Previous Article | Next Article 
Journal of Clinical Microbiology, January 2000, p. 18-21, Vol. 38, No. 1
0095-1137/0/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
Automated Specific Capture of Hepatitis C Virus RNA
with Probes and Paramagnetic Particle Separation
Hayato
Miyachi,1,*
Atsuko
Masukawa,2
Toshio
Ohshima,2
Toru
Hirose,3
Chaka
Impraim,4 and
Yasuhiko
Ando1
Department of Clinical
Pathology1 and Central Clinical
Laboratory,2 Bouseidai, Isehara, and
Roche Diagnostics K. K., Tokyo,3
Japan, and Roche Molecular Systems, Inc., Pleasanton,
California 94588-27224
Received 29 March 1999/Returned for modification 21 July
1999/Accepted 22 September 1999
 |
ABSTRACT |
We developed and evaluated a prototype automated specimen
preparation instrument for the specific capture of hepatitis C virus (HCV) RNA with probes and magnetic bead-fluid separation. HCV RNA was
isolated from serum by lysis of virus particles with a chaotropic
agent, followed by hybridization of the RNA with biotinylated probes
and capture of the hybridized RNA with streptavidin-coated paramagnetic
particles. After washing of the hybrid-particle complexes to remove
nonspecifically bound materials, the particles were resuspended in a
specimen diluent and were then ready for amplification and detection
with a fully automated PCR system (COBAS AMPLICOR; Roche Diagnostic
Systems). The analytical sensitivity in the dilution series was 33 copies per ml or greater. Comparison of the test results with those
obtained by a manual method based on organic extraction and
precipitation of RNA (SepaGene RV-R; Sanko Junyaku Co., Ltd.) showed
93% (49 of 53 samples) sensitivity and 100% (12 of 12 samples)
specificity. There was 94% overall agreement between results. When RNA
was extracted by the manual method from serum containing
103 or 105 copies of HCV per ml in the presence
of heparin, there was an inhibitory effect on detection of both HCV RNA
and the internal control. In contrast, when RNA was extracted from the
serum by the automated method, there was no inhibitory effect. This
inhibitory effect of heparin on the manual method was also observed for
a series of serum specimens from a hemodialysis patient, but the inhibitory effect was eliminated by the automated specimen preparation method. In summary, a fully automated RNA extraction system for PCR
detection of HCV RNA by use of specific capture with probes and
magnetic bead-fluid separation was shown to have performance similar to
that of the conventional manual method. In addition, it successfully
eliminated the inhibitory effect of the heparin in the serum and
permitted the detection of HCV RNA in serum samples from a hemodialysis
patient. The prototype automated RNA extraction system is suitable as a
totally automated system, starting with RNA extraction to detection of
HCV, if it was combined with the fully automated COBAS AMPLICOR PCR system.
 |
INTRODUCTION |
Advances in molecular biology and
biotechnology have facilitated analyses for detection of DNA or RNA
sequences (16). New technological advances have led to the
automation of major portions of the assay process (2, 8).
Automated systems have been developed for amplification and detection
of the nucleic acid sequences of infectious agents by PCR. An example
of this is the COBAS AMPLICOR system (Roche Diagnostic Systems,
Branchburg, N.J.), which amplifies target nucleic acid sequences,
captures the biotinylated and amplified products on
oligonucleotide-coated paramagnetic microparticles, and detects the
products with an avidin-horseradish peroxidase conjugate system
(3, 7). The remaining portion of the process to be automated
is extraction of nucleic acid from clinical specimens.
Extraction of RNA from serum specimens for laboratory use has been
widely performed with commercially available kits. However, they are
not very user-friendly when one is handling numerous samples at once,
and the extraction efficiency varies among samples. Moreover, those
kits cannot eliminate some inhibitors of enzymatic amplification that
may be present in clinical specimens and that may cause false-negative
PCR results (1, 14). Heparin is a potent inhibitor that may
be present in blood specimens. It has been used as an anticoagulant for
blood collection and hemodialysis and for the treatment of disseminated
intravascular coagulation. Because the presence of heparin in a sample
makes the interpretation of results difficult, sera from patients
undergoing hemodialysis may not be appropriate for testing by PCR
(15). Since a high prevalence of hepatitis C virus (HCV)
infection has been reported in patients undergoing hemodialysis
(6, 13), accurate assays for the detection of HCV RNA in
those patients need to be developed. In the study described here, we
developed and evaluated a prototype automated specimen preparation
instrument for the specific capture of HCV RNA with probes and magnetic
bead-fluid (B-F) separation. In particular, we addressed the questions
of whether the extraction system could eliminate the inhibitory effects
of heparin and whether it was suitable for automated RNA extraction
followed by detection of HCV with the fully automated COBAS AMPLICOR
PCR system.
