were detected by hybridization to a
biotin-labelled probe, followed by incubation with enzymatically labelled streptavidin and detection in a colorimetric assay. Samples were tested with kit lots H0315 and G1722 (version 1.0) and H0526 and
H011 (version 2.0).
For the in-house limiting-dilution method (8, 15), viral RNA
was extracted from 100 µl of plasma by incubation with proteinase K
(1 mg/ml) and sodium dodecyl sulfate (0.55%) in the presence of
poly(A) (40 mg/ml) and purified by phenol-chloroform extraction as
previously described (3). HCV RNA was reversed transcribed and amplified by nested PCR with primers specific for the 5'NCR. Limiting dilution of cDNA was then carried out by using an assumed efficiency of 5% for the reverse transcription step, as previously established (8, 20).
Sequence analysis.
Sequence analysis of the 5'NCR was
performed by cycle sequencing in accordance with the instructions
supplied with the Thermo Sequenase kit from Amersham International.
Quantitation of HCV RNA transcripts.
HCV transcripts of all
six genotypes were provided by J. Detmer (Bayer Diagnostics,
Emeryville, Calif.). The methods used for their synthesis and
quantitation have been described elsewhere (4, 7). Briefly,
to assess the quality of the preparations, HCV RNA transcripts were
electrophoresed on 1.5% formaldehyde gels and scanned on an Ambis 400 radioanalytic imager (Ambis, Inc., San Diego, Calif.). The preparations
of the HCV RNA transcripts used contained less than 3% free
nucleotides and were composed of at least 80% full-length transcripts.
Three independent analytical methods were used to quantify the
transcripts. These included phosphate determination (6),
measurement of A260, and hyperchromicity analysis (17). Transcripts were quantified by the bDNA-2 and limiting-dilution assays and both Roche Monitor assays.
Statistical analysis.
Differences in the distribution of
quantitative values between genotypes were detected by the Mann-Whitney
U test. The correlation between assays in the quantitation of HCV RNA
was measured by using the Spearman nonparametric test.
 |
RESULTS |
Quantitation of HCV by four different assays.
HCV RNA in
samples from a total of 96 blood donors infected with genotypes 1 to 6 was quantified by using the bDNA-2 assay in replicate in accordance
with the manufacturer's instructions, once by the Roche Monitor 1.0 and 2.0 assays and in at least eight replicates at limiting dilution by
the in-house assay (Table 1). The cutoff
sensitivity of the bDNA-2 assay was 0.2 × 106 copies
of RNA per ml, whereas the Roche Monitor and limiting-dilution assays
had a cutoff sensitivity of approximately 1,000 copies of RNA per ml.
Of 35 type 4 and 6 samples, 4 (11.4%) were below the cutoff of the
bDNA-2 assay and a different four samples were below the cutoff of the
Roche Monitor 1.0 assay, but all samples were positive by the Roche
Monitor 2.0 assay. Samples below the cutoff of each assay were assigned
the virus load of the cutoff. This approximation did not affect the
nonparametric methods used to analyze the data.
Correlations between the different assays for virus load measurement
were obtained. Spearman correlation coefficients ranged from 0.706 for
virus loads detected in the bDNA-2 and limiting-dilution assays to
0.351 between Roche Monitor versions 1.0 and 2.0. Higher correlation
coefficients were obtained for samples of genotype 1 than for those of
other genotypes (0.747 for bDNA-2 with limiting dilution and 0.821 and
0.847 with Roche Monitor versions 1.0 and 2.0, 0.611 and 0.540 for
limiting dilution with Roche Monitor versions 1.0 and 2.0, and 0.714 between Roche Monitor versions 1.0 and 2.0). Correlation coefficients
ranged from 0.445 to 0.731 (genotype 2), from 0.432 to 0.756 (genotype
3), from 0.246 to 0.756 (genotype 4), and from
0.313 to 0.903 (genotype 6) in pairwise comparisons of the quantitative results of the
four assays.
Virus load and genotype.
