Previous Article | Next Article 
Journal of Clinical Microbiology, September 2001, p. 3092-3098, Vol. 39, No. 9
0095-1137/01/$04.00+0 DOI: 10.1128/JCM.39.9.3092-3098.2001
Copyright © 2001, American Society for Microbiology. All rights reserved.
Enhancing the Specificity of the COBAS AMPLICOR CT/NG Test for
Neisseria gonorrhoeae by Retesting Specimens with
Equivocal Results
Barbara
Van Der Pol,1
David H.
Martin,2,3
Julius
Schachter,4
Thomas C.
Quinn,5,6
Charlotte A.
Gaydos,6
Robert B.
Jones,1
Kimberly
Crotchfelt,6
Jeanne
Moncada,4
D.
Jungkind,7
Buffy
Turner,8
Cynthia
Peyton,8
James F.
Kelly,9
Judith B.
Weiss,9 and
Maurice
Rosenstraus9,*
Indiana University School of Medicine,
Indianapolis, Indiana1; Louisiana State
University2 and City of New Orleans'
Delgado Clinic,3 New Orleans, Louisiana;
University of California
San Francisco, San
Francisco,4 and Roche Molecular Systems,
Pleasanton,9 California; National
Institute of Allergy and Infectious Diseases, National Institutes
of Health, Bethesda,5 and Johns Hopkins
University, Baltimore,6 Maryland;
Thomas Jefferson University Hospital, Pittsburgh,
Pennsylvania7; and University of
Texas Medical Branch, Galveston, Texas8
Received 7 December 2000/Returned for modification 28 April
2001/Accepted 20 June 2001
 |
ABSTRACT |
The COBAS AMPLICOR CT/NG test for Neisseria gonorrhoeae
cross-reacts with certain strains of nonpathogenic
Neisseria species. In some strains, the target sequence is
identical to that of N. gonorrhoeae, whereas other strains
have a small number of mismatches within the regions recognized by the
primers or probe used in the COBAS AMPLICOR NG test. These
cross-reactive strains are occasionally present in urogenital
specimens, causing false-positive results in the COBAS AMPLICOR NG
test. Analysis of the data generated in a large multicenter clinical
trial showed that 2.9% of the specimens gave signals between
A660s of 0.2 and 3.5 but that one-half of these
equivocal specimens did not contain N. gonorrhoeae. Most of
these equivocal specimens were correctly classified as true positive or
true negative by retesting in duplicate and defining a PCR-positive
result as two of three results with an A660 of
2.0. If specimens had been classified as positive or negative based
on a single test result using a cutoff of an
A660 of 0.2, specificity would have ranged from
96.2 to 98.9% depending on specimen type, sex, and presence of
symptoms. By employing the equivocal zone-retesting algorithm,
specificity increased to 98.6 to 99.9% with little effect (0.1 to
4.9% decrease) on sensitivity in most specimen types, enabling the
test to achieve a positive predictive value of at least 90% in
populations with a prevalence of 4% or higher. In
lower-prevalence populations, the test could be used to screen for
presumptive infections that would have to be confirmed by an
independent test.
 |
INTRODUCTION |
The COBAS AMPLICOR CT/NG test
provides a powerful diagnostic tool for screening for both chlamydial
and gonococcal infections. We and others have observed that the NG
portion of the test (COBAS AMPLICOR NG) cross-reacts with some isolates
of certain nonpathogenic Neisseria species (4).
When the original, recommended cutoff (A660 of
0.2) is used, the COBAS AMPLICOR NG test can produce false-positive
results for Neisseria gonorrhoeae, presumably due to the
presence of cross-reactive Neisseria species in some
urogenital specimens. In one population, approximately 26% of COBAS
AMPLICOR NG-positive results were false positives, which corresponded
to approximately 3% of the total population (4). In
contrast, the same laboratory observed fewer than 1% false-positive
results among urogenital specimens from a second population
(4).
In this paper, we confirm that the test does cross-react with some
isolates of Neisseria subflava (4) and
Neisseria cinerea and compare the target sequences in these
cross-reactive species with that of N. gonorrhoeae. Using
data from a multicenter trial (n = 4,173 patients;
prevalence = 13.1%) conducted at six sites in the United
States (8), we show how test sensitivity and specificity vary with the cutoff value used. We then assess whether both sensitivity and specificity can be optimized by establishing a
large equivocal zone and using an algorithm that involves additional testing of specimens yielding equivocal results. Implementation of the
equivocal zone-retesting algorithm identified by this analysis produced
good specificity (98.8 to 99.9%) without sacrificing sensitivity in
the multicenter trial (8).
 |
MATERIALS AND METHODS |
Clinical samples.
