Previous Article | Next Article ![]()
Journal of Clinical Microbiology, July 2001, p. 2590-2593, Vol. 39, No. 7
Department of Infection, Guy's, King's
College and St. Thomas' School of Medicine, St. Thomas' Hospital,
London SE1 7EH,1 and Osmetech plc,
Electra House, Crewe CW1 6WZ,2 United Kingdom
Received 11 January 2001/Returned for modification 17 February
2001/Accepted 22 April 2001
The Osmetech Microbial Analyzer (OMA) is an automated headspace
analyzer fitted with a novel detector system consisting of an array of
polymer sensors, each of which responds to different volatile organic
compounds. The system can be used for screening clinical urine
specimens for significant bacteriuria by sampling urine headspace and
subjecting the output of the multiple-detector response to principal
component analysis. The OMA readily distinguished artificially infected
urine samples from sterile controls. The OMA was then used to analyze
534 unselected clinical urine specimens, of which 21.5% had
significant bacteriuria (containing >105 CFU of
bacteria/ml). The sensitivity and specificity of the OMA compared with
conventional culture were 83.5 and 87.6%, respectively. The OMA is a
promising automated system for the rapid routine screening of urine
specimens, and further clinical trials are in progress.
Urines for bacterial culture are
among the most common specimens submitted to clinical microbiology
laboratories. In our own laboratory we receive up to 500 specimens a
day, but only 10 to 20% of these are subsequently found to be positive
for bacteria. A rapid screening method to exclude probable negatives
would save time and money as well as provide an improved clinical
service. Many rapid screening methods have been proposed, including the chemical detection of products of bacterial metabolism.
Analysis of these bacterial compounds has usually been performed by gas
chromatography (GC) or GC-mass spectrometry (GC-MS). Nonvolatile
compounds have been analyzed after chemical derivatization and more
volatile ones have been analyzed after organic extraction. Headspace
analysis is an adaptation of the latter strategy which eliminates the
need for extraction and simplifies sample handling. However, detection
in clinical samples (without culture) is difficult since bacterially
derived chemicals are present at low concentrations and easily swamped
by the chemical noise from the patient's body fluid. Previous work in
this area using GC headspace analysis (2-4, 6-8, 11) has
not led to a practical application of this method, and the more
sensitive and specialized technique of GC-MS has not been applied to
urine screening.
Industrial methods of direct headspace analysis have been improved with
the introduction of new types of conducting polymer sensors. When used
in multiple arrays and combined with computer pattern analysis of the
output data, these instruments can discriminate complex volatile
mixtures (14), and they have been used for the diagnosis
of a variety of clinical infections. Parry et al. (12, 13)
could identify the presence of In this study, we investigated the use of one such instrument, the
Osmetech Microbial Analyzer (OMA), for the analysis of infected and
uninfected human urine. The device samples the headspace above the
surface of the specimen and detects volatile compounds by using an
array of four conducting polymer sensors. Each sensor interacts with
different adsorbed volatile chemicals, depending on their size, shape,
and functional groups. We report here the results of analyses of
samples of reconstituted human urine (RHU) experimentally infected with
common urinary bacterial pathogens and of 534 clinical urine specimens
sent to the clinical laboratory for investigation of suspected bacteriuria.
Bacterial strains used in preliminary studies.
These were
strains of Escherichia coli, Klebsiella pneumoniae, Proteus
mirabilis, Staphylococcus aureus, Staphylococcus saprophyticus, and Enterococcus faecalis. There was one reference strain
from each species (shown in Table 1),
together with five recent clinical isolates of each species obtained
from patients with urinary tract infection. Organisms were identified
by standard laboratory methods and by using an API 20E system
(bioMerieux, Basingstoke, Hampshire, United Kingdom).
0095-1137/01/$04.00+0 DOI: 10.1128/JCM.39.7.2590-2593.2001
Copyright © 2001, American Society for Microbiology. All rights reserved.
