Journal of Clinical Microbiology, December 2001, p. 4302-4308, Vol. 39, No. 12
Institute of Transfusion Medicine and
Immunohematology, Red Cross Blood Donor Service, Frankfurt am Main,
Germany
Received 21 March 2001/Returned for modification 19 July
2001/Accepted 14 September 2001
Screening of blood donors for human immunodeficiency virus type 1 (HIV-1) infection by PCR permits the earlier diagnosis of HIV-1
infection compared with that by serologic assays. We have established a high-throughput reverse transcription (RT)-PCR assay based on 5'-nuclease PCR. By in-tube detection of HIV-1 RNA with a fluorogenic probe, the 5'-nuclease PCR technology (TaqMan PCR) eliminates the risk of carryover contamination, a major problem in PCR
testing. We outline the development and evaluation of the PCR assay
from a technical point of view. A one-step RT-PCR that targets the
gag genes of all known HIV-1 group M isolates was developed. An internal control RNA detectable with a heterologous 5'-nuclease probe was derived from the viral target cDNA and was packaged into MS2 coliphages (Armored RNA). Because the RNA was protected against digestion with RNase, it could be spiked into patient
plasma to control the complete sample preparation and amplification
process. The assay detected 831 HIV-1 type B genome equivalents per ml
of native plasma (95% confidence interval [CI], 759 to 936 HIV-1 B
genome equivalents per ml) with a To date, serologic enzyme
immunoassay-based antibody testing plus confirmation by Western
blotting are "gold standards" for the screening of blood donors for
human immunodeficiency virus (HIV) type 1 (HIV-1) infection. Due to the
lack of detectable antibody early after exposure to HIV-1, these assays
fail to detect a recent HIV infection. They leave a diagnostic window
during which the viral load is high (on the order of
106 to 109 genome
equivalents per ml of donor plasma) (1, 2, 4, 15, 16).
Transmission of HIV-1 by blood products from individuals who donated
blood during this preseroconversion window period has been reported
previously (23). Schreiber et al. have calculated the
remaining residual risk of HIV-1 transmission from serologically negative blood to range from 1/202,000 to 1/2,778,000
(20). Screening for HIV-1 by PCR is expected to reduce the
window period before seroconversion to positivity for HIV by 11 days
(20).
Although the feasibility of screening of blood donors for HIV-1 by
reverse transcription (RT)-PCR has been demonstrated, difficulties with
post-PCR processing, assay stability, and interpretation of results
restrict its use to specialized laboratories (18). It was
our aim to establish a high-throughput PCR assay for screening for
HIV-1 infection that provides reliable results without the need for a
sophisticated means of interpretation of results. We have used
the 5'-nuclease PCR technology (TaqMan PCR), in which a fluorescent
signal is generated in the course of amplification and the result is
reported through the closed cap of the reaction tube. A probe labeled
with a reporter dye and a quencher dye is added to the reaction mixture
prior to amplification. Fluorescence in the intact probe is quenched by
trans-molecular energy transfer (22). Due to its intrinsic
5'-nuclease activity, Taq DNA polymerase digests the probe
after it specifically anneals to the PCR product generated in each
cycle. This liberates the reporter dye from the quencher dye, causing
an increase in reporter dye-specific fluorescence during PCR. The risk
of post-PCR contamination is thereby eliminated because the reaction
tubes do not have to be opened for product detection (8,
12).
When large numbers of patient samples are routinely tested by PCR,
amplification failures due to inhibition of the reaction must be
efficiently monitored to avoid false-negative results. Inhibition of
PCR is usually detected by spiking a patient's sample with a weakly
positive control after the nucleic acids are prepared or while the
nucleic acids are being prepared (3, 10, 17). This
procedure does not allow control of the complete nucleic acid
preparation process, including lysis of the pathogen. To implement
monitoring of the complete preparation, we have used a phage-packaged
internal control RNA sequence (Armored RNA) that is coextracted with
the viral RNA from plasma (14), coamplified in the same
reaction tube, and detected with a heterologous probe.
