Previous Article | Next Article 
Journal of Clinical Microbiology, March 2000, p. 1113-1120, Vol. 38, No. 3
0095-1137/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
Quantitative and Cost Comparison of Ultrasensitive
Human Immunodeficiency Virus Type 1 RNA Viral Load Assays: Bayer
bDNA Quantiplex Versions 3.0 and 2.0 and Roche PCR Amplicor Monitor
Version 1.5
Tarek
Elbeik,1,2,*
Edwin
Charlebois,3
Patricia
Nassos,1,2
James
Kahn,3,4
Frederick M.
Hecht,3,4
David
Yajko,1,2
Valerie
Ng,1,2 and
Keith
Hadley1,2
Departments of Laboratory
Medicine1 and
Medicine,3 University of California, San
Francisco, and Clinical Laboratories2
and Positive Health Program,4 San
Francisco General Hospital, San Francisco, California
Received 15 July 1999/Returned for modification 22 October
1999/Accepted 24 December 1999
 |
ABSTRACT |
Quantification of human immunodeficiency virus type 1 (HIV-1) RNA
as a measure of viral load has greatly improved the monitoring of
therapies for infected individuals. With the significant reductions in
viral load now observed in individuals treated with highly active
anti-retroviral therapy (HAART), viral load assays have been adapted to
achieve greater sensitivity. Two commercially available ultrasensitive
assays, the Bayer Quantiplex HIV-1 bDNA version 3.0 (bDNA 3.0) assay
and the Roche Amplicor HIV-1 Monitor Ultrasensitive version 1.5 (Amplicor 1.5) assay, are now being used to monitor HIV-1-infected
individuals. Both of these ultrasensitive assays have a reported lower
limit of 50 HIV-1 RNA copies/ml and were developed from corresponding
older generation assays with lower limits of 400 to 500 copies/ml.
However, the comparability of viral load data generated by these
ultrasensitive assays and the relative costs of labor, disposables, and
biohazardous wastes were not determined in most cases. In this study,
we used matched clinical plasma samples to compare the quantification
of the newer bDNA 3.0 assay with that of the older bDNA 2.0 assay and
to compare the quantification and costs of the bDNA 3.0 assay and the
Amplicor 1.5 assay. We found that quantification by the bDNA 3.0 assay was approximately twofold higher than that by the bDNA 2.0 assay and
was highly correlated to that by the Amplicor 1.5 assay. Moreover, cost
analysis based on labor, disposables, and biohazardous wastes showed
significant savings with the bDNA 3.0 assay as compared to the costs of
the Amplicor 1.5 assay.
 |
INTRODUCTION |
Management of human immunodeficiency
virus type 1 (HIV-1)-infected individuals has greatly improved with the
introduction of quantitative viral load testing and highly active
anti-retroviral therapy (HAART). With successful HAART, viral load
measurements have dropped below the limit of detection of previously
available commercial assays (<400 to <500 HIV-1 RNA copies/ml)
(3, 5, 10, 12, 18). This degree of reduction in HIV-1 viral
load, and the general consensus that HAART therapy should aim to
suppress HIV-1 replication as fully as possible in order to attain
durable virologic responses, has prompted the need for even more
sensitive viral load quantification assays (2).
Consequently, a number of manufacturers, including Bayer Diagnostics
(formerly Chiron Diagnostics, Emeryville, Calif.) and Roche Molecular
Systems, Inc. (Somerville, N.J.), have adapted their existing viral
load assays to permit a lower limit of detection of 50 HIV-1 RNA
copies/ml.
Most HIV-1-infected patients are now monitored by ultrasensitive viral
load assays. However, it is unclear whether ultrasensitive assays,
which use different technologies, generate comparable quantitative
values. The comparability of viral load data is an important issue for
effective management of HIV-1-infected individuals since in many cases,
with no apparent choice, patient samples may be switched from one assay
to another at any given time as a result of insurance coverage or test
accessibility. This poses a significant problem since studies have
shown that the Quantiplex HIV-1 bDNA version 2.0 (bDNA 2.0) assay and
the Amplicor HIV-1 Monitor Standard version 1.0 (Amplicor 1.0) assay do
not generate comparable quantification results (1, 7, 9,
11). Another factor to consider in comparing the ultrasensitive
viral load assays is assay cost. In addition to the cost of assay
reagents, costs associated with labor, disposables, and biohazardous
waste should be considered when evaluating ultrasensitive viral load assays for routine use in the clinical laboratory.
This study compared HIV-1 viral load quantification and associated
costs of two commercially available ultrasensitive HIV-1 viral load
assays
Quantiplex HIV-1 bDNA version 3.0 assay (bDNA 3.0; Bayer
Diagnostics) and Amplicor HIV-1 Monitor Ultrasensitive version 1.5 assay (Amplicor 1.5; Roche Molecular Systems, Inc.). Using clinical
plasma samples, we evaluated the correlation and mean difference in
HIV-1 RNA copy number between the older and newer versions of the
Quantiplex HIV-RNA assay, bDNA 2.0 and bDNA 3.0, respectively, and
between bDNA 3.0 and Amplicor 1.5. We also conducted a cost analysis
based on labor, disposables, biohazardous wastes, and assay components
between bDNA 3.0 and Amplicor 1.5.
 |
MATERIALS AND METHODS |
Clinical plasma samples.
Excess uncharacterized plasma from
HIV-1-infected subjects remaining after routine testing by bDNA 3.0 (or
by bDNA 2.0 until July 1998) at the San Francisco General Hospital
(SFGH) Clinical Laboratory was used. Plasma from EDTA-anticoagulated
whole blood was processed and frozen to
80°C within 6 h
following blood draw. For routine testing, frozen plasma samples
(approximately 5-ml aliquots) were transferred from
80°C to cold
tap water for 10 to 15 min until thawed and then transferred to wet
ice. Plasma samples were maintained on wet ice for approximately 45 min
during preparation of aliquots for viral load testing. The excess
plasma was returned to
80°C until viral load testing was
successfully completed. Excess plasma samples then were transferred to
the University of California, San Francisco (UCSF) Clinical
Microbiology Research Laboratory and stored at
80°C for further analysis.
A total of 1,220 plasma samples, previously tested with bDNA 2.0 at the
SFGH Clinical Laboratory, were retested with bDNA 3.0 at both the UCSF
Clinical Microbiology Research Laboratory and the Bayer Reference
Testing Laboratory (BRTL; Emeryville, Calif.). These 1,220 samples were
selected to cover a wide viral load range (determined with bDNA 2.0) as
follows: 1,035 samples at <500 copies/ml, 64 at 500 to 10,000 copies/ml, 57 at 10,001 to 100,000 copies/ml), and 64 at >100,000
copies/ml.
