Previous Article | Next Article 
Journal of Clinical Microbiology, March 2000, p. 1036-1041, Vol. 38, No. 3
0095-1137/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
Failure of an Automated Blood Culture System To
Detect Nonfermentative Gram-Negative Bacteria
H.-G.
Klaerner,1
U.
Eschenbach,1
K.
Kamereck,1
N.
Lehn,2
H.
Wagner,1 and
T.
Miethke1,*
Institute of Medical Microbiology, Immunology
and Hygiene, Technical University of Munich, 81675 Munich,1 and Institute of Medical
Microbiology and Hygiene, University of Regensburg, 93053 Regensburg,2 Germany
Received 27 August 1999/Returned for modification 10 November
1999/Accepted 7 December 1999
 |
ABSTRACT |
During a 1-year study we observed that both aerobic and anaerobic
blood culture bottles from patients were negative by the BacT/Alert
system during a 7-day incubation period. However, upon subcultivation
of negative bottles, growth of Pseudomonas aeruginosa was
detectable. In an attempt to explain this observation, aerobic BacT/Alert Fan bottles were seeded with a defined inoculum (0.5 McFarland standard; 1 ml) of Escherichia coli,
Klebsiella pneumoniae, Serratia marcescens,
P. aeruginosa, Stenotrophomonas maltophilia, or
Acinetobacter baumannii. Half of the inoculated bottles
were loaded into the BacT/Alert system immediately, and the remainder were preincubated for 4, 8, 16, and 24 h at 36°C. With
preincubation all bottles seeded with the
Enterobacteriaceae signaled positive during the next
1.5 h. Organisms in bottles seeded with the nonfermentative species P. aeruginosa and A. baumannii remained
undetected by the BacT/Alert system for 7 days. S. maltophilia was detected if the preincubation time was equal or
less than 8 h. Without preincubation all bottles seeded with the
Enterobacteriaceae or nonfermentative species signaled
positive. Since nonfermentative species seem to enter a state of
bacteriostasis within the preincubation period, we reasoned that an
unknown factor is consumed. Accordingly, a smaller inoculum should
allow the detection of nonfermentative species, even after
preincubation, and serial dilutions of P. aeruginosa were
detected in preincubated bottles. In this case preincubated bottles
signaled positive faster than bottles without preincubation. We
conclude that all bottles from clinical settings should be subcultured
prior to loading to avoid false negatives. An alternative may be
preincubation at room temperature.
 |
INTRODUCTION |
Detection of bacteremia is one of
the most important functions of clinical microbiology laboratories. Due
to the high morbidity and mortality associated with this disease
process, rapid detection and identification of clinically relevant
microorganisms in blood cultures remain most important (1).
In response to these demands, the development of continuously
monitoring blood culture systems took place. The BacT/Alert system
(Organon Teknika, Eppelheim, Germany) is one of the fully automated
blood culture systems (8). The BacT/Alert software examines
the readings from each bottle to determine if there is evidence for
bacterial growth. The detection system accomplishes this by using three
types of measurements of the CO2 level: (i) the present
CO2 level in the bottle, (ii) the rate of CO2
production, and (iii) the acceleration of this rate. The combination of
the three measurements allows the system a high degree of accuracy and
specificity (8).
Ideally, blood culture bottles should be entered into a continuously
monitoring blood culture system as soon as possible. Delays in entry of
blood culture bottles may be due to collection from outlying hospitals,
limited instrument capacity, or collection of blood samples during the
night. Under these circumstances, it is recommended that the blood
culture bottles should be stored at 36°C until arrival in the
laboratory. However, this treatment of blood culture bottles prior to
loading into the BacT/Alert raises the question of whether the system
is still able to detect positive blood culture bottles.
During a 1-year study involving 8,107 blood culture bottles, we
observed that aerobic and anaerobic blood culture bottles from patients
were negative by the BacT/Alert system during a 7-day incubation period
but that subcultivation on blood agar plates resulted in growth of
Pseudomonas aeruginosa in 15 cases. In an attempt to find an
explanation for this observation, we speculated that the delayed entry
and preincubation at 36°C is responsible for the failure of the
BacT/Alert computer algorithm to detect certain positive cultures.