The data analyzed here were obtained
during a multicenter trial designed to evaluate the performance of the
COBAS AMPLICOR NG test (8). Briefly, specimens were
collected from consecutive, consenting individuals visiting sexually
transmitted disease clinics or family planning centers in six
geographical regions. Two endocervical swab specimens from women and
two urethral swab specimens from men were collected by standard
procedures. The first swab was processed for gonorrhea culture
according to each laboratory's standard procedure. Gram-negative
diplococci were confirmed as N. gonorrhoeae by glucose
utilization profiles or antibody reactivity. The second swab was placed
in chlamydial culture transport medium, which was then used for PCR
testing (12). Ten to 50 ml of first-catch urine was also
collected from both men and women. Four aliquots of each culture
transport medium swab and urine specimen were stored at
20°C for
retesting and for use in discrepant analysis. NG culture results were
available for all patients. PCR results for both swab and urine
specimens were available for 4,200 of 4,277 patients. One PCR result
was missing for the remaining 77 patients because the sample either was
not collected or was inhibitory when tested.
PCR testing.
Each specimen was processed and tested in the
COBAS AMPLICOR NG test as previously described (8). For
each processed specimen, the Chlamydia trachomatis, N. gonorrhoeae, and internal control (IC) target DNAs were
simultaneously amplified in a single reaction that contained two primer
pairs, one specific for C. trachomatis and the IC and one
specific for N. gonorrhoeae. The resulting amplification
products were detected separately by hybridization to magnetic
microparticles coated with N. gonorrhoeae-, C. trachomatis-, and IC-specific oligonucleotide probes.
All specimens with N. gonorrhoeae signals above an
A660 of 0.2 were retested in duplicate. Where
indicated, the results of the repeat testing were used to classify
specimens as COBAS AMPLICOR NG positive or negative (see below).
Classification of COBAS AMPLICOR results.
Specimens yielding
COBAS AMPLICOR NG signals above an A660 of 0.2 were classified as presumptive COBAS AMPLICOR NG positive. To identify
a cutoff that produced the best combination of sensitivity and
specificity, these presumptive positive specimens were reclassified as
COBAS AMPLICOR NG positive or negative using various cutoff values.
Where indicated, the combined results of initial and repeat testing
were compared at various cutoff values to determine the COBAS AMPLICOR
result. Specimens yielding COBAS AMPLICOR NG signals below an
A660 of 0.2 were interpreted as negative,
provided that the IC signal was above an A660 of
0.2. If the IC signal was below an A660 of 0.2, another aliquot of the original specimen was processed and tested.
Resolution of presumptive results.
Specimens from
culture-negative patients that yielded COBAS AMPLICOR NG signals above
an A660 of 0.2 were also tested with a PCR assay
for an alternative target DNA sequence located within the N. gonorrhoeae 16S rRNA gene (S. Y. Lu, S. Y. Kao, S. Silver, A. Purohit, M. Longiaru, and T. J. White, Abstr. 91st
Gen. Meet. Am. Soc. Microbiol. 1991, abstr. C-115, p. 361, 1991). If
the specimen was negative in the 16S rRNA test, the other specimen type
(swab or urine) was also tested for 16S rRNA to provide evidence for N. gonorrhoeae infection; this test was performed
regardless of whether the second specimen type had
originally tested positive for N. gonorrhoeae.
Definition of N. gonorrhoeae-positive and N. gonorrhoeae-negative specimens.
Specimens were classified as
positive or negative for N. gonorrhoeae based on a
combination of culture, COBAS AMPLICOR NG, and, where necessary, NG 16S
rRNA PCR results. For the purpose of this study, a specimen was
considered N. gonorrhoeae positive if (i) it was obtained
from a patient who was positive by culture or (ii) the initial COBAS
AMPLICOR NG test yielded a signal at an A660 of
0.2 and either of the two specimen types from the patient tested
positive by the NG 16S rRNA PCR assay. Specimens from culture-negative
patients were classified as N. gonorrhoeae negative if they
either (i) yielded a COBAS AMPLICOR NG signal at an
A660 of <0.2 or (ii) yielded an initial COBAS
AMPLICOR NG signal at an A660 of
0.2 that was
not confirmed by a positive NG 16S rRNA PCR result on either the swab
or urine specimen.
 |
RESULTS |
Cross-reactivity with nonpathogenic neisseriae.