In three of the four assays, a wide
range of virus loads were observed among donors infected with each
genotype (Table 1 and Fig. 1). However,
the range of virus loads obtained by using Roche Monitor 2.0 was much
narrower than that obtained with the other assays. The median values
calculated for these distributions differed between different
quantitation assays. For example, among the type 1 samples, median
values of 3.415, 2.065, 1.041, and 0.273 were observed for the bDNA-2,
limiting-dilution, and Roche Monitor version 1.0 and 2.0 methods,
respectively, whereas for type 4 samples, medians of 2.239, 1.84, 0.038, and 0.147 were obtained. When using the Mann-Whitney U test for
pairwise comparisons between genotypes, we observed no significant
difference in virus load between donors infected with types 1, 2, 3, and 4 when samples were analyzed with the bDNA-2 and limiting-dilution
assays. However, virus loads of type 6 samples were lower than those of
type 1 samples in bDNA-2 (median of 0.439 compared with 3.415), a
difference that approached statistical significance (P = 0.08). However, no difference in virus load between type 6 and 1 samples was detected by the limiting-dilution assay (medians of 2.320 and 2.065, respectively; P = 0.787).

View larger version (24K):
[in this window]
[in a new window]
|
FIG. 1.
Range of HCV RNA levels in plasma samples from blood
donors infected with genotypes 1 to 6 as detected by the bDNA-2 (A),
limiting-dilution (B), Roche Monitor version 1.0 (C), and Roche Monitor
version 2.0 (D) quantitative assays. The cutoff value of the bDNA-2
assay (0.2 × 106 copies/ml) is indicated.
|
|
Differences in virus loads between genotypes were observed with the
Roche Monitor version 1.0 assay. Virus loads of non-type 1 genotypes
were significantly lower than those of the type 1 samples (P
values were all
0.001), apart from the type 2 samples, where the
difference did not reach statistical significance. Small differences
between genotype 1, 2, and 3 viral loads were observed when the Roche
Monitor version 2.0 assay was used. However, type 4 samples showed
statistically significantly lower levels of RNA than type 1 samples
(median, 0.147; P = 0.013) while type 6 sample RNA
levels were higher (median, 0.568; P = 0.005).
The ratios of virus loads individually measured for each sample by the
bDNA-2 and Roche Monitor version 1.0 and 2.0 assays to those measured
by the limiting-dilution assay were calculated (Table
2). A particularly wide range of ratios
were observed with the Roche Monitor version 1.0 assay versus the
limiting-dilution assay, with values for types 3, 4, and 6 consistently
lower that those obtained for genotype 1 (P < 0.01).
In contrast, no significant difference in the ratios obtained for the
limiting-dilution assay with the bDNA-2 assay were observed for
genotypes 1, 2, 3, and 4 (Table 2). However, the median ratio of
genotype 6 samples to type 1 samples was 0.623, significantly lower
(P = 0.03) than in the reference test. In contrast, the
ratio of Roche Monitor version 2.0 to limiting-dilution assay results
for type 6 samples was much higher than those calculated for the other
genotypes (2.950). These samples had been tested at a 1-in-10 dilution
in the former assay because of the limited sample volume available. Subsequent experiments (see below) suggested that the difference in
quantitation between samples of type 6 and those of other genotypes might not have been observed had it been possible to test them undiluted.
Sequence analysis of the 5'NCRs of types 4 and 6.
Current
GenBank entries do not provide sequence information upstream of the
5'NCR primers used in the Roche and limiting-dilution assays. In order
to determine if sequence variation within the binding sites of the
primers used for amplification could account for the differences in
viral load detected in the genotype 4 and 6 samples, we obtained
nucleotide sequences by PCR using a novel primer from the extreme 5'
end of the HCV genome in a heminested PCR. No sequence variability in
binding sites was detected in either the type 4 or 6 samples by either
the Roche or limiting-dilution assays.
Quantitation of HCV RNA transcripts.
To compare the absolute
quantitations of RNA sequences provided by the different assays, four
different concentrations of RNA transcripts of HCV genotypes 1 through
6 were analyzed by Roche Monitor test versions 1.0 and 2.0 and the
bDNA-2 and limiting-dilution assays (Fig.