Therefore, we investigated the effect of delayed entry and how
preincubation at 36°C influences the sensitivity of the BacT/Alert blood culture system.
 |
MATERIALS AND METHODS |
Study design.
We used aerobic BacT/Alert Fan bottles
(Organon Teknika) throughout the experimental study. The bottles
contain brain heart infusion (2.8% wt/vol), sodium
polyethanolsulfonate (0.05%), pyridoxine-HCl (0.001%, wt/vol),
menadione (0.00005%, wt/vol), hemin (0.0005%, wt/vol), activated
charcoal (8.5%), L-cysteine and other complex amino acids,
and carbohydrate substrates in purified water. Bottles were labeled in
duplicate with incubation time, temperature, and delayed-entry time. A
suspension (equivalent to a 0.5 McFarland standard [referred to
hereafter as McFarland 0.5]) of the organism to be tested
(Escherichia coli, Klebsiella pneumoniae,
Serratia marcescens, P. aeruginosa,
Stenotrophomonas maltophilia, or Acinetobacter baumannii) was prepared, and 1 ml of each organism suspension was
added to the specific bottles. One hundred microliters per bottle was
removed and plated on 5% Columbia blood agar plates (Becton Dickinson,
Heidelberg, Germany) to determine the number of CFU per bottle. After
overnight incubation at 36°C, colony counts were performed. Half of
the bottles were loaded into the BacT/Alert immediately, and the other
half were incubated at 36°C for 4, 8, 16, and 24 h (this was
recommended by the manufacturer for delayed-entry bottles). All bottles
were vented just prior to inoculation. Bottles remained on the
BacT/Alert until they signaled positive or until 7 days elapsed, and
the time from loading to detection was determined for each bottle.
Bottles that failed to signal were subcultured on blood agar plates
after being unloaded.
In a second experiment a suspension (McFarland 0.5) of the organisms to
be tested (E. coli, P. aeruginosa, S. maltophilia, or A. baumannii) was prepared, and serial
dilutions (1:100, 1:1,000, and 1:10,000) were performed. BacT/Alert Fan
bottles were inoculated with 1 ml of each dilution, and 100 µl of
suspension from each bottle was removed and subcultured on blood agar
plates to determine the number of CFU. Half of the inoculated bottles
were loaded into the BacT/Alert immediately, and the remainder were
preincubated for 8 h at 36°C. The rest of the procedure was the
same as described for the first experiment.
In a third experiment a suspension (McFarland 0.5) of
P. aeruginosa was prepared, and dilutions of 1:1,000, 1:10,000, and
1:100,000 plus 10 ml of blood from normal human volunteers were
inoculated in each blood culture bottle. One hundred microliters
of
suspension from each bottle was removed and subcultured on
blood agar
plates to determine the number of CFU. Half of the
inoculated bottles
were loaded into the BacT/Alert immediately,
and the remainder were
preincubated for 8 h at 36°C. The rest
of the procedure was the
same as described for the first
experiment.
Evaluation of specimens.
This study was conducted from May
1998 until April 1999 at the University Hospital "Klinikum rechts der
Isar," Technical University of Munich. Aerobic and anaerobic
BacT/Alert Fan bottles were used on all wards of the hospital. We
included anaerobic bottles in the clinical study to obtain the whole
spectrum of clinically relevant bacteria. The composition of the
anaerobic medium is identical to that of the aerobic medium described
above. The difference between aerobic and anaerobic bottles is that
aerobic bottles are enriched with O2, whereas anaerobic
bottles are enriched with CO2 and N2. The time
needed for the blood culture bottles to arrive in the laboratory ranged
from 30 min to 12 h. It is common within the clinic to incubate
the blood culture bottles at 36°C in case they cannot be immediately
delivered to the laboratory. On arrival in the laboratory, aliquots
were taken from each blood culture bottle (aerobic and anaerobic
bottles) obtained from the medical and surgical care wards and
subcultured onto adequate medium (blood agar incubated at 36°C) prior
to loading into the BacT/Alert system. Bottles remained on the
BacT/Alert until they signaled positive or until 7 days elapsed.