A total of 159 additional uncharacterized clinical plasma samples,
previously tested with bDNA 3.0 by the SFGH Clinical Laboratory, were
tested again with bDNA 3.0 and the Amplicor 1.5 at the UCSF Clinical
Microbiology Research Laboratory. These samples were randomly selected
from approximately 500 banked samples that previously were collected
over 2 weeks and covered a wide viral load range (determined with bDNA
3.0) as follows: 48 samples at <50 copies/ml, 46 at 50 to 1,000 copies/ml, 33 at 1,001 to 10,000 copies/ml, and 32 at 10,001 to 72,942 copies/ml. These clinical samples were concurrently dispensed as 1.0- and 0.5-ml aliquots in 1.5-ml Sarstedt microcentrifuge tubes for bDNA
3.0 and Amplicor 1.5, respectively, and returned to
80°C. In this
way, samples for testing with both assays were handled in the same
manner and subjected to the same number of freeze thaw cycles (no more
than three).
Banked and characterized serial plasma samples were obtained from 24 individuals from the Options study at UCSF (F. M. Hecht, M. A. Chesney, M. P. Busch, B. D. Rawal, S. I. Staprans,
and J. O. Kahn, Abstr. 5th Conf. Retrovir. Opportun. Infect.,
abstr. 582, 1998; F. M. Hecht, B. D. Rawal, J. O. Kahn,
M. Swanson, M. A. Chesney, J. A. Levy, and P. Busch, Abstr.
6th Conf. Retrovir. Opportun. Infect., abstr. 178, 1999). These
individuals had been recently infected with HIV-1 (as documented by a
positive HIV-1 RNA and a negative or indeterminate HIV-1 antibody test
or documented HIV-1 antibody seroconversion within 6 months) and
initiated on HAART. The serial samples, which previously had been
tested with bDNA 2.0, included two baseline samples (collected
immediately prior to HAART) as well as samples collected at 4- to
8-week intervals from each subject. Stored specimens were retested with
bDNA 3.0 at the UCSF Clinical Microbiology Research Laboratory.
Committee on human research.
This study was performed in
accordance with the guidelines of the UCSF committee on human research.
In order to preserve patient confidentiality, specimens were unlinked
to information such as date, patient identifier, patient name, and
health center identification.
Study sites and technical training.
All comparative analyses
between bDNA 2.0 and bDNA 3.0 were performed at the SFGH Clinical
Laboratory and the BRTL. All viral load results and cost analyses
(labor, disposables, and biohazardous waste) for comparison between
bDNA 3.0 and Amplicor 1.5 were generated at the UCSF Clinical
Microbiology Research Laboratory at San Francisco General Hospital.
One operator performed all viral load determinations for the study
comparing bDNA 3.0 to Amplicor 1.5. He successfully completed all
training requirements by Roche Diagnostics (Branchburg, N.J.) and Bayer
Diagnostics (on site) and was issued certificates of training from both
companies. He then was retrained at the study evaluation site at San
Francisco General Hospital.
Amplicor 1.5 assay.
The Amplicor 1.5 assay is based on
reverse transcriptase (RT)-PCR, with a reported dynamic range from 50 to 75,000 HIV-1 RNA copies/ml (15). This assay has been
shown to be equivalent to the Amplicor HIV-1 Monitor version 1.0 assay
(Amplicor 1.0) (17), which has been approved by the United
States Federal Drug Administration (FDA). The Amplicor 1.5 assay was
performed according to manufacturer's instructions, and all samples
were run singly.
Samples were processed for the Amplicor 1.5 assay following a
three-step work flow: (i) specimen preparation (including
centrifugation and aspiration of clarified plasma, viral lysis, RNA
precipitation, RNA washing, and suspension of purified RNA in buffer),
(ii) amplification, and (iii) detection. According to the training
provided by Roche Diagnostics (Branchburg, N.J.), the Amplicor 1.5 assay (like the Amplicor 1.0 assay) required four separated areas
dedicated to sample processing (HIV-1 concentration, RNA extraction and
purification, and combination of RNA and master mix solution), master
mix preparation (RT-PCR reagent preparation), amplification (thermal
cycling and reaction termination), and detection (plate hybridization,
plate incubation, plate wash, detection). The Amplicor 1.5 assay setup was strictly unidirectional, and the operator was required to change
gloves and gown when entering each work area.
bDNA 2.0 and 3.0 assays.
The bDNA 2.0 and 3.0 assays are
based on branched DNA (bDNA) signal amplification technology. Both
assays utilize synthetic oligonucleotide probes (capture probes, target
probes, preamplifier probes, amplifier probes, and label probes) to
generate signal amplification through hybridization. In the earlier
version, bDNA 2.0, nonspecific hybridization between these probes
increased background noise, which in turn limited the sensitivity of
the assay. In the present version, bDNA 3.0, the nonnatural bases isocytidine (isoC) and isoguanidine (isoG) are incorporated in all
probes except those that hybridize to the target HIV-1 RNA (capture
probes and target probes). Since isoC and isoG do not hybridize with
natural bases, the nonspecific hybridization between probes containing
these nonnatural bases is greatly reduced. With the incorporation of
isoC and isoG to reduce background noise and the development of a new
target probe set and optimized buffer conditions to increase signal
generation capacity, the bDNA 3.0 assay achieves a lower detection
limit of 50 copies/ml (4). The bDNA 2.0 and 3.0 assays were
performed according to manufacturer's instructions. All samples for
bDNA 2.0 were run in duplicate, whereas all samples for bDNA 3.0 were
run singly (4, 8). The second-generation bDNA 2.0 assay is
performed manually and has a dynamic range of 500 to 800,000 HIV-1 RNA
copies/ml (8, 16). The bDNA 3.0 assay was used in
conjunction with the semi-automated Quantiplex 340 system and has a
dynamic range of 50 to 500,000 HIV-1 RNA copies/ml (4).
Samples were processed for the bDNA 3.0 assay following a three-step
work flow: (i) specimen preparation (including centrifugation and
aspiration of clarified plasma), (ii) hybridization, and (iii) detection. The bDNA 3.0 assay required two work areas
one area for
sample processing (HIV-1 concentration and viral lysis) and one area
for hybridization and detection (heating block for hybridization and
Quantiplex 340 for additional hybridization steps, washing, and
detection). This assay allowed for bidirectional movement during assay
performance, and the operator could wear the same gown and gloves
between both sites for all but the final detection step, which required
a new set of gloves.