Diagnosis of Bacteriuria by Detection of Volatile Organic
Compounds in Urine Using an Automated Headspace Analyzer with
Multiple Conducting Polymer Sensors
![]()
ABSTRACT
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
![]()
INTRODUCTION
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
-hemolytic streptococci by analyzing
contact dressings from chronic leg ulcers; Greenwood et al.
(5) showed that the pattern of volatile compounds released from dressings of infected chronic wounds could be used to monitor the
progress of wound healing; Chandiok et al. (1) analyzed volatile compounds from high vaginal swabs and, in a small group of
patients, were able to distinguish between patients with and without
bacterial vaginosis.
![]()
MATERIALS AND METHODS
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
TABLE 1.
PCA scores for reference and clinical bacterial isolates
cultured in RHU
Chemicals and reagents. The following items were purchased from Sigma Aldrich Co. Ltd. (Dorset, United Kingdom): 0.1 M hydrochloric acid, high-performance liquid chromatography-grade water, 0.1 M sodium hydroxide solution, 0.1 M ammonium hydroxide solution, and sodium sulfate. Nutrient broth, nutrient agar, and phosphate-buffered saline were obtained from Oxoid Ltd. (Basingstoke, United Kingdom) and made up according to the manufacturer's instructions. Control chemicals for calibrating the analyzer (USBlk, US1, US2, US3) were supplied by Osmetech plc (Crewe, United Kingdom).
RHU. Protein lyophilizate and urinary metabolite lyophilizate of male urine were obtained from Sigma Aldrich Co. Ltd. RHU was made by reconstituting these lyophilized fractions in an appropriate volume of sterile water as recommended by the supplier, pooling the two components, and finally filtering the resulting solution through a 0.2-µm-pore-size microbiological filter (Gelman Science, Northampton, United Kingdom).
Reconstituted urine specimens. Cultures of the reference strains and clinical isolates described above were made up in RHU to produce artificial urine specimens for testing. The bacterial strains were grown in nutrient broth overnight at 37°C without shaking. Serial dilutions of the cultures were made in phosphate-buffered saline to obtain a concentration of ~102 CFU/ml. One hundred microliters of this suspension was added to 10 ml of sterile RHU and cultured overnight without shaking. The bacterial concentration (total viable count) was determined by serial dilution and spread plate culture prior to analyzing in the OMA system. Each organism was cultured four times, and the results were expressed as an average of the four replicates. Blank control specimens were uninfected samples of RHU. Four blank controls were run with each organism.
Clinical urine specimens.
Clinical urine specimens were sent
to the microbiology laboratory from patients with suspected urinary
tract infection at Guy's and St. Thomas' Hospitals. Conventional
diagnosis of urinary tract infection was made by using standard methods
of microscopy and semiquantitative culture. Negative specimens were
defined as those containing <105 CFU of any organism per
ml as shown by conventional culture. Positive specimens contained
105 CFU of one or more strains of bacteria per ml. These
specimens were stored at 4°C and processed with the OMA within
24 h of culturing. Preliminary experiments showed excellent
reproducibility when clinical specimens containing
105
CFU of E. coli per ml were repeatedly analyzed. For this
reason, clinical samples were analyzed only once.
Instrumentation and operation.
The OMA consists of a sample
carousel which maintains sample vials at a constant temperature of
30 ± 0.5°C and presents the headspace to a sensor array for
analysis. The system is computer controlled, and data are captured to
files on a computer hard disk. Urine samples are analyzed in the
following way. One milliliter of culture or urine is transferred to a
22-ml sample vial containing 0.2 g of sodium sulfate and 0.1 ml of
1 M HCl. The vial is capped with a polytetrafluoroethylene-lined
silicone septum, placed in the carousel of the machine, and allowed to
equilibrate at 30°C for 5 min. The machine then automatically inserts
a needle through the sample vial septum, in order to analyze the
headspace. Nitrogen gas at 50% relative humidity is introduced above
the surface of the urine via the inner lumen of the coaxial needle. The
outer needle lumen allows the sample headspace to be delivered across the sensor array for 3 min at a flow rate of 60 ml/min. The sensor is
then allowed to recover before humid nitrogen gas is passed over the
sensor for a 4-min wash. The resistance of each of the polymer sensors
is measured during the sampling period, and the change
(
R) from the initial resistance (base resistance
R) is calculated. The needle is then removed, the carousel
moves the next sample into position, and the process is repeated. Each
reconstituted urine specimen in this study was analyzed four times, and
the results were recorded as the means of the four replicates. The clinical specimens were analyzed once.