 |
MATERIALS AND METHODS |
Clinical specimens.
The serum specimens used in this study
were obtained from 65 patients referred to Tokai University Hospital
for chronic liver diseases. To determine the limit of detection of the
assay, serial dilutions of serum from a patient containing 530 copies
of HCV per ml were prepared in HCV-negative serum. To assess the effect of heparin, a series of serum samples was obtained from a hemodialysis patient who was anticoagulated with 3,400 U of heparin. When needed, HCV RNA was quantitatively measured by the AMPLICOR HCV MONITOR test
(Roche Diagnostic Systems) (5). All samples were separated from clots within 4 h of collection, divided into aliquots, and stored at
80°C until the RNA was extracted.
RNA extraction.
HCV RNA was isolated from serum with a
prototype instrument for an automated system consisting of the reagents
developed by Roche Molecular Systems (Pleasanton, Calif.) and a robotic
processor developed by Precision System Science Co. Ltd. (Tokyo, Japan) (Fig. 1). Briefly, HCV RNA was isolated
from 300 µl of serum by lysis of virus particles with 500 µl of
guanidinium thiocyanate solution at 60°C for 20 min. The RNA was
hybridized with biotinylated probes (KY78) that were specific to the 5'
untranslated region of the HCV genome (11) and that were
identical to the downstream primer for amplification. The hybridized
RNA was then captured with streptavidin-coated paramagnetic particles.
The internal control (40 copies) was introduced into the specimen
during the lysis reaction. After washing of the hybrid-particle
complexes to remove nonspecifically bound materials, the particles were resuspended in 50 µl of a specimen diluent and were then ready for
amplification and detection by the COBAS AMPLICOR HCV test.

View larger version (23K):
[in this window]
[in a new window]
|
FIG. 1.
Assay format. (a) Virus is lysed to release nucleic
acid. (b) Nucleic acid is bound to biotinylated (B) probe. (c)
Biotinylated (B) probe is captured onto streptavidin (SA)-coated
paramagnetic particles. (d) Paramagnetic particles are separated and
washed on a magnet. 5' UTR, 5' untranslated region.
|
|
RNA was also isolated from 300 µl of serum by a manual method based
on guanidinium thiocyanate lysis and isopropanol precipitation (SepaGene RV-R; Sanko Junyaku Co., Ltd., Tokyo, Japan). Briefly, 300 µl of guanidinium thiocyanate solution was added to the specimen, and
the phases were then separated by adding 300 µl of sodium acetate and
600 µl of chloroform-agglutination solution, followed by
centrifugation at 12,000 × g for 15 min. The RNA was
precipitated from the upper phase with isopropanol and was resuspended
in 50 µl of a specimen diluent. The internal control (20 copies) was added to the specimen diluent prior to amplification.
COBAS AMPLICOR HCV test.
The COBAS AMPLICOR HCV test has
been described previously (3, 7). The optical density of
each reaction mixture at 660 nm was measured. Optical density readings
of >0.200 for HCV RNA were considered positive, and those of 0.100 to
0.200 were considered equivocal. Optical density readings of >0.200
for the internal control were considered positive.
All samples were extracted once, and a single amplification and
detection were performed with each extract unless stated otherwise. One
positive control and two negative controls provided with the kit were
run with each batch of patient specimens.
Interference by heparin.