Comparative cost analysis.
For comparative analysis of costs
of disposables and labor, the routine assay procedures were followed
for both the bDNA 3.0 and the Amplicor 1.5 assays, as described in the
package inserts. Costs for disposables were based on list prices as of
October 1998 from various vendors in the United States. No discounted pricing was used. For the assessment of labor, each procedure was timed
from the start to the end. Except for the bDNA 3.0 overnight incubation
and the Amplicor 1.5 thermal cycler step, all other "hands off"
intervals that required the presence of the operator (i.e., short
incubation steps) were included in this analysis. Since technical steps
within each procedure were staggered, there was little hands off time
to permit the operator to attend to unrelated projects.
The Amplicor 1.5 assay was performed in one run with 14 plates per run
(yielding 169 tests, or 159 reportable results). Costs of disposables
for the Amplicor 1.5 assay were determined for each of the three
workflow steps. Costs of sample preparation disposables were calculated
for the following: 182 aerosolized 200-µl tips ($80/960), 672 aerosolized 1,000-µl tips ($80/720), 672 fine-tip transfer pipets
($57/500), 14 5-ml pipets ($64/200), 14 10-ml pipets ($67/200), 168 1.5-ml screw-cap tubes ($235/1000), 14 50-ml tubes ($78/200), 154 ml of
reagent-grade ethanol ($90/4 liters), 101 ml of 2-propanol ($67/4
liters), and 12 pairs of gloves ($65/50 pair). Costs for disposables
for amplification were calculated for the following: 350 aerosolized
200-µl tips ($80/960), 14 25-µl Eppendorf dispensers ($112/100), 2 25-ml reagent reservoirs ($58/100), 168 caps ($75/2,400), 168 reaction
tubes ($195/2,000), and 12 pairs of gloves ($65/50 pair). Costs for disposables for detection were calculated for the following: 1,008 aerosolized 200-µl tips ($80/960), 56 25-ml reagent reservoirs ($58/100), 14 5-ml pipets ($64/200), 14 10-ml pipets ($67/200), 14 50-ml tubes ($78/200), and 7 pairs of gloves ($65/50 pair).
The bDNA 3.0 assay was performed either as two runs of one plate per
run (yielding 192 tests, or 168 reportable results) or as one run of
two plates per run (yielding 192 tests, or 168 reportable results).
Costs of disposables for the bDNA 3.0 assay were determined for each of
these two scenarios according to the three workflow steps. For the bDNA
3.0 assay performed in two runs of one plate per run, costs of sample
preparation disposables were calculated for the following: 192 aerosolized 1,000-µl tips ($90/800), 200 nonaerosolized 200-µl tips
($61/960), 2 25-µl Eppendorf dispensers ($112/100), 192 1.5-ml
screw-cap tubes ($235/1,000), and 8 pairs of gloves ($65/50 pair).
Costs of disposables for amplification (bDNA 3.0 in two runs, one plate
per run) were calculated for the following: 388 aerosolized 200-µl
tips ($90/960), 2 2-ml pipets ($125/500), 2 10-ml pipets ($67/200), 2 50-ml tubes ($78/200), and 4 pairs of gloves ($65/50 pair). Costs of
disposables for detection (bDNA 3.0 in two runs, one plate per run)
were calculated for the following: 102 aerosolized 200-µl tips
($90/960), 2 aerosolized 1000-µl tips ($90/800), 2 100-µl Eppendorf
dispensers ($100.5/100), 8 55-ml reagent reservoirs ($96/80), 8 10-ml
pipets ($67/200), 8 50-ml tubes ($78/200), and 8 pairs of gloves
($65/50 pair).
For the bDNA 3.0 assay performed in one run of two plates per run,
costs of sample preparation disposables were calculated for the
following: 192 aerosolized 1,000-µl tips ($90/800), 200 nonaerosolized 200-µl tips ($61/960), 1 25-µl Eppendorf dispenser ($112/100), 192 1.5-ml screw-cap tubes ($235/1,000), and 8 pairs of
gloves ($65/50 pair). Costs of disposables for amplification (bDNA in
one run, two plates per run) were calculated for the following: 388 aerosolized 200-µl tips ($90/960), 1 2-ml pipet ($125/500), 1 10-ml
pipet ($67/200), 1 50-ml tube ($78/200), and 4 pairs of gloves ($65/50
pair). Costs of disposables for detection (bDNA in one run, two plates
per run) were calculated for the following: 51 aerosolized 200-µl
tips ($90/960), 1 aerosolized 1,000-µl tips ($90/800), 1 100-µl
Eppendorf dispenser ($110.50/100), 4 55-ml reagent reservoirs ($96/80),
4 10-ml pipets ($67/200), 4 50-ml tubes ($78/200), and 4 pairs of
gloves ($65/50 pair).
Statistical analysis.
For the purpose of analysis, samples
reported to be below the level of detection for a particular assay were
assigned HIV-1 RNA copy number values of one-half of the cutoff level
for that assay. Similarly, samples reported to be above the dynamic
range of an assay were assigned copy number values of one plus the
highest value of the dynamic range of the assay. In comparative
analyses, the lower cutoff value of the two assays was used for both
sets of data. Correlation and linear regression analyses were performed on log10-transformed HIV-1 RNA copy numbers. All analyses
were carried out with the SAS Statistical Analyses package (SAS
Institute, Cary, N.C.).
For the correlation analyses, both the Pearson correlation coefficient
and the nonparametric Spearman rank correlation coefficient, along with
their respective 95% confidence intervals, were calculated. Standard
statistical methods were used to test for significant correlation
(6). Linear regression models were created with least-squares linear regression on log10-transformed HIV-1
RNA copy numbers with the SAS regression procedure, allowing for an intercept term. Significance of the overall model was assessed with the
F statistic, and individual terms in the model were tested with the
t statistic.
 |
RESULTS |
Quantification comparison: bDNA 2.0 versus bDNA 3.0.
The
agreement between quantification values by the bDNA 2.0 and bDNA 3.0 assays is shown in Fig. 1. The
quantification values obtained with these assays were significantly
correlated with one another, as indicated by the Pearson correlation
coefficient (r value) of 0.982 (95% confidence interval,
0.976 to 0.986; P < 0.0001).