The COBAS
AMPLICOR CT/NG test for N. gonorrhoeae gave negative results
when multiple isolates of various nongonococcal Neisseria strains were tested at concentrations of 104 cells per
amplification reaction (Table 1). In
contrast, one of six N. cinerea isolates and two of
nine Neisseria subflava subsp. perflava isolates
gave strong positive signals (Table 1). In the two cross-reactive
N. subflava subsp. perflava isolates, the primer
and probe sequences, which are contained within the putative cytosine
DNA methyltransferase gene (9), were identical to that
found in N. gonorrhoeae; the remainder of the amplified target sequence contained no differences and a 1-bp substitution, respectively. In the cross-reactive N. cinerea isolate, the
probe sequence differed from the N. gonorrhoeae sequence by
1 bp; the primer sequences were not determined, and there were six
additional base pair substitutions in the remainder of the target
sequence. The species identification of these isolates was confirmed by biochemical testing and 16S rRNA gene sequencing.
Clinical trial results.
Specimens were obtained from 1,110 asymptomatic women, 1,155 symptomatic women, 721 asymptomatic men, and
1,291 symptomatic men; the corresponding prevalence of N. gonorrhoeae infection was 5.5, 7.7, 3.3, and 30.0%, respectively.
Results for matching swab and urine specimens were available for 4,200 of 4,277 patients to give a total of 8,477 COBAS AMPLICOR NG tests
performed. Of these tests, 1,093 were performed on N. gonorrhoeae-positive specimens and 7,384 were performed on
N. gonorrhoeae-negative specimens. Nine of the 77 patients
with a missing PCR result for one specimen type gave a COBAS AMPLICOR
NG-positive result for the other sample (three positive swab and six
positive urine specimens); the missing results were due to inhibition
in the alternate specimen. All six urine specimens and one of the three
swab specimens were confirmed positive by culture. Thus, only two swab
specimens may have been affected by the missing urine PCR
result. These two samples were classified as N. gonorrhoeae
negative (based on the criteria defined in Materials and Methods).
These two results would have been classified as false negative had the
urine specimen tested NG positive by COBAS AMPLICOR and been confirmed
by the 16S rRNA PCR test.
When COBAS AMPLICOR NG results were evaluated using a cutoff of an
A660 of 0.2,

92.9% sensitivity was observed
in all specimen
types except female urine (Table
2). In contrast, specificity
was
relatively low, ranging from 96.2 to 98.9% (Table
2). Sensitivity
was
very high for all test sites (Table
3).
Three test sites
obtained an overall specificity of approximately 99%,
whereas
specificity at the other three sites was approximately 3%
lower
(Table
3). Presumably, lower specificity was due to the presence
of cross-reactive
Neisseria species in a small fraction of
the
specimens. To enhance specificity, we sought a testing algorithm
that could distinguish specimens containing cross-reactive
Neisseria species from those that contained
N. gonorrhoeae.
View this table:
[in this window]
[in a new window]
|
TABLE 2.
Effect of using an expanded equivocal zone on test
sensitivity and specificity for different specimen types
|
|
The distribution of COBAS AMPLICOR NG signals suggested that using a
higher cutoff could enhance specificity (Fig.
1). Fifty-three
percent of the specimens
with signals between
A660s of 0.2 and
3.5 were
N. gonorrhoeae negative. Within this absorbance range,
the
proportion of
N. gonorrhoeae-negative specimens tended to
decrease as the signal increased. Similar distributions were obtained
when results for each specimen type (urine, urethral swab, and
endocervical swab) were analyzed separately (data not shown).
Likewise,
the distribution of signals was similar for each of
the six sites (data
not shown). However, the proportion of
N. gonorrhoeae-negative specimens with signals at an
A660 of

0.2
was approximately 3% higher at
three sites (Table
3).

View larger version (26K):
[in this window]
[in a new window]
|
FIG. 1.
Histogram showing the distribution of signals produced
in the initial COBAS AMPLICOR NG test by swab and urine specimens
obtained from symptomatic and asymptomatic men and women. Separate
distributions are shown for N. gonorrhoeae-negative
specimens and N. gonorrhoeae-positive specimens.
Specimens were classified as N. gonorrhoeae negative or
N. gonorrhoeae positive as described in Materials and
Methods. Within each group, specimens had signals equal to or greater
than the lower value and less than the upper value of the range shown
on the x axis. The left- and rightmost bars are off scale;
the numbers of specimens in these two categories are shown above the
bars. The percentages of N. gonorrhoeae-positive and
N. gonorrhoeae-negative specimens that give COBAS AMPLICOR
NG-negative (A660 < 0.2), -equivocal
(A660 0.2 and < 3.5), and
-positive (A660 3.5) results are shown
beneath the graph. ND, no data.
|
|
Given this distribution of results, cumulative distribution analysis
was performed to determine how sensitivity and specificity
varied using
different cutoff values (Fig.