2). The results obtained for each sample
were deemed valid by the kit instructions, although both Roche Monitor
test versions 1.0 and 2.0 produced unexpectedly low quantitation values above concentrations of 106 copies of RNA/ml (Fig. 2) and
were unable to accurately quantify samples with greater than
106 HCV RNA copies per ml. In contrast, the bDNA-2 assay
was relatively accurate in detecting HCV RNA levels over a range of
transcript concentrations of 5 × 105 to 5 × 107 but was unable to detect HCV RNA levels below the
stated quantitation limit of 2 × 105 copies/ml. The
assay was less sensitive toward genotype 6a transcripts in comparison
with the other genotypes. Surprisingly, HCV RNA levels detected with
the in-house limiting-dilution assay were consistently higher than the
amount of HCV RNA transcript added to the test for all genotypes, with
the exception of genotype 3a (within the error margin); these findings
contrast with the equivalence of virus loads between this assay and the
bDNA-2 assay when plasma samples are tested. This may have occurred
because RNA transcripts may be more efficiently reverse transcribed
than full-length HCV genomic sequences (see Discussion).

View larger version (42K):
[in this window]
[in a new window]
|
FIG. 2.
Measurement of the absolute efficiencies of quantitative
assays by using RNA transcripts of genotypes 1a, 2b, 3a, 4a, 5a, and
6a. The ratio of the amount of HCV RNA detected by each assay to a
known amount of transcript added was calculated with increasing amounts
of RNA transcript to determine the linear dynamic range of each
assay.
|
|
Quantitation range of the Roche Monitor version 2.0 assay.
Following the observation of consistent underquantitation of RNA
transcripts at concentrations of >106 copies/ml, we
investigated whether a similar "saturation" effect influences the
quantitation of clinical specimens. Using samples of genotypes 1, 2, and 3, we compared the viral loads detected by Roche Monitor version
2.0 in undiluted samples with those obtained after 1-in-10 dilution
prior to testing and where the measured virus loads were multiplied by
10 (Table 3). The values of optical density at 450 nm (OD450) obtained with the Roche Monitor
version 2.0 assay of undiluted samples all fell within the operating
range of the assay (OD450,
0.15 and
2.0). However,
virus loads in the samples diluted 1 in 10 were consistently higher
than those of undiluted samples and were closer to those measured by
the bDNA-2 or limiting-dilution assay. (This is consistent with the finding reported in Table 1 for higher virus loads in the type 6 samples.) For example, samples of genotype 1 showed a median viral load
ratio of Roche Monitor version 2.0 to limiting dilution of 0.138 when
tested undiluted, compared with 0.774 for samples diluted 1 in 10.
View this table:
[in this window]
[in a new window]
|
TABLE 3.
Comparison of quantitation of HCV RNA in clinical
specimens at two dilutions by Roche Monitor version 2.0 with that by
other assays
|
|
In order to determine if the Roche Monitor assay version 2.0 was
operating outside its optimal range, a different set of samples of
genotypes 1, 2, 3, and 4 were diluted to an estimated viral load of
n × 104 HCV copies per ml according to
that calculated by limiting-dilution assay and retested by Roche
Monitor version 2.0. The results obtained, once corrected for the
dilution factor, were compared to the viral loads obtained from
undiluted aliquots of the same samples (Fig. 3). Consistently higher viral loads were
obtained when samples were diluted prior to testing for clinical
samples with viral titers higher than 105 to
106 HCV copies per ml of plasma.

View larger version (25K):
[in this window]
[in a new window]
|
FIG. 3.
Efficiencies of detection of HCV by the Roche Monitor
version 2.0 and limiting-dilution assays for genotypes 1 (A), 2 (B), 3 (C), and 4 (D). Five samples of each genotype were tested undiluted and
diluted to n × 104 according to the viral
load calculated by limiting dilution to within the dynamic range of the
Roche Monitor version 2.0 assay.
|
|
To investigate the difference in quantitation over a wide range of
dilutions, three further samples (1 each of genotypes 1 to 3) were
serially diluted and retested by the Roche Monitor version 2.0 assay
(Fig. 4; Table
4). Plotting of the ratio of the amount
of HCV RNA detected by Roche Monitor version 2.0 at various dilutions
to the input, determined by limiting dilution, against the amount of
HCV RNA added, showed that the measured HCV RNA levels increased in
proportion to the dilution. For example, the type 2 sample virus load
increased from 2.94 × 105 HCV copies per ml of plasma
undiluted to 1.43 × 107 HCV copies per ml at the
final dilution (104.0), giving ratios of the amount
detected to the amount added of 0.012 and 0.596, respectively. For each
of the samples, the Roche Monitor version 2.0 assay measured virus
loads comparable to those measured by limiting dilution only at
greater-than-100-fold dilutions.

View larger version (22K):
[in this window]
[in a new window]
|
FIG. 4.
