Previous Article | Next Article 
Journal of Clinical Microbiology, June 2000, p. 2117-2121, Vol. 38, No. 6
0095-1137/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
Accuracy and Precision of Quantitative Calibrated
Loops in Transfer of Bronchoalveolar Lavage Fluid
J. A.
Jacobs,1,*
E. I. G. B.
De Brauwer,1
E. I. M.
Cornelissen,1 and
M.
Drent2
Departments of Medical
Microbiology1 and
Pulmonology,2 University Hospital
Maastricht, Maastricht, The Netherlands
Received 12 November 1999/Returned for modification 9 January
2000/Accepted 21 March 2000
 |
ABSTRACT |
Quantitative cultures of bronchoalveolar lavage (BAL) fluid are
important in the diagnosis of ventilator-associated pneumonia, and
calibrated loops are commonly used to set up these cultures. In this
study, the performances of calibrated 0.010- and 0.001-ml loops in the
transfer of BAL fluid were determined. Five loops of one lot from seven
manufacturers were tested. Calibrations were performed by the
gravimetric method (0.010-ml loops) and the colorimetric method
(0.001-ml loops). Most of the 0.010-ml loops displayed a precision that
was less than 10%, but six of them showed very poor accuracies as they
transferred a deficiency (nichrome loops) or an excess (disposable
loops) of BAL fluid that exceeded ±10%. The mean maximum and minimum
BAL fluid volumes delivered by the 0.010-ml loops differed by a factor
3. The 0.001-ml loops displayed acceptable precision. Five of them
showed inaccuracies of
±10%, and mean maximum and minimum BAL fluid
volumes had a range of a factor of 2. For all loops, the volumes of BAL
fluid sampled were larger than the volumes of reagent-grade water
sampled. Results of the colony counting experiments confirmed these
findings and revealed a high intra-assay variability for the 0.001-ml
loops. We conclude that, when BAL fluid samples are cultured with
calibrated loops, (i) proper verification of the calibration of these
loops is mandatory, (ii) calibrations should be performed with BAL
fluid as the test solution, and (iii) borderline quantitative culture results should be interpreted with knowledge of the inaccuracy values
of these loops.
 |
INTRODUCTION |
Quantitative cultures of
bronchoalveolar lavage (BAL) fluid are used in the diagnosis of
ventilator-associated pneumonia (VAP). As the dilution of the lung
secretions in the BAL fluid is 10- to 100-fold, a colony count of
104 CFU/ml represents a bacterial load of 105
to 106/ml in the collection site, which is indicative of
bacterial pneumonia (3, 15). Conversely, a BAL fluid colony
count below the 104-CFU/ml threshold points to
oropharyngeal contamination. This theoretical concept has been
validated in numerous clinical studies, and quantitative culture of BAL
fluid specimens is consequently recommended as the reference method for
the diagnosis of VAP (6).
For quantitative cultures of BAL fluid, two approaches are used: the
serial dilution method and the calibrated loop method (3).
In the serial dilution method, 0.100-ml aliquots of the raw BAL fluid
and two serial 100-fold dilutions are inoculated onto the agar plate
surfaces. After incubation, counts are made from the dilution that
contains the greatest number of bacteria without confluence or
overcrowding. Many microbiological laboratories perceive this method as
too cumbersome and labor-intensive (15). Therefore, they
prefer the calibrated loop method, which they are familiar with.
Calibrated loops are routinely used to set up quantitative urine
cultures (5, 17). Calibrated loops are designed to transfer
a well-defined sample volume to agar plates, omitting the need for
dilutions. They may be reusable (loops made of nichrome or platinum) or
disposable (loops made of plastic). For BAL fluid samples, quantitative
calibrated loops designed for the delivery of 0.010 and 0.001 ml are
used. After incubation, the colonies are counted on the plates and the
number of CFU per milliliter is determined by multiplying the number of
colonies by the dilution factor. Calibrated loops are widely used for
quantitative BAL fluid cultures in the diagnostic and research settings
(4, 7, 8, 11, 12, 19, 20, 21, 23).