Data handling.
The percentage resistance change output from
the sensors was measured and recorded every second. An average of the
R values over a 30-s optimal time period (for example,
the period of 150 to 180 s) was used as the raw data.
(i) Calibration of the sensor. The performance of the sensor array was characterized by running a set of controls. In preliminary experiments with artificially infected RHU, the controls were sterile (uninoculated) RHU samples. In subsequent tests on clinical specimens using a more sensitive instrument, the controls were chemical standards provided by the manufacturer, which had been shown to produce results identical to those of a more expensive sterile RHU control and allowed calibration of the sensors. The sensor raw data for the controls were transformed into a reference map using principal component analysis (PCA) (9, 15). PCA reduces the control chemical data matrix into a set of scores and loading vectors. Once the system had been calibrated with the controls, scores for subsequent samples were calculated by multiplying the sample analysis data by the reference loadings calculated in the calibration step.
(ii) Classification of samples. In the preliminary studies, artificially infected RHU samples were classified by comparing their projected PCA scores against thresholds set by 24 negative-control RHU samples. Thresholds were set on the PC1 and PC2 axes at 3 standard deviations (SD) from the average PC1 or PC2 scores of the control RHU. Samples outside either or both ranges were classified as positive. In subsequent studies, clinical specimens were classified according to their projected response on a reference map that had been calibrated with chemical controls provided by the manufacturer. Sensitivities and specificities were calculated on the premise that the culture result was the true result.
| |
RESULTS |
|---|
|
|
|---|
Artificially infected RHU.
These all had bacterial
concentrations of
107 CFU/ml. Table 1 shows the PCA
scores for the RHU samples; each result for cultured urines is the mean
of four replicates. The blank controls had a mean PC1 score of 0.99 and
a mean PC2 score of 0.13. This gives a PC1 threshold of
0.6 to +2.58
and a PC2 threshold of
0.65 to +0.91. The RHU medium inoculated with
reference or clinical bacterial isolates gave mean PC1 and PC2 scores
ranging from
39.7 to 11.0, respectively, for P. mirabilis,
and
2.1 to +0.12, respectively, for S. saprophyticus. The
other species gave mean PC1 and PC2 scores ranging from approximately
14 to
8 and
0.4 to
0.2. When these results were plotted on the
PC1 and PC2 axes, all the infected RHU results fell outside the
thresholds delineated by the controls.
Clinical urine specimens.
A total of 534 clinical samples were
analyzed both by culture and by a more sensitive OMA instrument than
that used to analyze the artificially infected urine. When significant
bacteriuria was defined as
105 CFU/ml by conventional
culture, there were 115 positive and 419 negative samples, giving a
positive prevalence of 21.5% (Table 2).
When specimens were classified by the OMA as positive or negative
relative to thresholds set on the PC1 and PC2 axes by the control
chemical calibrators, the sensitivity and specificity of the instrument
were 83.5 and 87.6%, respectively. As would be expected, when the
CFU-per-milliliter cutoff was lowered, the sensitivity fell and the
specificity rose. Thus, when significant bacteriuria was defined as
104 CFU/ml, the sensitivity and specificity were 72.3 and
89.4%, respectively. The detailed results are shown in Table
3.
|
|
| |
DISCUSSION |
|---|
|
|
|---|
The detection of volatile organic compounds in urine by gas-liquid chromatography (GLC) was used some years ago to detect bacteriuria. Manja and Rao (11) performed conventional GLC on urine samples incubated with appropriate supplements and showed that E. coli could be identified by the production of ethanol from lactose and Klebsiella spp. by the production of ethanol from adonitol. Hayward and colleagues (7-8) and Coloe (2, 3) utilized headspace GLC to identify volatile bacterial metabolites in artificial cultures and urine. Proteus spp. characteristically produced dimethyl disulfide and methyl mercaptan from L-methionine and trimethylamine from acetylcholine; E. coli and other coliforms produced ethanol from lactose or arabinose. This system was moderately successful in distinguishing infected and noninfected urine by direct analysis, but better results were obtained after incubation with arabinose and acetylcholine.
