Previous Article | Next Article 
Journal of Clinical Microbiology, November 2000, p. 4015-4020, Vol. 38, No. 11
0095-1137/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
Quality of Human Immunodeficiency Virus Viral
Load Testing in Australia
Susan J.
Best,*
Anthony P.
Gust,
Elizabeth I. M.
Johnson,
Catherine H.
McGavin, and
Elizabeth
M.
Dax
National Serology Reference Laboratory,
Australia, at St Vincent's Institute of Medical Research, Fitzroy,
Victoria 3065, Australia
Received 2 May 2000/Returned for modification 29 June 2000/Accepted 30 August 2000
 |
ABSTRACT |
This study determined the proficiencies of laboratories measuring
human immunodeficiency virus type 1 (HIV-1) viral loads and the
accuracies of two assays used for HIV-1 viral load measurement in
Australia and investigated the variability of the new versions of these
assays. Quality assessment program panels containing (i) dilutions of
HIV-1 subtype B, (ii) replicates of identical samples of HIV-1 subtype
B, and (iii) samples of subtype E and B were tested by laboratories.
Total variability (within and between laboratories) was tested with
quality control samples. The coefficients of variation (CVs) for the
Roche AMPLICOR HIV-1 MONITOR version (v) 1.0 and Chiron Quantiplex bDNA
2.0 assays ranged from 53 to 87% and 22 to 31%, respectively. The
widespread occurrence of invalid runs with the AMPLICOR HIV-1 MONITOR
1.0 assay was identified. The CVs of the new versions of the assays
were 82 to 86% for the AMPLICOR HIV-1 MONITOR v 1.5 assay and 16 to
23% for the Quantiplex bDNA 3.0 assay. For virus dilution samples, all
but 5 of 19 laboratories obtained results within 2 standard deviations
of the mean. The Quantiplex bDNA 2.0 assay reported values lower than
those reported by the AMPLICOR HIV-1 MONITOR version 1.0 assay for
samples containing HIV-1 subtype B, whereas the reverse was true for
subtype E. Identification and resolution of the problem of invalid runs
markedly improved the quality of HIV-1 viral load testing. The
variability observed between laboratories and between assays, even the
most recent versions, dictates that monitoring of viral load in an
individual should always be by the same laboratory and by the same
assay. Results for an individual which differ by less than 0.5 log10 HIV-1 RNA copy number/ml should not be considered
clinically significant.
 |
INTRODUCTION |
The quantification of human
immunodeficiency virus (HIV) to establish the viral load or amount of
measurable virus in the blood has become essential to the clinical
management of HIV infection. It has been demonstrated that the viral
load is the most reliable predictor of the course of infection if it is
estimated once a steady state in viremia is achieved following
seroconversion (8). Viral load increases with the
development of resistance to antiretroviral therapy or with a
detrimental alteration in clinical status (5). It has been
advocated that the aim of current antiviral therapy for HIV infection
should be to maintain the viral load below the level of detection of
the available assays (3, 7).
The majority of commercial assays for the quantification of viral load
use techniques involving amplification either of target nucleic acid
(AMPLICOR HIV-1 MONITOR [Roche Molecular Systems, Pleasanton,
Calif.]) or a hybridization signal (Quantiplex HIV RNA [Chiron
Corporation, Emeryville, Calif., now Bayer Diagnostics], abbreviated
in the text as Quantiplex bDNA). The majority of the data in this
report were generated with version (v) 1.0 of the AMPLICOR HIV-1
MONITOR assay and the Quantiplex bDNA 2.0 assay. The report presents
limited data from tests with the latest versions of these assays (v 1.5 and 3.0, respectively) over the period encompassed by the present
study. The NucliSens HIV-1 QT assay (Organon Teknika, Boxtel, The
Netherlands), which uses amplification of nucleic acid sequences, is
not widely used for therapeutic monitoring in Australia.
It is a condition of registration of all HIV assays for use in
Australia that laboratories that use the assay participate in the
quality assurance (QA) programs run by the National Serology Reference
Laboratory (NRL), Melbourne, Australia. This report presents the
findings of the HIV type 1 (HIV-1) viral load QA program in Australia
for the AMPLICOR HIV-1 MONITOR and Quantiplex bDNA assays. We
investigated the proficiencies of laboratories measuring viral load
across a range of dilutions (range of viral RNA copy number per
milliliter), the reproducibility of determinations for replicate
samples, and whether the performances of the assays were consistent
between HIV-1 subtypes. The report also addresses the implication of
reporting of data from invalid runs.
 |
MATERIALS AND METHODS |
Nineteen laboratories (testing sites) in Australia and one in
New Zealand participate in the NRL's QA program for HIV viral load
testing. The program was instituted in August 1996 and consists of two
elements: a quality control (QC) program and a quality assessment
program (QAP). All laboratories return QAP and QC program data to the NRL.