In our hospital, quantitative culture of BAL fluid is the standard
method for the microbiological diagnosis of VAP (9). We
prefer the calibrated loop method and use reusable nichrome loops. By
repetition, we observed that colony counts on plates inoculated by use
of 0.010-ml loops did not reach a 10-fold range of those obtained by
use of 0.001-ml loops. We therefore decided to determine the
performance characteristics of different types of calibrated loops.
 |
MATERIALS AND METHODS |
Calibrated loops included and transfer procedures.
The
different quantitative loops that were studied are listed in Table
1. They included both reusable nichrome
loops and disposable plastic loops. For each loop type and
manufacturer, five loops of one lot were tested. Calibrations were
performed by the gravimetric method for the 0.010-ml loops and by the
colorimetric method for the 0.001-ml loops (2).
Reagent-grade water (Milli-Q plus system; Millipore, Etten-Leur, The
Netherlands) and freshly obtained BAL fluid specimens were used as the
test solutions.
All procedures were carried out in a standardized way, and the
different aliquots were delivered with uniform timing and motion. All
containers were widemouthed (diameter, 27 mm), nonadhesive, polypropylene, conical, 50-ml tubes containing 25 ml of test solution (product no. 227.261; Greiner, Alphen aan de Rijn, The Netherlands). The loops were held vertically and were inserted into the test solution
just below the liquid's surface, thereby taking care not to immerse
the shank. After a check to ensure that no bubbles were within the
loops, they were gently lifted straight up and the sample was
transferred. Before the first transfer, the nichrome loops were rinsed
in a separate flask of reagent-grade water, flamed, and cooled.
Together with the loops tested, the volumes delivered by a pipette were
determined. For the delivery of a 0.010-ml and a 0.001-ml volume, an
adjustable air-displacement pipette (Pipetman P20; Gilson,
Villiers-le-Bel, France) was used.
Calibration with reagent-grade water.
For the gravimetric
method (0.010-ml loops), weight determinations were performed with an
analytic balance with a readability of 0.1 mg (AE 166; Mettler-Toledo
Ltd., Tiel, The Netherlands). For each 0.010-ml loop, 20 loopfuls of
reagent-grade water were successively added to filter paper on the
balance pan. All filter papers were handled with forceps, and the
weighings were quickly completed in order to avoid any effect of
evaporation. The ambient temperature (22 ± 1°C) was checked for
each test procedure.
For the colorimetric method (0.001-ml loops), a 0.75% (wt/vol) stock
solution of Evan's blue dye (Gurr, Essex, United Kingdom) was prepared
in reagent-grade water and was strained through filter paper (product
no. 311620; Schleicher & Schuell, Dassel, Germany). The maximal
absorbance wavelength was checked with a Kontron Uvikon 680 spectrophotometer (Beun de Ronde Ltd., Abcoude, The Netherlands). Of
this stock solution, working dilutions of 0.2:100, 0.4:100, 0.6:100,
0.8:100, 1.0:100, 1.2:100, and 1.4:100 were prepared. Of each working
solution, an aliquot of 100 µl was transferred to a flat-bottom
microtiter plate (Nunc, Roskilde, Denmark). The absorbances were
subsequently read at 600 nm with a microtiter plate reader (SLT 340 ATCC; Beun de Ronde), and a calibration curve was prepared by plotting
the absorbance on the vertical scale versus the volume of Evan's blue
solution in the working dilutions on the horizontal scale. This curve
was checked for linearity and reproducibility. For each 0.001-ml loop,
12 loopfuls of the Evan's blue stock dye solution were successively
transferred to 100 µl of reagent-grade water in the flat-bottom
microtiter plate. After thorough mixing, absorbances were recorded by
the microplate reader, plotted on the calibration curve, and converted into the corresponding volumes.
Calibration with BAL fluid specimens.
