Journal of Clinical Microbiology, December 2001, p. 4407-4412, Vol. 39, No. 12
Department of Molecular Biology, Laboratory
Dr. Schiwara and Partners, 28357 Bremen,
Germany1; Department of Virology,
University Medical Center Utrecht, Utrecht,2 and
Department of Virology, Regional Public Health Laboratory,
Groningen,3 The Netherlands; and Shield
Diagnostics Limited, Dundee,4 and
Department of Virology, Manchester Royal Infirmary,
Manchester,5 United Kingdom
Received 30 April 2001/Returned for modification 14 August
2001/Accepted 27 September 2001
External quality control of hepatitis B virus (HBV) DNA detection
remains an important issue. This study reports and compares the results
obtained from two different proficiency panels for both the qualitative
and quantitative assessment of HBV DNA. The panels were designed by the
European Union Quality Control Concerted Action, prepared by Boston
Biomedica, Inc., and distributed in May 1999 (panel 1) and February
2000 (panel 2). Each contained two negative samples and six positive
samples with 103 to 107 copies/ml (panel 1) or
103 to 2 × 106 copies of HBV DNA per ml
(panel 2). For panel 1, 42 laboratories submitted 20 qualitative (all
in-house PCRs) and 37 quantitative (87% commercial assays) data sets.
For panel 2, 51 laboratories submitted 25 qualitative (all in-house
PCRs) and 47 quantitative (94% commercial assays) data sets. Five data
sets (8.8%) in panel 1 and two data sets (2.8%) in panel 2 contained
totals of six and two false-positives, respectively, corresponding to
false-positive result rates of 5.3% for panel 1 and 1.4% for panel 2. The false-negative result rates of 10.5% for panel 1 and 17.4% for
panel 2 were dependent on the detection levels of the assays employed
as well as panel composition. In the qualitative analysis of all data
sets, 47.4% (panel 1) and 51.4% (panel 2) had all samples correct. An
adequate or better score (all correct or only the weak-positive sample missed) was obtained with 77.2% of the panel 1 samples and 68.1% of
the panel 2 samples. In the quantitative analysis, 57.1% (panel 1) and
42.6% (panel 2) of the data sets achieved an adequate or better score
(positive results within the acceptable range of the geometric
mean ± 0.5 log10 of all positive results). These results demonstrate that while the qualitative performance of HBV
detection has considerably improved compared to that of a previously
published HBV proficiency study, the detection levels of many
commercial quantitative assays are still too high to allow adequate
quantitation of all relevant clinical samples.
Direct detection and quantitation of
hepatitis B virus (HBV) DNA in plasma or serum are now used routinely
to evaluate viremia in HBV-infected persons, to identify infectious
chronic carriers, and to predict and monitor the efficacy of antiviral
therapy (2, 8, 11). Since the early 1980s, a variety of
molecular detection and quantitation methods have been developed,
including dot and slot blot hybridization with radioactive and
nonradioactive DNA probes (19-21), chemiluminescent
detection of HBV DNA-RNA hybrids (14), PCR amplification
of HBV DNA followed by hybridization to probes bound to a microwell
plate (10, 12, 22) or magnetic beads (13),
branched DNA (bDNA) signal amplification of an HBV DNA-DNA hybrid
(7), transcription-mediated amplification
(9), and fluorescent real-time detection of amplified HBV
DNA (1). Each method, calibrated uniquely, exhibits its
own sensitivity, specificity, and dynamic range. Standardization is
ongoing (5, 6).
To assess the relative value of these methods in detecting and
quantitating HBV DNA, international proficiency studies with well-characterized, simulated clinical samples would be required. In
the first and only such study published to date (17), 39 laboratories analyzed 22 samples, including 12 undiluted samples with
and without HBV DNA. (The lowest positive sample contained 3.5 pg/ml,
or approximately 980,000 copies/ml.) Only 27.9% of the data sets had
all 12 samples correct, and 34.9% showed false-positive results.
Clearly, a majority of the participating laboratories had problems with
both sensitivity and specificity.