QC program.
A QC program provides a mechanism to monitor the
run-to-run performance of an assay. For the HIV viral load QC program,
NRL prepared and distributed the QC samples. The between-run,
between-laboratory, and between-lot-number performances of the assays
were monitored for laboratories returning data for QC samples to NRL.
QC samples were prepared in a class II biohazard cabinet by dilution of
a stock HIV-1 subtype B (HIV-1 Ba-L) culture supernatant in pooled,
defibrinated and delipidated normal human plasma (Basematrix; Boston
Biomedica, Inc., West Bridgewater, Mass.). The dilutions were mixed in
a sterile bottle, on a slow vortex, overnight at 4°C. Convenient
single-use volumes (250 µl for the AMPLICOR HIV-1 MONITOR assays,
1,250 µl for the Quantiplex bDNA assay) were aliquoted into sterile
screw-cap microcentrifuge tubes (Sarstedt, Nümbrecht, Germany),
and the tubes and their contents were frozen at
70°C prior to
dispatch to the laboratories on dry ice. Laboratories were required to
include the QC samples in at least every fourth assay run and with a
change of lot number or a change of operator. QC sample data were
returned to NRL on a monthly basis and included not only the viral
loads measured in the QC samples but also the numbers of invalid runs,
the reasons for invalid runs, as well as the numbers of invalid patient results.
Since the institution of the HIV viral load QC program, three sets of
coded QC samples have been circulated by the NRL. In
the most recent
set, samples QC105 and QC105C were designed to
have loads that fall
into the lower range (400 to 5,000 copies/ml),
with QC105 to be used in
the AMPLICOR HIV-1 MONITOR assay and
QC105C to be used in the
Quantiplex bDNA assay. Preparation of
separate QC samples to monitor
the lower ranges of the assays
was necessary because the Quantiplex
bDNA assay always returned
lower values for the subtype B culture
supernatant used for preparation
of the QC samples than the AMPLICOR
HIV-1 MONITOR assay did. Two
samples with loads in the range of
10
3 copies/ml were prepared to ensure that the lower target
value
was achievable in both assays. The detection ranges stated by
the
manufacturers were 400 to 750,000 copies/ml for the AMPLICOR
HIV-1
MONITOR v 1.0 assay and 500 to 500,000 copies/ml for the
Quantiplex
bDNA 2.0 assay. In the present report, only the results
for the first
set (QC101 and QC102; August 1996 to April 1997)
and most recent set
(QC105, QC105C, and QC106; February 1998 to
June 1999) of QC samples
are presented. The results obtained with
the other QC set were similar
to those obtained with the QC101
and QC102 set and have been omitted to
avoid unnecessary repetition
of results. Preliminary results obtained
with a separate set of
QC samples (108P and 109P) with low viral loads
designed for use
with the AMPLICOR HIV-1 MONITOR v 1.5 assay with
Ultrasensitive
sample preparation are also reported. These samples were
prepared
in normal human plasma because samples prepared in Basematrix
proved unsuitable for ultracentrifugation. We have previously
found
that centrifugation in Basematrix led to inconsistent pelleting
of the
RNA and frequent loss of the pellet when the supernatant
was
removed.
Run validity.
Run validity was determined according to the
manufacturer criteria. Two positive controls and one negative control
are provided by the manufacturer for the AMPLICOR HIV-1 MONITOR assay,
with a designated range into which the results for those controls must fall for an assay run to be valid. The Quantiplex bDNA assays include a
standard curve with four standards, each of which is tested in
duplicate in every run. The software accompanying the detection
luminometer determines whether the linearities and variabilities of the
values obtained for the standards and controls are acceptable for a
valid run.
QAP.
HIV viral load QAP panels were distributed twice yearly
to the 20 participating laboratories and consisted of up to eight samples, coded to disguise their identities. Laboratories were provided
with a results sheet on which to record run information and results.
Data were returned to the NRL for collation and analysis. To maintain
confidentiality, each laboratory was assigned a unique code number
under which results were reported (note that some laboratories may
appear with more than one laboratory code in this study).