In addition to
reagent-grade water, freshly obtained BAL fluid samples were used as
the test liquid by the gravimetric method for the assessment of the
0.010-ml loops. BAL fluid samples were obtained by instillation of four
50-ml aliquots of 0.9% NaCl during bronchoscopy, as described
previously (10). The BAL fluid samples were gently mixed on
a roller mixer (Coulter Electronics Ltd., Luton, England) before each
test. Likewise, 1 ml of a stock solution of 7.5% (wt/vol) Evan's blue
in reagent-grade water was diluted in 19 ml of a BAL fluid sample, and
this solution was used as the test liquid for the 0.001-ml loops by the
colorimetric method.
Description of the test characteristics.
For quantification
of the performance of the loops, the parameters accuracy and precision
were used (2). Accuracy is the closeness of agreement
between the stated (expected) volume of the calibrated loop and the
mean volume obtained during repeated, controlled deliveries. Accuracy
is numerically expressed as inaccuracy, which can be thought of as the
difference between the stated volume and the mean measured volume.
Inaccuracy is expressed in percent; a positive value and a negative
value indicate the delivery of an excess and a deficiency,
respectively, compared to the expected volume. Precision expresses the
agreement between replicate measurements and can be regarded as
intra-assay variability. Precision is numerically expressed as
imprecision, which can be considered the coefficient of variation of
replicate, controlled measurements.
Statistical analysis.
For the gravimetric method, the mean
weight was calculated for each loop. This value was converted into the
mean volume by using a standard Z value of 1.0032, which
corresponds to an air pressure of 720 mm Hg at a temperature of 22°C
(2). For the colorimetric method, the mean volume of each
loop was calculated from the individual absorbance determinations. For
each loop, inaccuracy values were calculated by using the mean volume
and the nominal (expected) volume. Imprecision values were calculated by using the individual weights, the corresponding mean weight, and the
number of weighings for the gravimetric method and by using the
corresponding data on volumes for the colorimetric method. Both
inaccuracy and imprecision values were expressed as percentages (2).
Quantitative cultures of BAL fluid.
For quantitative
cultures, a freshly obtained sterile BAL fluid specimen was used as the
test liquid, and Staphylococcus aureus ATCC 29123 was
selected as the test organism. From an overnight culture, a suspension
with a density of a no. 1 McFarland standard was made in 0.9% NaCl. A
10
5 dilution from this suspension was made in BAL fluid
until a final concentration of approximately 104 CFU/ml was
achieved, and this solution was sampled with the 0.001- and 0.010-ml
loops and pipettes. For three loops of each type, three successive
samplings were performed. Between the samplings with the individual
loops, the test solution was well mixed on a roller mixer. The aliquots
were transferred to cystine lactose electrolyte-deficient agar (Becton
Dickinson Microbiology Europe, Meylan Cedex, France) and were spread
over the entire surface of the agar. After incubation at 35°C for
24 h, the colonies were counted. Mean ± standard deviation
(SD) colony counts were determined for each loop type, and the
intra-assay variability for each loop was calculated.
 |
RESULTS |
The test characteristics assessed for the 0.010-ml loops in the
gravimetric method are listed in Table 2.