The present report describes two recent HBV proficiency panels (lowest
viral load of 1,000 copies/ml) designed by the European Union Concerted
Action on Quality Control (EU QCCA) of Nucleic Acid Amplification in
Diagnostic Virology and prepared by Boston Biomedica, Inc. (BBI; West
Bridgewater, Mass.). The results obtained with these panels demonstrate
that while the qualitative detection of HBV DNA has significantly
improved, the detection levels of many commercial quantitative assays
are still too high to allow adequate quantitation of the clinical
samples seen in routine diagnostic laboratories.
Panels. (i) Preparation.
Panels were prepared by BBI from
human plasma containing HBV DNA of subtype ad or
ay by appropriate dilution in sterile filtered defibrinated plasma (Basematrix) with 0.09% sodium azide as
preservative in accordance with the ISO 9001 Quality System Standards
and the 21CFR 820 "Good Manufacturing Practice for Medical
Devices: General." Plasma units were obtained from Food
and Drug Administration-licensed facilities that comply with the
applicable federal regulations (21CFR, part 600).
0095-1137/01/$04.00+0 DOI: 10.1128/JCM.39.12.4407-4412.2001
Copyright © 2001, American Society for Microbiology. All rights reserved.
European Proficiency Testing Program for Molecular
Detection and Quantitation of Hepatitis B Virus DNA
![]()
ABSTRACT
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
![]()
INTRODUCTION
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
![]()
MATERIALS AND METHODS
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
70°C until
shipment to the participants in May 1999 (panel 1) and February 2000 (panel 2).
(ii) Composition. Each panel consisted of eight coded samples. Six samples contained HBV DNA with approximate target levels of 103 to 107 copies/ml (panel 1) and 103 to 2 × 106 copies/ml (panel 2). Two samples contained no virus and served as negative controls. To evaluate interassay reproducibility, three identical samples were included in both panels: 2 × 106 copies/ml each for ad and ay and 2 × 105 copies/ml for ad. To assess a possible effect of HBV subtype, each panel contained two pairs of samples with identical viral loads, but different subtypes.
Participants. The panels were distributed on dry ice by courier service to 45 laboratories (panel 1) and 61 laboratories (panel 2). The recipient laboratories were asked to report the arrival and condition of the panel immediately by fax and to return the results as soon as possible, but within 6 weeks (panel 1) or 4 weeks (panel 2). A code number, known only to the Neutral Office, University of Manchester, Manchester, United Kingdom, identified each laboratory. Laboratories participating in both proficiency studies were assigned the same code for both panels. A questionnaire was also sent to obtain technical information on the procedures employed. To ensure confidentiality, all laboratories sent their results to the Neutral Office. The results were analyzed anonymously at the Department of Molecular Biology, Laboratory Dr. Schiwara and Partners, Bremen, Germany.
After the closing date for each panel, each participating laboratory was sent a certificate of participation, the code of the panel for individual performance assessment, and a written report summarizing all results.Qualitative analysis. For qualitative analysis, the results from the quantitative data sets were converted to qualitative data (i.e., positive or negative) and considered together with the true qualitative data sets. To assess performance, the following scoring system was applied: 1 point was given for each correct result for the true-positive and true-negative samples. A point was deducted for each false-positive or false-negative result, with the exception of the weak-positive sample containing 103 copies/ml. Thus, the maximum score to be obtained was 8 points, which was qualified as "good." Scores of 7 and 6 points were considered "adequate" and "mediocre," respectively, while <6 points was considered "poor."
Quantitative analysis. For quantitative analysis, the overall geometric mean (GM) and standard deviations (SD) were calculated for each (positive) sample for all assays, as well as separately according to the method used. To assess performance, the following scoring system was utilized: 1 point was awarded for each viral load result that was within the range of ±0.5 log10 of the overall GM of each sample. Scoring was for positive samples only. For all other results, i.e., each positive result outside this range, false-positive results, or negative results on positive samples, no point was given or deducted. The maximum score of 6 points was considered "good," 5 and 4 points were considered "adequate" and "mediocre," respectively, and <4 points was considered "poor."
| |
RESULTS |
|---|
|
|
|---|
Participants and methods.
In panel 1, 42 laboratories from 19 countries submitted 20 qualitative (all in-house PCRs) and 37 quantitative (87% commercial assays) data sets. In panel 2, 51 laboratories from 18 countries submitted 25 qualitative (all in-house
PCRs) and 47 quantitative (94% commercial assays) data sets. The
methods utilized and the corresponding detection limits as reported by
the participants are listed in Table 1.
|
Qualitative analysis. (i) Panel 1.