View larger version (30K):
[in this window]
[in a new window]
|
FIG. 1.
Results of HIV-1 RNA quantification of 185 samples by
the older bDNA 2.0 assay and the more sensitive bDNA 3.0 assay. The
solid line represents the identity line, where all determinations
should fall if a perfect correlation between the two assays was
achieved.
|
|
The distribution of differences between bDNA 2.0 and bDNA 3.0 results
over the range of samples tested is shown in Table
1. Overall, HIV-1 RNA copy numbers were
0.239 log10 higher for bDNA 3.0 as compared to those of the
bDNA 2.0 assay; however, this difference varied over the range of
samples tested. For example, samples with fewer than 10,000 HIV RNA
copies/ml by bDNA 3.0 showed a significantly higher shift from bDNA 2.0 results than did samples with >10,000 HIV RNA copies/ml. Table 1 also
shows the results of the linear regression of the
log10-transformed copy numbers with bDNA 3.0 results as the
dependent variable and bDNA 2.0 results as the independent variable
(i.e., prediction of bDNA 3.0 results from bDNA 2.0 results). This
analysis indicated a strong linear relationship between the results
given by the two assays over the range of samples tested (r = 0.98). Within the assay ranges tested, the <10,000 HIV-1 RNA
copies/ml range showed a weaker relationship between assay values than
did the two ranges above 10,000 HIV-1 RNA copies/ml (r = 0.91 and r = 0.90, respectively).
View this table:
[in this window]
[in a new window]
|
TABLE 1.
Relationship between log10-transformed
quantification values obtained with the bDNA 2.0 and bDNA
3.0 assaysa
|
|
To further assess differences in quantification values for samples
tested with bDNA 2.0, an additional 1,035 clinical samples below the
detection limit of the bDNA 2.0 assay (<500 HIV-1 RNA copies/ml) were
retested with the bDNA 3.0 assay at the BRTL. As shown in Table
2, approximately 80% of samples
initially reported by bDNA 2.0 to have <500 copies/ml were confirmed
to have <500 copies/ml by bDNA 3.0 (50.7% had <50 copies/ml, 29.6%
had 51 to 500 copies/ml). The remaining samples were quantified above
500 copies/ml by bDNA 3.0 (9.8% had 501 to 1,000 copies/ml, 8.6% had 1,001 to 2,000 copies/ml, 1.2% had 2,001 to 3,000 copies/ml, 0.1% had
>3,001 copies/ml).
View this table:
[in this window]
[in a new window]
|
TABLE 2.
bDNA 3.0 assay results for 1,035 clinical plasma samples
previously quantified as <500 copies/ml by bDNA 2.0
|
|
A comparison of bDNA 2.0 and bDNA 3.0 results for serial samples from
24 individuals newly infected with HIV-1 is shown in Fig.
2. When serial samples were tested with
bDNA 2.0, the average time for viral load to be reduced to
nondetectable levels (<500 copies/ml) was approximately 4 weeks after
the initiation of HAART. However, when matched samples were retested
with the more sensitive bDNA 3.0 assay, additional viral load data were
generated. Using the bDNA 3.0 assay, the average time for the viral
load to be reduced to nondetectable levels (<50 copies/ml) was
approximately 24 weeks after the initiation of HAART.

View larger version (34K):
[in this window]
[in a new window]
|
FIG. 2.
Viral load results for 24 HIV-infected patients
monitored for up to 36 weeks as measured with bDNA 2.0 (A) and bDNA 3.0 (B). The horizontal lines indicate the lower limits of detection for
the bDNA 2.0 and bDNA 3.0 assays at 500 and 50 copies/ml, respectively.
Data points falling below the limit of detection are plotted at half
the limit of detection for each assay. The shaded area indicates the
difference between the detection limits of the bDNA 2.0 and bDNA 3.0 assays.
|
|
Quantification comparison: bDNA 3.0 versus Amplicor 1.5.
The
agreement between quantification values measured by the bDNA 3.0 and
Amplicor 1.5 assays is shown in Fig. 3.
The quantification values obtained with these assays were significantly
correlated with one another, as indicated by the Pearson correlation
coefficient of 0.984 (95% confidence interval, 0.978 to 0.984;
P < 0.0001).

View larger version (25K):
[in this window]
[in a new window]
|
FIG. 3.
Results of HIV-1 RNA quantification of 159 samples by
the bDNA 3.0 assay based on bDNA signal amplification technology and
the PCR-based Amplicor 1.5 assay. The solid line represents the
identity line, where all determinations should fall if a perfect
correlation between the two assays was achieved.
|
|
The correlation and mean difference between bDNA 3.0 and Amplicor 1.5 results are shown in Table 3. For samples
above 10,000 HIV-1 RNA copies/ml, the two assays were approximately
90% correlated, and the mean difference in log10 copy
number between the two assays was significantly different from zero.
For samples with 10,000 or fewer HIV-1 RNA copies/ml, the correlation
between bDNA 3.0 and Amplicor 1.5 results was less strong
(approximately 75% correlated), and the mean difference in
log10 copy number between the two assays was not
significantly different from zero.
View this table:
[in this window]
[in a new window]
|
TABLE 3.
Relationship between log10-transformed
quantification values obtained with the bDNA 3.0 and Amplicor
1.5 assaysa
|
|
When the results of bDNA 3.0 and Amplicor 1.5 were dichotomized at 50 HIV-1 RNA copies/ml, there was excellent agreement between the two
assays for which samples contained less than 50 copies/ml and those
that contained more than 50 copies/ml or more (Table 4). Of the 159 samples tested, the two
assays disagreed on only eight
two samples reported by Amplicor 1.5 as
having <50 copies/ml were reported by bDNA 3.0 as having >50
copies/ml, and six samples reported by Amplicor 1.5 as having >50
copies/ml were reported by bDNA 3.0 as having <50 copies/ml. Moreover,
relatively low viral levels were measured for the discrepant samples.
The six samples reported by bDNA 3.0 as having <50 copies/ml were
quantified by Amplicor 1.5 at 61, 62, 74, 109, 125, and 155 copies/ml,
while the two samples reported by Amplicor 1.5 as having <50 copies/ml were quantified by bDNA 3.0 at 57 and 77 copies/ml.
Assay component cost comparison: bDNA 3.0 versus Amplicor 1.5.