2). Maximum
sensitivity
was achieved with a cutoff of an
A660 of 0.2. Sensitivity decreased
slowly as the
cutoff was increased to an
A660 of 3.0 and then
decreased more rapidly when the cutoff was further increased to
an
A660 of 3.5. Specificity increased continuously
as the cutoff
was increased from an
A660 of 0.2 to 3.5. We were unable to identify
a cutoff value that produced a
combination of greater than 90%
sensitivity and 99.5% specificity,
although a cutoff of an
A660 of 3.0 nearly met
these criteria with a sensitivity of 89.9% and
a specificity of
99.6%.

View larger version (17K):
[in this window]
[in a new window]
|
FIG. 2.
Cumulative distribution curves for initial test results
showing the sensitivity and specificity that would be obtained for
various cutoffs. Performance was calculated from the data shown in Fig.
1.
|
|
Identification of an equivocal zone and retesting algorithm.
Since cumulative distribution analysis (Fig. 2) did not identify a
single cutoff that yielded acceptable performance, we evaluated whether
employing an equivocal zone plus additional testing of the original
specimen in duplicate could enhance performance. Specimens were
classified as COBAS AMPLICOR NG negative if the A660 was below 0.2, positive if the
A660 was
3.5, and equivocal if the
A660 was between 0.2 and 3.5.
For the entire population, 85.9% of specimens were classified as COBAS
AMPLICOR NG negative (i.e.,
A660 < 0.2),
of which the
overwhelming majority (99.3%) were
N. gonorrhoeae negative (Fig.
1). Eleven percent of specimens were
classified as COBAS AMPLICOR
NG positive (i.e.,
A660 
3.5), and the vast majority
(98.2%)
of these were
N. gonorrhoeae positive (Fig.
1).
Only 2.9% of specimens
were classified as equivocal, with
approximately equal numbers
being
N. gonorrhoeae positive
and
N. gonorrhoeae negative. Thus,
only a small portion of
all samples fell into the equivocal zone
and had to be adjudicated by
additional
testing.
Initially equivocal specimens were interpreted as COBAS AMPLICOR NG
positive or COBAS AMPLICOR NG negative if at least two
of the three
tests (initial plus additional test results) yielded
signals above or
below the selected cutoff, respectively. Cumulative
distribution
analysis showed that specificity increased and sensitivity
decreased as
the cutoff used for repeat testing of these equivocal
specimens
increased (Fig.
3). Based on these data,
we selected
a cutoff of an
A660 of 2.0 for
resolving equivocal results. Use
of a cutoff value greater than an
A660 of 2.0 would yield somewhat
fewer
false-positive results, but this would come at the cost
of a
substantial increase in false-negative results (Fig.
3).
For cutoff
values less than an
A660 of 2.0, the number of
equivocal
specimens giving false-positive results was too high to
achieve
acceptable specificity.

View larger version (14K):
[in this window]
[in a new window]
|
FIG. 3.
Cumulative distribution curves for the combination of
initial and repeat test results showing the sensitivity and specificity
that would be obtained for various cutoffs. All specimens with an
initial A660 equal to or greater than 3.5 were
interpreted as COBAS AMPLICOR NG positive. The x-axis value
represents the cutoff used to interpret the three results for those
specimens with initial optical densities of between 0.2 and 3.5. These
initially equivocal specimens were interpreted as COBAS AMPLICOR NG
positive if at least two of the three results were above the value
shown on the x axis.
|
|
Performance with the equivocal zone-retesting algorithm.
The
equivocal zone-retesting algorithm increased specificity by 0.8 to
2.5% compared to the specificity achieved when each sample type
(gender, presence of symptoms, and urine versus swab) was scored as
COBAS AMPLICOR NG positive or negative based on a single test result
and a cutoff of an A660 of 0.2 (Table 2). The
final specificity achieved using this algorithm ranged from 98.6% for
urethral swab specimens from symptomatic men to 99.9% for urine
specimens from asymptomatic men and women (Table 2). At the three study
sites that exhibited 99% specificity for a cutoff of an
A660 of 0.2, the equivocal zone-retesting
algorithm increased specificity to 99.9% (Table 3). Use of the
algorithm increased specificity to similarly high levels at two of the
three sites that had exhibited lower specificity using a cutoff
of an A660 of 0.2 (Table 3). However, at
the remaining site, the specificity increased to only 98.3%
(Table 3).