Correct results for the two
negative samples were reported in 52 of the 57 data sets (91.2%), and
a false-positive result was obtained in 5 data sets, 1 of which
contained two false-positive results (Table
2). Of these six false-positive results
(6 of 114 negative samples = 5.3%), three were obtained with
commercial kits (two by Digene Hybrid Capture Systems and one by Roche
Monitor Assay). The weak-positive sample (103
copies/ml) was correctly reported positive in 30 data sets. The 27 data
sets with false-negative results were obtained with the Digene Hybrid
Capture Systems (13 of 13), an in-house PCR (7 of 15), an in-house
nested PCR (4 of 10), the Roche Monitor Assay (1 of 16), the Bayer bDNA
Assay (1 of 2), and the BAG AcuGen Test (1 of 1). Another low-positive
sample (2 × 105 copies/ml) was correctly
reported positive in 48 data sets. The nine false-negative results with
this sample were in data sets obtained with the Digene Hybrid Capture
Systems (7 of 13) and an in-house PCR (2 of 15). All other positive
samples (
2 × 106 copies/ml) were
correctly identified as positive in all data sets. The overall
false-negative result rate was therefore 10.5% [36/(6 × 57]
(Table 2).
|
|
(ii) Panel 2. Correct results for the two negative samples were reported in 70 data sets. A false-positive result was obtained in two qualitative data sets (both in-house PCRs); none of the quantitative data sets contained false-positive results (Table 2). Thus, the false-positive result rate was 1.4% (2 of 144 negative samples).
The weak-positive sample (103 copies/ml) was correctly reported positive in 40 of the 72 data sets (55.6%). The 32 negative results were obtained with all five Bayer bDNA assays, all three Digene Hybrid Capture System I tests, most (9 of 10) Digene Hybrid Capture System II tests, the 1 BAG AcuGen Test, 7 of 21 in-house PCRs, 1 of 7 in-house nested PCRs, and 2 of 20 Roche Monitor Assay tests. For the other positive samples, the negative result rates decreased with increasing sample viral load. The overall false-negative result rate was therefore 17.4% [75/(6 × 72] (Table 2). A total of 37 data sets (51.4%) obtained the maximum score of 8 points, 12 (16.7%) had a score of 7 points, 1 (1.4%) had a score of 6 points, and 22 (30.6%) had <6 points (Table 3).Quantitative analysis. (i) Panel 1.
Quantitative HBV data were
reported in 37 data sets, whereby results from 2 data sets (1 in-house
PCR, all results listed as <8 × 103;1
in-house nested PCR, all positive results listed as
>106) could not be included in the calculations.
Most data sets (87%) were derived from commercial kits. The overall GM
and SD were calculated for each (positive) sample from all assays
(Table 4) as well as separately for the
16 Roche data sets, the 13 Digene data sets, and the 6 remaining data
sets taken together (data not shown). The overall GM compared well with
the target levels. For most samples,
89% of the positive results
were within the range of GM ± 0.5 log10.
The GM values obtained with the Roche and Digene assays were remarkably
similar and likewise corresponded well to the target levels (except for
the weak-positive sample, which could not be detected by the Digene
assays), although the SD values for the Roche assay were consistently
larger than those for the Digene assays (data not shown). The GM values
for the six remaining data sets calculated together were consistently lower than the GM for the Roche and Digene assays and showed
consistently larger SD values (data not shown).
|
|
(ii) Panel 2.
Quantitative HBV data were reported in 47 data
sets, most (94%) derived from commercial kits. The overall GM and SD
were calculated for each (positive) sample from all assays (Table 4),
as well as separately according to the method used (data not shown).
The overall GM corresponded well with the target levels. For most samples,
83% of the positive results were within the range of GM ± 0.5 log10. As in panel 1, the GM for
the Roche and Digene assays were similar but with larger SD values for
the Roche assays (data not shown). The GM for the remaining assays
tended to be lower than for the Roche and Digene assays, with the
largest SD values obtained with the in-house quantitative PCRs (data
not shown).
