Previous Article | Next Article ![]()
Journal of Clinical Microbiology, July 2000, p. 2763-2765, Vol. 38, No. 7
Division of Bacterial and Mycotic Diseases,
National Center for Infectious Diseases, Centers for Disease Control
and Prevention, Atlanta, Georgia,1 and
Ministry of Health and Long Term Care, Toronto, Ontario,
Canada2
Received 11 January 2000/Returned for modification 23 March
2000/Accepted 1 May 2000
The Binax and the Biotest urinary antigen kits for the detection of
Legionnaires' disease caused by organisms other than Legionella pneumophila were compared by testing 45 urine samples from
non-Legionella pneumophila serogroup 1 patients previously
positive in a broad-spectrum enzyme-linked immunosorbent assay (ELISA).
Eighteen were positive with the Binax kit, and 13 were positive with
the Biotest. Although neither kit is as sensitive as ELISA, these
results extend the number of serogroups and species of
Legionella that can be diagnosed with the Binax or Biotest kit.
The usefulness of urinary antigen
detection for the diagnosis of Legionnaires' disease has been well
documented. The reported sensitivity of the Binax urinary antigen test
was 56% for all cases classified as definitive in a study of
community-acquired pneumonia (9) and 80% for
culture-positive Legionella pneumophila serogroup 1. In a
limited number of studies, a positive result was obtained for patients
with culture-confirmed Legionnaires' disease caused by strains other
than L. pneumophila serogroup 1 (2, 8). However,
there are no reports of positive urinary antigen results with
Legionnaires' disease caused by other species of
Legionella. The Biotest urine antigen enzyme immunoassay
(EIA) (Biotest AG, Dreieich, Germany) has been recently introduced and according to the manufacturer has a wide range of cross-reactivity to L. pneumophila serogroups and other Legionella
species (Biotest Legionella urine antigen EIA instructions,
Biotest AG). The broad-spectrum enzyme-linked immunosorbent assay
(ELISA) antigen is an intragenus urine assay reported by Tang and Toma
(12) to detect Legionnaires' disease caused by numerous
serogroups of L. pneumophila (13) and additional
species of Legionella (11). We compared the
abilities of the Binax Legionella urinary antigen
radioimmunoassay (RIA) and/or EIA kit and the Biotest urine antigen kit
to detect Legionella in urine samples that were positive in
the broad-spectrum EIA and had been confirmed by either culture or the
indirect immunofluorescence assay.
Urine samples were submitted to the Laboratory Branch, Ontario Ministry
of Health and Long Term Care, for testing between 1986 and 1998. If
available, lower respiratory tract specimens were obtained for culture
and direct fluorescent antibody staining by using standard procedures
for Legionella detection (1). In addition, paired
serum samples were obtained for antibody determination by the indirect
immunofluorescence assay using formalized antigens (1).
Forty-five samples that were positive in the broad-spectrum enzyme-linked immunosorbent assay were further evaluated by using the
Binax Equate Legionella urinary antigen RIA and/or the EIA kit. A sufficient quantity of sample was available for 42 of these samples from 39 patients who were tested with the Biotest
Legionella urine antigen EIA (Biotest AG). All samples were
tested according to each manufacturer's directions.
The results of testing the 45 urine samples with the Biotest and Binax
RIA and/or EIA are shown in Table 1.
Twenty-six of these samples were from patients with pneumonia caused by
L. pneumophila non-serogroup 1. There were 19 samples from
patients diagnosed with Legionnaires' disease from a species other
than L. pneumophila. Although the samples were tested with
both the Binax RIA and EIA, the results of the EIA were used for
comparison to the Biotest. The samples had been stored at
0095-1137/00/$04.00+0
Evaluation of the Binax and Biotest Urinary Antigen
Kits for Detection of Legionnaires' Disease Due to Multiple
Serogroups and Species of Legionella
![]()
ABSTRACT
Top
Abstract
Text
References
![]()
TEXT
Top
Abstract
Text
References
70°C for
more than 6 months prior to testing with the EIA kit. Of the 45 samples
that were previously positive with the broad-spectrum ELISA, 10 were
positive with the Binax Equate Legionella urinary antigen
EIA kit. Nine of the L. pneumophila non-serogroup 1 samples
were positive. There was one patient with culture-confirmed disease due
to L. pneumophila serogroup 6 with a ratio of 2.5 or higher,
which would have been considered probable by the criteria suggested by
Hackman et al. (5). One of the 19 samples from patients with
disease due to species of Legionella other than L. pneumophila were positive. This case was due to Legionella
hackeliae serogroup 2. There were eight samples which had been
positive in the RIA which were negative in the EIA after storage.
TABLE 1.