Three plasma samples of genotypes 1, 2, and 3 were
subjected to serial dilution prior to testing by the Roche Monitor
version 2.0 assay. The ratio of the amount detected by the Roche
Monitor version 2.0 assay to the amount added, according to limiting
dilution (see Table 4), was plotted against the dilution factor.
|
|
 |
DISCUSSION |
This study is an extension of the previous study (8)
which used a range of quantitative assays to measure virus load in individuals infected with HCV genotypes 1, 2, and 3. The existence of
genotype-specific differences in the sensitivities of these assays, as
documented for the original bDNA-1 assay (2, 4, 7) and the
Roche Monitor version 1.0 assay (2), affects their
quantitative accuracy, making it difficult to properly assess virological and host factors that influence disease severity and response to interferon therapy and interferon-ribavirin combination therapy. For example, underestimation of the virus load in type 2-infected individuals may lead to erroneous attribution of the virus
load to their greater response to treatment, even though it has been
documented extensively that the median virus loads of type 1- and
2-infected individuals are similar (12, 16, 18) and that
(poorly understood) virological differences between the two genotypes
contribute additionally and independently of the virus load to the
likelihood of an antiviral response.
Among currently available commercial assays for virus load measurement,
the Roche Monitor version 1.0 competitive assay shows the greatest
differences in sensitivity to genotypes 1 and 2 (times a factor of 9)
and 3 (times a factor of 12) (8). In the current study,
correction factors of times 53 for type 4 samples and times 15 for type
6 samples might be applied to produce values as calibrated to the
limiting-dilution and bDNA-2 assays, as proposed previously for types 2 and 3, although such corrections are unlikely to be accurate for
individual samples (8). The new version of the Roche Monitor
assay was modified to reduce the differences in sensitivity to non-type
1 genotypes. For low concentrations of RNA transcripts, detection
efficiencies were equivalent between genotypes and produced
quantitative values corresponding closely to the amount added; these
findings stand in contrast in both aspects to those of the previous
assays of type 1 to 3 transcripts (8). However, the
narrowness of the quantitative range was demonstrated by transcripts
and by testing of clinical specimens at different input dilutions (Fig.
3, 4, and 5). Unfortunately for clinical use, measured virus loads from
clinical specimens might vary over nearly 2 logs, depending on the test
concentration (Fig. 4), with no indication from the test results of
whether the virus load was correct or not. For example, undiluted type 2 and 3 samples produced incorrect virus loads of 2.94 × 105 and 3.16 × 105 from OD450
values of 0.515 and 0.563 (internal control values, 0.842 and 0.865)
yet they met the manufacturer's acceptance criteria (Table 4).
Equal sensitivities for all of the genotypes in the Roche Monitor
version 2.0 assay appeared to be supported by the equivalence in median
virus loads of clinical specimens from individuals infected with
genotypes 1 to 4. However, this latter observation is likely to also
have been an artifact of the inability to quantify higher virus loads
(the median virus load of the type 6a samples tested at a 1:10 dilution
was greater than those of the other genotypes). The extremely narrow
range of quantitation was the probable reason for the restricted range
of virus loads measured in the clinical specimens compared with other
assays (Fig. 1) and the poor correlation coefficients with results from
other assays. The distribution of virus loads differed principally from
the other assays, including the original Roche Monitor assay, in
lacking virus load values of greater than 106/ml (Fig. 1).
In contrast to the Roche Monitor version 2.0 assay, the bDNA-2 assay
had a range of quantitation that was more appropriate for virus load
testing of untreated study subjects, with 85% of the test specimens
having virus loads in the linear range of the assay (2 × 105 to 5 × 107 copies/ml, as evaluated by
using the RNA transcripts). However, while several studies have
documented the equivalence of quantitation of HCV genotypes 1, 2, and 3 (7-9), both transcripts and testing of clinical specimens
in the current study revealed a measurable reduction in sensitivity for
type 6a sequences compared with the limiting-dilution assay (twofold to
eightfold). The extent to which type 6a samples may be underquantified
requires further investigation with a larger number of samples.