From the data in Table 2, it is clear that the transfers of both
reagent-grade water and BAL fluid specimens by the adjustable pipette
were accurate and precise. The calibrated loops displayed less
favorable test characteristics, and both inaccuracy and imprecision
varied according to the fluid transferred. For the transfer of
reagent-grade water, most calibrated loops delivered a volume that was
too small, resulting in negative inaccuracy values. The nichrome loops
delivered the smallest volumes; i.e., instead of the expected 10-µl
volume, mean volumes of only 4.7 and 5.2 µl were delivered by the
Pro-Lab and the Medical Wire loops, respectively. The Italiana loop was the only loop that transferred an excess of reagent-grade water, and
the Sarstedt loop delivered exactly the expected volume. With BAL fluid
samples as the test fluid, all loops transferred larger volumes than
those of reagent-grade water transferred, and the inaccuracies varied
accordingly. The difference was most marked for the Greiner loop, which
transferred mean volumes of 6.0 and 10.8 µl for reagent-grade water
and BAL fluid, respectively. For the transfer of BAL fluid, only one of
seven loop types tested (i.e., Greiner) displayed inaccuracy values of
±10%. Furthermore, the range between the inaccuracies of the
different loops for the transfer of BAL fluid was high. The difference
between the mean BAL fluid volume transferred by the Pro-Lab nichrome
loop, on the one hand, and the Italiana loop, on the other, was more than 10 µl, or, in other words, the Italiana 0.010-ml loop delivered threefold the volume sampled by the Pro-Lab homologue. The imprecision values for all different types of loops for the transfer of BAL fluid
were fairly good relative to our other findings (i.e., precision variation of >15 to 20%), and the imprecision values for BAL fluid transfer were better than those obtained for the transfer of
reagent-grade water. To exclude factors related to the BAL fluid
specimen applied in the assay as the cause of the differences observed,
both nichrome loops and the disposable Greiner and Simport loops were
assessed with four additional BAL fluid specimens with different
protein contents and total cell counts (Table
3). The results were similar to those
obtained with the initial BAL fluid specimen.
View this table:
[in this window]
[in a new window]
|
TABLE 2.
Test characteristics of the 0.010-ml loops assessed by
the gravimetric method with reagent-grade water and BAL fluid as
the test solutionsa
|
|
View this table:
[in this window]
[in a new window]
|
TABLE 3.
Test characteristics of the 0.010-ml loops assessed by
the gravimetric method with five different BAL fluid samples as the
test solutionsa
|
|
The test characteristics for the 0.001-ml loops are listed in Table
4. In line with the results for the
0.010-ml loops, the volume of BAL fluid sampled was larger than the
corresponding volume of reagent-grade water. The nichrome loops
displayed the smallest inaccuracies in the transfer of reagent-grade
water, but they performed less well in the transfer of BAL fluid. For the transfer of BAL fluid, better inaccuracy values were found compared
to those for the 0.010-ml loops. Three of seven types of loops tested
showed inaccuracy values of
5%. The highest difference between the
inaccuracies of the 0.001-ml loops in the transfer of BAL fluid was
found between the Pro-Lab and the Simport loops, and this difference
reached 65%; that is, the Pro-Lab loop delivered approximately twofold
the volume of BAL fluid transferred by the Simport loop. As for the
0.010-ml loops, the imprecision values for all 0.001-ml loops were
satisfactory when BAL fluid was used as the test fluid, and the
inaccuracy values were similar for BAL fluid samples with different
protein contents and total cell counts (data not shown).
View this table:
[in this window]
[in a new window]
|
TABLE 4.
Test characteristics of the 0.001-ml loops assessed by
the colorimetric method with reagent-grade water and BAL fluid as the
test solutionsa
|
|
The culture results (mean colony counts) obtained with the 0.010-ml and
the 0.001-ml loops are listed in Table 5.
For most of the 0.010-ml loops, the differences in mean colony counts
reflected the calibrated differences. The nichrome reusable loops
consistently generated low colony counts, and three disposable loop
types yielded higher colony counts compared to those achieved by use of
the pipette. The mean maximum and minimum colony counts delivered by
the loops differed by a factor of 2.8. As can be read from the SD
values, the variations between the mean colony counts obtained with the
individual loops were higher than those recorded by use of volume
calibration. In addition, intra-assay variabilities exceeding 10% were
observed for the triplicate samplings for 13 of 21 individual loops
tested.
View this table:
[in this window]
[in a new window]
|
TABLE 5.