To assess the effect of heparin, 3 to 13 U of sodium heparin per ml was added to a serum sample containing
103 or 105 copies of HCV per ml. The RNA was
extracted from the serum by the SepaGene RV-R method and the specific
capture with probes and magnetic B-F separation. To assess the effect
of heparin treatment in vivo, a series of serum specimens containing
5.1 × 105 copies of HCV per ml, as quantified by the
AMPLICOR HCV MONITOR test, was serially obtained from a patient before
and at 1, 3, and 5 h after hemodialysis in which 3,400 U of
heparin was used as an anticoagulant. RNA was extracted from the serum,
and then HCV RNA was amplified by the COBAS AMPLICOR HCV test. To
demonstrate the presence of inhibitors, the series of serum samples
obtained after hemodialysis from the hemodialysis patient was diluted
1:10 in HCV-negative serum, and RNA was then extracted from the serum by the SepaGene RV-R method.
 |
RESULTS |
Assay performance.
The analytical sensitivity based on testing
of the serial dilutions was 33 copies of HCV per ml or greater (Table
1). Comparison of test results with those
obtained by the manual method showed 93% (49 of 53 samples)
sensitivity and 100% (12 of 12 samples) specificity. There was 94%
overall agreement between the results obtained by the automated method
and those obtained by the manual method (Table
2).
Elimination of inhibitory effect of heparin.
Heparin was added
at up to 13 U/ml to serum containing 103 or 105
copies of HCV per ml, and then RNA was extracted. When RNA was extracted from both of these serum samples by the manual method, the
PCR results were negative (Table 3).
Analysis of the results for the internal control indicated that there
was PCR inhibition for the PCR-negative samples. In contrast, when RNA
was extracted by the automated method, there was a lack of inhibition.
Elimination of inhibitory effects on PCR with serum from a
hemodialysis patient.
When RNA was extracted by the manual method
from serum obtained at various times after hemodialysis, the HCV PCR
assay was negative at least until 3 h after hemodialysis (Fig.
2A). Analysis of the results for the
internal control indicated that there was PCR inhibition for these
PCR-negative samples. Dilution of the sample 1:10 and a repeat assay
produced positive PCR results for both HCV RNA and the internal control
(Fig. 2B). In contrast, with extraction by the automated method, the
inhibitory effect on detection of both HCV RNA and the internal control
was abolished for the serum sample obtained immediately after
hemodialysis (Fig. 2C).

View larger version (11K):
[in this window]
[in a new window]
|
FIG. 2.
Elimination of inhibitory effects on PCR of serum from a
hemodialysis patient by the automated specific capture with probes and
magnetic B-F separation. HCV RNA-positive serum was obtained from a
hemodialysis patient, who was anticoagulated with 3,400 U of heparin,
at 0, 1, 3, and 5 h following hemodialysis. Serum samples were
diluted 1:10 in HCV RNA-negative serum. RNA was extracted from
undiluted (A) and diluted (B) aliquots by the manual method (the
SepaGene RV-R method) and then amplified for HCV RNA and the internal
control with the COBAS AMPLICOR system. HCV RNA was extracted from
undiluted aliquots by the automated extraction method based on probe
capture and magnetic B-F separation, and then amplified for HCV RNA
with the COBAS AMPLICOR system (C). Data are means of duplicate assays.
, HCV RNA; , internal control.
|
|
 |
DISCUSSION |
In the study described in this report, we developed and evaluated
a prototype automated specimen preparation instrument for the specific
capture of HCV RNA with probes and magnetic B-F separation. Extracted
RNA was successfully used in an automated PCR assay for the detection
of HCV RNA with the COBAS AMPLICOR system. The analytical sensitivity
of the automated extraction procedure was similar to that of the
SepaGene RV-R method. There was 94% overall agreement between the
results obtained by the automated method and those obtained by the
manual method. The automated RNA extraction system would be suitable as
a totally automated system starting with RNA extraction to detection of
HCV if it was combined with a fully automated PCR system.