Finally, and separately from the genotype issue, quantitation of RNA
transcripts and HCV genomic RNA in clinical specimens revealed subtle
differences in their efficiency of detection by PCR-based and
hybridization-based assays. Absolute quantitation with the
limiting-dilution assay requires an empirical measurement of the
efficiency of reverse transcription in order to convert the measured
frequency of cDNA sequences to an RNA concentration. For clinical
specimens, we and others have used an efficiency of 5% as a conversion
factor for HCV (and human immunodeficiency virus type 1) PCR. In the
current study and in our previous study, a 5% efficiency of reverse
transcription leads to equivalence of virus loads to those detected by
the bDNA-2 assay. However, use of this value for quantitation of RNA
transcripts produced consistent overestimation of the RNA concentration
(apart from type 3a transcripts). To obtain equivalence to the amount
of RNA transcript actually added and to the results of the bDNA-2 assay would necessitate a substantially greater efficiency of reverse transcription (25% for genotypes 1a, 2b, 4a, 5a, and 6a).
Both reverse transcription efficiency and hybridization to the probes
used in the bDNA-2 assay may be influenced by the secondary structure
of the RNA target. The RNA transcripts used for assay calibration used
in this and previous studies (7, 8) were 823 bases in length
and may therefore show internal base pairing in this region different
from that of full-length sequences. Structural differences that affect
the accessibility of internal sequences to hybridization to the probes
used in the bDNA-2 assay or to reverse transcription may therefore
account for the differences in relative quantitation by PCR- and
hybridization-based assays. Potentially, this could prevent the use of
subgenomic RNA transcripts for the calibration of assays for the
quantitation of HCV genomic RNA in clinical specimens.
In summary, this study describes the equivalent or nearly equivalent
quantitation of HCV genotypes by two commercially available assays that
operate over quite different ranges of virus loads, which necessarily
limits the applications for which the respective assays could be used clinically.
 |
ACKNOWLEDGMENTS |
We are very grateful to F. Davidson, Scottish National Blood
Transfusion Service; L. Prescott, Department of Medical Microbiology, University of Edinburgh; G. M. Dusheiko, Department of Medicine, Royal Free Hospital, London; and C. K. Lin, Hong Kong Red Cross, Kowloon, for providing samples of genotypes 4 and 6 for quantitative analysis. We are also indebted to J. Detmer and J. Kolberg for providing RNA transcripts of genotypes 1 to 6.
This study was funded in part by a research grant to J.M. and P.S. from
Nucleic Acid Systems, Chiron Corporation.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: Department of
Medical Microbiology, University of Edinburgh, Teviot Place, Edinburgh EH8 9AG, United Kingdom. Phone: 44 131 650 3138. Fax: 44 131 650 6531. E-mail: Peter.Simmonds{at}ed.ac.uk.
 |
REFERENCES |
| 1.
|
Berger, A.,
M. von Depka Prondzinski,
H. W. Doerr,
H. Rabenau, and B. Weber.
1996.
Hepatitis C plasma viral load is associated with HCV genotype but not with HIV coinfection.
J. Med. Virol.
48:339-343[Medline].
|
| 2.
|
Chan, C. Y.,
S. D. Lee,
S. J. Hwang,
R. H. Lu,
C. L. Lu, and K. J. Lo.
1995.
Quantitative branched DNA assay and genotyping for hepatitis C virus RNA in Chinese patients with acute and chronic hepatitis C.
J. Infect. Dis.
171:443-446[Medline].
|
| 3.
|
Chan, S.-W.,
F. McOmish,
E. C. Holmes,
B. Dow,
J. F. Peutherer,
E. Follett,
P. L. Yap, and P. Simmonds.
1992.
Analysis of a new hepatitis C virus type and its phylogenetic relationship to existing variants.
J. Gen. Virol.
73:1131-1141[Abstract/Free Full Text].
|
| 4.
|
Collins, M. L.,
C. Zayati,
J. J. Detmer,
B. Daly,
J. A. Kolberg,
T. A. Cha,
B. D. Irvine,
J. Tucker, and M. S. Urdea.
1995.
Preparation and characterization of RNA standards for use in quantitative branched DNA hybridization assays.
Anal. Biochem.
226:120-129[Medline].
|
| 5.
|
Craxi, A.,
S. Magrin,
C. Fabiano,
C. Linea, and P. Almasio.
1995.
Host and viral features in chromic HCV infection: relevance to interferon responsiveness.
Res. Virol.
146:273-278[Medline].
|
| 6.
|
Daly, J. A., and G. Ertingshausen.
1972.