GC detection of bacteriuria, with or without MS, was not developed into
a practical diagnostic tool. In the present study we investigated a new
type of instrument fitted with a novel multiple polymer sensor array
and an automated headspace sampler for the direct detection of volatile
bacterial compounds in clinical urine specimens. In order to encourage
volatile compounds to enter the headspace, specimens were acidified and
were maintained at 30°C. The addition of sodium sulfate further
enhanced volatility and increased sensitivity (10). The
analyzer delivered a sample of headspace gas to an array of sensors,
and the mean
R of each of the sensors was analyzed by
PCA. Threshold PCA values were set by analysis of a series of controls,
and these were used to classify test specimens as positive (infected)
or negative (noninfected).
This technique was used to analyze artificially infected and uninfected
RHU. At this early stage of development of the instrument, commercial
pooled RHU was chosen as the test medium because, unlike clinical urine
specimens, different samples from the same RHU batch have a uniform
chemical composition, and other workers can use this medium to repeat
our experiments. The instrument we used initially was less sensitive
than the later one, and all the infected RHU specimens tested contained
107 CFU/ml. Infected specimens were readily distinguished
from negative controls, with PCA scores all more than 3 SD away from
the mean of the controls. The different PCA scores produced by the
different species suggest that with refinement the analyzer may have
the potential to distinguish different organisms associated with
urinary tract infection.
For the analysis of clinical urine specimens, a more sensitive instrument was used, and the sensors were calibrated each day with a set of chemical standards. It is not practical to use negative urine controls in routine clinical analyses, because clinical specimens vary greatly in their chemical composition.
When 534 clinical urine specimens were analyzed, the OMA was found to
have a sensitivity of 84% and a specificity of 88% relative to
standard culture results when significant bacteriuria was defined as
105 CFU/ml. When the cutoff was defined as
104 CFU/ml, not unexpectedly, the sensitivity fell and
the specificity rose.
There are a number of possible reasons for the false-negative
results. Firstly, natural volatile compounds in human urine, caused by disease or diet, might saturate the sensor detectors and
block the response to bacterial compounds by competitive inhibition. Secondly, bacterial volatile products might be lost, either by adsorption onto urinary cells or protein or by dissipation during any
prolonged delay between specimen collection and analysis. Finally, some
bacterial species may not produce volatile compounds that can be
detected by the present sensor array; this may be the case for the
group B
-hemolytic streptococci that were missed.
At the present time we do not know exactly to which of the volatile compounds out of the complex mixture in the headspace the instrument is responding; therefore, we cannot be sure that the present sensors are optimized for urine analysis. Furthermore, we do not know if there are other significant volatile compounds, presently undetected, which could be included in the analysis by the addition of other sensors to the array. Analysis of headspace by MS might reveal metabolic products that could be targeted by specially designed arrays in order to improve both the sensitivity and specificity of urinalysis. The present speed of analysis, although faster than that of GLC, is limited by the need for the sensors to recover after each sample. However, speed can be improved by using several sensor arrays in sequence, so that samples can be processed continuously without waiting for sensor recovery.
There is a need in the clinical laboratory to rapidly screen out culture-negative urine specimens, so that time and resources can be directed at further analysis of doubtful or possibly positive ones. The OMA shows considerable promise for automated screening, and more-extensive clinical trials with more-refined versions of the instrument are in progress.
| |
ACKNOWLEDGMENTS |
|---|
Financial support for this project was provided by Osmetech plc.