The overall increase in specificity was accompanied by a decrease in
sensitivity in all specimen types. The decrease in sensitivity
ranged
from 0.1 to 1.3% for urethral swab and urine specimens
from
symptomatic men and for endocervical swab specimens (Table
2). The
decrease in sensitivity was limited because a relatively
large fraction
of specimens with equivocal results gave signals
above an
A660 of 2.0 in two out of three tests. More
substantial
decreases in sensitivity were observed for urethral swab
and urine
specimens from asymptomatic men and for urine specimens from
women
(Table
2). Most likely, the very low organism load in these
specimen
types resulted in sample variation, accounting for the failure
to consistently obtain strong positive
signals.
Five of the six sites exhibited a small (0 to 4%) decrease in
sensitivity associated with the use of the equivocal zone-retesting
algorithm (Table
3). The more substantial (10%) decrease in
sensitivity
at the remaining site occurred because this site tested
only female
specimens. Thus, half of the results from this site were
obtained
from female urine
specimens.
Reduced COBAS AMPLICOR NG signals due to competition from amplification
of
C. trachomatis DNA in coinfected specimens were
not
responsible for the decrease in sensitivity. Among 117
N. gonorrhoeae-positive specimens that had an equivocal result in
the
COBAS AMPLICOR NG, 25 were coinfected with
C. trachomatis and 92 were not. Twenty (80%) of the 25 coinfected specimens were
interpreted as NG positive using the equivocal zone-retesting
algorithm
(4 were interpreted as NG negative and 1 was inhibitory
when retested).
For comparison, 52 (57%) of the 92 specimens infected
with NG only
were interpreted as NG positive using the equivocal
zone-retesting
algorithm.
 |
DISCUSSION |
The results of this study demonstrate that the specificity of the
COBAS AMPLICOR CT/NG test for N. gonorrhoeae is
substantially enhanced by employing an equivocal zone and resolving
equivocal results by retesting the original specimen. Without an
equivocal zone, specificity ranged from 96.2 to 98.9%, which would
result in a positive predictive value less than 90% when the
prevalence of infection is below 20%. Specificity ranged from 98.6 to
99.9% when the equivocal zone-retesting algorithm was employed,
enabling the test to achieve a positive predictive value of at least
90% when performed on urogenital specimens obtained from populations with a prevalence of 4% or higher. The test should not, however, be
used to test throat swab specimens, since up to 50% of samples will
give false-positive results (J. Weiss and M. Rosenstraus, unpublished results).
Without the equivocal zone, specificity was low (96%) at some sites
and high (99%) at others. In another study, a single lab observed a
high frequency of false-positive results in samples from one population
but a very low frequency in a different population (4).
The reasons for this variation are not clear. Our data show that the
equivocal zone-retesting algorithm enhances specificity to greater than
99.3% for most populations with a high rate of false-positive results
and to 99.9% for populations with a low rate of false positives. This
specificity approaches the 99.1 to 100.0% specificity previously
reported for the Abbott LCx N. gonorrhoeae test (1-3,
5, 10, 11, 13).
The increase in specificity was accompanied by a slight decrease in
sensitivity, but this did not impact the test's diagnostic utility.
Sensitivity still exceeded 95% for urethral swab and urine specimens
from symptomatic males and for endocervical swab specimens. Sensitivity
was more severely compromised for specimens from asymptomatic men and
for female urine specimens. Although the sensitivity of asymptomatic
male urine testing was somewhat lower (10 to 20%) than that of culture
(8), testing noninvasively collected urine specimens may
be preferable to collecting and culturing urethral swab samples. The
sensitivity decrease for female urine specimens did not impact utility;
the test has suboptimal sensitivity for female urine with or without
the equivocal zone.
The same equivocal zone-retesting algorithm is used in the microwell
plate-based AMPLICOR CT/NG test for N. gonorrhoeae. The COBAS and microwell plate formats exhibited similar performance characteristics when the tests were performed in parallel on the specimens used in the present study (8). With the
equivocal zone-retesting algorithm, the specificity of the microwell
plate format ranged from 98.4 to 99.7% depending on the specimen
type being analyzed (data not shown).
Although a wide equivocal zone is required to maximize specificity, the
additional testing burden will be small. Depending on the test
population, approximately 1 to 4% of N. gonorrhoeae-negative specimens will give equivocal results, while
20% of N. gonorrhoeae-positive specimens will give
equivocal results. Thus, only 2 to 7% of specimens will need to be
retested in most populations with an N. gonorrhoeae prevalence ranging from 5 to 20%.