The first QAP panel, panel VLQA 97-1, contained five coded samples,
four of which were members of a fivefold dilution series
of HIV-1 Ba-L
in Basematrix (subtype B), including one dilution
presented in
duplicate. The fifth sample contained diluent
only.
The second QAP panel, panel VLQA 98-1, investigated whether sites
achieved similar values within a run. Two samples (QC105
or QC105C and
QC106) were coded and were included four times each
in the panel.
Panels containing QC105 or QC105C were distributed
to laboratories that
use the AMPLICOR HIV-1 MONITOR v 1.0 or Quantiplex
bDNA 2.0 assay,
respectively, for the reasons given above. In
addition, laboratories
were also requested to run the QC samples
in the same run as the QAP
samples, when
practical.
The third panel, panel VLQA 98-2, contained five coded plasma samples
from subjects infected with HIV-1 subtype B (one sample)
or subtype E
(four samples) and a control sample of anti-HIV-negative,
pooled normal
human plasma. These samples were estimated not to
contain very high
viral loads (<20,000 RNA copies/ml), but their
loads were expected to
fall comfortably within the linear range
of the assays at the various
testing sites. Again, when practical,
laboratories were also requested
to run QC samples QC105 or QC105C
and QC106 (subtype B-spiked normal
human plasma) in the same run
as the QAP
samples.
Data presentation and statistical analysis.
For the QC data,
results for each QC sample are presented as the mean ± 1 standard
deviation (SD) HIV-1 RNA copy number per milliliter. The coefficient of
variation (CV) is given as a percentage and has been calculated as the
SD divided by the mean. For the QAP panel data, the mean ± 2 SD
was calculated from the raw data for the number of HIV-1 RNA copy
number per milliliter. When the data are presented as residuals, values
for copy number per milliliter were first transformed to
log10 values and residuals were derived as the difference
between the reported log10 copy number per milliliter and
the mean log10 copy number per milliliter for each sample.
Statistical analysis.
The results obtained for QC samples
were compared between the two assay types by the t test. The
incidence of invalid runs between different assay lot numbers was
analyzed by the chi-square test. Between-sample data for the QAP panels
were analyzed by two-way analysis of variance (ANOVA) to test variance
between samples in the panel and between testing sites. When
significance was observed across the ANOVA table, a t test
was applied to locate significance within the data set.
 |
RESULTS |
QC program.
The results from two separate sets of QC samples
(set QC101 and QC102 and set QC105, QC105C, and QC106) tested by
previous assay versions and one set tested by the new versions of the
AMPLICOR HIV-1 MONITOR and Quantiplex bDNA assays are presented in
Table 1. Only results obtained from valid
runs are included (specific data for invalid run rates in each assay
appear below). The number of observations for the AMPLICOR HIV-1
MONITOR v 1.0 assay is substantially greater than the number of
observations for the Quantiplex bDNA 2.0 assay due to the more
widespread use of the former: 17 laboratories used the AMPLICOR HIV-1
MONITOR assay and between 2 and 6 laboratories used the Quantiplex bDNA
assay (some laboratories used both assays for a period within the
study). Despite the smaller numbers of observations, the variability of the data from the Quantiplex bDNA 2.0 assay was consistently lower than
the variability of the data from the AMPLICOR HIV-1 MONITOR v 1.0 assay
across both QC sample sets. The data obtained by the AMPLICOR HIV-1
MONITOR v 1.0 assay had a reduced variability with the more recent QC
sample set (Table 1) compared to the data obtained a maximum of 2 years
previously with the earlier QC101 and QC102 sample set (Table 1).
Values for HIV-1 copy number per milliliter were consistently lower by
the Quantiplex bDNA 2.0 assay than by the AMPLICOR HIV-1 MONITOR v 1.0 assay with these QC samples containing subtype B (P < 0.001).
View this table:
[in this window]
[in a new window]
|
TABLE 1.
HIV-1 RNA load measured by AMPLICOR HIV-1 MONITOR v 1.0 and v 1.5 assays and Quantiplex bDNA 2.0 and 3.0 assays in three
separately distributed sets of QC samples
|
|
Data obtained to date with the new versions of each assay are shown in
Table
1. The CV appears to be lower for the Quantiplex
bDNA 3.0 assay
than for the earlier version of the assay. This
may be due to the fact
that the later version is almost entirely
automated. The CV does not
appear to have changed for the AMPLICOR
HIV-1 MONITOR assay between v
1.0 and v 1.5, but further data
may be required to establish this with
certainty.