Bottles that failed to signal were subcultured on blood agar plates
after being unloaded. For all positive bottles the bacterial species
and time from loading to detection were determined. In parallel, the
presence of antibiotics in clinical blood culture bottles was analyzed
by adding a drop of the blood culture medium to agar holes of
Mueller-Hinton plates inoculated with Bacillus subtilis.
After incubation overnight, the plates were checked for growth
inhibition of the indicator bacteria.
 |
RESULTS |
Clinical study.
A total of 8,107 blood culture bottles were
obtained during the study period. Of these, 1,078 blood culture bottles
(aerobic or anaerobic bottles) were positive as determined by the
instrument in combination with subcultures, and 605 bottles were
considered clinically relevant. Accordingly, cases where both bottles
for a patient were positive were counted only once, and bottles growing microorganisms like propionibacteria or lactobacilli were not counted.
Of the 605 clinically relevant species detected, most of them (62.9%)
were gram-positive bacteria; 35.2% were gram-negative bacteria, and
only 1.8% were Candida spp. (Table
1). All gram-positive bacteria and
species that belong to the Enterobacteriaceae (26.9%) could
be detected by the BacT/Alert system. In contrast, 27.3% of the
Candida spp. and 40.5% of all nonfermentative species
remained undetected by the system. A total of 46.9% of all P. aeruginosa-positive blood culture bottles were not detected by the
BacT/Alert system, but all clinical samples with these bacterial
species were positive on initial or terminal subculture. The false
negatives enumerated in Table 1 were instrument negative for both
aerobic and anaerobic blood culture bottles. Evaluation of these
patients showed that all of them were critically ill due to cancer,
sepsis, or large operations like hemicolectomy or total gastrectomy.
All patients suffered from fever, with a body temperature of between
38.5 and 40.5°C, several of them received antibiotics like
piperacillin-combactam, mezlocillin-combactam, gentamicin, vancomycin,
and others, which was documented for 10 of the 15 false-negative
P. aeruginosa cases. Evaluation of the false-negative blood
culture bottles for the presence of antibiotics revealed that three out
of nine tested cases contained antibiotics. In the other six
false-negative cases, the presence of antibiotics could not be
demonstrated. However, in 4 of 14 tested and true-positive cases, we
could also demonstrate the presence of antibiotics. In conclusion, the
instrument is unable to detect more than 46% of cases of clinically
relevant bacteremias due to P. aeruginosa.
Effect of delayed-entry time on detection of gram-negative
bacterial strains.
With preincubation times of 4 h (done only
with E. coli and nonfermentative bacteria) and of 8, 16, and
24 h at 36°C, all bottles seeded with E. coli,
K. pneumoniae, and S. marcescens signaled
positive during the next 1.5 h after loading into the BacT/Alert
system (Table 2). The average detection
times for E. coli were 1.0 h (4-h preincubation),
1.2 h (8- and 16-h preincubations), and 1.5 h (24-h
preincubation); those for K. pneumoniae were 1.2 h (8- and 16-h preincubations) and 1.1 h (24-h preincubation); and those
for S. marcescens were 1.2 h (8- and 16-h
preincubations) and 1.0 h (24-h preincubation). For all three
Enterobacteriaceae the average detection time after 8, 16, and 24 h of preincubation was 1.2 h. In contrast, organisms
in bottles seeded with P. aeruginosa and A. baumannii were not detected by the BacT/Alert system. Even with
shorter preincubation times, down to 4 h, bottles failed to signal
positive for these species. Organisms in bottles seeded with S. maltophilia were undetected with preincubation times of longer
than 8 h. After preincubation times of 8 and 4 h, the bottles signaled positive but the time to detection was significantly longer
than that for the tested Enterobacteriaceae described above. In each case, however, the specimen in the bottle was viable as determined by subculture of an aliquot on blood agar plates.