Our observation that false-positive specimens generally give weak
signals is consistent with an earlier study, which showed that 14 of 17 specimens with signals between A450s of 0.2 and 0.8 in the microwell plate-based AMPLICOR CT/NG test were
false positive (4). These specimens, which
presumably contain cross-reactive neisseriae, may yield relatively weak
signals for two reasons. First, the cross-reactive neisseriae are
likely to be present at a low concentration (6). Second,
some, but not all, of the cross-reactive strains contain mismatches in
the primer or probe regions, which could reduce the efficiency of
amplification or hybridization and result in weak signals when the
target is present at a low level. Finally, some of the specimens
classified as false positive could actually be N. gonorrhoeae infected. Specimens containing a low concentration of
N. gonorrhoeae could have produced some of the initial
positive results that were not confirmed by additional testing.
Low organism concentration and probe mismatches also explain how
retesting initially equivocal specimens can distinguish most of those
containing cross-reactive neisseriae from N. gonorrhoeae-containing specimens. Cross-reactive specimens tend to
give negative results when retested because the low target
concentration produces aliquot-to-aliquot fluctuation in the amount of
target tested. Those specimens that do give reproducible results will
often give low signals if the cross-reactive organism has a mismatch in
the probe region. Specimens containing substantial amounts of
cross-reactive organisms or strains that do not have mismatches in the
probe region probably account for the small number of N. gonorrhoeae-negative specimens that give signals above an
A660 of 3.5 when initially tested. These
specimens, along with the few equivocal, cross-reactive specimens that
are not identified by retesting, explain why 100% specificity was not achieved.
Because some cross-reactive isolates contain target
sequences identical to those found in N. gonorrhoeae,
we hypothesize that these isolates arose by recombination between
N. gonorrhoeae and the non-cross-reactive
neisseriae. Genetic exchange between neisseriae is a common phenomenon,
occurring both in genes coding for housekeeping and in genes coding for
pathogen-associated proteins, and may play a role in the generation of
new subspecies or species (7). Genetic drift subsequent to
the genetic exchange probably accounts for the isolates that have one
or more nucleotide differences compared to the N. gonorrhoeae sequence. These differences generally do not result in
amino acid changes.
In summary, urogenital specimens that initially give false-positive
results with the COBAS AMPLICOR CT/NG test for N. gonorrhoeae can be distinguished from most N. gonorrhoeae-containing urogenital specimens by retesting all
samples that give results within an equivocal range. The algorithm used
here is the one recommended by the manufacturer in the Food and Drug
Administration-cleared version of the COBAS Methods Manual. Specificity
ranges from 98.6 to 99.9% when the equivocal zone-retesting
algorithm is used, enabling the test to achieve a 90% or higher
positive predictive value when performed on urogenital specimens
obtained from populations with a prevalence of 4% or higher.
Finally, the approach of retesting specimens with relatively low
signals may prove useful for enhancing specificity in other N. gonorrhoeae assay systems.
 |
ACKNOWLEDGMENTS |
We thank James Williams and Laura Brandenburg for laboratory
assistance and the many clinicians at each site for assistance with
enrolling patients.
This work was supported by Roche Molecular Systems.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: Roche Molecular
Systems, 1080 U.S. Highway 202 South, Somerville, NJ 08876. Phone: (908) 253-7463. Fax: (908) 253-3318. E-mail:
maurice.rosenstraus{at}roche.com.
 |
REFERENCES |
| 1.
|
Buimer, M.,
G. J. J. Van Doornum,
S. Ching,
P. G. H. Peerbooms,
P. K. Plier,
D. Ram, and H. H. Lee.
1996.
Detection of Chlamydia trachomatis and Neisseria gonorrhoeae by ligase chain reaction-based assays with clinical specimens from various sites: implications for diagnostic testing and screening.
J. Clin. Microbiol.
34:2395-2400[Abstract].
|
| 2.
|
Carroll, K. C.,
W. E. Aldeen,
M. Morrison,
R. Anderson,
D. Lee, and S. Mottice.
1998.
Evaluation of the Abbott LCx ligase chain reaction assay for detection of Chlamydia trachomatis and Neisseria gonorrhoeae in urine and genital swab specimens from a sexually transmitted disease population.
J. Clin. Microbiol.
36:1630-1633[Abstract/Free Full Text].
|
| 3.
|
Ching, S.,
H. Lee,
E. W. Hook III,
M. R. Jacobs, and J. Zenilman.
1995.
Ligase chain reaction for detection of Neisseria gonorrhoeae in urogenital swabs.
J. Clin. Microbiol.
33:3111-3114[Abstract].
|
| 4.
|
Farrell, D.