Data from two laboratories that used Ultrasensitive sample preparation
in conjunction with the AMPLICOR HIV-1 MONITOR v 1.5
assay are shown in
Table
1. Data presented for the Quantiplex
bDNA v 3.0 assay in Table
1
were generated in only one
laboratory.
Invalid assay runs. (i) AMPLICOR HIV-1 MONITOR v 1.0 assay.
Early monitoring of HIV viral load tests revealed that approximately
half the laboratories did not achieve consistently valid runs. Invalid
runs were more likely to occur for particular lot numbers of AMPLICOR
HIV-1 MONITOR v 1.0. Figure 1 shows the
invalid run rate, according to the manufacturer's criteria, for six
different lots supplied in 1998. The variability in the number of
invalid runs between lots was not predictable from one lot to the next.

View larger version (36K):
[in this window]
[in a new window]
|
FIG. 1.
Incidence of invalid runs, by lot number, as proportion
of total runs (shown by the number within each bar) of the AMPLICOR
HIV-1 MONITOR assay in all laboratories performing HIV viral load
testing in Australia. *, P < 0.05.
|
|
(ii) Quantiplex bDNA 2.0 assay.
During the reporting period
for the first QC sample set for viral load (QC101 and QC102), the
Quantiplex bDNA 2.0 assay had fewer invalid runs, with 5 of 77 runs
being invalid. Results for a small percentage (<10%) of individual
samples were rejected according to the manufacturer's criteria,
because the CV between the mandatory two replicates per sample was
>40% for negative samples (<500 HIV-1 RNA copies per ml) and >30%
for positive samples. When repeated, these samples usually yielded
valid results.
In the reporting period with QC105C and QC106, a low rate of invalid
runs continued: 2 of 91 runs were invalid. There was
a period of
approximately 4 months in which one laboratory that
used the Quantiplex
bDNA 2.0 assay experienced an increase in
the number of invalid runs
(from August to November 1997) which
was not seen in other laboratories
using the assay. However, despite
intensive investigation by the
laboratory and the manufacturer,
no clear cause of the invalid runs was
established.
Insufficient data have been generated to permit analysis of the invalid
run rate with the new versions of each
assay.
QAP. (i) Panel VLQA 97-1: accuracy across a range of viral
concentrations.
The results of testing of panel VLQA 97-1 by
laboratories using the AMPLICOR HIV-1 MONITOR v 1.0 and Quantiplex bDNA
2.0 assays are shown in Fig. 2. Results
were received from 19 laboratories: 13 used the AMPLICOR HIV-1 MONITOR
v 1.0 assay only, 2 used the Quantiplex bDNA 2.0 assay only, and 4 used
both assays. The data for the fivefold dilution series are presented as
residuals (difference of the reported value of log10 HIV-1
RNA copy number per milliliter for each sample from the mean
log10 HIV-1 RNA copy number per milliliter) and are given
by testing site for each sample in the panel (dilution). The results
for the 1:25 dilution were averaged, as this dilution was presented in
duplicate in the panel. The panel also contained a negative sample with
diluent only, which is not included on the graph, giving a total of
three results for each testing site.

View larger version (56K):
[in this window]
[in a new window]
|
FIG. 2.
Results obtained by individual laboratories, identified
by code number, for QAP panel VLQA 97-1, a panel of samples with a
range of HIV-1 dilutions. The panel contained an undiluted sample
( ), a 1/5 dilution of a sample ( ), and a 1/25 dilution of a
sample in duplicate, shown as the average ( ). Data are presented as
residuals (difference of the reported value from the mean). The mean is
indicated by the horizontal line, and 2 SDs above and below the mean
are shown by the shaded region. (A) AMPLICOR HIV-1 MONITOR assay; (B)
Quantiplex bDNA assay.
|
|
All but four results produced at any site by the AMPLICOR HIV-1 MONITOR
v 1.0 assay were within 2 SDs of the mean (Fig.
2A).
The four reported
values which fell outside 2 SDs of the mean
came from four separate
testing sites and were for three different
samples; i.e., it was not
only one site or one sample concentration
which provided the widely
variant results. Only one value was
outside 2 SDs by the Quantiplex
bDNA 2.0 assay (Fig.
