Previous Article | Next Article 
Journal of Clinical Microbiology, April 2001, p. 1682-1682, Vol. 39, No. 4
0095-1137/01/$04.00+0 DOI: 10.1128/JCM.39.4.1682.2001
LETTERS TO THE EDITOR
Binax and Biotest Legionella Urinary Antigen Kits
 |
LETTER |
The article of Robert F. Benson et al. entitled "Evaluation of
the Binax and Biotest urinary antigen kits for detection of Legionnaires' disease due to multiple serogroups and species of Legionella" (1) states the following in the
abstract: "Eighteen were positive with the Binax kit, and 13 were
positive with the Biotest." This statement is misleading since data
in both the article and Table 1 show that of 45 urine samples
previously positive by the enzyme-linked immunosorbent assay of Tang
and Toma (3), 45 were tested with the Binax enzyme
immunoassay (EIA), and of these, 10 were positive and 35 were negative.
Forty-two of the 45 were tested with the Biotest EIA; of these, 13 were
positive and 29 were negative (3 were not tested). Obviously,
there was a retest with the Binax radioimmunoassay: 26 of the
45 samples were retested. Of these, 13 were positive and 13 were
negative (19 were not tested).
If one combines the results for the Binax radioimmunoassay (RIA) and
the Binax EIA shown in Table 1 in reference 1, it may be possible to
find 18 positive samples, but the abstract should then state that 18 were positive with the Binax EIA and RIA kits combined, since the
results of two test kits were being compared against those of one
Biotest EIA. The abstract still would be short enough, even if the
facts shown in the article were spelled out, i.e., 10 of 45 specimens
positive by the Binax EIA, 13 of 42 positive by the Biotest EIA, and 13 of 26 positive by the Binax RIA.
In a report by Harrison et al. (2), the eight urine
samples from culture-verified non-serogroup 1 patients were positive by
the Danish in-house assay only. This is not stated in the article but
was presented at the EWGLI Meeting, 1 to 3 June 1997, Lisbon, Portugal,
prior to publication of the article (2). Harrison et al.
also reported that the Biotest EIA was positive for 5 of 46 cases that
were all negative by the Binax EIA or by in-house tests, all cases
proven either by culture serology, direct fluorescent-antibody assay, or PCR results, so that the Biotest EIA provided an additional 11% positive results from urine samples negative by other tests but
from patients proven positive.
With regards to the lower cutoff value, the Biotest assay does have two
cutoffs: one is the mean of negative controls plus 0.200; the other is
the mean of negative controls plus 0.100. In the latter case, the
sample must test positive repeatedly (two times) to be considered
positive. Many samples from non-serogroup 1 patients are weak
positives: using the kit criteria, one usually finds twice as many
non-serogroup 1 samples positive by the Biotest assay as with the Binax
assay (as described by Dr. L. Helbig at the EWGLI Meeting, 27 to 29 June 1999, Dresden, Germany, and possibly by Benson et al.
[1], who stated that 20 of the 42 [48%] samples would
be positive by the Biotest EIA with a lower cutoff, although without
stating what kind of lower cutoff).
This information about the cutoff value would be interesting to the
reader of the abstract, certainly more so than the mention of 18 positive samples found by using the Binax kit. The lower ratio of
>2.0, which increased the Binax test sensitivity, was not accompanied
by a concurrent specificity study with 300 negative samples, 176 potentially cross-reactive samples from the manufacturer, and 123 potentially cross-reactive samples from a multicenter study
(2), as was the case with the Biotest assay with the lower
cutoff. Parallel specificity studies should always be performed when
cutoff values are altered since in low-prevalence situations, high
false-positive numbers may result.
 |
REFERENCES |
| 1.
|
Benson, R. F.,
P. W. Tang, and B. S. Fields.
2000.
Evaluation of the Binax and Biotest urinary antigen kits for detection of Legionnaires' disease due to multiple serogroups and species of Legionella.