Without preincubation, all bottles seeded with one of the bacterial
strains to be tested were positive. The time needed for
the instrument
to detect each of the
Enterobacteriaceae was significantly
longer than that for the preincubated bottle (Table
2). Bottles
seeded
with
E. coli signaled positive after 5.3 h, and those
seeded
with
K. pneumoniae or with
S. marcescens
signaled positive after
6.2 h. For all three
Enterobacteriaceae the average detection
time was 5.9 h. Although organisms in bottles inoculated with
the nonfermentative
species were detected, the detection time
was significantly longer than
that for the
Enterobacteriaceae.
Bottles seeded with
P. aeruginosa were positive after 7.6 h, those
seeded
with
A. baumannii were positive after 8.9 h, and those
seeded with
S. maltophilia were positive after 9.1 h.
For all
three nonfermentative species the average detection time was
8.5
h.
Effect of different colony numbers on the detection of
gram-negative bacterial strains.
To determine whether smaller
inocula would influence the detection capacity of the BacT/Alert blood
culture system, serial dilutions from each bacterial strain (McFarland
0.5) were performed. With a preincubation time of 8 h at 36°C,
the bottles seeded with the Enterobacteriaceae species
E. coli and the nonfermentative species S. maltophilia signaled positive at each colony number tested (Table
3). Growth of E. coli was
detected in bottles inoculated with 3.0 × 103 CFU/ml
(1:100 dilution), 275 CFU/ml (1:1,000 dilution), and less than 10 CFU/ml (1:10,000 dilution) after 1.2, 1.0, and 2.8 h, respectively. Although the same McFarland density was used to generate
the dilutions of S. maltophilia, only 10 CFU/ml
(1:100 dilution) and less than 10 CFU/ml (1:1,000 and 1:10,000
dilutions) were inoculated as determined by culture on blood agar
plates. Therefore, the times to detection, i.e., 19.4 h, 21.3, and
24.1 h, were on average more than 1 order of magnitude longer than those for E. coli. In contrast, organisms in bottles
inoculated with 1.4 × 104 CFU/ml (1:100 dilution) and
5.5 × 103 CFU/ml (1:1,000 dilution) of A. baumannii remained undetected by the BacT/Alert system. A
1:10,000 dilution (150 CFU/ml), however, allowed detection on average
after 9.2 h. The same dilutions of P. aeruginosa showed
bacterial counts of 900 CFU/ml (1:100 dilution), 300 CFU/ml (1:1,000
dilution), and 30 CFU/ml (1:10,000 dilution). An inoculum of 9.0 × 102 CFU of P. aeruginosa per ml could not be
detected by the BacT/Alert system; smaller colony numbers resulted on
average in detection times of 7.4 and 9.1 h, respectively.
View this table:
[in this window]
[in a new window]
|
TABLE 3.
Time to detection of different colony numbers of
gram-negative bacterial strains after 8 h of preincubation of
aerobic bottles
|
|
Without preincubation, all bottles seeded with the different bacteria
listed in Table
4 signaled positive and
the time to
detection was in reverse proportion to the number of
colonies
inoculated. Although the absolute number of colonies differs
from
one strain to another, the
Enterobacteriaceae species
E. coli needed a shorter time to detection throughout the
experiment (on
average 7.8, 9.0, and 10.3 h). Bottles seeded with
S. maltophilia needed approximately one-quarter more time
until detection compared
to the preincubated ones (on average 23.9, 25.8, and 30.0 h).
The detection times for
A. baumannii
were on average 11.4 and
16.5 h, at inocula where the system
failed to detect the bacterial
strain after preincubation. The smallest
colony number tested
resulted in a detection time of on average
17.7 h, nearly twice
as long as the time with preincubation.
Bottles seeded with
P. aeruginosa at the highest colony
number tested signaled positive
after on average 10.6 h; at this
inoculum, the BacT/Alert system
could not detect any bacterial growth
in the case of preincubation.
Smaller colony numbers tested resulted in
clearly longer detection
times (on average 12.7 and 14.3 h) than
for the preincubated bottles.
View this table:
[in this window]
[in a new window]
|
TABLE 4.