The cost of assay components also were compared for bDNA 3.0 and
Amplicor 1.5. As of November 1999, the list price for Amplicor 1.5 was
$2,500 per kit (U.S. pricing; Roche Diagnostics). Each Amplicor kit
contains two 96-well plates that are designed to test 21 clinical
samples and 3 controls, for a total of 24 tests per kit. As such, the
cost of assay components for Amplicor 1.5 was $119.00 per clinical
sample, or $104.17 per test (which includes controls). By comparison,
the list price for bDNA 3.0 was $10,500 per kit (U.S. pricing; Bayer
Diagnostics). Each bDNA 3.0 kit contains one 96-well plate that is
designed to test 84 clinical samples, 9 standards, and 3 controls, for
a total of 96 tests per kit. As such, the cost of assay components for
bDNA 3.0 was $125.00 per clinical sample, or $109.38 per test (which
includes standards and controls). Hence, the cost of assay components
is approximately 5% higher for bDNA 3.0 than for Amplicor 1.5. These
calculations assume that all wells are used for each assay run and thus
indicate the lowest per-sample or per-test costs for assay components. If fewer samples are included in each run (i.e., not all wells are
used), then the cost of assay components would be higher. Also, it is
important to note that these calculations are based on the U.S. list
prices for each of the kits and that any reduced pricing that may be
offered by the manufacturers would affect the per-sample or per-test
cost of assay components.
Overall cost comparison: bDNA 3.0 versus Amplicor 1.5.
The
overall costs for the Amplicor 1.5 and bDNA 3.0 are shown in Table
5. Costs for Amplicor 1.5 were evaluated
for the assay performed as a single run consisting of 14 plates (total
of 168 tests: 159 clinical samples and 9 controls). Costs for bDNA 3.0 were evaluated both for the assay performed in a single run consisting of two plates run simultaneously (total of 192 tests: 168 clinical samples and 24 controls) as well as a two separate runs consisting of
one plate per run (total of 192 tests: 168 clinical samples and 24 controls). In Table 5, the cost of the assay kits were determined
according to the U.S. list price (as of November 1999), although kit
prices may be reduced according to discounted pricing offered by the
manufacturer. Labor costs are noted in terms of time (minutes) since
different sites may have different pay schedules for technicians.
The costs of disposables and labor for the two assays were compared for
each of the three work flow steps: sample preparation, amplification/hybridization, and detection. The highest costs of
disposables (not including assay components) were incurred in the
sample preparation step for both assays. Labor time was highest during
the sample preparation step for Amplicor 1.5. Labor time for the bDNA
3.0 assay was fairly evenly distributed throughout all three work flow
steps when the assay was performed in a single run with two plates, but
was somewhat higher for the detection step when the assay was performed
in two runs at one plate per run.
The lowest costs (not including assay components) were incurred with
bDNA 3.0. Per-test costs for the bDNA 3.0 assay performed in two runs
were significantly less (63% lower costs for disposables, 30% less
labor, 50% less biohazardous waste) than those for Amplicor 1.5. A
further overall reduction in disposables, labor, and biohazard waste
for the bDNA 3.0 assay to less than half that of Amplicor 1.5 was
realized when the bDNA 3.0 assay was performed in a single run (two
plates run simultaneously, containing 24 controls and 168 clinical samples).
 |
DISCUSSION |
With the introduction of newer, more effective therapies such as
HAART that can greatly reduce viral load to below 400 to 500 copies/ml,
more laboratories are using ultrasensitive assays with lower detection
limits of 50 copies/ml to monitor HIV-1-infected individuals. Since it
is not uncommon for patients to have longitudinal viral load data from
more than one assay or from earlier versions of the same assay, it is
important to understand whether these ultrasensitive assays yield
comparable results. Concerns about the comparability of viral load data
have been supported by earlier studies. For example, in the Stadi
trial, in which the combination of stavudine and didanosine was
clinically evaluated, viral quantification obtained with the Amplicor
1.0 assay (standard and ultrasensitive) was significantly higher than
that obtained with the bDNA 2.0 assay (14). Other studies
also have showed discordance in viral load quantification by the
Amplicor 1.0 assay as compared to that of the bDNA 2.0 assay (7,
9, 11). The observed discordance between the results of different
assays has led to the recommendation that patients be monitored with a
single manufacturer's assay (7, 14). Although use of an
external standard can negate differences in the copy number estimates
made by different assays (1), external standards are not
routinely used in clinical practice. Therefore, it is important to
understand the comparability of viral load data generated by the
different assays in order to avoid complications in patient monitoring
should assays be changed. Moreover, it is important to document the
comparative costs of the different assays in order to give laboratories
the information needed to select the most cost-effective system. Not only should the cost of assay components be considered but also the
costs incurred for disposables, labor, and biohazardous waste. Assay
cost considerations are particularly important in these times in which
laboratories may be required to cut costs as a result of lower
operating budgets and lower reimbursements.
In this study, we compared the quantification and costs of the most
recent versions of commercially available ultrasensitive viral load
assays
Amplicor 1.5 and bDNA 3.0. The two manufacturers, Roche
Diagnostics and Bayer Diagnostics, respectively, have used different
approaches in adapting their assays to address the current clinical
need for more sensitive viral load quantification. The older Amplicor
1.0 assay (including both the standard and ultrasensitive extraction
procedures) has been approved by the FDA and thus is the "gold
standard" assay by which all other HIV-1 RNA assays must favorably
compare in terms of quantification. The newer Amplicor 1.5 assay
incorporates primer pairs designed to detect all HIV-1 clades
(17) and has replaced version 1.0 in most labs worldwide, except for those in North America. Studies have shown that the Amplicor
HIV-1 Monitor versions 1.5 and 1.0 assays generate comparable quantification on matched clinical samples using the standard and
ultrasensitive procedures (17), and as such fall in line with the FDA-approved product. In order to achieve enhanced sensitivity with the Amplicor assay, Roche Diagnostics modified the specimen preparation procedure to allow greater input of RNA from 10-fold more
plasma. Whereas the standard procedure for the Amplicor 1.0 or 1.5 assay requires direct lysis of 0.2 ml of plasma and resuspension of RNA
in 0.4 ml of buffer, the ultrasensitive procedure for these assays
requires ultracentrifugation of 0.5 ml of plasma, lysis of the viral
pellet, and resuspension of RNA in 0.1 ml of buffer. With the
incorporation of this high-speed centrifugation step to concentrate the
virus, the lower detection limit of the Amplicor assay drops from 400 to 50 copies/ml (15). By contrast, the sensitivity of the
bDNA assay was improved by modifying the chemistry of the assay. In
addition to developing a new target probe set and optimizing buffer
conditions to increase signal generation capacity, the nonnatural bases
isoC and isoG were incorporated into the binding regions of probes to
reduce nonspecific hybridization and thus lower background noise. With
these modifications to the assay chemistry, the bDNA 3.0 assay achieves
a lower detection limit of 50 copies/ml (4).