Direct method for determining inorganic phosphate in serum with the "CentrifiChem."
Clin. Chem.
18:263-265[Abstract].
|
| 7.
|
Detmer, J.,
R. Lagier,
J. Flynn,
C. Zayati,
J. Kolberg,
M. Collins,
M. Urdea, and R. Sanchez-Pescador.
1996.
Accurate quantification of hepatitis C virus (HCV) RNA from all HCV genotypes by using branched-DNA technology.
J. Clin. Microbiol.
34:901-907[Abstract].
|
| 8.
|
Hawkins, A.,
F. Davidson, and P. Simmonds.
1997.
Comparison of plasma virus loads among individuals infected with hepatitis C virus (HCV) genotypes 1, 2, and 3 by Quantiplex HCV RNA assay versions 1 and 2, Roche Monitor assay, and an in-house limiting dilution method.
J. Clin. Microbiol.
35:187-192[Abstract].
|
| 9.
|
Lau, J. Y. N.,
P. Simmonds, and M. S. Urdea.
1995.
Implications of variations of "conserved" regions of hepatitis C virus genome.
Lancet
346:425-426[Medline].
|
| 10.
|
Lu, R. H.,
S. J. Hwang,
C. Y. Chan,
F. Y. Chang, and S. D. Lee.
1998.
Quantitative measurement of serum HCV RNA in patients with chronic hepatitis C: comparison between Amplicor HCV Monitor system and branched DNA signal amplification assay.
J. Clin. Lab. Anal.
12:121-125[Medline].
|
| 11.
|
Mahaney, K.,
V. Tedeschi,
G. Maertens,
A. M. Dibisceglie,
J. Vergalla,
J. H. Hoofnagle, and R. Sallie.
1994.
Genotypic analysis of hepatitis C virus in American patients.
Hepatology
20:1405-1411[Medline].
|
| 12.
|
Martinot Peignoux, M.,
P. Marcellin,
M. Pouteau,
C. Castelnau,
N. Boyer,
M. Poliquin,
C. Degott,
I. Descombes,
V. Lebreton,
V. Milotova,
J. P. Benhamou, and S. Erlinger.
1995.
Pretreatment serum hepatitis C virus RNA levels and hepatitis C virus genotype are the main and independent prognostic factors of sustained response to interferon alfa therapy in chronic hepatitis C.
Hepatology
22:1050-1056[Medline].
|
| 13.
|
Matsumoto, A.,
E. Tanaka,
T. Suzuki,
H. Ogata, and K. Kiyosawa.
1994.
Viral and host factors that contribute to efficacy of interferon-alpha(2a) therapy in patients with chronic hepatitis C.
Dig. Dis. Sci.
39:1273-1280[Medline].
|
| 14.
|
Orito, E.,
M. Mizokami,
T. Nakano,
H. Terashima,
O. Nojiri,
H. Sakakibara,
M. Mizuno,
M. Ogino,
M. Nakamura,
Y. Matsumoto,
K. I. Miyata, and J. Y. N. Lan.
1994.
Serum hepatitis C virus RNA level as a predictor of subsequent response to interferon-alpha therapy in Japanese patients with chronic hepatitis C.
J. Med. Virol.
44:410-414[Medline].
|
| 15.
|
Simmonds, P.,
P. Balfe,
J. F. Peutherer,
C. A. Ludlam,
J. O. Bishop, and A. J. Brown.
1990.
Human immunodeficiency virus-infected individuals contain provirus in small numbers of peripheral mononuclear cells and at low copy numbers.
J. Virol.
64:864-872[Abstract/Free Full Text].
|
| 16.
|
Smith, D. B.,
F. Davidson,
P. L. Yap,
H. Brown,
J. Kolberg,
J. Detmer,
M. Urdea,
P. Simmonds,
J. Mellor,
J. Neville,
L. Prescott,
B. C. Dow,
E. A. C. Follett,
T. Skuldamrongpanich,
S. Tanprasert,
C. Nuchaprayoon,
C. K. Lin,
M. C. Kew,
R. Crookes,
J. D. Conradie,
M. Lin,
C. Seed,
G. A. B. Deolim,
I. A. Martins,
G. A. Medgyesi,
T. Krusius,
A. A. Saeed,
D. Teo, and S. H. Lee.
1996.
Levels of hepatitis C virus in blood donors infected with different viral genotypes.