In this report, we outline the development and evaluation of the assay
from a technical point of view, focusing on analytical sensitivity
and specificity and on technical feasibility for the qualitative
high-throughput testing of single and pooled samples. The report first
provides, however, preliminary data on clinical performance obtained by
testing a panel of plasma samples in which the HIV-1 load was
quantified, a panel of plasma samples from patients infected
with different subtypes of HIV-1 group M, and two panels of follow-up
plasma from patients who have seroconverted.
HIV-1 sequence alignment.
The 1996 alignment of consensus
HIV-1 sequences of the Los Alamos National Laboratory was downloaded
from ftp://ncbi.nlm.nih.gov.
HIV-1-positive plasma samples.
Plasma samples from 39 individual patients with a confirmed diagnosis of HIV-1 infection were
kindly provided by H. Rabenau, Institute of Virology, University of
Frankfurt. The viral loads in all samples had previously been
determined by the Amplicor Monitor (version 1.5) HIV-1 RT-PCR assay
(Roche Molecular Systems, Branchburg, N.J.). The range of viral loads
in the samples was <400 to 106 genome
equivalents (geq)/ml, with 13 samples having loads below 20,000 geq/ml
and 8 samples having loads below 2,000 geq/ml.
Reference samples.
HIV-1-positive plasma samples in which
the HIV-1 loads were quantified were obtained from INSTAND e.V.,
Düsseldorf, Germany, a national reference center that provides
material for proficiency testing. The material contained 40,000 geq of
HIV-1 subtype B per ml of plasma.
0095-1137/01/$04.00+0 DOI: 10.1128/JCM.39.12.4302-4308.2001
Copyright © 2001, American Society for Microbiology. All rights reserved.
TaqMan 5'-Nuclease Human Immunodeficiency Virus
Type 1 PCR Assay with Phage-Packaged Competitive Internal Control for
High-Throughput Blood Donor Screening

![]()
ABSTRACT
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
95% probability of a
positive result, as determined by probit regression analysis. A
detection limit of 1,195 genome equivalents per ml of (individual) donor plasma (95% CI, 1,014 to 1,470 genome equivalents per ml of
plasma pooled from individuals) was achieved when 96 samples were pooled and enriched by centrifugation. Up to 4,000 plasma samples
per PCR run were tested in a 3-month trial period. Although data from
the present pilot feasibility study will have to be complemented by a
large clinical validation study, the assay is a promising approach to
the high-throughput screening of blood donors and is the first
noncommercial test for high-throughput screening for HIV-1.
![]()
INTRODUCTION
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
![]()
MATERIALS AND METHODS
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
Seroconversion panels. Two HIV-1 seroconversion panels were purchased from BioClinical Partners, Inc., Franklin, Mass. The patient identification numbers were 60722 and 62238. All material had been pretested with PRISM and AXSYM automated anti-HIV-1/2 assays (Abbott, Inc., Abbott Park, Ill.).
The viral loads of the samples were quantified with the Roche Amplicor Monitor HIV-1 RT-PCR assay system for patient 62238 and with an independent quantitative RT-PCR system for patient 60722. The sensitivity limits of the quantitative PCR assays were 400 and 100 geq/ml, respectively.Dilution buffer for Armored RNA phages. The TSM Armored RNA dilution buffer contained 10 mM Tris (pH 7.0), 100 mM NaCl, 1 mM MgCl2, 0.3% sodium azide, and 0.1% bovine gelatin (tissue culture grade). All substances were purchased from Sigma-Aldrich, Munich, Germany.
Nucleic acid extraction. By using the viral RNA kit from Qiagen, Hilden, Germany, 100 µl of serum was processed according to the manufacturer's instructions. For the elution of RNA, the columns were incubated with 75 µl of RNase-free water at 80°C. Elutes were stored at 4°C until analysis.
Plasma pooling and virus concentration. After pooling of 100-µl aliquots of 96 blood donor samples on an automated pipetting station (Genesis; TECAN, Crailsheim, Germany), the complete 9.6 ml was centrifuged at 48,000 × g for 60 min. The supernatant was decanted and the virus pellets were resuspended in 100 µl of water. Further procedures were the same as those used for the plasma samples from single donors (18).