Reproducibility.
Interpanel reproducibility could be evaluated
from the results obtained with three samples represented in both panels
(Table 6). For the ay subtype
sample with 2 × 106 copies/ml, the
qualitative detection rates were virtually the same: 100 and 99%,
respectively. However, the percentage of positive results within the
range of ±0.5 log10 of the GM for that sample was higher in panel 1 (91%) than in panel 2 (85%). The results for
the ad subtype sample with the same viral load showed the same pattern: nearly identical qualitative detection rates (100 and
99%, respectively), but a higher percentage of positive results within
the range of ±0.5 log10 of the GM in panel 1 (89%) compared to that in panel 2 (83%).
|
| |
DISCUSSION |
|---|
|
|
|---|
To assess the value of currently available methods as utilized by diagnostic laboratories for detecting and quantitating HBV DNA, international proficiency studies with well-characterized, simulated clinical samples are required. In the first such study performed in conjunction with the European Expert Group on Viral Hepatitis (17), 39 laboratories submitted 43 data sets for 22 samples, including 12 undiluted samples (7 positive, 5 negative) with viral loads from 3.5 pg/ml (approximately 980,000 copies/ml) to 222 pg/ml (approximately 62,160,000 copies/ml). Viral loads were determined by a single assay (Genostics liquid hybridization test from Abbott). All but one laboratory used the PCR technique; none used the uracil N-glycosylase system to prevent contamination from previously amplified products. Of the 43 data sets, 15 (35%) showed false-positive results for the five negative samples. Furthermore, despite the relatively high value of the weakest sample (980,000 copies/ml), 16 of 43 data sets (37%) showed false-negative results. Only 12 of 43 data sets (28%) had all 12 samples correct. No quantitative analysis of the data was performed. The first study, however, clearly demonstrated the large number of laboratories with sensitivity and specificity problems in detecting HBV DNA.
Similar problems with sensitivity and specificity were reported from
the early proficiency studies for Mycobacterium tuberculosis (16) and for hepatitis C virus (HCV) (24)
(29% false-positive results and 39% false-negative results on 10 undiluted samples
4 positive and 6 negative
in 31 data sets). In the
second HCV proficiency study, performed 3 years later (4),
little improvement was reported (21% false-positive results, but 63%
false-negative results for 10 undiluted samples
again 4 positive and 6 negative
in 136 data sets).
Compared to the first HBV proficiency study (17), the present study shows considerable improvement both in panel conception and results obtained. First, the samples in both panels were well characterized, having been defined by the manufacturers of several quantitative methods (Roche, Bayer, and Digene) and by three reference laboratories utilizing various assays. The close approximation of the GM of all test results from the participating laboratories supports the accuracy of the assigned target viral load. Secondly, the sample with the lowest viral load in both panels contained only 1,000 copies of HBV DNA per ml, reflecting more accurately the viral loads from patients undergoing antiviral therapy. Third, this study involved a large number of participants (42 and 51 in panels 1 and 2, respectively) and data sets (57 and 72 in panels 1 and 2, respectively). In addition, the participating laboratories are now utilizing a variety of methods for detecting and quantitating HBV DNA in addition to PCR, allowing such methods to be assessed in an international proficiency panel. Fourth, the data were analyzed not only qualitatively, but also quantitatively, by using a simple algorithm permitting comparison of the seven different quantitative assays employed. Finally, the description of two independent proficiency panels permits interpanel reproducibility testing.