J. Clin. Microbiol.
38:2763-2765[Abstract/Free Full Text].
|
| 2.
|
Harrison, T. G.,
S. Uldum,
S. Alexiou-Daniel,
J. Bangsborg,
S. Bernander,
V. Drasar,
J. Etienne,
J. Helbig,
D. Lindsay,
I. Lochman,
T. Marques,
F. deOry,
I. Tartakovskii,
G. Wewalka, and F. Fehrenbach.
1998.
A multicenter evaluation of the Biotest Legionella urinary antigen EIA.
Clin. Microbiol. Infect.
4:359-365[Medline].
|
| 3.
|
Tang, P. W., and S. Toma.
1986.
Broad-spectrum enzyme-linked immunosorbent assay for detection of Legionella soluble antigens.
J. Clin. Microbiol.
24:556-558[Abstract/Free Full Text].
|
| | | | |
Juergen Horn
Biotest AG Postfach 40 11 08 D-63276 Dreieich, Germany
|
 |
AUTHORS' REPLY |
We appreciate Dr. Horn's comments and offer the following response. In
the abstract of our article (1), the number of samples listed as positive for the Binax kit was the total for both the EIA and
RIA kits. This point was not stated directly in the abstract, but, as
Dr. Horn acknowledges, this is described in detail in the text and in
Table 1 of the article. We also stated that only the results of the
Binax EIA were used for comparison to the Biotest EIA. In regards to
the literature cited, we were not aware of the results presented at the
EWGLI meeting in 1997 but only of the results presented in the article
by Harrison et al. (2). We appreciate Dr. Horn's
clarification of the results published by Dr. Harrison. Our results
tend to support Dr. Horn's statement that the Biotest EIA is more
sensitive than the Binax EIA in detecting 13 of 42 versus 10 of 45 urine samples. Dr. Horn states that the Biotest EIA has two cutoff
values for positive results. This information is not stated in the
product insert we received with the kit. If this is the case, the
product insert should be changed to indicate that a lower cutoff can be
used. Our observations of the positive samples compared to the negative
samples is supported by Dr. Horn's statement. The purpose of our study
was to show that both tests, while not as sensitive as the
broad-spectrum assay, are capable of detecting nonserogroup-1
Legionnaires' disease.
 |
REFERENCES |
| 1.
|
Benson, R. F.,
P. W. Tang, and B. S. Fields.
2000.
Evaluation of the Binax and Biotest urinary antigen kits for detection of Legionnaires' disease due to multiple serogroups and species of Legionella.
J. Clin. Microbiol.
38:2763-2765.
|
| 2.
|
Harrison, T.,
S. Uldum,
A. Alexiou-Daniel,
J. Bangsborg,
S. Bernander,
V. Drasar,
J. Etienne,
J. Helbig,
D. Lindsay,
I. Lochman,
T. Marques,
F. deOry,
I. Tartakovskii,
G. Wewalka, and F. Fehrenbach.
1998.
A multicenter evaluation of the Biotest Legionella urinary antigen EIA.
Clin. Microbiol. Infect.
4:359-365.
|
| | | | |
Robert F. Benson
Barry S. Fields
Division of Bacterial and Mycotic Diseases National Center for Infectious Diseases Centers for Disease Control and Prevention Atlanta, Georgia 30333
|
Journal of Clinical Microbiology, April 2001, p. 1682-1682, Vol. 39, No. 4
0095-1137/01/$04.00+0 DOI: 10.1128/JCM.39.4.1682.2001
This article has been cited by other articles:
-
Luck, P. C., Schneider, T., Wagner, J., Walther, I., Reif, U., Weber, S., Weist, K.
(2008). Community-acquired Legionnaires' disease caused by Legionella pneumophila serogroup 10 linked to the private home. J Med Microbiol
57: 240-243
[Abstract]
[Full Text]
-
Dirven, K., Ieven, M., Peeters, M. F, van der Zee, A., De Schrijver, K., Goossens, H.
(2005). Comparison of three Legionella urinary antigen assays during an outbreak of legionellosis in Belgium. J Med Microbiol
54: 1213-1216
[Abstract]
[Full Text]