Primer and probe design. Oligonucleotides were designed with Primer Express software, supplied by Perkin-Elmer, Weiterstadt, Germany. The degree of nucleotide sequence homology of all oligonucleotides to sequences other than the HIV-1 sequence was checked by using the EMBL, GenBank, and DDBJ databases and the BLAST algorithm (www.ncbi.nlm.nih.gov/BLAST), which searches the EMBL, GenBank, and DDBJ databases.
Oligonucleotides in screening assay. Primers cdis (ATC AAG CAG CCA TGC AAA TGT T and ACC AGG CAG CTA TGC AAA TGT T, equal amounts) and cdia (TG AAG GGT ACT AGT AGT TCC TGC TAT GTC and CTG AAG GGT ACT AGT AGT TCC TGC TAT ATC; equal amounts) were used to amplify a conserved 152-bp region spanning nucleotides 578 to 730 of the HIV-1 subtype B gag gene (the numbering is according to the Los Alamos National Laboratory HIV-1 consensus sequence). Probe cdso29 (ACC ATC AAT GAG GAA GCT GCA GAA TGG GA; positions 607 to 636) was used to detect HIV-1 PCR products. It was labeled with 6-carboxyfluorescein (FAM) at its 5' end and 6-carboxy-N,N,N',N'-tetramethylrhodamine (TAMRA) at its 3' end. Probe cdsti (ACG ACG GAC CAC ACT GAC CAG TCA) was used to detect an alternative probe binding site in the internal control sequence, which had been generated by replacement of nucleotides 605 to 637 by site-directed mutagenesis. It was labeled with 6-carboxy-4,7,2',7'-tetrachlorofluorescein (TET) at its 5' end and TAMRA at its 3' end. Both of the probes were phosphorylated at their 3' ends to prevent elongation during PCR.
Construction of internal control sequence. Moloney murine leukemia virus Superscript reverse transcriptase (Life Technologies, Karlsruhe, Germany) was used for synthesis of cDNA from HIV-1 subtype B RNA, according to the manufacturer's instructions, with primer Bas (GT TCC TGA AGG GTA CTA GTA GTT CCT GCT ATG TC) as an RT primer. Primers As (G CTT TCA GCC CAG AAG TAA TAC CCA TG) and Bas were used to amplify a 307-bp fragment that included genome positions 485 to 729. Gene splicing by overlap extension was carried out for construction of an internal control sequence (5, 7, 19) by using primers Aas (CTC TGG TAG TTA CTC CTT CGA CGT CTT ACC CTA TTT TAA CAT TTG CAT GGC TGC TTG ATG) and Bs (TAG GGT AAG ACG TCG AAG GAG TAA CTA CCA GAG AGA TTG CAT CCA GTG CAT GCA GG) to introduce an alternative probe binding site into the PCR product at nucleotide positions 605 to 637. The resulting sequence constructs were cloned into pCR II vectors (Invitrogen, Carlsbad, Calif.) to generate plasmid pHIVSTD1.11. The integrities of the plasmid insert sequences, including confirmation of the presence of the alternative probe binding site, were confirmed with the Taq dye deoxy terminator cycle sequencing kit (Perkin-Elmer) and an ABI 310 automatic DNA sequencer (Applied Biosystems, Weiterstadt, Germany) according to the manufacturer's instructions. Primer S400 (GCT TTC AGC CCA GAA GTA ATA CCC ATG; nucleotides 485 to 511) was used as a sequencing primer. Packaging of internal control RNA into MS2 coliphages was performed by B. Pasloske, Ambion, Inc., Austin, Tex., with plasmid pHIVSTD1.11 as the template. The preparation procedure and the properties of the Armored RNA have been described previously (14). The Armored RNA stock solution contained 1014 particles per ml after preparation. For use as an internal control sequence, Armored RNA was spiked into plasma or resuspended virus pellets before nucleic acid extraction. The Armored RNA solution was custom designed for this work and is not identical to a similar product available from the same company.