2B). Because
fewer testing sites used the
Quantiplex bDNA 2.0 assay, the SD
was expected to have been greater
than that seen with the AMPLICOR
HIV-1 MONITOR assay. In spite of this,
the log
10 SD obtained by
the Quantiplex bDNA 2.0 assay was
approximately half that found
by the AMPLICOR HIV-1 MONITOR v 1.0 assay
for each
dilution.
(ii) Panel VLQA 98-1: within-run reproducibility.
QAP panel
VLQA 98-1 was decoded, and the values reported by each testing site for
samples QC105, QC105C, and QC106 are shown in Fig.
3. When testing sites had included the
matching samples for quality control, they were reported in the data
set, giving five datum points for some testing sites. For sample QC106
tested by the AMPLICOR HIV-1 MONITOR v 1.0 assay, 14 of 81 results
(17%) were greater than 7.5 × 105 copies/ml, the
upper detection limit of the assay. To calculate the mean ± SD
(Fig. 3B), these results have been included in the data set as 7.5 × 105 copies/ml, as it was not considered valid to
extrapolate values beyond the upper limit of detection cited by the
manufacturer. In practice, this is the value which would be reported
clinically for a sample with a viral load which exceeded the upper
limit of detection.

View larger version (35K):
[in this window]
[in a new window]
|
FIG. 3.
Results obtained by individual laboratories, identified
by code number, for QAP panel VLQA 98-1, a panel of replicates of HIV-1
QC samples QC105, QC105C, and QC106. The mean is indicated by the
horizontal line, and 2 SDs above and below the mean are shown by the
shaded region. (A and B) Data for QC105 and QC106, respectively,
obtained by the AMPLICOR HIV-1 MONITOR assay; (C and D) data for QC105C
and QC106, respectively, obtained by the Quantiplex bDNA assay. Note
the 10-fold difference in scale between panels B and D (results for
QC106).
|
|
No significant difference was found between the results obtained for
any of the samples in the panel by either the AMPLICOR
HIV-1 MONITOR v
1.0 assay or the Quantiplex bDNA 2.0 assay. While
most laboratories
achieved clustered results within 2 SDs of the
mean when testing
replicate samples (Fig.
3) using the AMPLICOR
HIV-1 MONITOR v 1.0 assay, results from four testing sites were
spread over a wide range (a
wide spread is defined as some of
the results being greater than 2 SDs
from the mean when other
results from the same site fell on the mean or
on the opposite
side of the mean). These were testing sites 2 and 127, with the
sample with a viral load in the lower range (QC105) (Fig.
3A),
and sites 32 and 36, with the sample with a viral load in the
upper
range (QC106) (Fig.
3B). Testing sites 9 and 45 produced
clustered
results for QC106, all or some of which were greater
than 2 SDs above
the mean (Fig.
3B).
By the Quantiplex bDNA 2.0 assay, five of six laboratories reported
results that were within 2 SDs of the mean for the sample
with a viral
load in the lower range (sample QC105C). One laboratory
reported 2 of 5 results for QC105C greater than 2 SDs of the mean
(Fig.
3C). Results
reported from this site were spread over a
wider range than those
reported by other users for replicates
of both QC105C and QC106 (Fig.
3C and D), although no site reported
results greater than 2 SDs of the
mean for sample QC106, which
had a viral load in the upper range (Fig.
3D).
(iii) Panel VLQA 98-2: detection of HIV-1 subtypes.
The
results obtained from testing sites for panel VLQA 98-2 are presented
in Table 2. Results were received from 19 laboratories. Only one laboratory used both the AMPLICOR HIV-1 MONITOR
v 1.0 and the Quantiplex bDNA 2.0 assays. Two laboratories used the Quantiplex bDNA 2.0 assay only, but in one of those laboratories, the
run containing the QAP samples was invalid. Therefore, those results
were excluded from the analysis. The remaining 16 laboratories used the
AMPLICOR HIV-1 MONITOR v 1.0 assay only. All testing sites except site
128 included samples QC105 or QC105C and QC106 in the run. Site 39 included only QC106. Data for the HIV-1-negative sample were <400
copies/ml for the AMPLICOR HIV-1 MONITOR v 1.0 assay and <500
copies/ml for the Quantiplex bDNA 2.0 assay (data not shown). For the
same sample, the averaged results obtained for the samples with HIV-1
subtype by the AMPLICOR HIV-1 MONITOR v 1.0 assay were consistently
lower than those obtained from either laboratory by the Quantiplex bDNA
2.0 assay. In contrast, for the samples with HIV-1 subtype B, the
AMPLICOR HIV-1 MONITOR v 1.0 assay viral load results were higher than
the Quantiplex bDNA 2.0 assay results. This was true for both the
subtype B-containing sample in the panel and QC106. Although the viral
loads in QC105 and QC105C fell within the same range, they are
different samples, one being used in each assay. Thus, it is not valid
to compare the values for the latter samples.