Colony counts for three loops of each type, tested in
triplicate, with a control culture of S. aureus in BAL fluid
as the test solution
|
|
Overall, the 0.001-ml loops and the pipette generated higher colony
counts compared to those obtained with the 0.010-ml loops. The
differences in mean colony counts between the 0.001-ml loops of various
manufacturers were higher than those observed by use of the calibrated
volumes. The triplicate samplings for the 21 loops tested generated
intra-assay variabilities that exceeded 10% for 19 loops and 30% for
6 loops. The mean maximum and minimum colony counts delivered by the
loops differed by a factor of 1.9.
 |
DISCUSSION |
In the present study, we assessed the performances of both
reusable and disposable quantitative calibrated loops in the transfer of reagent-grade water and BAL fluid. The 0.010-ml calibrated loops
displayed good precision, but most of them showed poor accuracies and
delivered volumes of BAL fluid that were too small (nichrome loops) or
delivered an excess volume (disposable loops). The 0.001-ml calibrated
loops displayed good precision, and four of seven types tested
delivered a BAL fluid volume with an inaccuracy of <±10%. The mean
maximum and minimum BAL fluid volumes delivered by the 0.010- and
0.001-ml loops of various manufacturers differed by factors of 3 and 2, respectively. The results of the colony counting experiments confirmed
these findings but revealed a high intra-assay variability for
replicate samplings for the 0.001-ml loops.
According to leading textbooks, quantitative calibrated loops may be
used for various purposes such as preparation of inocula for
antimicrobial tests, preparation of serum dilutions, addition of
ingredients to media and reagents to test procedures, and setting up of
quantitative cultures (2, 22). For the quantitative culture
of BAL fluid specimens, the 0.010- and 0.001-ml calibrated loops are
very attractive because at final dilutions of 1:100 (0.010-ml loop) and
1:1,000 (0.001-ml loop), a minimum concentration of 102
CFU/ml can be detected and a maximum concentration of more than 105 CFU/ml can be discerned without piling up of the
colonies on the agar plate's surface. In this way, the consensus
culture cutoff point of 104 CFU/ml for the diagnosis of VAP
(13) is within the reach of this technique. The calibrated
loop method is more amenable to a routine clinical microbiology
laboratory than the serial dilution method, and its feasibility allows
performance during off hours. For inoculation of clinical samples in
the microbiological laboratory, calibrated loops are preferred over
pipettes because the latter are generally slower and may present
problems in terms of decontamination, cross-contamination, and
infectious aerosol control (5). Calibrated loops are also
used for quantitative cultures of other respiratory specimens such as
protected specimen brush (PSB) samples (14, 16) and
bronchial aspirates (18).
The results of the present study demonstrated poor accuracies for the
majority of the 0.010-ml loops. The 0.001-ml loops performed better,
and several manufacturers are marketing 0.001-ml loops that displayed
excellent accuracy for the transfer of BAL fluid. It should be noted
that we tested only loops that belonged to single lots and that we do
not have further grounds to prefer one manufacturer's loop over
another. Irrespective of the loop that is chosen, we underscore the
need for determination of its accuracy before use and at any time that
a new lot is used, as recommended by Baron (2). Although not
required for calibrated loops (2), we elected to determine
their precision. When applied to set up a quantitative BAL fluid
culture, calibrated loops (including disposable ones) are repeatedly
used for inoculation of several enriched and selective agar plates, and
for this reason we were interested in precision as a test
characteristic. In the volume calibration experiments, the imprecision
values were considered satisfactory, but these values were not
reflected by low intra-assay variabilities in the colony counting
experiments. Replicate transfers by the 0.001-ml pipette also
yielded higher variations in colony counts, and the ranges in
colony counts observed for the 0.001-ml loops were higher than
those observed for the 0.010-ml loops. Therefore, it is tempting
to speculate that sampling error has been the cause of the high
intra-assay variability: irregular distribution of organisms and the
presence of cellular material, mucus, and debris may have influenced
the number of organisms present in the sample transferred. Procedures
such as vortexing or filtration may facilitate further homogenization
of the BAL fluid specimen, but because they alter the morphologies of
the cells in the BAL fluid specimen, we prefer to mix the BAL fluid samples on a roller mixer designed for peripheral blood samples.