5'-Nuclease HIV-1 RT-PCR. PCR was performed in a volume of 25 µl containing 50 mM KCl; 10 mM Tris-HCl (pH 8.3); 0.01 mM EDTA; 60 nM 6-carboxy-x-rhodamine; 0.2% (vol/vol) polyethylene glycol 6000 (Fluka Chemicals, Munich, Germany); 0.05% bovine gelatin (Sigma-Aldrich); 3 mM MgCl2; dATP, dCTP, dGTP, and dTTP each at a concentration of 200 µM; primers cdis and cdia each at a concentration of 400 nM; 300 nM probe cdso29; 100 nM probe cdsti; 6 U of RNase-out RNase inhibitor; 6 U of Moloney murine leukemia virus Superscript reverse transcriptase (Life Technologies); and 0.75 U of AmpliTaq Gold DNA polymerase. A total of 10 µl of nucleic acid extract was added to complete the reaction volume of 25 µl. Thermal cycling in a Perkin-Elmer 7700 thermal cycler involved preincubation at 40°C for 20 min, denaturation at 95°C for 7 min, 10 cycles of 95°C for 35 s and 62°C for 45 s with an annealing temperature decrement of 0.2°C per cycle, and 40 cycles of 93°C for 10 s and 58°C for 30 s with an annealing temperature increment of 0.2°C per cycle. Calculation and interpretation of fluorescence data either during or immediately after PCR have been described previously (5). A reaction was considered positive for wild-type HIV sequences when there was a signal from the FAM emission wavelength (518 nm), negative when there was a signal from the TET emission wavelength only (538 nm), and nonvalid when there was no signal. Reagents were purchased from Perkin-Elmer unless stated otherwise. The contamination precautions of Kwok and Higuchi were strictly complied with (11).
Statistical analysis. The relationship between the proportion of 31 (direct testing) or 24 (testing of pooled samples) positive samples in replicate determinations with different concentrations of HIV-1 RNA as inputs was examined by probit analysis as a model for nonlinear regression (6). The analysis was performed with the Statgraphics plus (version 5.0) software package (Statistical Graphics Corp.).
| |
RESULTS |
|---|
|
|
|---|
Sensitivity and specificity for HIV-1 in plasma samples from single
patients and sensitivity for various HIV-1 subtypes.
To determine
the analytical sensitivity of the assay, we used plasma from a national
proficiency study in which the HIV-1 subtype B load was quantified. The
plasma was diluted in pooled negative plasma to 2,500, 1,000, 800, and
600 geq/ml for replicate amplification reactions by the protocol given
in the Materials and Methods section. Thirty-one replicate reactions
were performed with each RNA input concentration on 2 different days to
minimize the effect of handling errors. The proportions of
positive results obtained for each input concentration were subjected
to probit regression analysis. The software determined a continuous
95% confidence interval (CI) of the probability that a positive result would be achieved at any given input concentration within the concentration range of the experiment. Table
1 summarizes the experimental results,
and Fig. 1A depicts the results of probit analysis. The concentration at which 95% of results are expected to be
positive was calculated to be 831 geq/ml (95% CI, 759 to 936 geq/ml).
The percent deviance (for the proportion of samples with positive
results) explained by the model was 99.273, proving the adequacy
of the model.
|
|
Implementation of Armored RNA internal control sequence.
For
implementation of an internal amplification control reaction, which
would be useful for continuous monitoring of the sample preparation
process, we determined the lowest possible number of copies of Armored
RNA internal control phages necessary to avoid detection
failures due to statistical distribution effects (Poisson
distribution). We spiked decreasing amounts of Armored RNA diluted in
TSM buffer into 100 µl of plasma before nucleic acid extraction.
Nucleic acid extracts were amplified by the PCR protocol given above.
Table 2 summarizes the results.
|
Sensitivity for pooled plasma samples from 96 donors.
Analytical sensitivity after pooling of 96 plasma samples from blood
donors was determined with the same quantified HIV-1 subtype B-positive
plasma sample mentioned above. Dilution series of HIV-1 subtype
B-positive plasma were prepared independently on 3 different days by
three different technicians. Samples from each of six concentration
steps were then spiked into eight replicate plasma pools of 9.5 ml each. Viral particles were concentrated from pools as
described in the Materials and Methods section. Pellets were
resuspended in 100 µl of water, inoculated with 750 particles of the
Armored RNA internal control, and extracted with the Qiagen viral RNA
kit. The proportions of 24 replicate reactions positive at each input
RNA concentration were subjected to probit regression analysis as
described above. Table 3 summarizes the experimental results, and Fig. 1B shows the results of probit analysis.