J. Infect. Dis.
173:727-730[Medline].
|
| 17.
|
Warshaw, M. M., and I. Tinico.
1966.
Optical properties of sixteen dinucleoside phosphates.
J. Mol. Biol.
20:29-38[Medline].
|
| 18.
|
Yamada, G.,
R. Takatani,
F. Kishi,
M. Takahashi,
T. Doi,
T. Tsuji,
S. Shin,
M. Tanno,
M. S. Urdea, and J. A. Kolberg.
1995.
Efficacy of interferon alfa therapy in chronic hepatitis C patients depends primarily on hepatitis C virus RNA level.
Hepatology
22:1351-1354[Medline].
|
| 19.
|
Zein, N. N.,
J. Rakela, and D. H. Persing.
1995.
Genotype-dependent serologic reactivities in patients infected with hepatitis C virus in the United States.
Mayo Clin. Proc.
70:449-452[Medline].
|
| 20.
|
Zhang, L. Q.,
P. Simmonds,
C. A. Ludlam, and A. J. Leigh Brown.
1991.
Detection, quantitation and sequencing of HIV-1 virus from the plasma of seropositive individuals and from factor VIII concentrates.
AIDS
5:675-681[Medline].
|
Journal of Clinical Microbiology, August 1999, p. 2525-2532, Vol. 37, No. 8
0095-1137/99/$04.00+0
Copyright © 1999, American Society for Microbiology. All rights reserved.
This article has been cited by other articles:
-
Forman, M. S., Valsamakis, A.
(2008). Performance Characteristics of a Quantitative Hepatitis C Virus RNA Assay Using COBAS AmpliPrep Total Nucleic Acid Isolation and COBAS TaqMan Hepatitis C Virus Analyte-Specific Reagent. J. Mol. Diagn.
10: 147-153
[Abstract]
[Full Text]
-
Forman, M. S., Valsamakis, A.
(2004). Verification of an Assay for Quantification of Hepatitis C Virus RNA by Use of an Analyte-Specific Reagent and Two Different Extraction Methods. J. Clin. Microbiol.
42: 3581-3588
[Abstract]
[Full Text]
-
Schuttler, C. G., Thomas, C., Discher, T., Friese, G., Lohmeyer, J., Schuster, R., Schaefer, S., Gerlich, W. H.
(2004). Variable Ratio of Hepatitis C Virus RNA to Viral Core Antigen in Patient Sera. J. Clin. Microbiol.
42: 1977-1981
[Abstract]
[Full Text]
-
Morishima, C., Chung, M., Ng, K. W., Brambilla, D. J., Gretch, D. R.
(2004). Strengths and Limitations of Commercial Tests for Hepatitis C Virus RNA Quantification. J. Clin. Microbiol.
42: 421-425
[Abstract]
[Full Text]
-
Quarleri, J. F., Bussy, M. V., Mathet, V. L., Ruiz, V., Iacono, R., Lu, L., Robertson, B. H., Oubina, J. R.
(2003). In Vitro Detection of Dissimilar Amounts of Hepatitis C Virus (HCV) Subtype-Specific RNA Genomes in Mixes Prepared from Sera of Persons Infected with a Single HCV Genotype. J. Clin. Microbiol.
41: 2727-2733
[Abstract]
[Full Text]
-
White, P. A., Pan, Y., Freeman, A. J., Marinos, G., Ffrench, R. A., Lloyd, A. R., Rawlinson, W. D.
(2002). Quantification of Hepatitis C Virus in Human Liver and Serum Samples by Using LightCycler Reverse Transcriptase PCR. J. Clin. Microbiol.
40: 4346-4348
[Abstract]
[Full Text]
-
Schirm, J., van Loon, A. M., Valentine-Thon, E., Klapper, P. E., Reid, J., Cleator, G. M.
(2002). External Quality Assessment Program for Qualitative and Quantitative Detection of Hepatitis C Virus RNA in Diagnostic Virology. J. Clin. Microbiol.
40: 2973-2980
[Abstract]
[Full Text]
-
Beld, M., Sentjens, R., Rebers, S., Weegink, C., Weel, J., Sol, C., Boom, R.
(2002). Performance of the New Bayer VERSANT HCV RNA 3.0 Assay for Quantitation of Hepatitis C Virus RNA in Plasma and Serum: Conversion to International Units and Comparison with the Roche COBAS Amplicor HCV Monitor, Version 2.0, Assay. J. Clin. Microbiol.