For the 0.010-ml loops as well as for the 0.001-ml loops, we found a
large range between the mean maximum and minimum volumes transferred by
the loops of different manufacturers. From Table 5, it is clear that a
single BAL fluid sampled with loops from different manufacturers may
yield colony counts spread out on either side of the
104-CFU/ml cutoff point for VAP. Therefore, BAL fluid
samples with quantitative culture results that approximate the
threshold value should not be strictly categorized as VAP positive or
VAP negative by exclusive dependence on their colony counts. Borderline
quantitative culture results should be interpreted together with
information on the patient's clinical condition and on the use of
antibiotics prior to bronchoscopy and together with data on the BAL
fluid's cytology (3, 9).
Apart from the performance characteristics linked to the calibration of
the loop, many other variables may cause sampling errors. Less fluid is
sampled when containers with small diameters are used, and a larger
volume is sampled when the loop is inserted and withdrawn at a 45°
angle or when the shank of the loop is immersed (1, 2).
Containers with small diameters may cause the loop to pick up a smaller
volume since plastic-liquid (adhesive) forces are greater than
liquid-liquid (cohesive) forces (1). When the shank of the
loop is wetted by deep immersion into the fluid, excess liquid drains
down and enlarges the volume transferred (2). In the present
study, all these factors were carefully controlled, but in daily
practice, they are additional reasons for a loop's inaccuracy and they
may account for the high degree of variability in the volume of BAL
fluid sampled.
It is of interest to compare the present results with those of a
previous study on the use of calibrated loops in respiratory specimens.
With PSB samples immersed in laboratory stock cultures as well as with
clinically obtained PSB samples, Middleton and coworkers
(15) compared the culture results obtained by both the
serial dilution and the calibrated loop methods. When their results are
compared with the present findings, several factors need to be
considered. First, Middleton and coworkers used platinum loops, and
these loops were not included in the present study. To our knowledge,
platinum loops are not commonly used because of their high costs (their
price approaches that of a pipette). Next, the PSB samples cultured in
the previous study were diluted 1:1,000 in lactated Ringer's solution,
and this solution may have cohesive and adhesive forces different from
those for the BAL fluid specimens in the present study. Furthermore, in
the previous study, 6 of 14 PSB samples cultured by the serial dilution
method yielded colony counts far above the diagnostic threshold
(103 CFU/ml) for PSB samples, and 1 sample yielded no
growth. The correlation between the serial dilution method and the
calibrated loop method for these samples was excellent, although the
colony counts obtained with the 0.010-ml loop were lower than those
obtained with the 0.001-ml loop and the serial dilution method. If,
however, PSB specimens with colony counts obtained by the serial
dilution method just above the culture cutoff point were considered,
three of six samples did not reach the 103-CFU/ml threshold
when cultured with the 0.010-ml loop, suggesting a deficiency of the
volume transferred by the 0.010-ml platinum loop. In line with the
findings of the present study, the 0.001-ml platinum loop in the
previous study (15) performed better, with no threshold
discrepancy between the culture results obtained with this loop and by
the serial dilution method.
In conclusion, the results of the present study indicate that for the
quantitative culture of BAL fluid specimens, most of the 0.001-ml loops
displayed good accuracy, but the majority of the 0.010-ml calibrated
loops did not meet acceptable accuracy limits for the transfer of BAL.
If calibrated loops are considered for use in the setup of quantitative
BAL fluid cultures, proper verification of their calibration is
mandatory. Calibrations should be performed with BAL fluid samples and
not with reagent-grade water. Borderline quantitative BAL fluid culture
results obtained by the calibrated loop method should be interpreted
with knowledge of the inaccuracy values for these loops.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: Department of
Medical Microbiology, University Hospital of Maastricht, P.O. Box 5800, 6202 AZ Maastricht, The Netherlands. Phone: 31-43-387 46 44. Fax: 31-43-387 66 43. E-mail: JJA{at}ms-azm-3.azm.nl.
 |
REFERENCES |
| 1.
|
Albers, A. C., and R. D. Fletcher.
1983.
Accuracy of calibrated-loop transfer.