The RNA input concentration at which 95% of tests are expected to be
positive was 1,195 geq/ml (95% CI, 1,014 to 1,470 geq/ml). Notably,
the datum point obtained for 1,250 geq/ml appeared to be an outlier,
giving a falsely low value. The zero value was overweighted by a
factor of 10 to obtain a probit curve that fit the data instead
of excluding the outlier, which otherwise would have caused a bias
toward a lower sensitivity limit. The percent deviance (for the
proportion of positive samples) explained by the model was
86.32.
|
Technical feasibility and specificity of procedure for day-to-day application. To assess whether the TaqMan HIV-1 PCR is technically suitable for day-to-day application, the assay was used in parallel with our established blood donor screening PCR (18) to screen pools of plasma in a 3-month test period. During this time, 1,791 pools were tested in 89 individual test runs, with a maximum of 4,000 individual plasma samples included in one run. In each run, two low-positive control plasma samples (2,500 geq of HIV-1 RNA per ml) were included, one of which was treated as single patient plasma sample and the other of which was inoculated into a negative plasma pool and concentrated as described above. All of the resulting 178 positive controls were detected.
In 51 of 1,791 reactions the internal control failed to be amplified, with the result being that 2.85% (95% CI, 2.13 to 3.73%) of pools had to be retested. Upon retesting of the same nucleic acid extract, the internal control was detected in 45 of these reactions. In the remaining six plasma pools the internal control was repeatedly undetectable. Thus, inhibitory substances were present in or an insufficient amount of Armored RNA was recovered from only 0.34% (95% CI, 0.12 to 0.73%) of plasma pools tested. When the complete preparation process for these remaining six pools from which no internal control was recovered was repeated starting from the pooling of the original samples, the internal control was detected in all pools. Consequently, inhibitory substances did not originate from patient plasma but came from the sample preparation process. Five false-positive reactions occurred among the 1,791 pools tested. The results for all five pools could be resolved to be true negative by four independent negative retests: two by the same method and two by the established routine PCR. The specificity was therefore 99.72% (95% CI, 99.48 to 99.97%). To complement the specificity data obtained from the trial period, the oligonucleotides used in the assay were tested for their alignments with sequences in DNA databases. No significant primer or probe sequence homologies other than with HIV or HIV-human fusion sequences were found with sequences in the EMBL, GenBank, and DDBJ databases by a search with the BLAST program, available at the National Center for Biotechnology Information website.Detection of low levels of HIV-1 RNA in early infection.
To
compare the performance of our HIV-1 RT-PCR to those of established
quantitative PCR systems, we tested two different seroconversion panels
from U.S. plasma donors in which the HIV load had been quantified by
PCR. An aliquot of 100 µl of plasma from each patient on each
individual bleeding date was inoculated into a pool of plasma from 95 individuals pretested and found to be HIV-1 negative by RT-PCR.
The total volume of 9.6 ml was processed under routine conditions for
screening of plasma from blood donors by PCR, as described above. The
results were also compared to those obtained by the Abbott PRISM and
AXSYM anti-HIV-1/2 assays and provided by the supplier of the
seroconversion panels. These assays are commonly used by blood
transfusion services. Viral loads and the data obtained by serologic
testing of each sample are summarized in
Table 4. Viral RNA could not be detected by our pooling assay in one low-positive sample (patient 62238; bleed date, 9 September 1996) that contained 1,180 geq/ml of plasma, but viral RNA was detected
with the Roche Amplicor Monitor HIV-1 RT-PCR assay system upon testing
of a single sample. For patients 60772 and 62238, PCR detection of
HIV-1 RNA was achieved 9 and 12 days earlier, respectively, than
anti-HIV-1 detection by both serologic assays.
|
| |
DISCUSSION |
|---|
|
|
|---|
High-throughput screening for HIV-1 by RT-PCR requires handling by experienced laboratory personnel and tremendous laboratory effort. In particular, the risk of contamination during product detection after thermal cycling restricts diagnostic PCRs to experienced laboratories.
Our HIV-1 TaqMan 5'-nuclease PCR provides a screening assay that does not require time-consuming and contamination-prone post-PCR processing steps; reaction vials stay closed after amplification and are discarded, with no need to open them.