We are grateful to S. Sundaralingam for excellent technical assistance.
| |
FOOTNOTES |
|---|
* Corresponding author. Mailing address: Department of Infection, King's College, St. Thomas' Campus, St. Thomas' Hospital, London SE1 7EH, United Kingdom. Phone: 44 (0) 207 928 9292, ext. 3244. Fax: 44 (0) 207 928 0730. E-mail: gary.french{at}kcl.ac.uk.
| |
REFERENCES |
|---|
|
|
|---|
| 1. |
Chandiok, S.,
B. A. Crawley,
B. A. Oppenheim,
P. R. Chadwick,
S. Higgins, and K. C. Persaud.
1997.
Screening for bacterial vaginosis: a novel application of artificial nose technology.
J. Clin. Pathol.
50:790-791 |
| 2. |
Coloe, P. J.
1978.
Ethanol from arabinose: a rapid method for detecting Escherichia coli.
J. Clin. Pathol.
31:361-364 |
| 3. |
Coloe, P. J.
1978.
Headspace gas liquid chromatography for rapid detection of Escherichia coli and Proteus mirabilis in urine.
J. Clin. Pathol.
31:365-369 |
| 4. | Davies, T., and N. J. Hayward. 1984. Volatile products from acetylcholine as markers in the rapid urine test using headspace gas-liquid chromatography. J. Chromatogr. 307:11-21[Medline]. |
| 5. | Greenwood, J. E., B. A. Crawley, S. L. Clark, P. R. Chadwick, D. A. Ellison, B. A. Oppenheim, and C. N. McCollum. 1997. Monitoring wound healing by odor. J. Wound Care 6:219-221[Medline]. |
| 6. | Hayward, N. J. 1983. Headspace gas-liquid chromatography for the rapid laboratory diagnosis of urinary tract infections caused by Enterobacteria. J. Chromatogr. 274:27-35[Medline]. |
| 7. |
Hayward, N. J., and T. H. Jeavons.
1977.
Assessment of technique for rapid detection of Escherichia coli and Proteus species in urine by head-space gas-liquid chromatography.
J. Clin. Microbiol.
6:202-208 |
| 8. |
Hayward, N. J.,
T. H. Jeavons,
A. J. C. Nicholson, and A. G. Thornton.
1977.
Development of specific tests for rapid detection of Escherichia coli and all species of Proteus in urine.
J. Clin. Microbiol.
6:195-201 |
| 9. | Jackson, J. E. 1981. Principal component and factor analysis part 1. Principal components. J. Qual. Technol. 13:46-58. |
| 10. |
Larsson, L.,
P.-A. Mårdh, and G. Odham.
1978.
Detection of alcohols and volatile fatty acids by head-space gas chromatography in identification of anaerobic bacteria.
J. Clin. Microbiol.
7:23-27 |
| 11. |
Manja, K. S., and K. M. Rao.
1983.
Gas-chromatographic detection of urinary tract infections caused by Escherichia coli and Klebsiella spp.
J. Clin. Microbiol.
17:264-266 |
| 12. |
Parry, A. D.,
P. R. Chadwick,
D. Simon,
B. A. Oppenheim, and C. N. McCollum.
1995.
Leg ulcer odor detection identifies -haemolytic streptococcal infection.
J. Wound Care
4:404-406[Medline].
|
| 13. |
Parry, A. D.,
P. R. Chadwick,
D. Simon,
B. A. Oppenheim, and C. N. McCollum.
1995.
Detection of -haemolytic streptococcal infection by analysis of leg ulcer odor, p. 134-137.
In
G. W. Cherry, D. J. Leaper, J. C. Lawrence, and P. Milward (ed.), Proceedings of the 4th European Conference on Advances in Wound Management Macmillan Magazines, London, United Kingdom.
|
| 14. | Persaud, K. C., and P. J. Travers. 1997. Arrays of broad specificity films for sensing volatile chemicals, p. 563-592. In E. Kress-Rogers (ed.), Handbook of biosensors and electronic noses, medicine, food and the environment. CRC Press Inc., Boca Raton, Fla. |
| 15. | Wold, S., K. Esbensen, and P. Geladi. 1987. Principal component analysis. Chem. Int. Lab. Syst. 2:35-37. |
This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»