One of the most significant results derived from this study is the lowest false-positive rate reported to date for any large proficiency panel. Similar low false-positive rates have been found in the EU QCCA HCV RNA and human immunodeficiency virus proficiency studies carried out simultaneously with this study (J. Schirm, A. M. van Loon, E. Valentine-Thon, J. Reid, P. E. Klapper, and G. M. Cleator, submitted for publication; A. M. van Loon, J. Schirm, E. Valentine-Thon, J. Reid, P. E. Klapper, and G. M. Cleator, unpublished data) and in the EU QCCA enterovirus proficiency study carried out several months earlier (23). Indeed, the rate of 8.8% (5 of 57) false-positive data sets or 5.3% (6 of 114) false-positive results for panel 1 improved even further to 2.8% (2 of 72) false-positive data sets or 1.4% (2 of 144) false-positive results for panel 2. These low rates may simply reflect the greater expertise of the participating laboratories compared to several years ago. The improvement may also result from the increasing use of commercial kits, many of which contain contamination control enzyme systems. Indeed, 56% of data sets in panel 1 and 61% of data sets in panel 2 were obtained from commercial kits. On the other hand, the use of commercial kits does not safeguard against false-positive results: while the only false-positive results in panel 2 were obtained with in-house PCRs, 3 of 6 false-positive results in panel 1 were obtained with commercial kits. Finally, the lower false-positive rate in panel 2 compared to that in panel 1 may also represent a beneficial effect of participation in this proficiency program: of the four laboratories generating false-positive results in panel 1, three participated in panel 2 without producing false-positive results. Furthermore, the two laboratories generating false-positive results in panel 2 had not participated in panel 1.
In contrast to the low rate of false-positives in this study, the rate
of negative results for (true) positive samples was remarkably high:
for panel 1, 43.9% (25 of 57) of data sets, or 10.5% [36/(57 × 6] of results; for panel 2, 45.8% (33 of 72) of data sets, or 17.4%
[75/(72 × 6] of results. Obviously, this high rate reflects the
deliberate low viremic composition of the samples, in particular in
panel 2 (
2 × 106 copies/ml) coupled with
the high detection limits of some commercial kits still in use (i.e.,
Hybrid Capture System I with 1.4 × 106
copies/ml or Bayer bDNA with 0.7 × 106
copies/ml). (While the negative results obtained with low viremic samples by using such kits could arguably be excluded from the category
"false-negative," we have chosen to include them in this category
for the sake of simplicity.) As a result, for both qualitative and
quantitative analyses, the percentage of data sets achieving scores of
adequate or better decreased in panel 2 compared to that in panel 1. While it is recognized that such a panel composition "penalizes"
users of commercial kits with high detection limits, the increasing
clinical requirement for low-level dynamic ranges dictates this
necessity. A potentially relevant assay is the newly introduced Cobas
Amplicor HBV Monitor Test from Roche Diagnostics. However, while its
low detection limit (200 copies/ml) is suitable for monitoring patients
undergoing therapy, its upper dynamic range of only 200,000 copies/ml
requires pretest dilution of HBsAg- and HBeAg-positive samples,
concomitantly increasing costs (13, 15). Alternatively,
transcription-mediated amplification followed by hybridization of two
probes with different specific activities allows a broad detection
range of 5 × 103 to 5 × 108 copies/ml (9). In contrast to
Cobas Amplicor, however, this method contains no internal control for
inhibition and is not yet automated.
In addition to low detection levels and a broad dynamic range, appropriate quantitation assays should be calibrated to internationally defined reference standards, as now available for HCV RNA (18). In the present study, the quantitative analysis showed variations ranging from 6 to 35% (percent positive results outside the defined range) in the actual copy numbers assigned to samples. Similar method-related deviations in quantitation have been reported by others (3, 5) In 1999, the Eurohep Pathobiology Group established two international reference plasma preparations, each containing approximately 2.6 × 109 copies/ml, thereafter defined as 109 Eurohep units (6). These Eurohep samples have been used for the evaluation of commercial kits (10, 14), and one of these may be the basis of a World Health Organization reference sample. In the present study, a panel 1 sample (ay subtype) with 107 copies/ml could also serve as a candidate reference standard, because the GM ± SD log10 for all 35 assays was 7.05, and 94% of the positive results with this sample were within the range of GM ± 0.5 log10.
Finally, both qualitative and quantitative assays must yield
reproducible results. The interpanel reproducibility of the three samples represented in both panels was excellent: overall qualitative detection rates were nearly identical. However, for the two samples with 2 × 106 copies/ml, the percentage of
positive results within the range of ±0.5 log10
of the GM for that sample was higher in panel 1 than in panel 2, while
for the lower viremic sample (2 × 105
copies/ml), the opposite was true. Apparently, the increased use of
commercial kits in panel 2 led to an increase in "within-range" detection rates in the lower dynamic range
at the expense, however, of
a decreased within-range detection rate in the upper dynamic range.