In addition to contamination, unspecific by-products infrequently generated by PCR can contribute to false-positive screening results by misinterpretation of DNA bands in an agarose gel (18). These errors can be eliminated by hybridization with a specific probe. However, classical hybridization procedures, like Southern blotting or dot blotting, increase the assay time and the contamination risk even more. The TaqMan 5'-nuclease PCR overcomes this problem by in-tube hybridization, which is suitable for high-throughput screening. As a result, the specificity of 99.72% that was achieved is superior to that of our established routine method (18).
Use of an internal control sequence enables verification of the sensitivity for each individual reaction tube, while it circumvents the need for additional labor-intensive and costly external control reactions (3, 10, 17). Because the target DNA sequences with different base compositions undergo different amplification conditions in the PCR (9), we used a competitive internal control sequence to monitor PCR failures. The base composition of the internal control sequence including the primer binding sites is nearly identical to that of wild-type HIV-1 RNA; hence, the amplification properties for the internal control sequence exactly reflect those of the wild type. Packaging of the internal control RNA in a phage (Armored RNA) makes the control RNA resistant to RNase degradation (14). Because the Armored RNA can be spiked into donor plasma like a virus without being degraded by ubiquitous RNases, this format enables monitoring of the complete nucleic acid extraction and amplification process for each specimen. The concentration of Armored RNA inoculated into each sample is about 10 times the detection limit of the assay. Thus, a 10-fold drop in efficiency of any step of the preparation or amplification procedure will likely be detected. The observed proportion of tests that had to be repeated due to failure of the internal control (2.85%) does not prohibit the day-to-day application of the assay. It appeared that the amplification part of the procedure is more susceptible to errors than the preparation part of the procedure, since only 11.8% of controls that failed to be amplified were repeatedly negative when the same RNA extract was retested. In contrast to earlier findings from tests with single samples (17), confirmed failure of detection resulted from the nucleic acid preparation process and not from the samples themselves. This is attributable to the fact that each patient plasma sample is diluted by inclusion in a pool of 96 samples; any inhibitors possibly present in single samples are similarly diluted.
The analytical sensitivity for testing of plasma from single donors and
pools of plasma has been determined by probit regression analysis of
the proportions of samples positive in replicate reactions with
different input concentrations. This calculation has proved adequate
for description of the sensitivity limits of PCR assays (5,
21). Expressed as the RNA input concentration at which
95% of
the test reactions are positive, sensitivity limits were 831 geq/ml for
plasma from single donors and 1,195 geq/ml for pooled plasma (with
respect to a single positive sample in a pool). The small reduction in
sensitivity observed despite the pooling of plasma is attributable to
the high efficiency of concentration of HIV-1 particles by
centrifugation. It is consistent with the similarly high concentration
efficiency achieved by the ultrasensitive protocol of the Roche
Amplicor Monitor assay. By the Roche assay, samples are centrifuged at
only 20,000 × g for 1 h. We used 48,000 × g instead because the same pools are used to test for
hepatitis C and B viruses, and concentration of hepatitis C virus in
particular has proved to require high centrifugation forces
(personal observation).
It has been shown by others that screening of plasma from blood donors by PCR with a sensitivity limit in the range of 103 geq/ml can significantly reduce the window during which HIV-1 infection fails to be diagnosed by serological assays (16, 23). Murthy et al. have indicated that this may be sufficient to completely eliminate the window (13). The reductions of the window periods for two seroconverting patients obtained in the present study (9 and 12 days, respectively) are in concordance with the calculations of Schreiber et al., who proposed a reduction of 11 days (20).