This study showed a statistically significant correlation between the
bDNA 3.0 and the Amplicor 1.5 assays over a wide range of viral load
(50 to 75,000 copies/ml) when tested with matched samples. Also, this
study showed that the lower limit of detection was equivalent for both
assays when tested on matched clinical samples. Discordant samples from
both assays had viral loads within the lower end of the dynamic range
(61 to 155 copies/ml) and were well within the reported coefficient of
variation for the assays. Assay performance in this study was within
acceptable limits since the standard deviations for the operator, the
assays, and the environment in which this study was conducted were
within assay specifications (within threefold, or less than a 0.5-log
difference) based upon controls. Moreover, the performance of the
operator was evaluated and approved by both companies by using
certified test panels so as to rule out potential bias against either
assay. Hence, viral load data generated by either assay yielded
comparable results.
The close agreement between the more recent versions of the Amplicor
and bDNA assays observed in this study, as compared to the findings of
earlier studies of older assay versions (7, 11) may be
explained, at least in part, by quantification differences between the
2.0 versus 3.0 versions of the bDNA assay. We found a two- to sevenfold
difference in quantification between bDNA 2.0 and 3.0 for samples
containing above 500 HIV-1 RNA copies/ml, with the greatest difference
occurring at the lower end of the bDNA 2.0 dynamic range (<10,000
HIV-1 RNA copies/ml). It is perhaps not surprising that this anomaly
was not noted earlier, since bDNA 2.0 was introduced before the
introduction of HAART and thus prior to the significant reductions in
viral load now commonly observed. Indeed, discrepancies between bDNA
2.0 and Amplicor 1.0 were noticed only after viral levels in the
majority of HAART-treated individuals began to decline. Our study also
showed that the use of a more sensitive viral load assay in monitoring
serial samples from recently infected individuals beginning HAART
provided at least an additional 20 weeks of detectable viral load as
compared to the older, less sensitive assay. This added viral load
information improves the management of infected individuals since it
provides critical information on the efficacy of HAART and helps to
rapidly identify the emergence of viral breakthrough in the absence of genotypic or phenotypic analysis.
Given the comparability of assay results, other factors, such as assay
components, labor, disposables, and biohazardous waste, may influence
the selection of a viral load assay in a clinical setting. Based on
U.S. list prices as of November 1999, the per-test cost for the bDNA
3.0 assay was approximately 5% higher than that of the Amplicor 1.5 assay. However, our study demonstrated a significant savings in labor
(~30 to 53%), disposables (~63 to 67%), and biohazardous waste
(~50%) for the bDNA 3.0 assay as compared to costs of the Amplicor
1.5 assay. It is important to consider that although manufacturers may
offer various discounted price plans for kits based on the number of
kits purchased, these discounts may not offset the additional hidden
costs incurred for labor, disposables, and biohazardous waste.
Discounts in the prices of disposables would lead to additional cost
savings, but these discounts would be realized for both assays, and the
relative difference in the cost of disposables must still be taken into
account. The time involved in labor also differed between the two
assays. The bDNA 3.0 assay required fewer steps and involved less
repeat pipetting and lower complexity (i.e., RNA extraction) than the
Amplicor 1.5 assay. The most significant difference in labor between
these assays was the sample preparation step
only one manipulation, lysis of the pellet, is required for the bDNA 3.0 assay, whereas Amplicor 1.5 requires four manipulations
one for the viral pellet and
three for the RNA pellet. The complexity of the sample preparation step
for the Amplicor 1.5 assay adds significantly to the cost of
disposables and labor, although automation of this step should significantly reduce the labor component. The other steps (i.e., amplification/hybridization and detection) for bDNA 3.0 and Amplicor 1.5 are easily performed and require relatively minimal labor. Use of
the fully automated COBAS system for Amplicor 1.5 will simplify the
amplification/hybridization steps. Thus, in selecting the most
cost-effective approach, laboratories must consider not only the price
of assay components but also all of the costs incurred in HIV-1 viral
load testing.
In addition to the costs associated with viral load testing described
in this study (assay components, labor, disposables, and biohazardous
waste), HIV-1 viral load testing laboratories need to consider the
dynamic ranges of the assays for additional potential costs savings.
For example, the two specimen preparation procedures of the Amplicor
1.0 and 1.5 assays allow for dynamic ranges of 50 to 75,000 copies/ml
and 400 to 750,000 copies/ml for the ultrasensitive and standard
procedures, respectively. However, additional costs for viral load
testing may result from the need to reflex from one specimen
preparation procedure to another. This issue was addressed in a recent
study (13) in which it was proposed that, depending on the
virologic response, the standard procedure, the ultrasensitive
procedure, or a combination of both procedures could be used as the
most cost-effective strategy for the Amplicor 1.0 assay. Whereas two
extraction procedures are needed to realize the full dynamic range of
the Amplicor assay, this approach is not necessary with the bDNA 3.0 assay since the dynamic range of this assay is from 50 to 500,000 copies/ml.
Different individuals working in the management of HIV-1 infection may
be impacted differently by the findings of this study on the
comparability of viral load data and the relative costs of the bDNA 3.0 and Amplicor 1.5 assays. For the health care worker, the correlation
between the quantification values of the bDNA 3.0 and Amplicor 1.5 assays may provide more flexibility in monitoring patients should
patients be switched from one assay to another since longitudinal viral
load data collected with either or both assays should still be
meaningful. For the laboratory director, the availability of two
commercial ultrasensitive assays that are comparable allows greater
flexibility in selecting the most cost-effective approach for viral
load testing. It is essential that comparative studies continue to be
performed as new assays for the management of HIV-1 infection are
introduced. Moreover, such studies should be supported and encouraged
by the manufacturers. As such, any claims can rapidly be confirmed by
independent groups in a timely manner.
 |
ACKNOWLEDGMENTS |
We thank Roche Molecular Systems, Roche Diagnostics, and Bayer
Diagnostics for providing kits and disposables.