Time to detection of different colony numbers of
gram-negative bacterial strains without preincubation of
aerobic bottles
|
|
Effect of different colony numbers of P. aeruginosa in
the presence of blood on detection.
To determine whether blood
would influence the detection capacity of the BacT/Alert system, serial
dilutions of P. aeruginosa (McFarland 0.5) were performed.
With a preincubation time of 8 h, bottles inoculated with 340 CFU/ml (1:1,000 dilution) signaled positive after 5.7 days. A dilution
of 1:10,000 (20 CFU/ml) resulted in a detection time of 6 h in one
bottle; in the other one, the same number of CFU remained undetected by
the system. A final dilution of 1:100,000 (5 CFU/ml) was detected after
7.1 h.
Without preincubation organisms in all bottles were detected, although
the time to detection increased as the colony number
got smaller. The
dilutions of 1:1,000 (340 CFU/ml) and 1:100,000
(5 CFU/ml) were
detected after 12.1 and 15.9 h, respectively.
Organisms in one
bottle inoculated with 20 CFU/ml (1:10,000 dilution)
were detected
after 2.8 days; those in the other one were detected
after 13.7
h.
Analysis of growth curves.
Growth curves for E. coli and P. aeruginosa (McFarland 0.5) with or without
preincubation (8 h) were analyzed. Every 10 min the BacT/Alert system
measured the absolute CO2 concentration and its
acceleration during each time period in the bottle as determined by the
color change of the lacmus paper at the bottom of the bottle. The
reflection unit represents a combination of both determinations. A
threshold exists for both (the absolute CO2 concentration
at 3,300 reflection units and the acceleration of the concentration at
32 reflection units/10 min), and above these thresholds the computer
system would determine the bottle as positive. Eight hours of
preincubation of E. coli resulted in greater than 4,600 reflection units after 1.2 h until determination as positive (Fig.
1A). Without preincubation, 3,500 reflection units were reached after 5.2 h, which was enough for
detection of bacterial growth (Fig. 1A). In contrast, 8 h of
preincubation of P. aeruginosa resulted in less than 2,400 reflection units over 7 days, which was not enough for determination
as positive (Fig. 1B). Without preincubation, 2,600 reflection units
after 7.4 h caused the BacT/Alert system to signal positive for
bacterial growth in the bottle (Fig. 1B).

View larger version (11K):
[in this window]
[in a new window]
|
FIG. 1.
Effect of delayed-entry time and different colony
numbers on the detection of gram-negative bacterial strains. (A) Growth
curves of E. coli (McFarland 0.5) preincubated for 8 h
( ) or not preincubated ( · ). (B)
Growth curves of P. aeruginosa (McFarland 0.5) preincubated
for 8 h ( ) or not preincubated
( · ) (C) Growth curves of P. aeruginosa (300 CFU/ml) preincubated for 8 h ( ) or not
preincubated ( · ).
|
|
The growth curve of
P. aeruginosa, where bottles were
inoculated with only 300 CFU/ml (McFarland 0.5 diluted 1:100), reached
2,650 reflection units at 7.4 after loading into the BacT/Alert
computer (Fig.
1C). Approximately the same reflection units were
measured, but reached after 12.7 h, for detection of the same
inoculum without preincubation (Fig.
1C).
 |
DISCUSSION |
More than 200,000 cases of septicemia occur annually in the United
States (4). The five most common isolates from blood cultures were Staphylococcus aureus, coagulase-negative
staphylococci, E. coli, K. pneumoniae, and
Enterococcus species (6). Microorganisms that
almost always (>90% of isolates) represent true infection when
isolated from the blood include S. aureus, E. coli and other Enterobacteriaceae, P. aeruginosa, S. pneumoniae, and Candida albicans (5). In our study 35.2% of all positive blood
cultures were gram-negative bacteria and 6.1% belonged to the
nonfermentative group, where 40.5% remained undetected by the
BacT/Alert system. In each case these bacteria caused symptomatic
bacteremia in critically ill patients. Without routine subculture of
aliquots from each bottle set from these patients after arrival in the
laboratory, none of the nonfermentative species would be detected.
