Previous Article | Next Article 
Journal of Clinical Microbiology, October 2000, p. 3577-3580, Vol. 38, No. 10
0095-1137/00/$04.00+0
Evaluation of the MicroScan Rapid Neg ID3 Panel for
Identification of Enterobacteriaceae and Some Common
Gram-Negative Nonfermenters
Caroline Mohr
O'Hara* and
J. Michael
Miller
Centers for Disease Control and Prevention,
Atlanta, Georgia
Received 25 February 2000/Returned for modification 18 May
2000/Accepted 4 August 2000
 |
ABSTRACT |
The MicroScan Rapid Neg ID3 panel (Dade Behring, Inc., West
Sacramento, Calif.) is designed for the identification of gram-negative bacilli. We evaluated its ability to accurately identify
Enterobacteriaceae that are routinely encountered in a
clinical laboratory and glucose nonfermenting gram-negative bacilli.
Using 511 stock cultures that were maintained at
70°C and passaged
three times before use, we inoculated panels according to the
manufacturer's instructions and processed them in a Walk/Away
instrument using version 22.01 software. The time to identification was
2 h and 30 min. All panel identifications were compared to
reference identifications previously determined by conventional tube
biochemicals. At the end of the initial 2.5-h incubation period, 405 (79.3%) identifications were correct. An additional 49 (9.6%)
isolates were correctly identified after required additional off-line
biochemical tests were performed. Thus, at 24 h, 88.8% of the 511 strains tested were correctly identified. Twenty-two (4.3%) were
identified to the genus level only. Twenty-six (5.1%) strains were
misidentified. Because the system is based on fluorogenics, there are
no conventional tests readily available with which to compare possibly
incorrect reactions. Of the 28 Salmonella strains that were
tested, 5 were incorrectly reported. The 21 remaining errors were
scattered among the genera tested. Testing on nine strains gave a
result of "no identification" (very rare biotype). The Rapid Neg
ID3 panel in this study approached 89% accuracy for the identification
of gram-negative organisms encountered in the hospital laboratory.
 |
INTRODUCTION |
Automated identification and
susceptibility testing methods have become the mainstay of clinical
microbiology laboratories and have provided laboratory staff with an
efficient means of organism classification, thereby giving physicians
the availability of rapid and generally accurate results that is
necessary to guide therapy. Errors do occur, however, and as competing
resources put more demands on shrinking laboratory staffs, expertise in recognizing errors and evaluating questionable results from these instruments is diminishing. This places a greater demand on the manufacturer to provide an instrument and a database that are as free
from error as possible. However, the laboratory is no less accountable
for undetected errors that may or may not result in an adverse outcome
for patients. As new versions of databases become available, it is
critical that appropriate evaluations be conducted in order to provide
laboratory staff with unbiased data from which to make purchasing
decisions. The update that resulted in the MicroScan Rapid Neg ID3
(RNID3) panel was designed to improve accuracy in the identification of
common human clinical taxa by modifying the biochemical tests in the
panel by replacing 10 tests on the RNID2 panel with new substrates. The
new panel also eliminated the need for a mineral oil overlay on the
decarboxylase tests and increased the shelf life from 6 months to 1 year when stored at 2 to 8°C. In this study, the MicroScan RNID3
panel was evaluated using members of the Enterobacteriaceae
and some common gram-negative nonfermenters. (Part of this research was
presented at the 99th General Meeting of the American Society for
Microbiology, Chicago, Ill., 30 May to 3 June 1999.)
 |
MATERIALS AND METHODS |
Bacterial strains.
Table 1 lists the 475 biochemically
typical and atypical Enterobacteriaceae and Table 2 lists
the 36 commonly isolated biochemically typical glucose nonfermenting
gram-negative bacilli that were used in this study. All strains were of
human origin. This set also contained larger numbers of genera and
species not likely to be found in routine work but which were contained
in the database, thereby presenting a true challenge to the system.
Stock cultures from the collection of the Centers for Disease Control
and Prevention were maintained in defibrinated sheep blood at
70°C
and were passed two times on trypticase soy agar with 5% sheep blood
(TSA II; BD, Sparks, Md.) before inoculation onto MacConkey's agar for
testing in the panels.
Media and biochemical tests.
Biochemical tests for
identification were performed using conventional media and the methods
described by Edwards and Ewing (2), with some modifications
by Hickman and Farmer et al. (3, 4). Commercial media were
used wherever possible.
Identification panel.