Our sincere thanks to Lynette Sawyer at the Bayer Reference Testing
Laboratory for providing HIV-1 viral load results on clinical samples
used for this study. We also greatly thank Linda Wuestehube for
editorial assistance.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: San Francisco
General Hospital, Department of Laboratory Medicine, 1001 Potrero Ave., NH, Room 2M35, San Francisco, CA 94110. Phone: (415) 476-4604. Fax:
(415) 206-3045. E-mail: elbeik{at}itsa.ucsf.edu.
 |
REFERENCES |
| 1.
|
Brambilla, D.,
S. Leung,
J. Lew,
J. Todd,
S. Herman,
M. Cronin,
D. E. Shapiro,
J. Bremer,
C. Hanson,
G. V. Hillyer,
G. D. McSherry,
R. S. Sperling,
R. W. Coombs, and P. S. Reichelderfer.
1998.
Absolute copy number and relative change in determinations of human immunodeficiency virus type 1 RNA in plasma: effect of an external standard on kit comparisons.
J. Clin. Microbiol.
36:311-314[Abstract/Free Full Text].
|
| 2.
|
Carpenter, C. C.,
M. A. Fischl,
S. M. Hammer,
M. S. Hirsch,
D. M. Jacobsen,
D. A. Katzenstein,
J. S. Montaner,
D. D. Richman,
M. S. Saag,
R. T. Schooley,
M. A. Thompson,
S. Vella,
P. G. Yeni, and P. A. Volberding.
1998.
Antiretroviral therapy for HIV infection in 1998: updated recommendations of the International AIDS Society-USA Panel.
JAMA
280:78-86[Abstract/Free Full Text].
|
| 3.
|
Chun, T. W.,
L. Stuyver,
S. B. Mizell,
L. A. Ehler,
J. A. Mican,
M. Baseler,
A. L. Lloyd,
M. A. Nowak, and A. S. Fauci.
1997.
Presence of an inducible HIV-1 latent reservoir during highly active antiretroviral therapy.
Proc. Natl. Acad. Sci. USA
94:13193-13197[Abstract/Free Full Text].
|
| 4.
|
Collins, M. L.,
B. Irvine,
D. Tyner,
E. Fine,
C. Zayati,
C.-A. Chang,
T. Horn,
D. Ahle,
J. Detmer,
L.-P. Shen,
J. Kolberg,
S. Bushnell,
M. S. Urdea, and D. D. Ho.
1997.
A branched DNA signal amplification assay for quantification of nucleic acid targets below 100 molecules/ml.
Nucleic Acids Res.
25:2979-2984[Abstract/Free Full Text].
|
| 5.
|
Finzi, D.,
M. Hermankova,
T. Pierson,
L. M. Carruth,
C. Buck,
R. E. Chaisson,
T. C. Quinn,
K. Chadwick,
J. Margolick,
R. Brookmeyer,
J. Gallant,
M. Markowitz,
D. D. Ho,
D. D. Richman, and R. F. Siliciano.
1997.
Identification of a reservoir for HIV-1 in patients on highly active antiretroviral therapy.
Science
278:1295-1300[Abstract/Free Full Text].
|
| 6.
|
Fleiss, J. L.
1981.
Statistical methods for rates and proportions, vol. xviii. , p. 321.
Wiley, New York, N.Y.
|
| 7.
|
Hodara, V.,
A. Monticelli,
S. Pampuro,
H. Salomón,
H. Jauregui Rueda, and O. Libonatti.
1998.
HIV-1 viral load: comparative evaluation of three commercially available assays in Argentina.
Acta Physiol. Pharmacol. Ther. Latinoam.
48:107-113[Medline].
|
| 8.
|
Kern, D.,
M. Collins,
T. Fultz,
J. Detmer,
S. Hamren,
J. J. Peterkin,
P. Sheridan,
M. Urdea,
R. White,
T. Yeghiazarian, and J. Todd.
1996.
An enhanced-sensitivity branched-DNA assay for quantification of human immunodeficiency virus type 1 RNA in plasma.
J. Clin. Microbiol.
34:3196-3202[Abstract].
|
| 9.
|
Lin, H. J.,
L. Pedneault, and F. B. Hollinger.
1998.
Intra-assay performance characteristics of five assays for quantification of human immunodeficiency virus type 1 RNA in plasma.
J. Clin. Microbiol.
36:835-839[Abstract/Free Full Text].
|
| 10.
|
Natarajan, V.,
M. Bosche,
J. A. Metcalf,
D. J. Ward,
H. C. Lane, and J. A. Kovacs.
1999.
HIV-1 replication in patients with undetectable plasma virus receiving HAART.
Lancet
353:119-120[Medline].
|
| 11.
|
Nolte, F. S.,
J. Boysza,
C. Thurmond,
W. S. Clark, and J. L. Lennox.
1998.
Clinical comparison of an enhanced-sensitivity branched-DNA assay and reverse transcription-PCR for quantitation of human immunodeficiency virus type 1 RNA in plasma.
J. Clin. Microbiol.
36:716-720[Abstract/Free Full Text].
|
| 12.
|
Notermans, D. W.,
S. Jurriaans,
F. de Wolf,
N. A. Foudraine,
J. J. de Jong,
W. Cavert,
C. M. Schuwirth,
R. H. Kauffmann,
P. L. Meenhorst,
H. McDade,
C. Goodwin,
J. M. Leonard,
J. Goudsmit, and S. A. Danner.
1998.
Decrease of HIV-1 RNA levels in lymphoid tissue and peripheral blood during treatment with ritonavir, lamivudine and zidovudine. Ritonavir/3TC/ZDV Study Group.
AIDS
12:167-173[CrossRef][Medline].
|
| 13.
|
Raboud, J. M.,
E. Seminari,
S. L. Rae,
P. R. Harrigan,
R. S. Hogg,
B. Conway,
C. Sherlock,
M. T. Schechter,
M. V. O'Shaughnessy, and J. S. G. Montaner.
1998.
Comparison of costs of strategies for measuring levels of human immunodeficiency virus type 1 RNA in plasma by using Amplicor and Ultra Direct assays.
J. Clin. Microbiol.
36:3369-3371[Abstract/Free Full Text].
|
| 14.
|
Segondy, M.,
J. Izopet,
I. Pellegrin,
B. Montes,
B. Dumon,
C. Pasquier,
M. Peeters,
H. J. Fleury,
J. Puel, and J. Reynes.
1998.
Comparison of the Quantiplex HIV-1 RNA 2.0 assay with the Amplicor HIV-1 Monitor 1.0 assay for quantitation of levels of human immunodeficiency virus type 1 RNA in plasma of patients receiving stavudine-didanosine combination therapy.