40: 788-793
[Abstract]
[Full Text]
-
Germer, J. J., Heimgartner, P. J., Ilstrup, D. M., Harmsen, W. S., Jenkins, G. D., Patel, R.
(2002). Comparative Evaluation of the VERSANT HCV RNA 3.0, QUANTIPLEX HCV RNA 2.0, and COBAS AMPLICOR HCV MONITOR Version 2.0 Assays for Quantification of Hepatitis C Virus RNA in Serum. J. Clin. Microbiol.
40: 495-500
[Abstract]
[Full Text]
-
Lusida, M. I., Nagano-Fujii, M., Nidom, C. A., Soetjipto, , Handajani, R., Fujita, T., Oka, K., Hotta, H.
(2001). Correlation between Mutations in the Interferon Sensitivity-Determining Region of NS5A Protein and Viral Load of Hepatitis C Virus Subtypes 1b, 1c, and 2a. J. Clin. Microbiol.
39: 3858-3864
[Abstract]
[Full Text]
-
Bendinelli, M., Pistello, M., Maggi, F., Fornai, C., Freer, G., Vatteroni, M. L.
(2001). Molecular Properties, Biology, and Clinical Implications of TT Virus, a Recently Identified Widespread Infectious Agent of Humans. Clin. Microbiol. Rev.
14: 98-113
[Abstract]
[Full Text]
-
Lee, S. C., Antony, A., Lee, N., Leibow, J., Yang, J. Q., Soviero, S., Gutekunst, K., Rosenstraus, M.
(2000). Improved Version 2.0 Qualitative and Quantitative AMPLICOR Reverse Transcription-PCR Tests for Hepatitis C Virus RNA: Calibration to International Units, Enhanced Genotype Reactivity, and Performance Characteristics. J. Clin. Microbiol.
38: 4171-4179
[Abstract]
[Full Text]
-
Tokita, H., Kaufmann, G. R., Matsubayashi, M., Okuda, I., Tanaka, T., Harada, H., Mukaide, M., Suzuki, K., Cooper, D. A.
(2000). Hepatitis C Virus Core Mutations Reduce the Sensitivity of a Fluorescence Enzyme Immunoassay. J. Clin. Microbiol.
38: 3450-3452
[Abstract]
[Full Text]
-
Yu, M.-L., Chuang, W.-L., Dai, C.-Y., Chen, S.-C., Lin, Z.-Y., Hsieh, M.-Y., Wang, L.-Y., Chang, W.-Y.
(2000). Clinical Evaluation of the Automated COBAS AMPLICOR HCV MONITOR Test Version 2.0 for Quantifying Serum Hepatitis C Virus RNA and Comparison to the Quantiplex HCV Version 2.0 Test. J. Clin. Microbiol.
38: 2933-2939
[Abstract]
[Full Text]
-
Kleiber, J., Walter, T., Haberhausen, G., Tsang, S., Babiel, R., Rosenstraus, M.
(2000). Performance Characteristics of a Quantitative, Homogeneous TaqMan RT-PCR Test for HCV RNA. J. Mol. Diagn.
2: 158-166
[Abstract]
[Full Text]
-
Martinot-Peignoux, M., Le Breton, V., Fritsch, S., Le guludec, G., Labouret, N., Keller, F., Marcellin, P.
(2000). Assessment of Viral Loads in Patients with Chronic Hepatitis C with AMPLICOR HCV MONITOR Version 1.0, COBAS HCV MONITOR Version 2.0, and QUANTIPLEX HCV RNA Version 2.0 Assays. J. Clin. Microbiol.
38: 2722-2725
[Abstract]
[Full Text]
-
Cartwright, C. P.
(1999). Synthetic Viral Particles Promise To Be Valuable in the Standardization of Molecular Diagnostic Assays for Hepatitis C Virus. Clin. Chem.
45: 2057-2059
[Full Text]
-
WalkerPeach, C. R., Winkler, M., DuBois, D. B., Pasloske, B. L.
(1999). Ribonuclease-resistant RNA Controls (Armored RNA) for Reverse Transcription-PCR, Branched DNA, and Genotyping Assays for Hepatitis C Virus. Clin. Chem.
45: 2079-2085
[Abstract]
[Full Text]