Preliminary data on the clinical sensitivity of our assay in its
intended application (testing of German blood donors) have been
obtained with a small panel of plasma samples from German HIV-1
RNA-positive patients. The low viral loads (
800 geq/ml) in the three
samples for which the assay failed to detect HIV-1 RNA, as well as the
viral load in one sample from the seroconversion panels for which the
assay failed to detect HIV-1 RNA (1,180 geq/ml), are in concordance
with the analytical sensitivity limits determined for our assay. This
also implies that false-negative reactions due to primer-probe
mismatches occurred in none of the remaining samples. The observation
that HIV-1 RNA was detected in plasma samples containing subtypes A, B,
C, D, E, and F is another indicator of the good diagnostic performance
of the assay. However, for a comprehensive determination of the
clinical sensitivity of the assay, a large validation study is
required. This would include comparative testing of various
preseroconversion plasma panels by our method and established
commercial procedures, e.g., the Roche AmpliScreen or GenProbe TIGRIS
assays. Furthermore, testing of low-positive samples from patients
infected with HIV-1 clades that are rare in Germany, including HIV-1
subtype O, will be necessary. Moreover, validation of the assay for the
quantification of viral loads in patients, which should be feasible,
given the technical possibilities of real-time PCR, may be profitable
for clinical laboratories.
However, the data obtained in the present pilot feasibility study are promising with respect to clinical application of the assay. Our data suggest that recent advances in PCR technology make it feasible to screen for viral pathogens like HIV-1 with high sample throughput without being dependent on commercially provided testing formats. To our knowledge, this is the first noncommercial TaqMan HIV-1 RT-PCR assay available for this purpose.
| |
ACKNOWLEDGMENTS |
|---|
We are grateful to B. Rüster for helpful discussions and to C. Weis and S. Buhr for expert technical assistance.
| |
FOOTNOTES |
|---|
* Corresponding author. Mailing address: Institut für Transfusionsmedizin und Immunhämatologie, Blutspendedienst des DRK Hessen, Sandhofstr. 1, D-60528 Frankfurt am Main, Germany. Phone: 0049-69-6782-251. Fax: 0049-69-6782-256. E-mail: wroth{at}bsdhessen.de.
Present address: Bernhard Nocht Institute of Tropical Medicine,
Hamburg, Germany.
| |
REFERENCES |
|---|
|
|
|---|
| 1. | Busch, M. P., L. L. Lee, G. A. Satten, D. R. Henrard, H. Farzadegan, K. E. Nelson, S. Read, R. Y. Dodd, and L. R. Petersen. 1995. Time course of detection of viral and serologic markers preceding human immunodeficiency virus type 1 seroconversion: implications for screening of blood and tissue donors. Transfusion 35:91-97[CrossRef][Medline]. |
| 2. | Clark, S. J., M. S. Saag, W. D. Decker, S. Campbell-Hill, J. L. Roberson, P. J. Veldkamp, J. C. Kappes, B. H. Hahn, and G. M. Shaw. 1991. High titers of cytopathic virus in plasma of patients with symptomatic primary HIV-1 infection. N. Engl. J. Med 324:954-960[Abstract]. |
| 3. |
Cone, R. W.,
A. C. Hobson, and W. H. Meei-Li.
1992.
Coamplified positive control detects inhibition of polymerase chain reaction.
J. Clin. Microbiol.
30:3185-3189 |
| 4. | Daar, E. S., T. Moudgil, R. D. Mayer, and D. D. Ho. 1991. Transient high levels of viremia in patients with primary human immunodeficiency virus type 1 infection. N. Engl. J. Med. 324:961-964[Abstract]. |
| 5. | Drosten, C., M. Weber, E. Seifried, and W. K. Roth. 2000. Evaluation of a new PCR assay with competitive internal control sequence for blood donor screening. Transfusion 40:718-724[CrossRef][Medline]. |
| 6. | Finney, D. J. 1971. Probit analysis, 3rd ed. Cambridge University Press, London, United Kingdom. |
| 7. | Higuchi, R. 1989. Using PCR to engineer DNA, p. 63-69. In H. A. Ehrlich (ed.), PCR technology. Principles and applications for DNA amplification. Stockton Press, New York, N.Y. |
| 8. |
Holland, P. M.,
R. D. Abramson,
R. Watson, and D. H. Gelfand.
1991.
Detection of specific polymerase chain reaction product by utilizing the 5' to 3'-exonuclease activity of Thermus aquaticus DNA polymerase.
Proc. Natl. Acad. Sci. USA
88:7276-7280 |
| 9. | Innis, M. A., and D. H. Gelfand. 1990. Optimization of PCRs. In PCR protocols: a guide to methods and applications. Academic Press, Inc., San Diego, Calif. |
| 10. |
Kox, L. F.,
D. Rhienthong,
A. M. Miranda,
N. Udomsantisuk,
K. Ellis,
J. van Leeuwen,
S. van Heusden,
S. Kuijper, and A. H. Kolk.
1994.