In conclusion, this extensive proficiency study demonstrates considerable improvement in the qualitative performance of currently available HBV DNA assays. At the same time, however, the results emphasize the need for commercial viral load kits with sufficiently low detection levels to permit adequate HBV DNA detection and quantitation in clinical samples.
| |
ACKNOWLEDGMENTS |
|---|
The Blood-Borne Virus Working Party of the EU QCCA is grateful to all participating laboratories, reference laboratories (University Hospital Rotterdam, Department of Virology, Rotterdam, The Netherlands; University Medical Center Utrecht, Department of Virology, Utrecht, The Netherlands; and Roche Diagnostics, Basel, Switzerland) and manufacturers (Roche Diagnostics, Bayer Diagnostics, and Digene Diagnostics) for their contribution to this HBV DNA proficiency study.
The EU QCCA Program was supported with a grant from the EU Biomed 2 Program.
| |
FOOTNOTES |
|---|
* Corresponding author. Mailing address: Department of Molecular Biology, Laboratory Dr. Schiwara and Partners, Haferwende 12, 28357 Bremen, Germany. Phone: 0049-421-20720. Fax: 0049-421-2072-167. E-mail: Elizabeth.Valentine-Thon{at}Schiwara.de.
| |
REFERENCES |
|---|
|
|
|---|
| 1. |
Abe, A.,
K. Inoue,
T. Tanaka,
J. Kato,
N. Kajiyama,
R. Kawaguchi,
S. Tanaka,
M. Yoshiba, and M. Kohara.
1999.
Quantitation of hepatitis B virus genomic DNA by real-time detection PCR.
J. Clin. Microbiol.
37:2899-2903 |
| 2. | Beyer, A., J. Braner, H. W. Doerr, and B. Weber. 1998. Quantitation of viral load: clinical relevance for human immunodeficiency virus, hepatitis B virus and hepatitis C virus infection. Intervirology 41:24-34[CrossRef][Medline]. |
| 3. | Butterworth, L.-A., S. L. Prior, P. J. Buda, J. L. Faoagali, and W. G. E. Cooksley. 1996. Comparison of four methods for quantitative measurement of hepatitis B viral DNA. J. Hepatol. 24:686-691[CrossRef][Medline]. |
| 4. | Damen, M., H. T. Cuypers, H. L. Zaaijer, H. W. Reesink, W. P. Schaasberg, W. H. Gerlich, H. G. Niesters, and P. N. Lelie. 1996. International collaborative study on the second EUROHEP HCV-RNA reference panel. J. Virol. Methods 58:175-185[CrossRef][Medline]. |
| 5. | Gerlich, W. H., K. H. Heermann, R. Thomssen, and the Eurohep Group. 1995. Quantitative assays for hepatitis B virus DNA: standardization and quality control. Viral Hep. 1:53-57. |
| 6. |
Heermann, K. H.,
W. H. Gerlich,
M. Chudy,
S. Schaefer,
R. Thomssen, and the Eurohep Pathobiology Group.
1999.
Quantitative detection of hepatitis B virus DNA in two international reference plasma preparations.
J. Clin. Microbiol.
37:68-73 |
| 7. | Hendricks, D. A., B. J. Stowe, B. S. Hoo, J. L. Kolberg, B. D. Irvine, P. D. Neuwald, M. S. Urdea, and R. P. Perrillo. 1995. Quantitation of HBV DNA in human serum using a branched DNA (bDNA) signal amplification assay. Am. J. Clin. Pathol. 104:537-546[Medline]. |
| 8. | Jardi, R., M. Buti, F. Rodriquez-Frias, M. Cortina, R. Esteban, J. Guardia, and C. Pascual. 1996. The value of quantitative detection of HBV-DNA amplified by PCR in the study of hepatitis B infection. J. Hepatol. 24:680-685[CrossRef][Medline]. |
| 9. |
Kamisango, K.,
C. Kamogawa,
M. Sumi,
S. Goto,
A. Hirao,
F. Gonzales,
K. Yasuda, and S. Iino.
1999.
Quantitative detection of hepatitis B virus by transcription-mediated amplification and hybridization protection assay.