An automated system for PCR assays provides improvements not only in
labor efficiency but also in the accuracy of results. Major problems
with PCR assays are false-positive results because of carryover
contamination of previously amplified products and false-negative
results because of the amplification inhibitors present in clinical
specimens (4, 12). In the COBAS AMPLICOR system, the use of
dUTP and uracil-N-glycosylase reduces the risk of carryover
contamination (9). With regard to total quality control,
false-negative PCR results become more problematic. In particular, the
inhibitory effect of heparin on PCR has been problematic with samples
from hemodialysis patients (1). When RNA was extracted from
sera by the SepaGene RV-R method, there was an inhibitory effect on the
detection of both HCV RNA and the internal control at least until
3 h after hemodialysis. The inhibitory effect of heparin was
successfully eliminated when RNA was extracted by the automated probe
capture and magnetic B-F separation method. This extraction system
would be applicable to PCR assays of serum specimens from hemodialysis patients.
One of the major advantages of nucleic acid amplification over
conventional methods for the diagnosis of an infectious disease is
sensitive detection of agents directly from clinical specimens. However, clinical specimens may contain a variety of amplification inhibitors such as heparin, hemoglobin, heme, urea, and so on (4,
12). The nature of many inhibitors is still unknown. Therefore,
it is desirable to use an extraction method that can eliminate
inhibitors as much as possible and monitor the efficacy of extraction.
The extraction method used in the present study is theoretically
suitable for eliminating any inhibitors in the serum, because it is
based on specific capture with probes and magnetic B-F separation. On
the other hand, conventional manual methods such as the SepaGene RV-R
method are not able to eliminate heparin, which is thought to bind to
nucleic acids (1). In the present study, the internal
control was demonstrated to be useful for monitoring for the presence
of inhibitors in a patient under hemodialysis as well as in an in vitro
interference study. It could be used to determine which samples yield
invalid results and therefore need to be retested (10).
Since many kinds of clinical specimens, from body fluids to tissues,
are subjected to tests for direct detection of infectious agents,
development of automated systems is desirable for nucleic acid
extraction procedures that can be applied to a variety of clinical
specimens, in addition to serum specimens.
In summary, a fully automated RNA extraction system was developed. The
system was based on the specific capture with probes and magnetic B-F
separation, with the performance of the assay being comparable to that
of the conventional manual method. In addition to that, it successfully
eliminated inhibitors, such as heparin, from serum and can be used for
the assay of serum from patients undergoing hemodialysis. The automated
RNA extraction system is suitable for use as a totally automated
system, starting with RNA extraction to detection of HCV, when combined
with a fully automated PCR system.
 |
ACKNOWLEDGMENT |
We thank Takatoshi Kakuta of the Department of Internal Medicine,
Tokai University School of Medicine, for providing patient materials.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: Department of
Clinical Pathology, Bouseidai, Isehara, Kanagawa 259-1193 Japan. Phone: 81-463 (93) 1121. Fax: 81-463(93)8607. E-mail:
miyachi{at}is.icc.u-tokai.ac.jp.
 |
REFERENCES |
| 1.
|
Beutler, E.,
T. Gelbart, and W. Kuhl.
1990.
Interference of heparin with the polymerase chain reaction.
BioTechniques
9:166[Medline].
|
| 2.
|
Diamandis, E. P.
1996.
Automation of molecular diagnostics.
Clin. Chem.
42:7-8[Free Full Text].
|
| 3.
|
DiDomenico, N.,
H. Link,
R. Knobel,
T. Caratsch,
W. Weschler,
Z. G. Loewy, and M. Rosenstraus.
1996.
COBAS AMPLICORTM: fully automated RNA and DNA amplification and detection system for routine diagnostic PCR.
Clin. Chem.
42:1915-1923[Abstract/Free Full Text].
|
| 4.
|
Greenfield, L., and T. J. White.
1993.
Sample preparation methods, p. 122-137.
In
D. H. Persing, T. F. Smith, F. C. Tenover, and T. J. White (ed.), Diagnostic molecular microbiology: principles and applications. American Society for Microbiology, Washington, D.C.
|
| 5.
|
Ichijo, T.,
A. Matsumoto,
M. Kobayashi,
K. Furihata, and E. Tanaka.
1997.
Quantitative measurement of HCV RNA in the serum: a comparison of three assays based on different principles.
J. Gastroenterol. Hepatol.
12:500-506[Medline].
|
| 6.
|
Jeffers, L. J.,
G. O. Perez,
M. D. De Medina,
C. J. Ortiz-Interian,
E. R. Schiff,
K. R. Reddy,
M. Jimenez,
J. J. Bourgoignie,
C. A. Vaamonde,
R. Duncan,
M. Houghton,
G.-L. Choo, and G. Kuo.