Results for Binax RIA and EIA and Biotest urine
antigen test
The results of the Biotest are also listed in Table 1. Thirteen of 42 samples were positive. Twelve of these were from persons with Legionnaires' disease due to L. pneumophila non-serogroup 1 and one was from a patient with disease caused by L. hackeliae serogroup 2. This sample was also positive in the Binax test. Two of the L. pneumophila non-serogroup 1 samples that were positive in the Biotest and negative in the Binax EIA were positive in the RIA. The Biotest identified all the samples which were positive in the Binax EIA except one.
The initial testing of urine samples was performed using the RIA kit. The samples were subsequently stored and retested using both the Binax and Biotest EIA kits. For comparison only the results of the EIA were used, because previous reports have shown that some urine samples can become negative after long-term storage (10). There were eight samples that were positive with the Binax RIA but negative in the EIA. The Binax kit was not as sensitive as the broad-spectrum ELISA, detecting 10 of the 45 (22%) samples tested or 15 of 45 (33%) if a ratio of >2.0 is used, as suggested by Domínguez et al. (3, 4). Previous reports have shown that all urine samples with a ratio above 2.0 were positive after concentration of the soluble antigen. Previous reports of positive results with the Binax RIA and/or EIA kit for non-serogroup 1 disease have been associated with other serogroups of L. pneumophila, specifically serogroups 4 and 10 (8) and serogroup 5 (2).
The Biotest kit identified 13 of 42 (31%) samples tested. If a lower cutoff were used based on personal observation of the EIA plate, 20 (48%) samples would have been positive. There was agreement on nine of the samples, and the Biotest identified four samples that were negative in the Binax test. All of these samples were from patients with culture-confirmed disease due to L. pneumophila serogroup 6.
Previous comparison of the two tests by Domínguez et al. (3) was performed with urine samples from patients with pneumonia due to L. pneumophila confirmed by culture (no serogroup was given) or seroconversion and patients with a prior positive result using the Binax kit. Their evaluation included one sample from a patient with pneumonia due to L. longbeachae that was negative in both tests. However, both tests were able to detect soluble antigen from culture extracts from all L. pneumophila serogroups tested and L. bozemanii. These authors stated the major advantage of the Biotest kit over the Binax kit was the broad spectrum of cross-reactivity with antigens of other serogroups and species. Our evaluation showed that both kits were capable of detecting multiple serogroups of L. pneumophila.
In a multicenter evaluation of the Biotest compared to the Binax and an in-house-developed assay, the Biotest was negative for eight urine samples from culture-proven cases of non-serogroup 1 L. pneumophila. The eight samples were positive by using either the Binax or in-house assay. However, the number that were positive with the Binax kit alone was not provided (6). The Biotest was positive for two of three urine samples from patients with serological evidence of infection to L. pneumophila non-serogroup 1. In an evaluation of the Binax EIA kit, Kazandjian et al. (7) tested 12 samples from culture-confirmed non-serogroup 1 L. pneumophila legionellosis, and all were negative. They concluded that the Binax EIA can detect only L. pneumophila serogroup 1 antigen. Our results show that this is not the case, which has importance for surveillance and epidemiology. Currently most, if not all, investigators using the Binax test report a positive urine result as being positive for L. pneumophila serogroup 1. Considering all urine antigen-positive cases as being due to L. pneumophila serogroup 1 would result in a higher percentage of cases being reported as possibly being caused by this particular serogroup.
The broad-spectrum assay is more sensitive for detecting urine antigen both from L. pneumophila of different serogroups and from other Legionella species than either of the commercial assays. The broad-spectrum assay was previously reported to have a sensitivity of 70% and a specificity of nearly 100% (11). Further evaluations performed between 1992 and July 1996 for 1,397 patients with both a respiratory tract specimen and a urine specimen resulted in a specificity of 99.4%, a sensitivity of 77%, and a positive predictive value of 82%.
We did not have sufficient amounts of urine samples available to perform concentration experiments with the Binax and Biotest assays. This procedure has not been evaluated with the broad-spectrum ELISA. The enhanced sensitivity of both the Binax and Biotest when using concentrated urine from patients with L. pneumophila serogroup 1 has been shown (3, 4), and it is likely that this would apply to urine samples from cases caused by other Legionella species.