View this table:
[in this window]
[in a new window]
|
TABLE 2.
HIV-1 RNA levels in samples with different HIV-1 subtypes
(QAP panel VLQA 98-2 and QC samples) measured by AMPLICOR HIV-1
MONITOR and Quantiplex bDNA assays
|
|
 |
DISCUSSION |
The use of viral load assessments is relatively new in Australia
compared to the United States, where clinical trials (American Clinical
Trials Group) necessitated the establishment of a QA program at an
earlier date (13). Both the AMPLICOR HIV-1 MONITOR and
Quantiplex bDNA assays were first registered in Australia in 1996 and,
together with the Organon Teknika NucliSens HIV-1 QT, are the only
assays currently licensed for diagnostic viral load testing. In
contrast, a larger variety of assays is used in the United States,
including in-house assays set up within individual laboratories. There
is abundant evidence to show that without quality assurance, the
performance parameters of such assays are poor (3).
The institution of monitoring of HIV viral load tests in Australia
found that approximately half the laboratories performing the AMPLICOR
HIV-1 MONITOR v 1.0 assay had intermittent or continuing difficulty
achieving assay runs which were valid according to the manufacturer's
criteria. Consistently, the reason for invalidity was a failure to
achieve a value for the high-positive kit control which was within the
specified range. With cooperation between the manufacturer and
personnel in several laboratories, very close attention to equipment
calibration, and strict adherence to the assay protocol, the proportion
of invalid runs by the AMPLICOR HIV-1 MONITOR v 1.0 assay was reduced
from 16 to 3% over a period of 6 months across all laboratories.
However, although laboratories demonstrated this improved performance,
increased invalid run rates continue to occur periodically and can be
correlated with the supply of different assay lots. A combination of
factors, including minor manufacturing and laboratory variations,
probably contribute to an invalid run. Prior to the availability of the NRL's program, results from invalid runs were regularly reported to
clinicians in the belief that a single invalid control result did not
necessarily invalidate the entire run. This practice was justified on
the basis of the nature of the test and its cost rather than on the
criterion of accuracy. The NRL's QA program has shown that, when
analyzed overall, the numerical results for QC samples do not change
significantly in invalid runs (data not shown). However, in multiple
instances the high-positive control, provided by the manufacturer,
returned values greater than the upper limit of detection and the value
for the higher-load NRL QC sample was greater than 2 SDs from the mean.
Thus, while values reported from invalid runs may not affect the
collective interpretation of a large amount of data, this argument
cannot be used to defend reporting of values from invalid runs, as the
error in an individual run may be substantial.
Familiarity with the assays and attention to detail have also served to
reduce the between-run variability both within and between
laboratories, reflected by the reduction in CVs seen between the data
generated by the AMPLICOR HIV-1 MONITOR v 1.0 assay for the first set
of QC samples (samples QC101 and QC102) and the more recent set of
samples (samples QC105 and QC106). In the new version of the AMPLICOR
HIV-1 MONITOR assay, v 1.5, results collected to date suggest that
there is no difference in the CV from the previous version of the same
assay except when Ultrasensitive sample preparation is used. The high
CV observed when the modified sample preparation is used may reflect
the inexperience of the operators with the new technique but may also
highlight a difficulty obtaining precise results for samples with low
viral loads by the AMPLICOR HIV-1 MONITOR assay in conjunction with
Ultrasensitive sample preparation. As the new version does not differ
in practical terms from the previous version, it is not anticipated
that the CV will decrease with use of the assay but reflects inherent
assay variability. It should be noted that the variability observed with the Quantiplex bDNA 2.0 assay, and now with the Quantiplex bDNA
3.0 assay, has remained consistently less than 35% throughout the program.