Another possibility to reduce false negatives with preincubated blood culture bottles might be to perform preincubation at room temperature instead of 36°C. As shown by others (Organon Teknika, unpublished communication) and by ourselves (unpublished observation),
preincubation at room temperature does allow the detection of P. aeruginosa. This procedure offers the advantage that the workload
caused by use of subcultivation could be avoided. On the other hand,
there is no official recommendation for preincubation at room
temperature for the automatic blood culture system, and whether
fastidious organisms tolerate this procedure was not analyzed.
Although our facility is an on-site microbiology laboratory, blood
culture bottles are not received within a few minutes. Since we have no
automatic transportation system, the bottles are transported to the
laboratory by service personnel. The service personnel collect the
blood culture bottles from different wards and deliver them to the
laboratory. The time period between drawing of the blood sample and
arrival of the bottles in the laboratory ranges from 30 min to 12 h, and we think that this is not unusual for a hospital consisting of
around 1,200 beds. Intensive care units usually place blood culture
bottles in incubators until they are collected by service personnel.
The situation is worse during the night, when blood culture bottles are
incubated overnight within the clinic and are delivered to the
laboratory the following day. We are convinced that this is not a
unique situation and is encountered in many hospitals throughout the world.
A similar study was done at the Institute of Medical Microbiology and
Hygiene, University of Regensburg, Regensburg, Germany. This analysis
of the detection capacity of the BacT/Alert system showed that more
than 10% of all nonfermentative species remained undetected (N. Lehn,
personal communication). This hospital has the advantage of possessing
an automated transportation system, and we consider this a reason why
nonfermentative species were detected more often than in the study
presented here (Lehn, personal communication).
Recently, in a comparative study between the BACTEC 9240 (Becton
Dickinson, Heidelberg, Germany) and the BacT/Alert system, Ziegler et
al. reported 9 versus 14 false-negative blood cultures, respectively
(8). Although the overall false-negative rate for the
BacT/Alert was only 0.32% in that study, more than 64% of the
false-negative cultures were P. aeruginosa,
Acinetobacter spp., and S. maltophilia. The
false-negative rate for the BACTEC instrument was 0.20%, and only two
Pseudomonas isolates were included (22.2%). Regarding the
low false-negative rates, the authors concluded that blind terminal
subculture may not be necessary for both systems. Other investigators
showed high false-negative rates, up to 6%, and suggested that routine
subculturing of negative BACTEC cultures may be necessary (2,
3).
Our study showed that delayed entry of nonfermentative species into the
BacT/Alert system influences the detection capacity of the BacT/Alert
system. P. aeruginosa and A. baumannii were viable at each preincubation time at 36°C tested but remained undetected. These findings are confirmed by our clinical results, where
P. aeruginosa was undetected by the BacT/Alert system in 15 of 32 cases. However, the bacterial count of P. aeruginosa in all of these patients was high enough to cause symptomatic bacteremia; i.e., the patients had an elevated body temperature exceeding 38.5°C. Although without preincubation all
Enterobacteriaceae as well as nonfermentative species tested
were detected by the blood culture system, interestingly, the detection
times without preincubation were approximately fivefold longer for the
Enterobacteriaceae. In contrast, the BacT/Alert system was
able to detect the nonfermentative species when preincubation was not
done, but the time to detection was up to 9 h. These findings
support the contention that preincubation of blood culture bottles, in
terms of delay in entry into the BacT/Alert system, allows rapid
detection of Enterobacteriaceae but impedes the detection of
tested nonfermentative species, which of course are clinically relevant
bacteria in bloodstream infections.