The RNID3 panel is manufactured by
Dade Behring, Inc., MicroScan Inc., West Sacramento, Calif., and is
designed for use with the Walk/Away automated instrument. The 36 substrates contained on the panel include arabitol, adonitol, lysine,
and ornithine with the decarboxylase base, glucuronic acid, arabinose,
urea, inositol, mannitol, mannose, melibiose, raffinose, salicin,
sorbitol, sucrose, glucose, and tryptophan for indole production;
4-methylumbelliferyl (MeU)-
-D-xyloside,
MeU-
-D-galactopyranoside,
MeU-
-D-glucopyranoside, MeU-N-acetyl-
-D-glucosaminide,
MeU-
-D-glucopyranoside,
MeU-
-D-mannopyranoside, MeU-
-D-glucuronide,
MeU-
-D-N,N'-diacetylchitobioside,
MeU-
-D-galactopyranoside, MeU-phosphate (both acid and
alkaline pH); L-arginine-7-amido-4-methylcoumarin (AMC),
-L-glutamate-AMC,
N-glutaryl-glycyl-arginine-AMC,
-L-glutamic acid-AMC, L-pyroglutamic acid-AMC,
L-proline-AMC, and L-tyrosine-AMC. These
substrates work by one of the following mechanisms: hydrolysis of
fluorogenic substrates, pH changes following substrate utilization, production of specific metabolic byproducts, or the rate of production of specific metabolic byproducts after 2.5 h of incubation in the
Walk/Away. While this newer panel does require the automated addition
of a rapid indole reagent to the tryptophan well, it no longer requires
mineral oil overlays on the decarboxylases.
The data management software used in this study was version 22.01. The
current version of software is 22.26 (May 1999), but this version does
not include changes in the identification database that would
invalidate the results of this evaluation.
Definitions.
The category of "correct" implies that the
organism identification was correct to the genus and species levels at
the end of the 2.5-h incubation time. The category of "correct with
additional tests" implies that the organism identification was
correct to the genus and species levels after the performance of
additional off-line biochemical tests that are suggested by the
manufacturer. These tests are suggested when the probability of
accuracy of identifications of the first organisms listed is below 85%
(low probability). There are 48 possible additional tests that would extend the time of final identification to 24 h. The category of
"correct to genus" indicates that the genus identification was
correct but the species was incorrect when the database indicated that
the organism could be identified to the species level. The category of
"error" implies that both the genus and species identifications are
incorrect, and "no ID" indicates that no identification was given
by the system or that the system gave a choice of multiple answers that
were not resolved by additional testing.
Tests were repeated in duplicate when a response other than
"correct" or "correct with additional tests" occurred.
Statistical analysis.
Results were evaluated by chi-square
analysis with Yates' corrected coefficient to arrive at a P value.
 |
RESULTS AND DISCUSSION |
Table 1 shows the results of testing
475 strains of Enterobacteriaceae in the RNID3. Table
2 shows the accuracy of identification of
five glucose-fermenting and nonfermenting organisms that are commonly
isolated in clinical laboratories. At 24 h, after additional testing, the RNID3 accurately identified 88.8% of 511 strains evaluated in this study.
View this table:
[in this window]
[in a new window]
|
TABLE 2.
Accuracy of identification of commonly occurring
glucose-fermenting and glucose-nonfermenting organisms
using RNID3a
|
|
Table 3 lists the errors that occurred in
identification. Twenty-six (5.1%) strains were misidentified. Of these
26 errors, 21 were at probability levels of 88% or greater; 14 were at
probability levels of 95% or higher. Of the five Salmonella
misidentifications, one was at a probability level of 76%, two were at
a probability level of 90%, one was at a probability level of 95%,
and one was at a probability level of 99%, suggesting that an
inaccurate result might be submitted to the physician. Both of the
Shigella misidentifications were at probability levels of
98% or greater. Because the system is based on fluorogenics, there are
no conventional tests readily available with which to compare possibly
incorrect reactions. Testing on nine strains yielded a result of "no
identification" (very rare biotype).
Table 4 shows the results of testing 130 strains that more closely resemble the assortment of strains that might
be routinely found in local area hospital laboratories. After
additional testing was completed, 91.5% of the strains were correctly
identified.