J. Clin. Microbiol.
18:40-42[Abstract/Free Full Text].
|
| 2.
|
Baron, E. J.
1992.
Instrument maintenance and quality control, p. 12.1.1-12.22.6.
In
H. D. Isenberg (ed.), Clinical microbiology procedures handbook, vol. 2. American Society for Microbiology, Washington, D.C.
|
| 3.
|
Baselski, V. S., and R. G. Wunderink.
1994.
Bronchoscopic diagnosis of pneumonia.
Clin. Microbiol. Rev.
7:533-558[Abstract/Free Full Text].
|
| 4.
|
Cantral, D. E.,
T. G. Tape,
E. C. Reed,
J. R. Spurzem,
S. I. Rennard, and A. B. Thompson.
1993.
Quantitative culture of bronchoalveolar lavage fluid for the diagnosis of bacterial pneumonia.
Am. J. Med.
95:601-607[CrossRef][Medline].
|
| 5.
|
Clarridge, J. E.,
M. T. Pezzlo, and K. L. Vosti.
1987.
Cumitech 2A
Laboratory diagnosis of urinary tract infections. Coordinating ed., W. E. Sheperd. American Society for Microbiology, Washington, D.C.
|
| 6.
|
Ewig, S.
1996.
Diagnosis of ventilator-associated pneumonia: nonroutine tools for routine practice.
Eur. Respir. J.
9:1339-1341[CrossRef][Medline].
|
| 7.
|
Guerra, L. F., and R. P. Baugham.
1990.
Use of bronchoalveolar lavage to diagnose bacterial pneumonia in mechanically ventilated patients.
Crit. Care Med.
18:169-173[Medline].
|
| 8.
|
Humphreys, H.,
R. Winter,
M. Baker, and C. Smith.
1996.
Comparison of bronchoalveolar lavage and catheter lavage to confirm ventilator-associated lower respiratory tract infection.
J. Med. Microbiol.
45:226-231[Abstract].
|
| 9.
|
Jacobs, J. A., and E. I. G. B. De Brauwer.
1999.
Applications of BAL fluid cytology for the assessment of infectious lung disease.
Hosp. Med.
60:550-556[Medline].
|
| 10.
|
Jacobs, J. A.,
E. I. G. B. De Brauwer,
G. Ramsay,
N. A. M. Cobben,
S. S. Wagenaar,
A. J. A. M. van der Ven,
C. A. Bruggeman, and M. Drent.
1999.
Detection of noninfectious conditions mimicking pneumonia in the intensive care setting: usefulness of BAL fluid cytology.
Respir. Med.
93:571-578[CrossRef][Medline].
|
| 11.
|
Kollef, M. H.,
K. R. Bock,
R. D. Richards, and M. L. Hearns.
1995.
The safety and diagnostic accuracy of minibronchoalveolar lavage in patients with suspected ventilator-associated pneumonia.
Ann. Intern. Med.
122:743-748[Abstract/Free Full Text].
|
| 12.
|
Luna, C. M.,
P. Vujacik,
M. S. Niederman,
C. Vay,
C. Gherardi,
J. Matera, and E. Jolly.
1997.
Impact of BAL data on the therapy and outcome of ventilator-associated pneumonia.
Chest
111:676-685[Abstract/Free Full Text].
|
| 13.
|
Meduri, G. U., and J. Chastre.
1992.
The standardization of bronchoscopic techniques for ventilator-associated pneumonia.
Chest
102:S557-S564.
|
| 14.
|
Mertens, A. H.,
J. M. Nagler,
D. I. Galdermans,
H. R. Slabbynck,
B. Weise, and D. Coolen.
1998.
Quality assessment of protected specimen brush samples by microscopic cell count.
Am. J. Respir. Crit. Care Med.
157:1240-1243[Abstract/Free Full Text].
|
| 15.
|
Middleton, R. M., III,
W. Huff,
D. A. Brickey, and M. B. Kirkpatrick.
1996.