In the present study, we also demonstrated that smaller inocula allow
the detection of nonfermentative species by the BacT/Alert system even
when preincubation is done. Only a 1:10,000 dilution of A. baumannii (150 CFU/ml) and 300 CFU of P. aeruginosa per ml resulted in small enough colony numbers for detection. Without preincubation each colony number tested was detected, but the times to
detection were significantly longer than those for the preincubated
bottles throughout the experiment. Both experiments demonstrated that
the preincubation time and the colony number of P. aeruginosa and A. baumannii in the blood culture bottle influence the detection capacity of the BacT/Alert system, although preincubation might influence the system more. However, the number of
CFU per milliliter blood culture used was significantly higher than the
number of CFU encountered during bacteremia or sepsis. Therefore, we
evaluated the possibility that blood might influence the detection
capacity of the BacT/Alert system, and we seeded low colony numbers of
P. aeruginosa in the presence of 10 ml of blood per bottle.
This experiment showed that there might be a threshold at 20 CFU/ml of
blood culture, above which the detection of P. aeruginosa is
impeded in the case of preincubation. This colony number would be equal
to a bacteremia of 100 CFU/ml, which is a reasonable number of bacteria
causing bloodstream infections (7). Although the number of
bacteria causing bloodstream infections might be lower (down to 30 CFU/ml), the bactericidal properties of the blood and early treatment
with antibiotics are two factors that can lower the actual bacterial
number found. These results show that routine preincubation of blood
cultures may not be the optimal maneuver for nonfermentative species,
which are often responsible for symptomatic septicemia in critically
ill patients.
Despite its mechanism of action being ill defined, incubation of blood
culture bottles at 36°C, in case they cannot be entered into the
blood culture system right away, is recommended and frequently used. We
interpreted our data to indicate that nonfermentative species, when
preincubated in a clinically relevant colony number, were viable but
remained undetected by the BacT/Alert system. The possibility that
insufficient O2 concentrations in the preincubated blood
culture bottles, which were vented at the time of inoculation, were
responsible for this phenomenon was not proven. A separate experiment
showed that repeating ventilation of the bottles prior to loading did
not influence the detection capacity of the BacT/Alert system (data not
shown). Analysis of the growth curves showed that bacterial growth
occurs to a certain extent, but the detection system of the BacT/Alert
computer, which is based on three types of measurements of the
CO2 level in the bottle, was not sensitive enough. The
threshold of 3,300 reflection units and the acceleration rate of 32 reflection units/10 min were not reached; therefore, no bacterial
growth was determined. We posit that nonfermentative species after
preincubation of clinically relevant bacterial numbers
especially P. aeruginosa and A. baumannii
may grow only
weakly and reach a steady-state concentration where CO2
acceleration occurs at a relatively low basis. Alternatively,
nonfermentative species might enter stationary-phase growth after the
phase of logarithmic expansion during preincubation and therefore can
not be detected in the laboratory. Although it is not ruled out, we
consider this unlikely, at least for the clinical specimens, since much
higher inocula of P. aeruginosa can be detected by the
automatic blood culture system. Even if clinical specimens are
preincubated for several hours, they do not match the bacterial numbers
corresponding to 1 ml of a McFarland 0.5 solution, but these numbers
are detected by the blood culture system. Although the studies
presented here did not investigate the exact mechanism, the results
provide indirect support for the hypothesis that an unknown factor in
the medium of the blood culture bottles is consumed during
preincubation and therefore nonfermentative bacteria do not grow.
Further examinations of the growth of the nonfermentative species and
its interaction with the medium should allow better definition of the
mechanism by which nonfermentative species remained undetected by the
BacT/Alert system.
In summary, we have shown that the BacT/Alert system failed to detect
three nonfermentative species in clinically relevant bacterial counts.
These bacteria, in particular P. aeruginosa, are well known
to cause septicemia, especially in critically ill patients.
Although the exact mechanism of the failure to detect these
microorganisms remains to be elucidated, we show that
preincubation of blood culture bottles after inoculation with these
species plays an important role. These findings emphasize the
importance of the rapid transportation of blood cultures to the
laboratory, especially in cases where P. aeruginosa is
expected. In the laboratory the bottles have to be loaded in the
BacT/Alert as soon as possible. Samples collected during the night or
from outlying hospitals may be held at room temperature until they are
loaded into the BacT/Alert (Organon Teknika, unpublished
communication). Bottles which have been preincubated for 4 h
or longer should be subcultured prior to loading to avoid false negatives.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: Institute
of Medical Microbiology, Immunology and Hygiene, Technical University
of Munich, Trogerstrasse 9, 81675 Munich, Germany. Phone: 49 89 4140-4187. Fax: 49 89 4140 4868. E-mail:
Thomas.Miethke{at}lrz.tu-muenchen.de.
 |
REFERENCES |
| 1.
|
Pittet, D.,
N. Li,
R. F. Woolson, and R. P. Wenzel.
1997.