Each QAP is designed to ask a question specifically of the quality of
HIV viral load testing. In the present study, the questions were,
"Can laboratories consistently distinguish a 5-fold difference in
viral load?," "What is the within-run reproducibility of the same
sample?," and "Are assays equally effective for detection of
different HIV-1 subtypes?" Only 5 of 19 laboratories, using either
assay, produced one result for one of the dilution samples (a different
one in each case) which fell outside the range of acceptable
variability. The reproducibility of the results obtained with replicate
samples demonstrated greater variability, particularly in laboratories
using the AMPLICOR HIV-1 MONITOR assay, which may be expected for the
PCR technique. Of 17 laboratories, 3 had widely divergent results for
replicate samples with high and low viral loads. Only one of these
laboratories had problems with reproducibility for samples with both
high and low viral loads. The overall results indicate that while
variability is important, there are no systematic testing problems with
any laboratory in the detection of virus dilutions.
It has previously been recognized that quantification of viral load by
the different assays varies. A lower copy number by the Quantiplex bDNA
2.0 assay has consistently been reported (1, 11, 12) and is
also supported by the present study. Of greater significance, however,
is the fact that the genetic variability of HIV-1 can influence viral
load quantification by the various assays. The AMPLICOR HIV-1 MONITOR v
1.0 assay demonstrated poorer sensitivity for some HIV-1 non-B
subtypes, with the Quantiplex bDNA 2.0 assay more efficiently detecting
subtypes A and E (1, 4, 9, 11). In the present study the
values obtained by the AMPLICOR HIV-1 MONITOR v 1.0 assay for subtype E
were lower than those obtained for subtype B. The predominance of HIV-1
subtype B in Australia reduced the impact of this finding. However,
with the prevalence of subtype E in Southeast Asia (2), the
incidence of this subtype in Australia might reasonably be expected to
change. The updated version of the AMPLICOR HIV-1 MONITOR assay has
been designed to address the deficiency in detection of HIV-1 non-B subtypes seen in v 1.0 by the use of modified primers for nucleic acid
amplification. When alternative primers have been used, improved detection of subtype A has been observed (9). Furthermore, in a recent direct comparison, the updated versions of the AMPLICOR HIV-1 MONITOR and Quantiplex bDNA assays demonstrated very close correlations in the number of copies per milliliter, unlike the earlier
versions of the same assays (6). Insufficient data have been
obtained through the NRL QA program to establish whether improved
sensitivity to HIV-1 non-B subtypes has been achieved or whether a
better correlation of results is demonstrated by the newer versions of
the assays.
In conclusion, the results obtained from the NRL QA program support the
recommendation that testing for HIV-1 viral load in an individual
patient should always be conducted by the same test and by the same
laboratory. The continued variability observed with new versions of the
most commonly used assays suggests that this recommendation should
remain in place. As long as the same test is consistently used for
within-patient measures, the relative values obtained will indicate the
efficacy of antiretroviral treatment. The new versions of the viral
load assays will probably redress some of the problems associated with
the sensitivities of the assays for various HIV-1 subtypes. However
highly divergent subtypes may remain undetectable (10). In
addition, despite the increased levels of sensitivity claimed by each
manufacturer, problems of variability in the assays continue with low
viral loads, particularly in the Roche assay.
 |
ACKNOWLEDGMENTS |
NRL's HIV viral load QA program was partially funded by a grant
from the Diagnostics and Technology Branch of the Department of Health
and Aged Care (DHAC) of the Australian Government. NRL is funded by an
operational grant from the Population Health Division of DHAC.
We are grateful to A. Dunne and Professor S. Crowe of the Macfarlane
Burnet Centre for Medical Research for the generous supply of HIV-1
subtype B supernatants used in the preparation of the QC samples and to
S. Walker for assistance with data collection. In addition, we
acknowledge the cooperation of the participating laboratories.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: National
Serology Reference Laboratory, Australia, at St Vincent's Institute of
Medical Research, 41 Victoria Parade, Fitzroy, Vic 3065, Australia.
Phone: 61 3 9418 1111. Fax: 61 3 9418 1155. E-mail:
sue{at}nrl.gov.au.
 |
REFERENCES |
| 1.
|
Berger, A.,
J. Braner,
H. W. Doerr, and B. Weber.
1998.
Quantification of viral load: clinical relevance for human immunodeficiency virus, hepatitis B virus and hepatitis C virus infection.
Intervirology
41:24-34[CrossRef][Medline].
|
| 2.
|
Brown, A. E., and J. G. McNeil.
1998.
HIV vaccine development: a subtype E-specific strategy.