J. Clin. Microbiol.
37:310-314 |
| 10. | Kessler, H. H., K. Pierer, E. Dragon, H. Lackner, B. Santner, D. Stünzner, E. Stelzl, B. Waitzl, and E. Marth. 1998. Evaluation of a new assay for HBV DNA quantitation in patients with chronic hepatitis B. Clin. Diagn. Virol. 9:37-43[CrossRef][Medline]. |
| 11. | Lok, A. S. F. 1994. Treatment of chronic hepatitis B. J. Viral Hep. 1:105-124. |
| 12. |
Mantero, G.,
A. Zonaro,
A. Albertini,
P. Bertolo, and D. Primi.
1991.
DNA enzyme immunoassay: general method for detecting products of polymerase chain reaction.
Clin. Chem.
37:422-429 |
| 13. |
Marin, I. J.,
M. Poljak,
K. Seme,
J. Megli -Volkar,
M. Mati i ,
G. Le ni ar, and V. Brinovec.
2001.
Comparative evaluation of semiautomated COBAS AMPLICOR hepatitis B virus (HBV) MONITOR Test and manual microwell plate-based AMPLICOR HBV MONITOR Test.
J. Clin. Microbiol.
39:758-761 |
| 14. |
Niesters, H. G. M.,
M. Krajden,
L. Cork,
M. de Medina,
M. Hill,
E. Fries, and A. D. M. E. Osterhaus.
2000.
A multicenter study evaluation of the Digene Hybrid Capture II signal amplification technique for detection of hepatitis B virus DNA in serum samples and testing of EUROHEP standards.
J. Clin. Microbiol.
38:2150-2155 |
| 15. |
Noborg, U.,
A. Gusdal,
E. K. Pisa,
A. Hedrum, and M. Lindh.
1999.
Automated quantitative analysis of hepatitis B virus DNA by using the Cobas Amplicor HBV Monitor Test.
J. Clin. Microbiol.
37:2793-2797 |
| 16. |
Noordhoek, G. T.,
A. H. J. Kolk,
G. Bjune,
D. Catty,
J. W. Dale,
P. E. M. Fine,
P. Godfrey-Faussett,
S.-N. Cho,
T. Shinnick,
S. B. Svenson,
S. Wilson, and J. D. A. van Embden.
1994.
Sensitivity and specificity of PCR for detection of Mycobacterium tuberculosis: a blind comparison study among seven laboratories.
J. Clin. Microbiol.
32:277-284 |
| 17. | Quint, W. G. V., R. A. Heijtink, J. Schirm, W. H. Gerlich, and B. Niesters. 1995. Reliability of methods for hepatitis B virus DNA detection. J. Clin. Microbiol. 33:225-228[Abstract]. |
| 18. | Saldanha, J., N. Lelie, and A. Heath. 1999. Establishment of the first international standard for nucleic acid amplification technology (NAT) assays for HCV RNA. W.H.O. Collaborative Study Group. Vox Sang. 76:149-158[CrossRef][Medline]. |
| 19. | Scotto, J., M. Hadchouel, C. Hery, J. Yvart, P. Tiollais, and C. Brechot. 1983. Detection of hepatitis B virus DNA in serum by a single spot hybridization technique: comparison with results for other viral markers. Hepatology 3:279-282[Medline]. |
| 20. |
Valentine-Thon, E.,
J. Steinmann, and W. Arnold.
1990.
Evaluation of the commercially available HepProbe kit for detection of hepatitis B virus DNA in serum.
J. Clin. Microbiol.
28:39-42 |
| 21. | Valentine-Thon, E., J. Steinmann, and W. Arnold. 1991. Detection of hepatitis B virus DNA in serum with nucleic acid probes labelled with 32P, biotin, alkaline phosphatase or sulphone. Mol. Cell. Probes 5:299-305[CrossRef][Medline]. |
| 22. | Valentine-Thon, E. 1995. Evaluation of SHARP Signal System for enzymatic detection of amplified hepatitis B virus DNA. J. Clin. Microbiol. 33:477-480[Abstract]. |
| 23. |
Van Vliet, K.,
P. Muir,
J. M. Echevarria,
P. E. Klapper,
G. M. Cleator, and A. M. van Loon.
2001.
Multicenter proficiency testing of nucleic acid amplification methods for the detection of enteroviruses.