1990.
Hepatitis C infection in two urban hemodialysis units.
Kidney Int.
38:320-322[Medline].
|
| 7.
|
Jungkind, D.,
S. DiRenzo,
K. G. Beavis, and N. S. Silverman.
1996.
Evaluation of automated COBAS AMPLICOR PCR system for detection of several infectious agents and its impact on laboratory management.
J. Clin. Microbiol.
34:2778-2783[Abstract].
|
| 8.
|
Landegren, U.,
R. Kaizer,
C. T. Caskey, and L. Hood.
1988.
DNA diagnostics. Molecular techniques and automation.
Science
242:229-237[Abstract/Free Full Text].
|
| 9.
|
Longo, M. C.,
M. S. Berninger, and J. L. Hartley.
1990.
Use of uracil DNA glycosylase to control carry-over contamination in polymerase chain reactions.
Gene
93:125-128[CrossRef][Medline].
|
| 10.
|
Miyachi, H.,
A. Masukawa,
T. Ohshima,
H. Fusegawa,
T. Hirose,
C. Impraim, and Y. Ando.
1998.
Monitoring of inhibitors of enzymatic amplification in polymerase chain reaction and evaluation of efficacy of RNA extraction for the detection of hepatitis C virus using the internal control.
Clin. Chem. Lab. Med.
36:571-575[CrossRef][Medline].
|
| 11.
|
Nolte, F. S.,
C. Thurmond, and M. W. Fried.
1995.
Preclinical evaluation of AMPLICOR hepatitis C virus test for detection of hepatitis C virus RNA.
J. Clin. Microbiol.
33:1775-1778[Abstract].
|
| 12.
|
Rolfs, A.,
I. Schuller,
U. Finckh, and I. Weber-Rolfs.
1992.
PCR: clinical diagnostics and research, p. 51-67.
Springer-Verlag, Berlin, Germany.
|
| 13.
|
Silini, E.,
F. Bono,
A. Cerino,
V. Piazza,
E. Solcia, and M. U. Mondelli.
1993.
Virological features of hepatitis C virus infection in hemodialysis patients.
J. Clin. Microbiol.
31:2913-2917[Abstract/Free Full Text].
|
| 14.
|
Tasa, G.,
M. Uuskula,
A. Hirvonen, and A.-V. Mikelsaar.
1995.
Optimization of PCR to yield successful amplification from heparin-contaminated DNA.
Methods Mol. Cell. Biol.
5:122-124.
|
| 15.
|
Willems, M.,
H. Moshage,
F. Nevens,
J. Fevery, and S. H. Yap.
1993.
Plasma collected from heparinized blood is not suitable for HCV-RNA detection by conventional RT-PCR assay.
J. Virol. Methods
42:127-130[CrossRef][Medline].
|
| 16.
|
Young, F. E.
1987.
DNA probes: fruits of the new biotechnology.
JAMA
229:2404-2406.
|
Journal of Clinical Microbiology, January 2000, p. 18-21, Vol. 38, No. 1
0095-1137/0/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
This article has been cited by other articles:
-
Miyachi, H., Masukawa, A., Asai, S., Miura, T., Tamatsukuri, S., Hirose, T., Ando, Y.
(2003). Quantitative Assay of Hepatitis C Virus RNA Using an Automated Extraction System for Specific Capture with Probes and Paramagnetic Particle Separation. J. Clin. Microbiol.
41: 572-575
[Abstract]
[Full Text]
-
Meng, Q., Wong, C., Rangachari, A., Tamatsukuri, S., Sasaki, M., Fiss, E., Cheng, L., Ramankutty, T., Clarke, D., Yawata, H., Sakakura, Y., Hirose, T., Impraim, C.
(2001). Automated Multiplex Assay System for Simultaneous Detection of Hepatitis B Virus DNA, Hepatitis C Virus RNA, and Human Immunodeficiency Virus Type 1 RNA. J. Clin. Microbiol.
39: 2937-2945
[Abstract]
[Full Text]