1999.
Evaluation of AMPLICOR Neisseria gonorrhoeae PCR using cppB nested PCR and 16S rRNA PCR.
J. Clin. Microbiol.
37:386-390[Abstract/Free Full Text].
|
| 5.
|
Kehl, S. C.,
K. Georgakas,
G. R. Swain,
G. Sedmak,
S. Gradus,
A. Singh, and S. Foldy.
1998.
Evaluation of the Abbott LCx assay for detection of Neisseria gonorrhoeae in endocervical swab specimens from females.
J. Clin. Microbiol.
36:3549-3551[Abstract/Free Full Text].
|
| 6.
|
Knapp, J., and R. J. Rice.
1995.
Neisseria and Branhamella, p. 324-340.
In
P. R. Murray, E. J. Baron, M. A. Pfaller, F. C. Tenover, and R. H. Yolken (ed.), Manual of clinical microbiology, 6th ed. ASM Press, Washington, D.C.
|
| 7.
|
Maiden, M. C. J.,
B. Malorny, and M. Achtman.
1996.
A global gene pool in the neisseriae.
Mol. Microbiol.
21:1297-1298[CrossRef][Medline].
|
| 8.
|
Martin, D. H.,
C. Cammarata,
B. Van Der Pol,
R. B. Jones,
T. C. Quinn,
C. A. Gaydos,
K. Crotchfelt,
J. Schachter,
J. Moncada,
D. Jungkind,
B. Turner, and C. Peyton.
2000.
Multicenter evaluation of AMPLICOR and automated COBAS AMPLICOR CT/NG tests for Neisseria gonorrhoeae.
J. Clin. Microbiol.
38:3544-3549[Abstract/Free Full Text].
|
| 9.
|
Miyada, C. G., and T. L. Born.
1991.
A DNA sequence for the discrimination of Neisseria gonorrhoeae from other Neisseria species.
Mol. Cell. Probes
5:327-335[Medline].
|
| 10.
|
Smith, K. R.,
S. Ching,
H. Lee,
Y. Ohhashi,
H. Y. Hu,
H. C. Fisher III, and E. W. Hook, III.
1995.
Evaluation of the ligase chain reaction for use with urine for identification of Neisseria gonorrhoeae infections in females attending a sexually transmitted disease clinic.
J. Clin. Microbiol.
33:455-457[Abstract].
|
| 11.
|
Stary, A.,
S.-F. Ching,
L. Teodorowicz, and H. Lee.
1997.
Comparison of ligase chain reaction and culture for detection of Neisseria gonorrhoeae in genital and extragenital specimens.
J. Clin. Microbiol.
33:239-242[Abstract].
|
| 12.
|
Van Der Pol, B.,
T. C. Quinn,
C. A. Gaydos,
K. Crotchfelt,
J. Schachter,
J. Moncada,
D. Jungkind,
D. H. Martin,
B. Turner,
C. Peyton, and R. B. Jones.
2000.
Multicenter evaluation of the AMPLICOR and Automated COBAS AMPLICOR CT/NG tests for detection of Chlamydia trachomatis.
J. Clin. Microbiol.
38:1105-1112[Abstract/Free Full Text].
|
| 13.
|
Xu, K.,
V. Glanton,
S. R. Johnson,
C. Beck-Sague,
V. Bhullar,
D. H. Candal,
K. S. Pettus,
C. E. Farshy, and C. M. Black.
1998.
Detection of Neisseria gonorrhoeae infection by ligase chain reaction testing of urine among adolescent women with and without Chlamydia trachomatis infection.
Sex. Transm. Dis.
25:533-538[Medline].
|
Journal of Clinical Microbiology, September 2001, p. 3092-3098, Vol. 39, No. 9
0095-1137/01/$04.00+0 DOI: 10.1128/JCM.39.9.3092-3098.2001
Copyright © 2001, American Society for Microbiology. All rights reserved.
This article has been cited by other articles:
-
Moncada, J., Schachter, J., Liska, S., Shayevich, C., Klausner, J. D.
(2009). Evaluation of Self-Collected Glans and Rectal Swabs from Men Who Have Sex with Men for Detection of Chlamydia trachomatis and Neisseria gonorrhoeae by Use of Nucleic Acid Amplification Tests. J. Clin. Microbiol.
47: 1657-1662
[Abstract]
[Full Text]
-
Bachmann, L. H., Johnson, R. E., Cheng, H., Markowitz, L. E., Papp, J. R., Hook, E. W. III
(2009). Nucleic Acid Amplification Tests for Diagnosis of Neisseria gonorrhoeae Oropharyngeal Infections. J. Clin. Microbiol.