Southeast Asian J. Trop. Med. Public Health
29:377-382[Medline].
|
| 3.
|
Carpenter, C. C.,
M. A. Fischl,
S. M. Hammer,
M. S. Hirsch,
D. M. Jacobsen,
D. A. Katzenstein,
J. S. Montaner,
D. D. Richman,
M. S. Saag,
R. T. Schooley,
M. A. Thompson,
S. Vella,
P. G. Yeni, and P. A. Volberding.
1998.
Antiretroviral therapy for HIV infection in 1998.
JAMA
280:78-86[Abstract/Free Full Text].
|
| 4.
|
Dunne, A. L., and S. M. Crowe.
1997.
Comparison of branched DNA and reverse transcriptase polymerase chain reaction for quantifying six different HIV-1 subtypes in plasma.
AIDS
11:126-127[Medline].
|
| 5.
|
Henrard, D. R.,
E. Daar,
H. Farzadegan,
S. J. Clark,
J. Phillips,
G. M. Shaw, and M. P. Busch.
1995.
Virologic and immunologic characterization of symptomatic and asymptomatic primary HIV-1 infection.
J. Acquir. Immune Defic. Syndr. Hum. Retrovirol.
9:305-310[Medline].
|
| 6.
|
Highbarger, H. C.,
W. G. Alvord,
M. K. Jlang,
A. S. Shah,
J. A. Metcalf,
H. C. Lane, and R. L. Dewar.
1999.
Comparison of the Quantiplex version 3.0 assay and a sensitized Amplicor Monitor assay for measurement of human immunodeficiency virus type 1 RNA levels in plasma samples.
J. Clin. Microbiol.
37:3612-3614[Abstract/Free Full Text].
|
| 7.
|
Ho, D. D.,
A. U. Neumann,
A. S. Perelson,
W. Chen,
J. M. Leonard, and M. Markowitz.
1995.
Rapid turnover of plasma virions and CD4 lymphocytes in HIV-1 infection.
Nature
373:123-126[CrossRef][Medline].
|
| 8.
|
Mellors, J. W.,
C. R. Rinaldo,
P. Gupta,
R. M. White,
J. A. Todd, and L. A. Kinsley.
1996.
Prognosis in HIV-1 infection predicted by the quantity of virus in plasma.
Science
272:1167-1170[Abstract].
|
| 9.
|
Nkengasong, J. N.,
M. Kalou,
C. Maurice,
C. Bile,
M. Y. Borget,
S. Koblari,
E. Boateng,
M. Sassan-Morokro,
E. Anatole-Ehounou,
P. Ghys,
A. E. Greenberg, and S. Z. Wiktor.
1998.
Comparison of NucliSens and Amplicor Monitor assays for quantification of human immunodeficiency virus type 1 (HIV-1) RNA in plasma of persons with HIV-1 subtype A infection in Abidjan, Côte d'Ivoire.
J. Clin. Microbiol.
36:2495-2498[Abstract/Free Full Text].
|
| 10.
|
Oelrichs, R. B., and C. Workman.
1999.
Beside the limit how HIV-1 can mask its true viral load.
Victorian Infect. Dis. Bull.
2:33.
|
| 11.
|
Parekh, B.,
S. Phillips,
T. C. Granade,
J. Baggs,
D. J. Hu, and R. Respess.
1999.
Impact of HIV type 1 subtype variation on viral RNA quantitation.
AIDS Res. Hum. Retrovir.
15:133-142[CrossRef][Medline].
|
| 12.
|
Prud'homme, I. T.,
J. E. Kim,
R. G. Pilon,
T. Minkus,
N. Hawley-Foss,
W. Cameron, and E. W. Rud.
1998.
Amplicor HIV monitor, NASBA HIV-1 RNA QT and Quantiplex HIV RNA version 2.0 viral load assays: a Canadian evaluation.
J. Clin. Virol.
11:189-202[Medline].
|
| 13.
|
Yen-Lieberman, B.,
D. Brambilla,
B. Jackson,
J. Bremner,
R. Coombs,
M. Cronin,
S. Herman,
D. Katzenstein,
S. Leung,
H. J. Lin,
P. Palumbo,
S. Rasheed,
J. Todd,
M. Vahey, and P. Reichelderfer.
1996.
Evaluation of a quality assurance program for quantitation of human immunodeficiency virus type 1 RNA in plasma by the AIDS Clinical Trials Group virology laboratories.
J. Clin. Microbiol.
34:2695-2701[Abstract].
|
Journal of Clinical Microbiology, November 2000, p. 4015-4020, Vol. 38, No. 11
0095-1137/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.