J. Clin. Microbiol.
36:3392-3395[Abstract/Free Full Text].
|
| 15.
|
Sun, R.,
J. Ku,
H. Jayakar,
J. C. Kuo,
D. Brambilla,
S. Herman,
M. Rosenstraus, and J. Spadoro.
1998.
Ultrasensitive reverse transcription-PCR assay for quantitation of human immunodeficiency virus type 1 RNA in plasma.
J. Clin. Microbiol.
36:2964-2969[Abstract/Free Full Text].
|
| 16.
|
Todd, J.,
C. Pachl,
R. White,
T. Yeghiazarian,
P. Johnson,
B. Taylor,
M. Holdoniy,
D. Kern,
S. Hamren,
D. Chernoff, and M. Urdea.
1995.
Performance characteristics for the quantitation of plasma HIV-1 RNA using branched DNA signal amplification technology.
J. Acquir. Immune Defic. Syndr. Hum. Retrovirol.
10:S35-S44.
|
| 17.
|
Triques, K.,
J. Coste,
J. L. Perret,
C. Segarra,
E. Mpoudi,
J. Reynes,
E. Delaporte,
A. Butcher,
K. Dreyer,
S. Herman,
J. Spadoro, and M. Peeters.
1999.
Efficiencies of four versions of the AMPLICOR HIV-1 MONITOR test for quantification of different subtypes of human immunodeficiency virus type 1.
J. Clin. Microbiol.
37:110-116[Abstract/Free Full Text].
|
| 18.
|
Vandamme, A. M.,
K. Van Vaerenbergh, and E. De Clercq.
1998.
Anti-human immunodeficiency virus drug combination strategies.
Antivir. Chem. Chemother.
9:187-203[Medline].
|
Journal of Clinical Microbiology, March 2000, p. 1113-1120, Vol. 38, No. 3
0095-1137/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
This article has been cited by other articles:
-
Bendavid, E., Young, S. D., Katzenstein, D. A., Bayoumi, A. M., Sanders, G. D., Owens, D. K.
(2008). Cost-effectiveness of HIV Monitoring Strategies in Resource-Limited Settings: A Southern African Analysis. Arch Intern Med
168: 1910-1918
[Abstract]
[Full Text]
-
Zaric, G. S., Bayoumi, A. M., Brandeau, M. L., Owens, D. K.
(2008). The Cost-Effectiveness of Counseling Strategies to Improve Adherence to Highly Active Antiretroviral Therapy among Men Who Have Sex with Men. Med Decis Making
28: 359-376
[Abstract]
-
Westreich, D. J., Hudgens, M. G., Fiscus, S. A., Pilcher, C. D.
(2008). Optimizing Screening for Acute Human Immunodeficiency Virus Infection with Pooled Nucleic Acid Amplification Tests. J. Clin. Microbiol.
46: 1785-1792
[Abstract]
[Full Text]
-
Swanson, P., de Mendoza, C., Joshi, Y., Golden, A., Hodinka, R. L., Soriano, V., Devare, S. G., Hackett, J. Jr.
(2005). Impact of Human Immunodeficiency Virus Type 1 (HIV-1) Genetic Diversity on Performance of Four Commercial Viral Load Assays: LCx HIV RNA Quantitative, AMPLICOR HIV-1 MONITOR v1.5, VERSANT HIV-1 RNA 3.0, and NucliSens HIV-1 QT. J. Clin. Microbiol.
43: 3860-3868
[Abstract]
[Full Text]
-
Palmer, S., Wiegand, A. P., Maldarelli, F., Bazmi, H., Mican, J. M., Polis, M., Dewar, R. L., Planta, A., Liu, S., Metcalf, J. A., Mellors, J. W., Coffin, J. M.
(2003). New Real-Time Reverse Transcriptase-Initiated PCR Assay with Single-Copy Sensitivity for Human Immunodeficiency Virus Type 1 RNA in Plasma. J. Clin. Microbiol.
41: 4531-4536
[Abstract]
[Full Text]
-
Mani, I., Gilbert, P., Sankale, J.-L., Eisen, G., Mboup, S., Kanki, P. J.
(2002). Intrapatient Diversity and Its Correlation with Viral Setpoint in Human Immunodeficiency Virus Type 1 CRF02_A/G-IbNG Infection. J. Virol.
76: 10745-10755
[Abstract]
[Full Text]
-
Si-Mohamed, A., Andreoletti, L., Colombet, I., Carreno, M.-P., Lopez, G., Chatelier, G., Kazatchkine, M. D., Belec, L.
(2001). Quantitation of Human Immunodeficiency Virus Type 1 (HIV-1) RNA in Cell-Free Cervicovaginal Secretions: Comparison of Reverse Transcription-PCR Amplification (AMPLICOR HIV-1 MONITOR 1.5) with Enhanced-Sensitivity Branched-DNA Assay (Quantiplex 3.0). J. Clin. Microbiol.
39: 2055-2059
[Abstract]
[Full Text]
-
Kellogg, J. A., Atria, P. V., Sanders, J. C., Eyster, M. E.
(2001). Intra- and Interlaboratory Variabilities of Results Obtained with the Quantiplex Human Immunodeficiency Virus Type 1 RNA bDNA Assay, Version 3.0. CVI
8: 560-563
[Abstract]
[Full Text]
-
Swanson, P., Soriano, V., Devare, S. G., Hackett, J. Jr.
(2001). Comparative Performance of Three Viral Load Assays on Human Immunodeficiency Virus Type 1 (HIV-1) Isolates Representing Group M (Subtypes A to G) and Group O: LCx HIV RNA Quantitative, AMPLICOR HIV-1 MONITOR Version 1.5, and Quantiplex HIV-1 RNA Version 3.0. J. Clin. Microbiol.
39: 862-870
[Abstract]
[Full Text]
-
Murphy, D. G., Côté, L., Fauvel, M., René, P., Vincelette, J.
(2000). Multicenter Comparison of Roche COBAS AMPLICOR MONITOR Version 1.5, Organon Teknika NucliSens QT with Extractor, and Bayer Quantiplex Version 3.0 for Quantification of Human Immunodeficiency Virus Type 1 RNA in Plasma. J. Clin. Microbiol.
38: 4034-4041
[Abstract]
[Full Text]
-
(2000). Comparison of Ultrasensitive Versions of bDNA and RT-PCR. AIDS Clin Care
2000: 6-6
[Full Text]