A more reliable PCR for detection of Mycobacterium tuberculosis in clinical samples.
J. Clin. Microbiol.
32:672-678 |
| 11. | Kwok, S., and R. Higuchi. 1989. Avoiding false positives with PCR. Nature 339:237-238[CrossRef][Medline]. |
| 12. | Livak, K. J., S. J. Flood, J. Marmaro, W. Giusti, and K. Deetz. 1995. Oligonucleotides with fluorescent dyes at opposite ends provide a quenched probe system useful for detecting PCR product and nucleic acid hybridization. PCR Methods Appl. 4:357-362[Medline]. |
| 13. | Murthy, K. K., D. R. Henrard, J. W. Eichberg, K. E. Cobb, M. P. Busch, J. P. Allain, and H. J. Alter. 1999. Redefining the HIV-infectious window period in the chimpanzee model: evidence to suggest that viral nucleic acid testing can prevent blood-borne transmission. Transfusion 39:688-693[CrossRef][Medline]. |
| 14. |
Pasloske, B. L.,
C. R. Walkerpeach,
R. D. Obermoeller,
M. Winkler, and D. B. DuBois.
1998.
Armored RNA technology for production of ribonuclease-resistant viral RNA control and standards.
J. Clin. Microbiol.
36:3590-3594 |
| 15. | Petersen, L. R., G. A. Satten, R. Dodd, M. Busch, S. Kleinman, A. Grindon, and B. Lenes. 1994. Duration of time from onset of human immunodeficiency virus type 1 infection to development of detectable antibody. Transfusion 34:283-289[CrossRef][Medline]. |
| 16. | Piatak, M., Jr., M. S. Saag, L. C. Yang, S. J. Clark, J. C. Kappes, K. C. Luk, B. H. Hahn, G. M. Shaw, and J. D. Lifson. 1993. High levels of HIV-1 in plasma during all stages of infection determined by competitive PCR. Science 259:1749-1754. |
| 17. |
Rosenstraus, M.,
Z. Wang,
S. Y. Chang,
D. DeBonville, and J. P. Spadoro.
1998.
An internal control for routine diagnostic PCR: design, properties, and effect on clinical performance.
J. Clin. Microbiol.
36:191-197 |
| 18. | Roth, W. K., M. Weber, and E. Seifried. 1999. Feasibility and efficacy of routine PCR screening of plasma from donors for hepatitis C virus, hepatitis B virus, and HIV-1 in a blood bank setting. Lancet 353:359-363[CrossRef][Medline]. |
| 19. | Rüster, B., S. Zeuzem, and W. K. Roth. 1995. Quantification of hepatitis C virus by competitive reverse transcription and polymerase chain reaction using a modified hepatitis C virus transcript. Anal. Biochem 224:597-600[CrossRef][Medline]. |
| 20. |
Schreiber, G. B.,
M. P. Busch,
S. H. Kleinmann, and J. J. Korelitz.
1996.
The risk of transfusion-transmitted viral infections.
N. Engl. J. Med.
334:1685-1690 |
| 21. |
Smieja, M.,
J. B. Mahony,
C. H. Goldsmith,
S. Chong,
A. Petrich, and M. Chernesky.
2001.
Replicate PCR testing and probit analysis for detection and quantitation of Chlamydia pneumoniae in clinical specimens.
J. Clin. Microbiol.
39:1796-1801 |
| 22. | von Förster, T. 1948. Zwischenmolekulare Energiewanderung und Fluoreszenz. Ann. Physics (Leipzig) 2:55-57. |
| 23. | Ward, J. W., S. D. Holmberg, J. R. Allen, D. L. Cohn, S. E. Critchley, S. H. Kleinman, B. A. Lenes, O. Ravenholt, J. R. Davis, M. G. Quinn, and H. W. Jaffe. 1988. Transmission of human immunodeficiency virus (HIV) by blood transfusions screened as negative for HIV antibody. N. Engl. J. Med. 318:473-478[Abstract]. |