Comparison of quantitative cultures to semiquantitative loop cultures of bronchoscopic protected specimen brush samples.
Chest
109:1204-1209[Abstract/Free Full Text].
|
| 16.
|
Örtqvist, Ä.,
M. Kalin,
L. Lejdeborn, and B. Lundberg.
1990.
Diagnostic fiberoptic bronchoscopy and protected brush culture in patients with community-acquired pneumonia.
Chest
97:576-582[Abstract/Free Full Text].
|
| 17.
|
Pezzlo, M.
1992.
Aerobic bacteriology, p. 1.1.1-1.20.44.
In
H. D. Isenberg (ed.), Clinical microbiology procedures handbook, vol. 1. American Society for Microbiology, Washington, D.C.
|
| 18.
|
Pirali, F.,
M. Longo,
M. Gelmi,
D. Colombrita,
G. Ravizzola,
G. Pinsi,
L. Peroni,
N. Manca, and A. Turano.
1994.
Diagnosis of bronchopulmonary infections by quantification of microflora.
Eur. J Epidemiol.
10:703-706[CrossRef][Medline].
|
| 19.
|
Solé-Violán, J.,
F. Rodriguez de Castro,
A. Rey,
J. C. Martin-González, and P. Cabrere-Navarro.
1994.
Usefulness of microscopic examination of intracellular organisms in lavage fluid in ventilator-associated pneumonia.
Chest
106:889-894[Abstract/Free Full Text].
|
| 20.
|
Speich, R.,
M. Hauser,
T. Hess,
J. Wüst,
E. Grebski,
F. H. Kayser, and E. W. Russi.
1998.
Low specificity of the bacterial index for the diagnosis of bacterial pneumonia by bronchoalveolar lavage.
Eur. J. Clin. Microbiol. Infect. Dis.
17:78-84[CrossRef][Medline].
|
| 21.
|
Thorpe, J. E.,
R. P. Baughman,
P. T. Frame,
T. A. Wesseler, and J. L. Staneck.
1987.
Bronchoalveolar lavage for diagnosing acute bacterial pneumonia.
J. Infect. Dis.
155:855-861[Medline].
|
| 22.
|
Turano, A., and F. Pirali.
1988.
Quantification methods in microbiology, p. 8-14.
In
A. Balows, W. J. Hausler, M. Ohashi, and A. Turano (ed.), Laboratory diagnosis of infectious diseases.Principles and practice. Springer-Verlag, New York, N.Y.
|
| 23.
|
Violán, S. J.,
F. R. de Castro,
J. C. Luna,
A. B. Benítez, and J. L. M. Alonso.
1993.
Comparative efficacy of bronchoalveolar lavage and telescoping plugged catheter in the diagnosis of pneumonia in mechanically ventilated patients.
Chest
103:386-390[Abstract/Free Full Text].
|
Journal of Clinical Microbiology, June 2000, p. 2117-2121, Vol. 38, No. 6
0095-1137/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
This article has been cited by other articles:
-
Linssen, C. F M, Bekers, O., Drent, M., Jacobs, J. A
(2008). C-reactive protein and procalcitonin concentrations in bronchoalveolar lavage fluid as a predictor of ventilator-associated pneumonia. Ann Clin Biochem
45: 293-298
[Abstract]
[Full Text]
-
Jacobs, J. A., Stobberingh, E. E., Cornelissen, E. I. M., Drent, M.
(2005). Detection of Streptococcus pneumoniae Antigen in Bronchoalveolar Lavage Fluid Samples by a Rapid Immunochromatographic Membrane Assay. J. Clin. Microbiol.
43: 4037-4040
[Abstract]
[Full Text]
-
Halaby, T., Boots, H., Vermeulen, A., van der Ven, A., Beguin, H., van Hooff, H., Jacobs, J.
(2001). Phaeohyphomycosis Caused by Alternaria infectoria in a Renal Transplant Recipient. J. Clin. Microbiol.
39: 1952-1955
[Abstract]
[Full Text]