Bascomb et al. reported 97.1% accuracy with 405 fresh clinical
isolates and 90.7% accuracy with 247 stock isolates (1). Schreckenberger et al. reported an accuracy of 96.2% after additional testing was completed and an error rate of 3.8% using 317 fresh clinical isolates (P. C. Schreckenberger, S. Connell, J. Skinner, B. L. Zimmer, D. Glenn, D. A. Bruckner, J. M. Janda, and
S. L. Abbott, 98th Gen. Meet. Am. Soc. Microbiol. 1998, poster
C-153, 1998). One might expect that the accuracy of the system when
testing fresh clinical isolates might be higher than that observed with stock strains.
In an effort to determine the cause of the errors in identification,
the strains for which incorrect answers were given were retested in
duplicate. One panel was processed without oil overlays on the
decarboxylases; a matching panel had the decarboxylases overlaid with
mineral, as was the procedure with the RNID2 panels. One strain each of
Escherichia coli and Escherichia hermannii gave
correct identifications on the panels that were overlaid with oil,
while the answers from the panels with no oil overlays remained incorrect.
In this study, the RNID3 panel performed as well for the identification
of bacteria routinely isolated in hospital laboratories as it does for
the identification of a challenge set of strains (P > 0.05). In addition, the accuracy of the new panel in this study
was not significantly different (P > 0.05 [Yates'
correction]) from the RNID2 results previously published
(5), for which the same study set was used.
An in-depth study of the accuracy of this panel to correctly identify
less commonly encountered glucose nonfermenting gram-negative organisms
will be the subject of a future publication.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: Centers for
Disease Control and Prevention, Mailstop C16, Atlanta, GA 30333. Phone: (404) 639-2316. Fax: (404) 639-3241. E-mail: cmo1{at}cdc.gov.
 |
REFERENCES |
| 1.
|
Bascomb, S.,
S. L. Abbott,
J. D. Bobolis,
D. A. Bruckner,
S. J. Connell,
S. K. Cullen,
M. Daugherty,
D. Glenn,
J. M. Janda,
S. J. Lentsch,
G. B. Williams,
J. Wong, and B. L. Zimmer.
1997.
Multicenter evaluation of the MicroScan rapid gram-negative identification type 3 panel.
J. Clin. Microbiol.
35:2531-2536[Abstract].
|
| 2.
|
Edwards, P. R., and W. H. Ewing.
1972.
Identification of Enterobacteriaceae, 3rd ed.
Burgess Publishing Co., Minneapolis, Minn.
|
| 3.
|
Farmer, J. J., III,
M. A. Asbury,
F. W. Hickman,
D. J. Brenner, and the Enterobacteriaceae Study Group.
1980.
Enterobacter sakazakii: a new species of "Enterobacteriaceae" isolated from clinical specimens.
Int. J. Syst. Bacteriol.
30:569-584[CrossRef].
|
| 4.
|
Hickman, F. W., and J. J. Farmer, III.
1978.
Salmonella typhi: identification, antibiograms, serology, and bacteriophage typing.
Am. J. Med. Technol.
44:1149-1150[Medline].
|
| 5.
|
O'Hara, C. M., and J. M. Miller.
1992.
Evaluation of the autoSCAN-W/A system for rapid (2-hour) identification of members of the family Enterobacteriaceae.
J. Clin. Microbiol.
30:1541-1543[Abstract/Free Full Text].
|
Journal of Clinical Microbiology, October 2000, p. 3577-3580, Vol. 38, No. 10
0095-1137/00/$04.00+0
This article has been cited by other articles:
-
Janda, J. M., Abbott, S. L.
(2006). The Genus Hafnia: from Soup to Nuts. Clin. Microbiol. Rev.
19: 12-28
[Abstract]
[Full Text]
-
O'Hara, C. M.
(2005). Manual and Automated Instrumentation for Identification of Enterobacteriaceae and Other Aerobic Gram-Negative Bacilli. Clin. Microbiol. Rev.
18: 147-162
[Abstract]
[Full Text]
-
O'Hara, C. M., Sowers, E. G., Bopp, C. A., Duda, S. B., Strockbine, N. A.
(2003). Accuracy of Six Commercially Available Systems for Identification of Members of the Family Vibrionaceae. J. Clin. Microbiol.
41: 5654-5659
[Abstract]
[Full Text]
-
O'Hara, C. M., Miller, J. M.
(2002). Ability of the MicroScan Rapid Gram-Negative ID Type 3 Panel To Identify Nonenteric Glucose-Fermenting and Nonfermenting Gram-Negative Bacilli. J. Clin. Microbiol.
40: 3750-3752
[Abstract]
[Full Text]