These results demonstrate that neither the Binax nor Biotest assay is as sensitive as the broad-spectrum ELISA; however, both are capable of detecting Legionnaires' disease due to non-serogroup 1 L. pneumophila and other Legionella species.
| |
FOOTNOTES |
|---|
* Corresponding author. Mailing address: Respiratory Diseases Branch, Division of Bacterial and Mycotic Diseases, National Center for Infectious Diseases, Mailstop GO3, Centers for Disease Control and Prevention, Atlanta, GA 30333. Phone: (404) 639-3563. Fax: (404) 639-4215. E-mail: REB3{at}CDC.GOV.
| |
REFERENCES |
|---|
|
|
|---|
| 1. | Centers for Disease Control. 1987. Hospital-laboratory diagnosis of Legionella infections. Centers for Disease Control, Atlanta, Ga. |
| 2. | Chang, F.-Y., S. L. Jacobs, S. M. Colodny, J. E. Stout, and V. L. Yu. 1996. Nosocomial Legionnaires' disease caused by Legionella pneumophila serogroup 5: laboratory and epidemiologic implications. J. Infect. Dis. 174:1116-1119[Medline]. |
| 3. |
Domínguez, J. A.,
N. Galí,
P. Pedroso,
A. Fargas,
E. Padilla,
J. M. Manterola, and L. Matas.
1998.
Comparison of the Binax Legionella urinary antigen enzyme immunoassay (EIA) with the Biotest Legionella Urin antigen EIA for detection of Legionella antigen in both concentrated and nonconcentrated urine samples.
J. Clin. Microbiol.
36:2718-2722 |
| 4. | Domínguez, J. A., J. M. Manterola, R. Blavia, N. Sopena, F. J. Belda, E. Padilla, M. Giménez, M. Sabrià, J. Morera, and V. Ausina. 1996. Detection of Legionella pneumophila serogroup 1 antigen in nonconcentrated urine and urine concentrated by selective ultrafiltration. J. Clin. Microbiol. 34:2334-2336[Abstract]. |
| 5. | Hackman, B. A., J. F. Plouffe, R. F. Benson, B. S. Fields, and R. F. Breiman. 1996. Comparison of Binax Legionella urinary antigen EIA kit with Binax RIA urinary antigen kit for detection of Legionella pneumophila serogroup 1 antigen. J. Clin. Microbiol. 34:1579-1580[Abstract]. |
| 6. | Harrison, T., S. Uldum, S. Alexiou-Daniel, J. Bangsborg, S. Bernander, V. Drasar, J. Etienne, J. Helbig, D. Lindsay, I. Lochman, T. Marques, F. de Ory, I. Tartakovskii, G. Wewalka, and F. Fehrenbach. 1998. A multicenter evaluation of the Biotest Legionella urinary antigen EIA. Clin. Microbiol. Infect. 4:359-365[Medline]. |
| 7. | Kazandjian, D., R. Chiew, and G. L. Gilbert. 1997. Rapid diagnosis of Legionella pneumophila serogroup 1 infection with the Binax enzyme immunoassay urinary antigen test. J. Clin. Microbiol. 35:954-956[Abstract]. |
| 8. | Kohler, R. B., L. J. Wheat, M. L. V. French, P. L. Meenhorst, W. C. Winn, Jr., and P. H. Edelstein. 1985. Cross-reactive urinary antigens among patients infected with Legionella pneumophila serogroups 1 and 4 and the Leiden 1 strain. J. Infect. Dis. 152:1007-1012[Medline]. |
| 9. | Plouffe, J. F., T. M. File, R. F. Breiman, B. A. Hackman, S. J. Salstrom, B. J. Marston, B. S. Fields, and the Community Based Pneumonia Incidence Study Group. 1995. Reevaluation of the definition of Legionnaires' disease: use of the urinary antigen assay. Clin. Infect. Dis. 20:1286-1291[Medline]. |
| 10. | Rigby, E. W., J. F. Plouffe, B. A. Hackman, D. S. Hill, R. F. Benson, and R. F. Breiman. 1997. Stability of Legionella urinary antigens over time. Diagn. Microbiol. Infect. Dis. 28:1-3[CrossRef][Medline]. |
| 11. | Tang, P., and C. Krishnan. 1993. Legionella antigenuria: six-year study of broad-spectrum enzyme-linked immunosorbent assay as a routine diagnostic test, p. 12-13. In J. M. Barbaree, R. F. Breiman, and A. P. Dufour (ed.), Legionella: current status and emerging perspectives. American Society for Microbiology, Washington, D.C. |
| 12. |
Tang, P. W., and S. Toma.
1986.
Broad-spectrum enzyme-linked immunosorbent assay for detection of Legionella soluble antigens.
J. Clin. Microbiol.
24:556-558 |
| 13. |
Tang, P. W.,
S. Toma, and W. D. Rajkumar.
1989.
Detection of urinary antigens of Legionella pneumophila serogroup 12 by broad-spectrum enzyme-linked immunosorbent assay.
J. Clin. Microbiol.
27:783-784 |
This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»