Microbiological factors influencing the outcome of nosocomial bloodstream infections: a 6-year validated, population-based model.
Clin. Infect. Dis.
24:1068-1078[Medline].
|
| 2.
|
Shigei, J. T.,
J. A. Shimabukuro,
M. T. Pezzlo,
L. M. de la Maza, and E. M. Peterson.
1995.
Value of terminal subcultures for blood cultures monitored by BACTEC 9240.
J. Clin. Microbiol.
33:1385-1388[Abstract].
|
| 3.
|
Smith, J. A.,
E. A. Bryce,
J. H. Ngui-Yen, and F. J. Roberts.
1995.
Comparison of BACTEC 9240 and BacT/Alert blood culture systems in an adult hospital.
J. Clin. Microbiol.
33:1905-1908[Abstract].
|
| 4.
|
Washington, J. A., II, and D. M. Ilstrup.
1986.
Blood cultures: issues and controversies.
Rev. Infect. Dis.
8:792-802[Medline].
|
| 5.
|
Weinstein, M. P.
1996.
Current blood culture methods and systems: clinical concepts, technology, and interpretation of results.
Clin. Infect. Dis.
23:40-46[Medline].
|
| 6.
|
Weinstein, M. P.,
M. L. Towns,
S. M. Quartey,
S. Mirrett,
L. G. Reimer,
G. Parmigiani, and L. B. Reller.
1997.
The clinical significance of positive blood cultures in the 1990s: a prospective comprehensive evaluation of the microbiology, epidemiology, and outcome of bacteremia and fungemia in adults.
Clin. Infect. Dis.
24:584-602[Medline].
|
| 7.
|
Whimbey, E.,
B. Wong,
T. E. Kiehn, and D. Armstrong.
1984.
Clinical correlations of serial quantitative blood cultures determined by lysis-centrifugation in patients with persistent septicemia.
J. Clin. Microbiol.
19:766-771[Abstract/Free Full Text].
|
| 8.
|
Ziegler, R.,
I. Johnscher,
P. Martus,
D. Lenhardt, and H.-M. Just.
1998.
Controlled clinical laboratory comparison of two supplemented aerobic and anaerobic media used in automated blood culture systems to detect bloodstream infections.
J. Clin. Microbiol.
36:657-661[Abstract/Free Full Text].
|
Journal of Clinical Microbiology, March 2000, p. 1036-1041, Vol. 38, No. 3
0095-1137/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
This article has been cited by other articles:
-
Sautter, R. L., Bills, A. R., Lang, D. L., Ruschell, G., Heiter, B. J., Bourbeau, P. P.
(2006). Effects of Delayed-Entry Conditions on the Recovery and Detection of Microorganisms from BacT/ALERT and BACTEC Blood Culture Bottles. J. Clin. Microbiol.
44: 1245-1249
[Abstract]
[Full Text]
-
Seegmuller, I., Eschenbach, U., Kamereck, K., Miethke, T.
(2004). Sensitivity of the BacT/ALERT FA-medium for detection of Pseudomonas aeruginosa in pre-incubated blood cultures and its temperature-dependence. J Med Microbiol
53: 869-874
[Abstract]
[Full Text]
-
van der Vorm, E. R., Dondorp, A. M., van Ketel, R. J., Dankert, J.
(2000). Apparent Culture-Negative Prosthetic Valve Endocarditis Caused by Peptostreptococcus magnus. J. Clin. Microbiol.
38: 4640-4642
[Abstract]
[Full Text]
-
Meessen, N. E. L., de Vries-Hospers;, H. G., Miethke, T.
(2000). Failure of an Automated Blood Culture System To Detect Nonfermentative Gram-Negative Bacteria. J. Clin. Microbiol.
38: 2803-2804
[Full Text]