47: 902-907
[Abstract]
[Full Text]
-
Moncada, J., Donegan, E., Schachter, J.
(2008). Evaluation of CDC-Recommended Approaches for Confirmatory Testing of Positive Neisseria gonorrhoeae Nucleic Acid Amplification Test Results. J. Clin. Microbiol.
46: 1614-1619
[Abstract]
[Full Text]
-
Peter-Getzlaff, S., Luethy, J., Springer, B., Kaul, K. L., Mangold, K. A.
(2007). Diagnostic Value of Molecular Confirmation Assays for Neisseria gonorrhoeae. J. Clin. Microbiol.
45: 3856-3858
[Full Text]
-
Mangold, K. A., Regner, M., Tajuddin, M., Tajuddin, A. M., Jennings, L., Du, H., Kaul, K. L.
(2007). Neisseria Species Identification Assay for the Confirmation of Neisseria gonorrhoeae-Positive Results of the COBAS Amplicor PCR. J. Clin. Microbiol.
45: 1403-1409
[Abstract]
[Full Text]
-
Alary, M, Gbenafa-Agossa, C, Aina, G, Ndour, M, Labbe, A C, Fortin, D, Steele, M, Peeling, R W
(2006). Evaluation of a rapid point-of-care test for the detection of gonococcal infection among female sex workers in Benin. Sex. Transm. Infect.
82: v29-v32
[Abstract]
[Full Text]
-
Lowe, P., O'Loughlin, P., Evans, K., White, M., Bartley, P. B., Vohra, R.
(2006). Comparison of the Gen-Probe APTIMA Combo 2 Assay to the AMPLICOR CT/NG Assay for Detection of Chlamydia trachomatis and Neisseria gonorrhoeae in Urine Samples from Australian Men and Women.. J. Clin. Microbiol.
44: 2619-2621
[Abstract]
[Full Text]
-
Boel, C. H. E., van Herk, C. M. C., Berretty, P. J. M., Onland, G. H. W., van den Brule, A. J. C.
(2005). Evaluation of Conventional and Real-Time PCR Assays Using Two Targets for Confirmation of Results of the COBAS AMPLICOR Chlamydia trachomatis/Neisseria gonorrhoeae Test for Detection of Neisseria gonorrhoeae in Clinical Samples. J. Clin. Microbiol.
43: 2231-2235
[Abstract]
[Full Text]
-
Luijt, D. S., Bos, P. A. J., van Zwet, A. A., van Voorst Vader, P. C., Schirm, J.
(2005). Comparison of COBAS AMPLICOR Neisseria gonorrhoeae PCR, Including Confirmation with N. gonorrhoeae-Specific 16S rRNA PCR, with Traditional Culture. J. Clin. Microbiol.
43: 1445-1447
[Abstract]
[Full Text]
-
Lynn, F., Hobbs, M. M., Zenilman, J. M., Behets, F. M. T. F., Van Damme, K., Rasamindrakotroka, A., Bash, M. C.
(2005). Genetic Typing of the Porin Protein of Neisseria gonorrhoeae from Clinical Noncultured Samples for Strain Characterization and Identification of Mixed Gonococcal Infections. J. Clin. Microbiol.
43: 368-375
[Abstract]
[Full Text]
-
Boyadzhyan, B., Yashina, T., Yatabe, J. H., Patnaik, M., Hill, C. S.
(2004). Comparison of the APTIMA CT and GC Assays with the APTIMA Combo 2 Assay, the Abbott LCx Assay, and Direct Fluorescent-Antibody and Culture Assays for Detection of Chlamydia trachomatis and Neisseria gonorrhoeae. J. Clin. Microbiol.
42: 3089-3093
[Abstract]
[Full Text]
-
Tabrizi, S N, Chen, S, Cohenford, M A, Lentrichia, B B, Coffman, E, Shultz, T, Tapsall, J W, Garland, S M
(2004). Evaluation of real time polymerase chain reaction assays for confirmation of Neisseria gonorrhoeae in clinical samples tested positive in the Roche Cobas Amplicor assay. Sex. Transm. Infect.
80: 68-71
[Abstract]
[Full Text]
-
Diemert, D. J., Libman, M. D., Lebel, P.
(2002). Confirmation by 16S rRNA PCR of the COBAS AMPLICOR CT/NG Test for Diagnosis of Neisseria gonorrhoeae Infection in a Low-Prevalence Population. J. Clin. Microbiol.
40: 4056-4059
[Abstract]
[Full Text]